← Product Code [JDR](/productcode/JDR) · K102387

# MEDISCOPE COMPRESSIOPN STAPLE AND ACCESSORIES (K102387)

_Mediscope Manufacturing, Inc. · JDR · Oct 24, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102387

## Device Facts

- **Applicant:** Mediscope Manufacturing, Inc.
- **Product Code:** [JDR](/productcode/JDR.md)
- **Decision Date:** Oct 24, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Mediscope Asymmetrical Compression Staple is intended to be used for fixation such as: osteotomies in the forefoot, first metatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfoot arthrodeses or osteotomies, fixation of osteotomies for hallux valgus treatment (Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

## Device Story

Mediscope Asymmetrical Compression Staple is a single-patient-use metallic bone fixation device. Manufactured from nickel-free stainless steel; features barbs to prevent back-out; available in 5 sizes. Used by orthopedic surgeons in clinical settings to cross fusion sites of two bones. Operates by providing mechanical compression across osteotomy or arthrodesis sites to stabilize bone segments during healing. Output is physical fixation of bone. Benefits patient by maintaining alignment and promoting fusion in foot/ankle procedures.

## Clinical Evidence

Bench testing only. Testing performed per ASTM F564-02 (Standard Specification and Test Methods for Metallic Bone Staples) comparing subject device to predicate. Tests included elastic static bending, constant amplitude bending fatigue, and pull-out fixation strength. Biocompatibility testing conducted per ISO 10993-1.

## Technological Characteristics

Material: Nickel-free stainless steel. Design: Asymmetrical staple with barbs. Dimensions: 5 sizes. Testing standards: ASTM F564-02 (mechanical testing), ISO 10993-1 (biocompatibility).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Wright Medical Compression Staple and Simple Staple ([K043059](/device/K043059.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

!!

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENES

Date of Submission

Official Contact / Address of Manufacturing facility

August 9, 2010

Daniel Helme Chief Operating Officer Mediscope Manufacturing Inc. 2832 NW 22nd Terrace Pompano Beach, FL 33069

Phone: 954-975-9997 Fax: 954-975-0775 dhelme@mediscope-mfg.com

| Trade Name                       | Mediscope Asymmetrical Compression Staple and Accessories                                                   |
|----------------------------------|-------------------------------------------------------------------------------------------------------------|
| Common/Usual Name                | Asymmetrical Compression Staple                                                                             |
| Device Classification Name       | Staple, Fixation, Bone                                                                                      |
| Classification Reference         | 21 CFR 888.3030                                                                                             |
| Classification                   | Class II                                                                                                    |
| Appropriate Classification Panel | Orthopedic                                                                                                  |
| Product Code                     | JDR                                                                                                         |
| Predicate Devices                | The device is substantially equivalent to:<br>Wright Medical Compression Staple and Simple Staple (K043059) |

Reason for submission

New device

p. lof2

{1}------------------------------------------------

### Device Description

The Mediscope Compression Staple is a single patient use device made of surgical grade The Medisope Oompression Stapie is cross the fusion site of two bones from the adiacent position.

The design features of the Mediscope Compression Staple are summarized below:

- Manufactured from nickel-free stainless steel .
- 5 sizes .
- Barbs to prevent back out .

#### Indications for Use

The Mediscope Asymmetrical Compression Staple is intended to be used for fixation such as: r in wedisoope noymined. Ostedtical osteotomies in the forefoot, first metatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfoot arthrodeses or osteotomies, fixation of osteotomies for hallux valgus

treatment (Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

## Summary of Testing and Conclusion

The Mediscope Asymmetrical Compression Staple has the following similarities to the previously cleared predicate device:

- Same operating principle. ם
- Same technology. ロ
- O Same intended use.

Substantial equivalency of the Mediscope Asymmetrical Compression Staple and the Wright Medical Charlotte Staple was proven through testing as outlined in ASTM F564-02 (Standard Specification and Test Methods for Metallic Bone Staples). The following is a list of the tests conducted used to compare the two devices: (1) Elastic Static Bending of Metallic Bone Staples; (2) Constant Amplitude Bending Fatigue Tests of Metallic Bone Staples; and (3) Pull-Out Fixation Strength of Metallic Bone Staples.

Biocompatibility testing was conducted per ISO 10993-1 (2003)(E): "Biological Evaluation of Medical Devices. Part 1: Evaluation and Testing".

(End of Tab.)

P. 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mediscope Manufacturing, Inc. % Dr. Diane Sudduth Senior RA & OA Consultant 8282 Shadow Wood Boulevard Pompano Beach, Florida 33069

OCT 2 4 2011

Rc: K102387

Trade/Device Name: Mediscope Compression Staple and Accessories Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: September 16, 2011 Received: September 19, 2011

Dear Dr. Sudduth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Dr. Diane Sudduth

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/defaul/htm.

Sincerely yours,

Fan

Mark N. Melkefo Melkerson
Director

On
Division of Surgical, Orthopedic
and Restorative Devices

and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): *K102387*

Device Name: Mediscope Asymmetrical Compression Staple and Accessories

The Mediscope Asymmetrical Compression Staple is intended to be used for The Mediooope Acymholinshrodesis, mono or bi-cortical osteotomies in the fixaton Suon ao. Lior ano alangeal arthrodesis, Akin osteotomy, midfoot and foreloot, first motatures, fixation of osteotomies for hallux valgus hindloot arthrouses of Colon), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

Prescription Use ____________________________________________________________________________________________________________________________________________________________

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michael Rines for MRM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K902387

p. lotl

Page 113 of 188

---

**Source:** [https://fda-staging.innolitics.com/device/K102387](https://fda-staging.innolitics.com/device/K102387)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).