ELECSYS FERRITIN CALCHECK 5

K102267 · Roche Diagnostics · JJX · Sep 10, 2010 · Clinical Chemistry

Device Facts

Record IDK102267
Device NameELECSYS FERRITIN CALCHECK 5
ApplicantRoche Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateSep 10, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

The Elecsys Ferritin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Ferritin reagent on the indicated Elecsys and cobas e immunoassay analyzers.

Device Story

Elecsys Ferritin CalCheck 5 is an assayed quality control product consisting of five levels of human ferritin in a human serum matrix. It is used in clinical laboratories to verify the calibration and reportable assay range of the Elecsys Ferritin reagent on Modular Analytics E170/cobas e 601 and Elecsys 2010/cobas e 411 analyzers. The device provides target values for ferritin concentrations ranging from <2 ng/mL to 1970 ng/mL. Healthcare providers use the output to ensure the accuracy and linearity of the immunoassay system during routine testing, supporting reliable ferritin measurement for patient diagnosis and monitoring.

Clinical Evidence

No clinical data. Bench testing only. Value assignment was performed by duplicate runs on multiple analyzer platforms (Modular Analytics E170/cobas e 601 and Elecsys 2010/cobas e 411). Stability was validated via accelerated and real-time studies.

Technological Characteristics

Liquid control material; human liver ferritin in human serum matrix; five concentration levels; stored at 2-8°C; intended for use on Elecsys and cobas e immunoassay analyzers.

Indications for Use

Indicated for use as an assayed control for calibration verification and assay range verification of the Elecsys Ferritin reagent on specified Elecsys and cobas e immunoassay analyzers. For in vitro diagnostic and prescription use.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k102267 B. Purpose for Submission: New device C. Measurand: Calibration verification and assay range verification for Elecsys Ferritin Reagent D. Type of Test: Assayed quality control materials E. Applicant: Roche Diagnostics F. Proprietary and Established Names: Elecsys Ferritin CalCheck 5 G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJX | Class I, reserved | 862.1660 | Clinical Chemistry | H. Intended Use: 1. Intended use(s): Refer to indications for use below 2. Indication(s) for use: The Elecsys Ferritin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Ferritin reagent on the indicated Elecsys and cobas e immunoassay analyzers. {1} 3. Special conditions for use statement(s): For in vitro diagnostic use For prescription use 4. Special instrument requirements: For use with Elecsys Ferritin reagent on the Modular Analytics E170/cobas e 601 and Elecsys 2010/cobas e 411 analyzers. I. Device Description: The Elecsys Ferritin CalCheck 5 set contains five levels of ferritin from human liver in a human serum matrix. During manufacture, human ferritin is spiked into the matrix at the target concentrations listed below. CalCheck Level 1 is free of analyte. | Elecsys Ferritin CalCheck Level | Target Value (ng/mL) | | --- | --- | | Level 1 | Less than 2 | | Level 2 | 27.5 | | Level 3 | 915 | | Level 4 | 1550 | | Level 5 | 1970 | All human source materials were prepared exclusively from the blood of donors tested individually and found to be nonreactive for HBsAg, antibodies to HCV and HIV. The testing methods were FDA approved or cleared in compliance with the European Directive 98/79/EC. J. Substantial Equivalence Information: 1. Predicate device name(s): Elecsys CalCheck Ferritin 2. Predicate 510(k) number(s): k981281 (Elecsys Ferritin) 3. Comparison with predicate {2} | | Candidate Device | Predicate - Elecsys CalCheck Ferritin (k981281) | | --- | --- | --- | | Indications for Use | The Elecsys Ferritin CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys Ferritin reagent on the indicated Elecsys and cobas e immunoassay analyzers. | Same | | Analyte | Ferritin | Same | | Matrix | Human serum | Same | | Format | Aqueous | Same | | Number of levels per set | 5 | 3 | | Stability | Unopened: Store at 2 to 8°C until expiration date Opened: 4 hours at 20 to 25°C | Unopened: Same Opened: 5 hours at 20 to 25°C | K. Standard/Guidance Document Referenced (if applicable): None were referenced L. Test Principle: Not applicable M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: {3} Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): **Traceability** The sponsor claims traceability of ferritin used in this quality control set to National Institute for Biological Standards and Controls (NIBSC) reagent for Ferritin (human liver) 80/602. The Elecsys Ferritin assay was standardized against the Enzymun-Test Ferritin method. **Value Assignment** For each Elecsys Ferritin CalCheck 5 lot manufactured, value assignment for each level of CalCheck was performed by duplicate runs on at least three Elecsys Modular Analytics E170/cobas e 601 analyzers. Additionally, the same value assignment procedure was performed on the Elecsys 2010/cobas e 411 analyzers. The assigned target value of each CalCheck is the median of the observed values. The labeling states that obtained values should fall within the specified range provided on lot-specific value sheets and that laboratories should establish appropriate acceptance criteria when using this product for its intended use. **Stability** Stability testing protocols and acceptance criteria were reviewed and found to be acceptable. Closed (unopened) vial shelf life of 2 years at the recommended storage temperature (2 to 8°C) was demonstrated based on accelerated stability studies. Real-time closed-vial stability studies for the Elecsys Ferritin CalCheck 5 solutions are ongoing. Open vial stability of 4 hours was demonstrated at the recommended storage temperature of 20 to 25°C. Storage recommendations are provided in the labeling. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: {4} a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The expected values are provided in the labeling for each specific lot. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...