← Product Code [NBW](/productcode/NBW) · K102383

# MICRODOT & MICRODOT XTRA (K102383)

_Cambridge Sensors Limited · NBW · Jan 27, 2012 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102383

## Device Facts

- **Applicant:** Cambridge Sensors Limited
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** Jan 27, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The microdot Home Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use only), by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. Is intended for “single patient-home use” only and should not be shared. It should be used only for testing fresh capillary whole blood samples from a fingerstick or palm for glucose (sugar). Alternate site testing should be done only during steady- state times (when glucose is not changing rapidly). It is not intended for use on neonates or in the diagnosis of or screening for diabetes mellitus. The system consists of the microdot Home Blood Glucose Meter, microdot Home Test Strips, and microdot Control solutions. microdot ® Home Test Strips The microdot ® Home Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick or palm.

## Device Story

System measures glucose in fresh capillary whole blood via electrochemical biosensor technology. User applies blood sample to disposable test strip; meter detects current generated by enzymatic reaction (Glucose Dehydrogenase/NAD). Meter displays plasma-equivalent glucose concentration (mg/dL) after 10 seconds. Designed for single-patient home use; includes meter, test strips, and control solutions. Lancing device with clear cap enables alternate site testing (palm). Healthcare providers use results to monitor diabetes control effectiveness. Benefits include convenient, rapid glucose monitoring for patients. Device is auto-coding; requires no manual calibration by user.

## Clinical Evidence

User performance study with 152 diabetic hospital outpatients. Compared palm capillary blood glucose results to YSI reference method. Results for glucose > 75 mg/dL showed 98% within ± 20% of reference. Study confirmed performance during steady-state conditions.

## Technological Characteristics

Electrochemical biosensor; Glucose Dehydrogenase/NAD enzyme; 600 nanoliter sample volume; measurement range 25-525 mg/dL; operating range 10-40°C, 10-90% RH; auto-coding; plasma-equivalent calibration; single-patient use; disinfection validated for 375 cycles using bleach-based wipes.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- microdot® Blood Glucose Monitoring System (k070524)
- microdot® Xtra Blood Glucose Monitoring System (k083273)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k102383

B. Purpose for Submission:
Addition of alternative site testing on the palm to previously cleared device (k083273)

C. Measurand:
Capillary whole blood glucose

D. Type of Test:
Quantitative, amperometric glucose dehydrogenase (NAD)

E. Applicant:
Cambridge Sensors Limited

F. Proprietary and Established Names:
microdot Home Blood Glucose Monitoring System

G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345 Glucose Test System
2. Classification:
Class II
3. Product code:
NBW – Blood glucose test system, over the counter
LFR – Glucose dehydrogenase, glucose

{1}

4. Panel:
Chemistry 75

H. Intended Use:

1. Intended use(s):
See indications for use below.

2. Indication(s) for use:
The microdot Home Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use only), by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. Is intended for “single patient-home use” only and should not be shared. It should be used only for testing fresh capillary whole blood samples from a fingerstick or palm for glucose (sugar). Alternate site testing should be done only during steady- state times (when glucose is not changing rapidly). It is not intended for use on neonates or in the diagnosis of or screening for diabetes mellitus.

The system consists of the microdot Home Blood Glucose Meter, microdot Home Test Strips, and microdot Control solutions.

microdot ® Home Test Strips
The microdot ® Home Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick or palm.

3. Special conditions for use statement(s):
Not for neonatal use, not for screening or diagnosis of diabetes mellitus.

Not for use on critically ill patients, patients in shock, severely hypotensive individuals, dehydrated patients or individuals experiencing hyperglycemic-hyperosmolar state, with or without ketosis.

Alternative site testing (AST) should not be used to calibrate continuous glucose monitors (CGMs) nor for use in insulin dose calculations.

AST testing should only be done during steady-state times (when glucose is not changing rapidly).

Single-patient use only

4. Special instrument requirements:
microdot Home Blood Glucose Meter

{2}

I. Device Description:

The microdot Home Blood Glucose Monitoring System consists of the microdot Home Blood Glucose Meter, microdot Home Test Strips, one bottle of microdot Control Solution, lancing device and lancets. Two additional levels of control solutions may be purchased separately.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Microdot Xtra Blood Glucose Monitoring System

2. Predicate 510(k) number(s):

k083273

3. Comparison with predicate:

The modified device has the same intended use and performance characteristics as the predicate. The difference is the new device has the addition of alternative site testing for palm.

K. Standard/Guidance Document Referenced (if applicable):

ISO 15197: 2003 In vitro diagnostic test systems - Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus

L. Test Principle:

The test is based on the enzymatic conversion of glucose. Glucose dehydrogenase converts glucose in the sample to gluconolactone, with concomitant reduction of the enzyme cofactor NAD+ to NADH. The NADH is re-oxidised to NAD+ by the mediator compound which in turn becomes reduced; re-oxidation of the mediator by the meter induces a micro current to flow, and the size of this micro current is directly proportional to the amount of glucose in the sample.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Precision and reproducibility were evaluated in k070524.

b. Linearity/assay reportable range:

{3}

Linearity was evaluated in k070524.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The controls supplied with this device were cleared under k070524.

d. Detection limit:

The detection limit of the test system is 20 mg/dL, and was evaluated in k075024.

e. Analytical specificity:

Specificity was evaluated in k070524.

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Method comparison studies to demonstrate accuracy using capillary whole blood samples from the fingertip were summarized in k070524

b. Matrix comparison:

Not applicable. For use with capillary whole blood only.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Alternative site testing (AST) for blood glucose measurement on samples from the palm were evaluated during a user performance study with 152 diabetic hospital outpatients. Each subject was asked to perform a self test using only the provided English language labeling. Testing was performed by the lay user using the palm to show comparable performance at times of steady state conditions with the fingertip glucose results. The fingertip blood specimen was obtained by a healthcare

4

{4}

professional and the plasma was tested on the YSI reference method.

The range of glucose values for these samples was 48-367 mg/dL (by YSI).

The results are summarized in the tables below.

For glucose concentrations ≤ 75 mg/dL

|  within ± 5 mg/dL | within ± 10 mg/dL | within ± 15 mg/dL  |
| --- | --- | --- |
|  0/3 (0%) | 2/3 (67%) | 3/3 (100%)  |

For glucose concentrations &gt; 75 mg/dL

|  within ± 5 % | within ± 10 % | within ± 15 % | within ± 20 %  |
| --- | --- | --- | --- |
|  61/149 (41%) | 109/149 (73%) | 134/149 (90%) | 146/149 (98%)  |

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Expected blood glucose results for people without diabetes were cited from the literature and presented in the labeling as follows:

Before breakfast (fasting): &lt;100 mg/dL

Before meals: &lt;110 mg/dL

2 hours after meals: &lt;140 mg/dL

Bedtime: &lt;120 mg/dL

American Diabetes Association, Standards of medical care in diabetes, Diabetes Care, Volume 34, Supplement 1, S11-S69 (2011)

N. Instrument Name:

Microdot Home Blood Glucose Meter

O. System Descriptions:

1. Modes of Operation:

Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes ______ or No ☐ X

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ______ or No ☐ X

{5}

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X _______ or No ☐

3. Specimen Identification:

There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.

4. Specimen Sampling and Handling:

The glucose test is intended to be used with capillary whole blood from the finger or palm. The whole blood sample is applied directly to the test strip.

5. Calibration:

Calibration is factory set by the manufacturer. No coding by the user is required.

6. Quality Control:

The sponsor is providing one level of glucose control solution with the starter kit for the device. Two additional control levels are available for purchase separately. Recommendations on when to test the control materials are provided in the labeling. An acceptable range for each control level is printed on the test strip vial label.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above:

A study was conducted to evaluate the effect of altitude up to 10,000 feet on performance of the test system. Venous whole blood samples at three glucose concentrations (approximately 60, 120 and 350 mg/dL) were tested at sea level and 10,000 feet using a hypobaric chamber, with one lot of test strips and one meter. Each sample was also evaluated by the YSI method. At altitudes up to 10,000 feet, test results supported the claimed altitude.

The meter and lancing device are intended for single-patient use only. Dispatch Hospital Cleaner with Bleach Wipes with EPA registration # 56392-8 were validated for use with the meter and lancing device. Laboratory studies demonstrated that levels of virus were below the detection limit of the assay after disinfection. The sponsor also demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 375 cleaning and disinfection cycles designed to simulate 5 years of device use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.

{6}

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K102383](https://fda-staging.innolitics.com/device/K102383)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com