EASYMAX T1 SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS10042

K102345 · Eps Bio Technology Corp. · NBW · Jun 21, 2011 · Clinical Chemistry

Device Facts

Record IDK102345
Device NameEASYMAX T1 SELF MONITORING BLOOD GLUCOSE SYSTEM, MODEL EPS10042
ApplicantEps Bio Technology Corp.
Product CodeNBW · Clinical Chemistry
Decision DateJun 21, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The EasyMax T1 Self Monitoring Blood Glucose System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, forearm, and upper arm. It is intended for use by a single patient and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The EasyMax T1 Pro Self Monitoring Blood Glucose System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, forearm, and upper arm. It is intended for use by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

System measures glucose in capillary whole blood via electrochemical test strips; meter processes electrical signal to calculate glucose concentration; results displayed on screen for patient or clinician; aids diabetes management; T1 Pro intended for professional multi-patient use; T1 intended for single-patient home use; includes control solutions for accuracy verification.

Clinical Evidence

Bench testing only. Validation focused on device modification impact and disinfection durability. 11,000 cleaning/disinfection cycles (22,000 total wipes) performed to simulate 3-4 years of use; demonstrated no change in performance or external material integrity. Viral inactivation validated for PDI Super Sani-Cloth.

Technological Characteristics

Electrochemical glucose test system; consists of meter and single-use test strips; includes control solutions for quality verification.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from finger, palm, forearm, and upper arm. For use by people with diabetes (single patient) or healthcare professionals (professional setting) to monitor diabetes control. Not for diagnosis/screening of diabetes or use on neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIVD Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER: k102345 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) EPS Bio Technology Corporation, EasyMax N Self Monitoring Blood Glucose System - k083099. 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: A. The physical appearance of the meter has changed. B. The two button (M and S) functions are removed. C. The trade name of the meter has changed from EasyMax N to EasyMax T1. D. The memory function has changed from "a memory capacity of 480 test results" to "no memory capacity". E. The power supply source changed from "two alkaline 1.5V batteries" to "two CR2032 batteries". F. Print circuit board has changed. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis - risk analysis was FMEA. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. {1} The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. ## Reviewer's Comments: The device is intended for single patient home use (EasyMax T1 Self Monitoring Blood Glucose System) and multiple patient use in a professional healthcare settings (EasyMax T1 Pro Self Monitoring Blood Glucose System). PDI® SUPER SANI-CLOTH® Germicidal disposable wipes (EPA Reg. No: 9480-4) was validated demonstrating complete inactivation of live virus for use with the meter and lancing device. The sponsor also demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 11,000 cleaning/disinfection cycles (each cleaning/disinfection cycle includes 11,000 pre-cleaning and 11,000 disinfection wipes adding up to a total of 22,000 wipes) designed to simulate 3 years (EasyMax T1 Pro Self Monitoring Blood Glucose System) and 4 years (EasyMax T1 Self Monitoring Blood Glucose System) of device use. Labeling has been reviewed for adequate instructions in validated cleaning and disinfection procedures.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...