← Product Code [MQV](/productcode/MQV) · K102392

# MICROFUSE PUTTY AND MICROFUSE ST MIS (K102392)

_Globus Medical, Inc. · MQV · Dec 22, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102392

## Device Facts

- **Applicant:** Globus Medical, Inc.
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Dec 22, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

MicroFuse® Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. MicroFuse® Bone Void Filler may be combined with autogenous bone marrow aspirate or autograft. MicroFuse Putty is a bone graft extender to be used with autogenous bone marrow aspirate and autograft. MicroFuse® is intended to be gently packed into bony voids or gaps of the skeletal system (e.g. the spine, pelvis, and extremities). These osseous defects may be surgically created or created from traumatic injury to the bone. MicroFuse® provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

MicroFuse® is a resorbable bone graft scaffold composed of bonded poly(lactide-co-glycolide) or poly(lactic acid) microspheres; available as putty or ST MIS implants. Putty incorporates an inert biodegradable carrier. Device is packed into non-structural bony voids or gaps in the skeletal system (spine, pelvis, extremities) by surgeons during orthopedic procedures. It acts as a scaffold that resorbs over time, replaced by natural bone during healing. It may be combined with autogenous bone marrow aspirate or autograft to facilitate bone regeneration. Benefits include filling defects to support healing in non-load-bearing applications.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing (molecular weight, carrier degradation, pH, shelf life, biocompatibility) and animal testing.

## Technological Characteristics

Scaffold composed of poly(lactide-co-glycolide) or poly(lactic acid) microspheres. May contain barium sulfate and calcium sulfate. Available as putty or ST MIS implants. Resorbable material. No software or electronic components.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- MicroFuse® Bone Void Filler ([K071187](/device/K071187.md), [K082442](/device/K082442.md) and [K083232](/device/K083232.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K102392

# 510(k) Summary: MicroFuse® Additional Implants

| Company:        | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>(610) 415-9000                                                  |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Kelly J. Baker, Ph.D.<br>Director, Clinical Affairs & Regulatory                                                                           |
| Date Prepared:  | August 20, 2010                                                                                                                            |
| Device Name(s): | MicroFuse® Putty and MicroFuse® ST MIS                                                                                                     |
| Classification: | Per 21 CFR as follows:<br>§888.3045: Resorbable Calcium Salt Bone Void Filler<br>Product Code is MQV<br>Regulatory Class II Panel Code 87. |
| Predicate(s):   | MicroFuse® Bone Void Filler (K071187, K082442 and<br>K083232)                                                                              |

#### Purpose:

The purpose of this submission is to add MicroFuse® additional implants (Putty and ST MIS) to the MicroFuse® product line.

### DEVICE DESCRIPTION:

MicroFuse® Bone Void Filler is a porous bone graft scaffold composed of bonded poly (lactide-co-glycolide) or poly(lactic acid) microspheres. MicroFuse® is available with and without a combination of barium sulfate and calcium sulfate. MicroFuse® is provided in a variety of shapes and sizes, in the form of granules, sheets, pre-formed blocks, putty and ST MIS implants. MicroFuse® ST Granules and putty are designed to be gently packed into contained voids or defects. MicroFuse® Sheets are designed to be used with shallow bony defects, or as a bone graft onlay to cover a defect. MicroFuse® Blocks and ST MIS implants are designed to fill an entire defect. MicroFuse® Putty is composed of MicroFuse® ST Granules combined with an inert biodegradable carrier. MicroFuse® implants are available in short-term (ST), mid-term (MT), or long-term (LT) composition.

### Indications for Use:

MicroFuse® Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. MicroFuse® Bone Void Filler may be combined with autogenous bone marrow aspirate or autograft. MicroFuse® Putty is a bone graft extender to be used with autogenous bone marrow aspirate and autograft. MicroFuse® is intended to be gently packed into bony voids or gaps of the skeletal system (e.g. the spine, pelvis, and extremities). These osseous defects may be surgically created or created from traumatic injury to the

{1}------------------------------------------------

bone. MicroFuse® provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Performance Data:

MicroFuse® additional implants satisfy molecular weight testing, carrier degradation testing, biocompatibility testing, shelf life testing, pH testing, and animal testing, and special controls provided in the Guidance for Industry and FDA, Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device, Issued June 2, 2003,

## Basis of Substantial Equivalence:

The MicroFuse® additional implants are similar to the predicate device(s) with respect to design, indications for use and principles of operation. The information provided within this premarket notification supports substantial equivalence to the predicate device(s).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.

DEC 2 2 2010

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Globus Medical Inc. % Kelly J. Baker, Ph.D. Director, Clinical Affairs and Regulatory 2560 General Armistead Avenue Audubon, Pennsylvania 19403

ue

Re: K102392

Trade/Device Name: MicroFuse® Putty and MicroFuse® ST MIS Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: November 24, 2010 Received: November 26, 2010

Dear Dr. Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Kelly J. Baker, Ph.D.

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

A. B. Rute
for.

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

510(k) Number: K102392

MicroFuse® Putty and MicroFuse® ST MIS Device Name:

#### Indications:

MicroFuse® Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. MicroFuse® Bone Void Filler may be combined with autogenous bone marrow aspirate or autograft. MicroFuse Putty is a bone graft extender to be used with autogenous bone marrow aspirate and autograft. MicroFuse® is intended to be gently packed into bony voids or gaps of the skeletal system (e.g. the spine, pelvis, and extremities). These osseous defects may be surgically created or created from traumatic injury to the bone. MicroFuse® provides a bone void filler that resorbs and is replaced with bone during the healing process.

Prescription Use × (Per 21 CFR §801.109)

OR

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*(Division Sign-Off)*

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K102392

---

**Source:** [https://fda-staging.innolitics.com/device/K102392](https://fda-staging.innolitics.com/device/K102392)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).