The device was evaluated in a US hospital during routine clinical procedures to assess the effectiveness of the high-level disinfection process on endoscopes used in clinical practice.
In-use study in a US hospital; Observational/In-use evaluation
Flexible endoscopes used in clinical procedures; Sample Size: Three flexible endoscopes; Number of Sites: 1
Not applicable for this study
Microbiological recovery after processing (absence of organisms)
Indications for Use
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6-minute exposure sequence).
Device Story
The Reliance Endoscope Processing System is an electromechanical washer/disinfector for reprocessing flexible endoscopes. It processes up to two devices simultaneously. The system uses Reliance DG Dry Germicide, a two-part dry chemistry that generates peracetic acid when mixed with water. The processing cycle includes an optional enzymatic wash phase, a high-level disinfection phase (50-57°C), a rinse phase using 0.2-micron filtered water, and a HEPA-filtered air purge. A microprocessor-based controller manages cycles and performs self-decontamination (D-SHORT and D-LONG cycles). The device is operated by clinical staff in a healthcare setting. It provides a printed cycle summary for documentation, which can include operator, patient, and device identifiers via an optional barcode reader. The system ensures patient safety by automating disinfection parameters, monitoring chemical presence via process indicators, and verifying filter integrity. It benefits patients by providing standardized, validated high-level disinfection of heat-sensitive endoscopes, reducing the risk of cross-contamination.
Clinical Evidence
Bench testing included microbiological efficacy studies (AOAC sporicidal, tuberculocidal, bactericidal, fungicidal tests) and simulated-use studies demonstrating >6 log10 reduction of Mycobacterium terrae. In-use clinical evaluation in a US hospital processed three types of flexible endoscopes; no organisms were recovered after processing. Bioburden levels prior to HLD were up to 10^5 CFU/device. Material compatibility was validated over 300 cycles with no functional degradation.
Technological Characteristics
Electromechanical washer/disinfector; microprocessor-controlled. Uses Reliance DG Dry Germicide (peracetic acid-generating oxidative chemistry). Features 0.2-micron bacterial-retentive water filtration with automated integrity check and HEPA-filtered air purge. Certified to ANSI/UL 61010-1, CAN/CSA C22.2 61010-1, IEC 61010-1, IEC 61010-2-40, and IEC 61326-1. Materials evaluated for biocompatibility and device compatibility. Standalone system.
Indications for Use
Indicated for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices, including GI flexible endoscopes, bronchoscopes, and their accessories.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "STERIS" in large, bold, capital letters. Above the word "STERIS" is the text "K102244". To the right of the word "STERIS" is the text "p1 of 7". Below the word "STERIS" is a logo that consists of several horizontal lines.
DEC 3 2010
# 510(k) Summary For Reliance® Endoscope Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact:
Robert F. Sullivan Senior Director, Regulatory Affairs Telephone: (440) 392-7695 (440) 357-9198 Fax No:
October 14, 2010 Summary Date:
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{1}------------------------------------------------
<102^{244}
p^{2}f
#### Device Name 1.
| Trade Name: | Reliance Endoscope Processing System |
|----------------------|-------------------------------------------------------------|
| Common/usual Name: | Automated Endoscope Reprocessor |
| Classification Name: | 21 CFR 876.1500<br>Endoscope Cleaning Germicide Accessories |
#### 2. Predicate Device
Reliance Endoscope Processing System K040049
### 3. Description of Device
The Reliance Endoscope Processing System is an economical, easy-to-use high level disinfection system intended to wash and high level disinfect up to two manually cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and related accessories.
The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories.
- The Reliance Endoscope Processor is an electromechanical washer/high level . disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor self-decontamination cycles.
- . Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use oxidative chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor.
- Optional washing is provided through the automated delivery of Klenzyme . Enzymatic Presoak and Cleaner during the wash phase of the cycle.
- CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, . is used in one of the two self-decontamination cycles provided by the processor.
- Various accessories are available to accommodate the processing needs of . specific endoscopes and endoscopic accessories.
- Verify Reliance CI Process Indicator is available to monitor for the presence . of the Reliance DG active ingredient, peracetic acid.
{2}------------------------------------------------
## The Reliance Endoscope Processing Cycle has the following features:
- > The first part of this cycle is an optional programmable washing phase. This phase consists of a wash that uses Klenzyme, followed by a rinse. The washing phase can be programmed on or off. In the "on" mode, the user can choose either one or two washing phases per processing cycle, and the wash time can be adjusted to be between 5 and 10 minutes. The Reliance cleaning phase does not replace manual cleaning by the user.
K102244
- => The second part is a high level disinfection phase that is non-optional and the parameters cannot be changed by the user. In this phase, the proprietary Reliance DG components, provided in a single use container, are dissolved with water at ~50°C for four minutes of generation time and circulated throughout the processor and through device lumens for 6 minutes of high level disinfection solution exposure time.
- > Following the high level disinfection phase, the Reliance Endoscope Processor removes the high level disinfection solution through a rinse phase which is non-optional and the parameters cannot be changed by the user. The processor filters the rinse water (as well as all of the water used throughout the cycle) through a 0.2 micron bacterial-retentive filter. It also incorporates an automatic internal integrity check of this filter at the end of each processing cycle. If the integrity check fails, an alarm alerts the user, and the processor does not complete the cycle.
- > The last step in the processing cycle is an air purge phase using HEPAfiltered air. The air purge helps to remove excess rinse water from the processed devices. The final air purge is preset to run for 4 minutes; additional air purge time may be selected by the operator.
- > The processor will print a detailed cycle summary at the end of each cycle that includes information such as processor number, cycle date, start and stop times, as well as phase parameters. With an optional bar code reader, the printouts can also include identification numbers for the operator, patient, device, doctor and procedure.
The processor features two decontamination cycles that are to be used without endoscopes in the processor:
- > The first, called D-SHORT, consists of hot water circulating through the processor for 10 minutes, followed by a 10-minute hot air purge. This cycle is to be run every 54 hours. D-SHORT is intended to prevent biofilm from forming.
- > The second, called D-LONG, consists of a cycle in which CIP 200 Acid-Based Process and Research Cleaner is added to hot water. The cleaning solution is then circulated through the processor for 20 minutes; this is followed by three rinses to remove the solution from the processor and a 10minute hot air purge. D-LONG is to be used on those occasions when the D-SHORT cycle has not been run within the past 54 hours.
{3}------------------------------------------------
K102244
p4 of 7
#### 4. Intended Use
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6minute exposure sequence).
#### 5. Description of Safety and Substantial Equivalence
The Reliance Endoscope Processing System has the same technological characteristics as the predicate device.
A summary of performance testing applicable to both the proposed Reliance Endoscope Processing System and its predicate Reliance Endoscope Processing System is provided in the Table attached.
{4}------------------------------------------------
The Reliance Endoscope Processing System was developed and validated in accord with two primary FDA Guidance documents: Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants/High Level Disinfectants (2000), and · Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washers (1993). A summary of this testing follows: · Reliance DG Dry Germicide Efficacy: Reliance DG Dry Germicide was tested and shown to generate an effective high level disinfection solution using the standard array of microbiological tests for germicidal efficacy. The following testing was performed at conditions of use that were worst case with respect to contact time, circulation, water hardness, temperature, artificial soiling, and germicide concentration (at or below the minimum effective dose of 9000 mg/L PAA minutes): Sporicidal: Reliance DG Dry Germicide was proven to be sporicidal as defined by AOAC Sporicidal Activity Test with an in situ exposure time of 6 minutes. Confirmatory testing was completed successfully and supplemental confirmatory testing was completed in the Reliance Endoscope Processor. Potency was subsequently confirmed in the processor using Reliance DG Dry Germicide containers that were aged beyond the end of its shelf life. > Tuberculocidal: Reliance DG Dry Germicide was proven to be tuberculocidal as defined by the AOAC Tuberculocidal Activity Test on-clir with an exposure time of 6 minutes. Potency was subsequently confirmed using Reliance DG Dry Germicide aged beyond the end of its shelf life. > Virucidal: The Reliance Process was proven to reduce the viable population of poliovirus Type 1, adenovirus Type 5, and herpes simplex virus Type 1 by > 4 log10-> Bactericidal: Reliance DG Dry Germicide was proven to be bactericidal as defined by the AOAC Bactericidal Activity Test with an exposure time of 6 minutes at worst case conditions, whether performed in situ or in vitro. > Fungicidal: Reliance DG Dry Germicide was proven to be fungicidal as defined by the AQAC Fungicidal Activity Test with an exposure time of 6 minutes, whether performed in situ or in vitro. Simulated-Use: Reliance DG Dry Germicide, at the minimum recommended dose, reproducibly achieved greater than a 6 log10 reduction of Mycobacterium terrae in triplicate trials within the Reliance Endoscope Processor for selected clinically relevant flexible endoscopes and their accessories. The test articles represented the range of most challenging devices, accessories, and processing situations. In-Use: Reliance Endoscope Processing System was evaluated in an in use study in a US hospital. Three flexible endoscopes representing the range of types indicated in the product labeling were used in clinical procedures and processed according to instructions for use. In triplicate evaluations of each endoscope, no organisms were recovered after processing. Bioburden levels on the clinically used endoscopes after manual cleaning and before high level disinfection were determined to be as high as 105 CFU/device.
Appendix A: Page 6 of 8
1102244
p5077
{5}------------------------------------------------
Non-clinica Germicide Biocompatibili Material Compatibility
### Biocompatibility:
The Reliance chemical formulations, as supplied in packaging as well as in use dilutions, can be safely handled and used by customers. Residues that may remain on medical endoscopes and accessories are below established residue limits and do not pose a risk to patients. Safety statements in product labeling are appropriate to the potential risk.
56077
- Reliance DG Dry Germicide, its components, reaction products, and residuals remaining on medical devices were evaluated for biocompatibility and possible risks to users. Testing included acute oral and ocular toxicity tests, dermal irritation studies, in vitro bacterial mutation genotoxicity studies, sensitization tests, and in vitro cytotoxicity evaluations; literature reviews of raw material toxicity data were also performed. Certain components in the single-use container, which under normal use conditions never contact the user, have the potential for irritation or skin sensitization; therefore appropriate warnings and instructions are displayed on labeling for the unusual event of a spill or container breakage.
- Use dilution reaches non-cytotoxic levels with minimal dilution. îî
- > Biocompatibility testing of extracts from processed medical devices demonstrated that no toxic residuals remain on devices under worst case circumstances. The test data indicate that the worst case residue levels for the components of maximum potential risk are far below the allowable limits. Furthermore, the processor final rinse water was found to be non-cytotoxic.
### · Reliance Endoscope Processing System Material Compatibility:
The Reliance System was evaluated for its effect on intact medical devices. including flexible endoscopes and/or common materials of device construction. After 300 processing cycles, no deleterious effects were observed other than minor cosmetic changes. No functional changes in flexible endoscopes were observed.
### · Reliance DG Dry Germicide Stability:
Reliance DG was tested and found stable for 18 months in the unopened moisture-resistant packaging at the stated conditions for storage. Once opened, the containers within each pouch are to be used within 2 weeks, or by the expiration date on the container, whichever comes first.
{6}------------------------------------------------
| | <b>Reliance Endoscope Processor Performance</b> | |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The critical process parameters for the processor (water temperature and volume, fresh Reliance DG Dry Germicide container detection, boot pressure, delivery of washing solutions and high level disinfection solution and water filter integrity testing) were each evaluated in replicate under worst case conditions and found to be within required specifications. | |
| | Each processor phase or cycle was separately evaluated and documented to be effective under worst case conditions: | |
| | ⇒ Validation of the <b>washing phase</b> of the Endoscope Processing Cycle documented that after the shortest possible washing phase, devices pre-soiled with a combination of eggs, blood, mucin and serum in saline: 1) were visually clean, and 2) achieved greatly reduced yield of extractable protein per cm2 device area (assayed to be reduced from >173 to < 5 µg/cm2). | |
| Non-clinical Tests:<br>Processor Performance | ⇒ Validation of the <b>high level disinfection phase</b> of the Endoscope Processing Cycle was performed through a simulated-use study in challenging clinically relevant endoscopes, as well as in an in-use study. | |
| | ⇒ The <b>rinse phase</b> of the endoscope processing cycle was shown to be effective. Evaluations of extracts of devices exposed to worst case conditions in the Endoscope Processing Cycle documented that levels of residuals remaining on devices were far below allowable limits and were not cytotoxic. | |
| | ⇒ The <b>air purge phase</b> was validated to confirm the ability to remove rinse water from processed medical devices. | |
| | ⇒ The <b>filter integrity test</b> system of the processor was documented to reliably detect filter failure. | |
| | ⇒ The <b>two self-decontamination cycles</b> were shown to be effective as follows: | |
| | D-LONG cycle - can disinfect the Reliance Endoscope Processor after a high level challenge with <i>Pseudomonas aeruginosa</i> followed by a 5 day inactive period; D-SHORT cycle - can kill bacteria that have potential to form biofilm. | |
| Reliance Endoscope Processor is certified to the following electrical standards | ANSI/UL -61010-1, 2nd Ed., | Electrical equipment for measurement, control and laboratory use Part 1, 2nd Edition |
| | CAN/CSA C22.2 61010-1, 2nd Ed. | Electrical equipment for measurement, control and laboratory use Part 1, 2nd Edition |
| | IEC 61010-1:2001 | Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements, |
| | IEC 61010-2-40, 1st Ed. | Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat medical materials |
| | IEC 61326-1:2005, 1st Ed. | Electrical equipment for measurement, control and laboratory use, EMC requirements Part 1: General requirements |
K102244
.
{7}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Robert F. Sullivan Senior Director, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060
Re: K102244
Trade/Device Name: Reliance® Endoscope Processing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II DEC - 3 2010 Product Code: NZA Dated: October 26, 2010 Received: October 27, 2010
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your secures of the device is substantially equivalent (for the relerched doove and have a trennilosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interstate conmisered prof to may 20, 10, 10, 10, 10, 10, 20, 20, 2011, 10, 2019, 10, 2019, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 1 Amendinents, or to de riocs that navis Act (Act) that do not require approval of a premarket the Federal Food, Drug, and Cosment . At (Prec) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . approval uppireation (1 Milly). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, requirencens for annual rogistration, including and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (FINA), It may be subject to ademicined Regulations, Title 21, Parts 800 to 898. In device can be found in the Code of Pourial Regarmers, concerning your device in the Federal Register.
{8}------------------------------------------------
### Page 2- Mr. Sullivan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements incall that it DAT has made a determand regulations administered by other Federal agencies. Of the Act of ally I ederal banks in requirements, including, but not limited to: registration 1 ou must comps) with 807); labeling (21 CFR Part 801); medical device reporting allu listing (21 OF R F ar vor ); lated adverse events) (21 CFR 803); good manufacturing (reporting of modical device to the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerery yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Reliance® Endoscope Processing System
Indications for Use:
The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6-minute exposure sequence).
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elzath & Clamine-Wels
vision Sign-Off)
ivision of Anesthesiology, General Hospital
Anesthes
fection Control, Dental Devices
510(k) Number: K102244
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.