# GE LOGIQ I/E & VIVID E COMPACT DIAGNOSTIC ULTRASOUND (K102256)

_Ge Medical Systems China Co., Ltd. · IYN · Oct 5, 2010 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102256

## Device Facts

- **Applicant:** Ge Medical Systems China Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Oct 5, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The LOGIQ ile & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OG; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.

## Device Story

Compact, portable diagnostic ultrasound system (LOGIQ i, LOGIQ e, Vivid e) featuring integrated keyboard, LCD display, and interchangeable electronic-array transducers. Device acquires, processes, and displays ultrasound data; supports B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. Used in clinical settings by physicians for real-time imaging and fluid flow analysis. Output displayed on integrated GUI; assists clinicians in diagnostic evaluation of anatomy and pathology. Benefits include portable, point-of-care diagnostic capability for diverse clinical applications.

## Clinical Evidence

No clinical data required. Substantial equivalence established via non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, electromagnetic compatibility, and thermal/electrical/mechanical safety testing.

## Technological Characteristics

Portable ultrasound system with integrated keyboard, LCD, and electronic-array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, and harmonic imaging modes. Conforms to 21 CFR 892.1560, 892.1550, and 892.1570. Design conforms to 21 CFR 820, ISO 9001:2000, and ISO 13485. Connectivity includes digital acquisition and display.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ-ile & Vivid-e Compact Ultrasound ([K091374](/device/K091374.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**OCT** 5 2010

142

# Attachment B

510(k) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

#### GE Healthcare GE Medical Systems (China) Co., Ltd No. 19 Changjiang Road, National Hi-Tech Development Zone Wuxi, Jiangsu Province, CHINA 214028 Section a): 1. Submitter: GE Medical Systems (China) Co., Ltd. No. 19 Changjiang Road, National Hi-Tech Development Zone, Wuxi, Jiangsu Province, CHINA 214028 Contact Person: Yalan Wu, Manager, Safety and Regulatory Telephone: 86-510-85278652; Fax: 86-510-85227347 Date Prepared: June 20, 2010 2. Device Name: GE LOGIQ ille, Vivid e Ultrasound Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN Diagnostic Ultrasonic Transducer, 21 CFR 892.1570, 90-ITX 3. Marketed Device: GE LOGIQ-ile & Vivid-e Compact Ultrasound, 510(k) No: K091374 (90-IYO/IYN/ITX) A device currently in commercial distribution.

4. Device Description: The GE Compact Ultrasound is a very compact and portable diagnostic ultrasound system having three variations: LOGIQ e and Vivid e, each with options and features suited for its It has an integrated keyboard, LCD display and several interchangeable electronic-array market niche. transducers with an approximate size of 34 cm wide, 29 cm deep and 6 cm high in transport configuration and provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, an intuitive layout of specialized controls, color GUI display and Doppler audio.

5. Indications for Use: The subject modified device is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intraoperative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.

6. Companison with Predicate Device: The modified Compact Ultrasound devices are virually identical having the same design, construction, materials, brand names and intended uses. All technological characteristics and safety and effectiveness features are equivalent. The modified device has additional Transesophageal indication for use and Intima Media Thickness (IMT) measurement with Vascular.

#### Section b):

Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and 1. disinfection effectiveness, electromagnetic compatibility, as well as themal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

- 2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Healthcare that the GE LOGIQ ile Vivid e Ultrasound imaging device is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 2510 Street NW BUFFALO MN 55313

OCT 5 2010

Re: K102256

Trade/Device Name: GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasound pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: September 27, 2010 Received: September 28, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound, as described in your premarket notification:

|        | Transducer Model Number |        |
|--------|-------------------------|--------|
| 4C-RS  | 12L-RS                  | 6S-RS  |
| 8C-S   | 16L-RS                  | P2D    |
| E8C-RS | i12L-RS                 | 6Tc-RS |
| 8L-RS  | i/t739-RS               |        |
| 9L-RS  | 3S-RS                   |        |

{2}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely yours,

Mach. A.D.O. Then for

David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

# Indications for Use Form

510(k) Number (if known):

Device Name: __GE LOGIQ i, LOGIQ e and Vivid e Diagnostic Ultrasound

#### Indications For Use:

The LOGIQ ile & Vivid e is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OG; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transesophageal, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics

MuhaMD.D.M. David 6, Brown
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102056

Page 1 of 1

{4}------------------------------------------------

### GE Compact Ultrasound System

LOGIQ i, LOGIQ e, Vivid e

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|--|
|                                                    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Small Organ [2]                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Cardiac [3]                                        | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Thoracic/Pleural (specify)[4]                      | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Other [5]                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
| Transesophageal                                    | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N               | N     |  |
| Transrectal                                        | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                 |       |  |
| Transvaginal                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                 |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
| Intraoperative                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

{4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse is for digitally encoded harmonics.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Nuhal D. O'Kee for David C. Brown
Division Sign Off

510(k) K102256

Office of In Vitro Diagnostic Device Evaluation and Safety

{5}------------------------------------------------

# GE Compact Ultrasound with 4C-RS Transducer

#### LOGIQ i, LOGIQ e, Vivid e

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Small Organ [2]                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Thoracic/Pleural (specify) [4]                     | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Other[5]                                           | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse is for digitally encoded harmonics.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Mudand D. O'K. Go. David b. Brown
Division Sign-Off

Prescription Use (21 CFR 801 Subpart D)

Office of In Vitro Diagnostic Device Evaluation and Safe

510(k) K102256

{6}------------------------------------------------

### GE Compact Ultrasound with 8C-RS Transducer

#### LOGIQ i, LOGIQ e, Vivid e

intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Small Organ [2]                                    | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Neonatal Cephalic                                  | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac [3]                                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Thoracic/Pleural (specify) [4]                     | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Other[5]                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

(*) Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

] Coded Pulse is for digitally encoded harmonics .

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Mhll.D.O'K. for David C. Brown
Division Sign-Off

Prescription Use (21 CFR 801 Subpart D)

Office of In Vitro Diagnostic Device Evaluation and S

510(k) K102256

{7}------------------------------------------------

## GE Compact Ultrasound with E8C-RS Transducer

#### LOGIQ İ, LOGIQ e, Vivid e

|                                                     |   | Mode of Operation |               |               |                  |                    |                  |                   |                     |                |  |
|-----------------------------------------------------|---|-------------------|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|--|
| Clinical Application<br>Anatomy/ Region of interest | B | M                 | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |  |
| Ophthalmic                                          |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Fetal / Obstetrics                                  | P | P                 | P             |               | P                |                    | P                | P                 | P                   |                |  |
| Abdominal[1]                                        | P | P                 | P             |               | P                |                    | P                | P                 | P                   |                |  |
| Pediatric                                           |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Small Organ [2]                                     |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Neonatal Cephalic                                   |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Adult Cephalic                                      |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Cardiac                                             |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Peripheral Vascular                                 |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Musculo-skeletal Conventional                       |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Musculo-skeletal Superficial                        |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Thoracic/Pleural (specify) [4]                      |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Other[5]                                            | P | P                 | P             |               | P                |                    | P                | P                 | P                   |                |  |
| Exam Type, Means of Access                          |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Transesophageal                                     |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Transrectal                                         | P | P                 | P             |               | P                |                    | P                | P                 | P                   |                |  |
| Transvaginal                                        | P | P                 | P             |               | P                |                    | P                | P                 | P                   |                |  |
| Transuretheral                                      |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Intraoperative (specify)                            |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Intraoperative Neurological                         |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Intravascular                                       |   |                   |               |               |                  |                    |                  |                   |                     |                |  |
| Laparoscopic                                        |   |                   |               |               |                  |                    |                  |                   |                     |                |  |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

[*] Coded Pulse is for digitally encoded harmonics .

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Andrew D. Kirk for David G. Brown
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{8}------------------------------------------------

## GE Compact Ultrasound with 8L-RS Transducer

#### LOGIQ İ, LOGIQ e, Vivid e

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Small Organ [2]                                    | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac [3]                                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Thoracic/Pleural (specify) [4]                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other [5]                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)[5]                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

[1] Coded Pulse is for digitally encoded harmonics.

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Andrew D. Thoms for David C. Brown
Division Sign Off

Prescription Use (21 CFR 801 Subpart D)

Office of In Vitro Diagnostic Device Evaluation and Safe

510(k) K102256

{9}------------------------------------------------

## GE Compact Ultrasound with 9L-RS Transducer

LOGIQ İ, LOGIQ e, Vivid e

Mode of Operation Clinical Application PW CW Color Color M B Power Combined M Harmonic Coded Other Anatomy/Region of Interest Doppler Doppler Doppler Doppler Doppler Modes lmaging Pulse® Ophthalmic Fetal / Obstetrics Abdominal [1] P P P P P P P P Pediatric P P P P P P P P Small Organ 같 P P P P P P P P Neonatal Cephalic Adult Cephalic Cardiac (3) Peripheral Vascular P P P P P P P P Musculo-skeletal Conventional P P P P P P P P Musculo-skeletal Superficial Thoracic/Pleural (specify) [4] Other (5) Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transuretheral Intraoperative (specify)[5] P P P P P P P P Intraoperative Neurological Intravascular Laparoscopic

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

*] Coded Pulse is for digitally encoded harmonics .

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Michal D. O'Shea Lisa David G. Brown
Division Sign-Off

Prescription Use (21 CFR 801 Subpart D)

ce of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{10}------------------------------------------------

## GE Compact Ultrasound with 12L-RS Transducer

#### LOGIQ i, LOGIQ e, Vivid e

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Fetal / Obstetrics                                 |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Abdominal                                          |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Pediatric                                          | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Small Organ [2]                                    | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Cardiac                                            |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Peripheral Vascular                                | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Musculo-skeletal Conventional                      | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Musculo-skeletal Superficial                       | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Thoracic/Pleural (specify) [4]                     | P | P | P             |               | P                |                    | P                | P                 | P                   | P               |
| Other (specify) [5]                                |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Transrectal                                        |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Intraoperative [5] (specify)                       |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                   |                     |                 |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                   |                     |                 |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(*) Coded Pulse is for digitally encoded harmonics.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Andra D'Oliveira for David 6 Brown
Division Sign Off

510(k) K102256

Office of In Vitro Diagnostic Device Evaluation and Safety

{11}------------------------------------------------

## GE Compact Ultrasound with 16L-RS Transducer

### LOGIQ İ, LOGIQ e, Vivid e

Mode of Operation Clinical Application bM CM B Color Power Combined Color M M Harmonic Coded Anatomy/Region of Interest Doppler Doppler Doppler Doppler Modes Doppler Imaging Pulse® Ophthalmic Fetal / Obstetrics Abdominal Pediatric P P P p P P P P Small Organ 121 P P P P P P P P Neonatal Cephalic Adult Cephalic Cardiac Peripheral Vascular P P P P P P P P Musculo-skeletal Conventional P P P P P P P P Musculo-skeletal Superficial P P P P P P P P Thoracic/Pleural (specify) [4] Other (specify) 15) Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transuretheral Intraoperative [3] (specify) Intraoperative Neurological Intravascular Laparoscopic

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse is for digitally encoded harmonics.

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Aayyad Dothan Lan David 6. Boiso
Division Sign Off

office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{12}------------------------------------------------

## GE Compact Ultrasound with i12L-RS Transducer

#### LOGIQ i, LOGIQ e, Vivid e

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal [1]                                      | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Small Organ [2]                                    | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac [3]                                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Thoracic/Pleural (specify) [4]                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other [5]                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)[5]                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

{4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, 8/Power/PWD.

[1] Coded Pulse is for digitally encoded harmonics .

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

nce of CDRH, Office of In Vitro Diagnostics (OIVD)

Auhul 1052K for David b Brown
Division Sign-Off

fice of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{13}------------------------------------------------

## GE Compact Ultrasound with i/t739-RS Transducers

#### LOGIQ i, LOGIQ e, Vivid e

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Small Organ [2]                                    | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac[3]                                         | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Thoracic/Pleural (specify) (4)                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other (4)                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)[5]                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse is for digitally encoded harmonics.

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Ardd DDK for David b. Brown
Division Sign Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{14}------------------------------------------------

## GE Compact Ultrasound with 3S-RS Transducer

### LOGIQ i, LOGIQ e, Vivid e

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Fetal / Obstetrics                                 | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Abdominal[1]                                       | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Pediatric                                          | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Small Organ [2]                                    |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Adult Cephalic                                     | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Cardiac [3]                                        | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Thoracic/Pleural (specify)[4]                      | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Other [5]                                          | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transrectal                                        |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative (specify)                           |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                   |                     |                 |       |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

[*] Coded Pulse is for digitally encoded harmonics .

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics (OIVD)

Nuhad D. Khawam for David 6. Brown
Division Sign-Off

Prescription Use (21 CFR 801 Subpart D)

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102256

{15}------------------------------------------------

# GE Compact Ultrasound with 6S-RS Transducer

#### LOGIQ İ, LOGIQ e, Vivid e

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                 |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse¹ | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Small Organ [2]                                    |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Thoracic/Pleural (specify) [4]                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Other [5]                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               | P     |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative (specify)[5]                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN and Urological

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] For detection of fluid and pleural motion/sliding;

[5] Other use includes Urology/Prostate

["] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse is for digitally en…

---

**Source:** [https://fda-staging.innolitics.com/device/K102256](https://fda-staging.innolitics.com/device/K102256)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).