DASH KNEE

K102251 · Brainlab AG · OLO · May 17, 2011 · Neurology

Device Facts

Record IDK102251
Device NameDASH KNEE
ApplicantBrainlab AG
Product CodeOLO · Neurology
Decision DateMay 17, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee.

Device Story

DASH knee is an intraoperative image-guided surgery system for total knee replacement. It uses passive markers on surgical instruments tracked by a camera stand to map physical tools to a virtual 3D model of the patient's bone. The 3D model is generated intraoperatively by acquiring bone surface landmarks; no preoperative CT is required. The system features a main calculation unit integrated into the camera stand and an iPod touch acting as an embedded navigation display, communicating wirelessly. Surgeons use the system to navigate cutting guides, perform ligament balancing, and measure bone alignment. The device aids in accurate prosthesis placement, potentially improving surgical outcomes through minimally invasive navigation.

Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing, usability workshops (use labs) simulating OR setups with plastic bones, and pre-clinical validation in a cadaver lab. A literature search was performed to support the safety and effectiveness of the underlying BrainLAB computer-assisted knee replacement software.

Technological Characteristics

Image-guided stereotaxic instrument; passive marker tracking; wireless communication between camera-integrated calculation unit and iPod touch display. Software-based landmark registration and navigation. No preoperative imaging required.

Indications for Use

Indicated for patients undergoing orthopedic surgery, including total knee replacement and ligament balancing, where stereotactic navigation of rigid anatomical structures (skull, long bones, vertebrae) is appropriate.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 1 7 2011 # KIO2251 510 (k) Summary of Safety and Effectiveness for DASH knee Manufacturer: | Address: | BrainLAB AG | |-----------------|--------------------------| | | Kapellenstrasse 12 | | | 85622 Feldkirchen | | | Germany | | Phone: | +49 89 99 15 68 0 | | Fax: | +49 89 99 15 68 33 | | Contact Person: | Mr. Alexander Schwiersch | | Summary Date: | July 12, 2010 | #### Device Name: Trade name: DASH knee Common/Classification Name: BrainLAB DASH, BrainLAB Image Guided Surgery System / Instrument, Stereotaxic #### Predicate Device: BrainLAB Knee (K073615) Kolibri Image Guided Surgery System (K014256) Device Classification Name: Instrument, Stereotaxic Regulatory Class: Class II Regulation Number: 21 CFR 882.4560 Product Code: OLO ### Device Description: Dash is an image guided surgery system for total knee replacement surgery based on landmark based visualization of the femur and tibia. It is intended to enable operational navigation in orthopedic surgery. It links a surgical instrument, tracked by passive markers to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. DASH knee uses the registered landmarks to navigate the femoral and tibial cutting guides to the optimally position. DASH knee software registers the patient data needed for navigating the surgery intraoperatively. No preoperative CT-scanning is necessary. #### Intended Use: DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical {1}------------------------------------------------ structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee. #### Changes made to the predicate device: Reduced SW complexity: The spectrum of possible workflows has been reduced to one already contained universal express workflow. Within that express workflow registration steps have been combined with the help of an additional instrument. Additional instrument: Device for multiple point acquisition on femur condyles. Compared to the predicate device, where femoral landmarks have been calculated out of two surface acquisition steps, regarding landmarks are here acquired directly in one step. In combination with a regarding femoral acquisition page the amount of registration steps has been reduced for the user. Reduced Platform: Compared to predicate device, the main calculation unit is part of the camera stand. A smaller separated display, represented by an iPod touch, acts an embedded display in the instrumentation for navigation. Wireless communication between embedded display (iPod touch) and main calculation unit (integrated in camera stand). #### Completed verification activities: Following design verification activities have been per formed to ensure correct system functionality as it has been specified: The first part of the verification covered the instrument and system accuracy during registration and navigation. The registration values have been compared to external measured reference values. After the verification of the instruments in combination with the software the verification of the software algorithms itself has been performed. Part three of the verification includes the testing of all possible workflows to ensure the correct behavior of the system for all possible procedures. With the knowledge of the above named points the current device has been compared to the predicate device. The next step was the detailed verification of the signed specifications covering the detailed functionality of buttons for example. At last the measures aqainst the defined risks of the risk analysis have been tested. This strategy ensures the verification of basic software algorithms up to specific detailed functionality, the comparison to the predicate device and the safety of the defined measures of the risk analysis. All tests have been successfully completed. {2}------------------------------------------------ #### Completed validation activities Following design validation activities have been performed: A literature search has been performed to prove safety and effectiveness of BrainLAB computer assisted total knee replacement software. This applies directly to DASH knee, since DASH knee is derived from the previous marketed device BrainLAB knee (K073615), Non-clinical Validation has been performed to prove the system targets and supplement requirement specifications if necessary. With the help of usability workshops (use labs) OR setups and surgery proceedings have been simulated with plastic bones (sawbones). Pre-clinical Validation has been performed to confirm/ complete detailed specification for each requirement. Here OR setups and surgery proceedings have been simulated in a cadaver lab. Testing persons went through same procedure like for the non clinical use lab sessions. To prove that all validation issues are addressed, a final validation has been performed under non clinical conditions. #### Substantial equivalence: Dash knee has been verified and validated according to BrainLAB's procedures for product design and development. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device BrainLAB Knee (K073615) and Kolibri Image Guided Surgery System (K014256). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 BrainLAB AG % Mr. Alexander Schwiersch Kapellenstrasse 12 85622 Feldkirchen, Germany Re: K102251 MAY 1 7 2011 Trade/Device Name: DASH Knee Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: April 29, 2011 Received: May 06, 2011 Dear Mr. Schwiersch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Mr. Alexander Schwiersch comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. erely yours, Kof Petro N.C.P D.I Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number: K102251 Device Name: DASH knee Indications For Use: DASH knee is intended to be an intraoperative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to the anatomy. The system aids the surgeon to accurately navigate a knee prosthesis to the intraoperatively planned position. Ligament balancing and measurements of bone alignment are provided by DASH knee. Example orthopedic surgical procedures include but are not limited to: - · Total Knee Replacement - Ligament Balancing Prescription Use *X* (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | <i>Nilfo Dyden</i> for <i>mxm</i> | |---------------------|-----------------------------------| | (Division Sign-Off) | | Division of Surgical, Orthopedic, and Restorative Devices | 510(k) Number | K102251 | |---------------|---------| |---------------|---------| | Page | 1 of 1 | |------|--------| |------|--------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...