The Alco-Breath Tube is a device to test for alcohol in human breath. It is a disposable device designed for one-time use and is a screening test that gives preliminary results. This device provides a semi-quantitative estimate of alcohol levels in breath. The ABT-08 device gives readings between 0 and 0.08% BAC and the ABT-15 device gives readings between 0 and 0.15% BAC. The CheckPoint breath alcohol test is a device to test for alcohol in human breath. It is a disposable device designed for one-time use and is a screening test that gives preliminary results. The test is available at cut-offs of 0.02, 0.04, 0.05, and 0.08% BAC.
Device Story
Disposable, one-time use in vitro diagnostic devices for breath alcohol screening. Input: exhaled human breath collected via balloon or optional volumetric bag. Principle: chromogenic reaction where reagent-treated silica gel changes color upon exposure to breath alcohol at or above specific threshold levels. Output: visual color change indicating presence of alcohol; provides semi-quantitative estimate of alcohol concentration. Used in medical and therapeutic programs; operated by user. Results are preliminary and not for evidential use. Benefits: provides rapid, non-invasive screening for alcohol presence.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated across 600 trials per device type using Breath Alcohol Sample Simulator (BASS) under five lighting conditions. Analytical specificity confirmed for cigarette smoke and temperature variations (10°C to 40°C). Consumer study (n=60) compared device results to evidentiary breath alcohol tester; results showed correlation and confirmed user ability to interpret visual color changes.
Technological Characteristics
Glass tube/ampule containing reagent-treated silica gel; flexible plastic tube with porous filters; optional volumetric collection bag. Sensing principle: solid-phase chromogenic chemical reaction. Visual readout. No electronic components or software. Storage: 50–85°F. Stability: 18 months (Alco-Breath Tube) to 3 years (CheckPoint).
Indications for Use
Indicated for screening of alcohol in human breath for over-the-counter use. Provides semi-quantitative or qualitative estimates of blood alcohol concentration (BAC). Not for legal determination of alcohol presence or concentration; not for use in determining safety to drive.
Regulatory Classification
Identification
A breath-alcohol test system is a device intened to measure alcohol in the human breath. Measurements obtained by this device are used in the diagnosis of alcohol intoxication.
Predicate Devices
BreathScan Alcohol Detector (k060761)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k102225
B. Purpose for Submission:
New Devices
C. Measurand:
Breath Alcohol
D. Type of Test:
Semi-quantitative and qualitative visually read color change
E. Applicant:
AlcoPro, Inc
F. Proprietary and Established Names:
Alco-Breath Tube and CheckPoint® Breath Alcohol Test
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DJZ | Class I, reserved | 21 CFR 862.3050 | Toxicology (91) |
H. Intended Use:
1. Intended use(s):
Refer to indication for use below.
2. Indication(s) for use:
Alco-Breath Tube breath alcohol test
The Alco-Breath Tube is a device to test for alcohol in human breath. It is a disposable
{1}
device designed for one-time use and is a screening test that gives preliminary results. This device provides a semi-quantitative estimate of alcohol levels in breath. The ABT-08 device gives readings between 0 and 0.08% BAC and the ABT-15 device gives readings between 0 and 0.15% BAC.
CheckPoint breath alcohol test
The CheckPoint breath alcohol test is a device to test for alcohol in human breath. It is a disposable device designed for one-time use and is a screening test that gives preliminary results. The test is available at cut-offs of 0.02, 0.04, 0.05, and 0.08% BAC.
3. Special conditions for use statement(s):
Over-the-counter use
The sponsor includes the following in the Warnings, Precautions, and Limitations section of the labeling:
"Do not use this test to determine if it is safe for you to drive. The actual result may be significantly higher or lower than indicated by this product. The Alco-Breath Tube is not intended to legally determine the presence of alcohol or concentration of alcohol in a person. This product should be used only as a screening device and is only an indication of the possible presence of alcohol in the blood of the user. The exact level of alcohol in the blood cannot be accurately determined by using this product."
4. Special instrument requirements:
Not applicable.
I. Device Description:
Alco-Breath Tube breath alcohol test
The Alco-Breath Tube is a visually read semi-quantitative test for the detection of alcohol in exhaled breath. It indicates relative Blood Alcohol Concentration (BAC) between 0 and 0.08 (ABT -08) or 0.15 (ABT-15). The device consists of two parts, a glass tube containing treated silica gel and a balloon assembly.
CheckPoint breath alcohol test
The Breath Alcohol Test is a visually read semi-quantitative test for the detection of alcohol in exhaled breath. It indicates relative Blood Alcohol Concentration (BAC) at four different cut-off levels (0.02, 0.04, 0.05, and 0.08% BAC). The device is designed as
2
{2}
a glass ampule containing reagent treated silica gel and held inside a flexible plastic tube with porous filters. An optional volumetric bag is used to collect human breath for testing.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
BreathScan Alcohol Detector
2. Predicate 510(k) number(s):
k060761
3. Comparison with predicate:
| Feature | Predicate Device: BreathScan® Alcohol Detector (K060761) | New Device: Alco-Breath Tube | New Device: CheckPoint® Breath Alcohol Test |
| --- | --- | --- | --- |
| Intended Use | Detect presence of alcohol in exhaled breath | Same as predicate | Same as predicate |
| Target Population | Over the counter | Same as predicate | Same as predicate |
| Calibration/Accuracy Checks | None required | Same as predicate | Same as predicate |
| Methodology | Chromogenic reaction | Same as predicate | Same as predicate |
| Anatomical Site | Mouth | Same as predicate | Same as predicate |
| Test Sample | Exhaled human breath | Same as predicate | Same as predicate |
| Blowing Time | 12 seconds | 1 minute | Same as predicate |
| Test Time | 2 minutes | Same as predicate | Same as predicate |
| Result | Qualitative | Same as predicate; also provides semi-quantitative results | Same as predicate |
| Interpretive Method | Visual Color Change | Same as predicate | Same as predicate |
| Measuring Units | BAC % | Same as predicate | Same as predicate |
| Measurement Range | Separate devices available at different cut-off levels: 0.02%, 0.04%, 0.05%, and 0.08% | Separate devices available at different cut-off levels: 0.04% and 0.08% | Same as predicate |
| Mouthpiece | None Required | Same as predicate | Same as predicate |
| Collection Device | None Required | Balloon | Optional volumetric bag |
# K. Standard/Guidance Document Referenced (if applicable):
{3}
The sponsor states conformance to the following standards:
- Department of Transportation National Highway Traffic Safety Administration [NHTSA 73 FR 16956] Highway Safety Programs; Model Specifications for Screening Devices to Measure Alcohol in Bodily Fluids
L. Test Principle:
The Breath Alcohol Detector consists of crystals that employ a solid-phase chemistry system based on a chromogenic chemical reaction. Alcohol, if present in the exhaled breath, reacts with the coated crystals to produce a color change. This color change is proportional to the concentration of alcohol in the breath, which is an approximation of relative Blood Alcohol Concentration (BAC). The light yellow crystals produce a color change when alcohol vapors are oxidized to acid and the indicator chemical changes to chromic ion. Results are interpreted as positive when the majority of the crystals change from yellow to green. This indicates that alcohol vapors are present at a concentration approximately equal to the value indicated on the tube for Alco breath tube or greater than or equal to the cutoff for the Breath alcohol test.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Alco-Breath Tube:
The ABT-08 can measure between 0.00 - 0.08% BAC and the ABT-15 measures BAC between 0.00 - 0.15%. The ABT - 08 tester was evaluated at 0.00, 0.016, and 0.064% BAC. For this study, any result less than 0.04% BAC was interpreted as negative and any result greater than 0.04% BAC was interpreted as positive.
The ABT - 15 was evaluated at 0, 0.032, and 0.128% BAC. For this study, any result less than 0.08% BAC was interpreted as negative and any result greater than 0.08% BAC was interpreted as positive.
Blood Alcohol Concentrations were simulated by a Breath Alcohol Sample Simulator (BASS), which provides an alcohol-in-air test sample with known alcohol concentrations, flow rate, and air composition. Five different lighting conditions were tested: Fluorescent, daylight, incandescent, sodium vapor, and mercury vapor.
Testing was performed by 3 operators, 3 days, and up to 3 lots for a total of 600 results.. For the ABT-08 device, testing consisted of: 200 trials at a Blood Alcohol Concentration
4
{4}
(BAC) of 0.064, 200 trials at a BAC of 0.016, and 20 trials at a BAC of 0.000. For the ABT-15 device, testing consisted of 20 trials at a BAC of 0.128, 20 trials at a BAC of 0.032, and 20 trials at a BAC of 0.000.
Results were as follows:
| Lighting | Concentration | ABT-08 (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | 0.016% | 200/200 |
| | 0.064% | 200/200 |
| Incandescent | 0% | 200/200 |
| | 0.016% | 200/200 |
| | 0.064% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | 0.016% | 200/200 |
| | 0.064% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | 0.016% | 200/200 |
| | 0.064% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | 0.016% | 200/200 |
| | 0.064% | 200/200 |
| Lighting | Concentration | ABT-15 (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | 0.032% | 200/200 |
| | 0.128% | 200/200 |
| Incandescent | 0% | 200/200 |
| | 0.032% | 200/200 |
| | 0.128% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | 0.032% | 200/200 |
| | 0.128% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | 0.032% | 200/200 |
| | 0.128% | 200/200 |
{5}
6
| Sodium Vapor | 0% | 200/200 |
| --- | --- | --- |
| | 0.032% | 200/200 |
| | 0.128% | 200/200 |
CheckPoint® Breath Alcohol Test:
For each of the CheckPoint® Breath Alcohol Test device type (with cut-off levels of 0.02, 0.04, 0.05, and 0.08) testing was performed by 3 operators, over 3 days, and up to 3 lots for a total of 600 results. Blood Alcohol Concentrations are simulated by a Breath Alcohol Sample Simulator (BASS), which provides an alcohol-in-air test sample with known alcohol concentrations, flow rate, and air composition. The simulator is used to produce a simulated breath sample at - 60% cut-off and + 60% cut-off. Five different lighting conditions were tested: Fluorescent, daylight, incandescent, sodium vapor, and mercury vapor.
For devices with 0.02 cut-off level, tests consist of: 200 trials at a Blood Alcohol Concentration (BAC) of 0.032, 20 trials at a BAC of 0.008, and 200 trials at a BAC of 0.000. For devices with 0.04 cut-off level, tests consisted of 200 trials at a BAC of 0.064, 200 trials at a BAC of 0.016, and 200 trials at a BAC of 0.000. For devices with 0.05 cut-off level, tests consist of 200 trials at a BAC of 0.080, 200 trials at a BAC of 0.020, and 200 trials at a BAC of 0.000. For devices with 0.08 cut-off level, tests consisted of 200 trials at a BAC of 0.128, 200 trials at a BAC of 0.032, and 200 trials at a BAC of 0.000. Results were as follows:
| Lighting | Concentration | CP-02 w/bag
(expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
{6}
| Lighting | Concentration | CP-02 w/o bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Lighting | Concentration | CP-04 w/bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
{7}
| Lighting | Concentration | CP-04 w/o bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Lighting | Concentration | CP-05 w/bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
{8}
| Lighting | Concentration | CP-05 w/o bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Lighting | Concentration | CP-08 w/bag (expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
{9}
| Lighting | Concentration | CP-08 w/o bag
(expected result/observed result) |
| --- | --- | --- |
| Daylight | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Incandescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Fluorescent | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Mercury Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
| Sodium Vapor | 0% | 200/200 |
| | c/o - 60% | 200/200 |
| | c/o + 60% | 200/200 |
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
Alcohol standard solutions at various BAC concentrations were prepared from commercially available solutions and are verified on a certified evidential breath alcohol tester. These solutions were used to verify the cut-off concentration of the devices during manufacture.
Stability
The sponsor provided data to confirm that the CheckPoint breath alcohol test was stable for a period of up to three years from date of manufacture and the Alco-Breath Tube was stable for a period of 18 months from date of manufacture when stored at 50 – 85° F.
d. Detection limit:
{10}
The NHTSA guidelines do not specifically address the detection limit of breath alcohol devices but require testing at a % BAC of zero (blank reading) to assess the possibility of false positives. Precision studies included zero concentration samples in which nonalcoholic human breath was used as the sample. There were no false positives observed when measuring zero concentration samples. See above for precision testing.
e. Analytical specificity:
1. Cigarette smoke
Tests were conducted for each device to evaluate the potential interference on the test results from cigarette smoke. For the Alco-Breath Tube Test, both the ABT - 08 and ABT-15 testers were evaluated. For the CheckPoint® Breath Alcohol Test, tests were conducted for each device cut-off level, 0.02%, 0.04%, 0.05% and 0.08%.
A trained operator conducted the study with volunteers to produce a total of five trials. An alcohol-free volunteer was asked to smoke approximately one half of a cigarette and perform an alcohol test. The volunteer smoked another inhalation and repeated the test to produce a total of five trials. Results of this study show that the devices at each cut-off level produces no positive results from cigarette smoke.
| Alco-Breath Tube | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Device | Lot# | Test #1 | Test #2 | Test #3 | Test #4 | Test #5 |
| ABT .08 | 021910 | Negative | Negative | Negative | Negative | Negative |
| ABT .15 | 021910 | Negative | Negative | Negative | Negative | Negative |
| CheckPoint Breath Alcohol Test | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Device | Cut Off | Lot# | Test #1 | Test #2 | Test #3 | Test #4 | Test #5 |
| CP .02 | 0.02% | 082808 | Negative | Negative | Negative | Negative | Negative |
| CP .04 | 0.04% | 012209 | Negative | Negative | Negative | Negative | Negative |
| CP .05 | 0.05% | 072309 | Negative | Negative | Negative | Negative | Negative |
| CP .08 | 0.08% | PREEC6 | Negative | Negative | Negative | Negative | Negative |
2. Temperature:
Tests were conducted for each device to evaluate the potential interference on the test results from varying temperatures. For the Alco-Breath Tube Test both the ABT - 08 and ABT-15 testers were evaluated. For the CheckPoint® Breath Alcohol Test, tests were conducted for each device cut-off level, 0.02%, 0.04%, 0.05% and 0.08%.
Results were as follows:
{11}
| Alco-Breath Tube | | | | |
| --- | --- | --- | --- | --- |
| Device | 10° | | 40° | |
| | 0.016% (expected result/observed result) | 0.064% (expected result/observed result) | 0.016% (expected result/observed result) | 0.064% (expected result/observed result) |
| ABT .08 | 20/20 | 20/20 | 20/20 | 20/20 |
| Alco-Breath Tube | | | | |
| Device | 10° | | 40° | |
| | 0.032% (expected result/observed result) | 0.128% (expected result/observed result) | 0.032% (expected result/observed result) | 0.128% (expected result/observed result) |
| ABT .15 | 20/20 | 19/20 | 20/20 | 20/20 |
| --- | --- | --- | --- | --- |
| CheckPoint Breath Alcohol Test with volumetric bag | | | | | |
| --- | --- | --- | --- | --- | --- |
| Device | Cut Off Level | 10° | | 40° | |
| | | -60% Cut-Off (expected result/observed result) | +60% Cut-Off (expected result/observed result) | -60% Cut-Off (expected result/observed result) | +60% Cut-Off (expected result/observed result) |
| CP .02 | 0.02% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .04 | 0.04% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .05 | 0.05% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .08 | 0.08% | 20/20 | 20/20 | 20/20 | 20/20 |
| CheckPoint Breath Alcohol Test without volumetric bag | | | | | |
| --- | --- | --- | --- | --- | --- |
| Device | Cut Off Level | 10° | | 40° | |
| | | -60% Cut-Off (expected result/observed result) | +60% Cut-Off (expected result/observed result) | -60% Cut-Off (expected result/observed result) | +60% Cut-Off (expected result/observed result) |
| CP .02 | 0.02% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .04 | 0.04% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .05 | 0.05% | 20/20 | 20/20 | 20/20 | 20/20 |
| CP .08 | 0.08% | 20/20 | 20/20 | 20/20 | 20/20 |
# 3. Vibration:
Tests were conducted for each device to evaluate the potential interference on the test results from vibration. For the Alco-Breath Tube Test, both the ABT - 08 and ABT-15 testers were evaluated. For the CheckPoint® Breath Alcohol Test, tests were conducted for each device cut-off level, $0.02\%$ , $0.04\%$ , $0.05\%$ and $0.08\%$ . Results were as follows:
{12}
| Device | -60% Cut-Off (expected result/observed result) | +60% Cut-Off (expected result/observed result) |
| --- | --- | --- |
| ABT .08 | 20/20 | 20/20 |
| ABT .15 | 20/20 | 20/20 |
| Device | Cut Off Level | -60% Cut-Off (expected result/observed result) |
| --- | --- | --- |
| CP .02 | 0.02% | 20/20 |
| CP .04 | 0.04% | 20/20 |
| CP .05 | 0.05% | 20/20 |
| CP .08 | 0.08% | 20/20 |
| Device | Cut Off Level | -60% Cut-Off (expected result/observed result) |
| --- | --- | --- |
| CP .02 | 0.02% | 20/20 |
| CP .04 | 0.04% | 20/20 |
| CP .05 | 0.05% | 20/20 |
| CP .08 | 0.08% | 20/20 |
f. Assay cut-off:
The CheckPoint® devices produce a blue/green or green/blue color at 0.02%, 0.04%, 0.05%, and 0.08% cut-off levels.
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor performed a consumer study to determine if consumers could correctly use and interpret the devices using only the supplied User's Manual, and to compare the results to a professional quantitative device. For this study, the test was conducted by trained operator(s) and volunteer subject(s). The Alco Breath Tube Test and the CheckPoint® Breath Alcohol Test were compared against an evidentiary alcohol test device.
Thirty (30) drinking subjects were paired with thirty (30) non-drinking subjects. The nondrinking subjects administered breath tests throughout the evening to the drinking
{13}
subjects. The tests were performed using the manufacturer's test protocol supplied with the device.
# Alco-Breath Tube Comparison study
Sixty-two results as measured by the evidential device fell within the $0 - 0.08\%$ BAC measuring range of the ABT-08 tester. Simple linear regression produces a slope of 0.91, an intercept of 0 and a correlation coefficient of 0.86.
Seven results read greater than $0.08\%$ BAC with the evidential device and read $0.08\%$ BAC (the highest possible reading) on the ABT device.
Four results read greater than $0.08\%$ BAC with the evidential device but less than $0.08\%$ BAC on the ABT device, as follows:
| ABT | Evidential |
| --- | --- |
| 0.04 | 0.110 |
| 0.07 | 0.131 |
| 0.07 | 0.138 |
| 0.04 | 0.167 |
The sponsor includes the following in the Warnings, Precautions, and Limitations section of the labeling:
"Do not use this test to determine if it is safe for you to drive. The actual result may be significantly higher or lower than indicated by this product. The Alco-Breath Tube is not intended to legally determine the presence of alcohol or concentration of alcohol in a person. This product should be used only as a screening device and is only an indication of the possible presence of alcohol in the blood of the user. The exact level of alcohol in the blood cannot be accurately determined by using this product."
CheckPoint with Blow Bag Comparison Study
| CheckPoint with Blow Bag 0.04% Tester | 0.00 to 0.010% | Low Negative >0.010 to 0.020% | Near Cut-off Negative (>0.020 to <0.040%) | Near Cut-off Positive (0.040 to 0.060%) | High Positive (>0.060) |
| --- | --- | --- | --- | --- | --- |
| Positive | 0 | 1 | 9 | 13 | 23 |
| Negative | 23 | 6 | 4 | 0 | 0 |
CheckPoint without Blow Bag Comparison Study
| CheckPoint without Blow Bag 0.04% Tester | 0.00 to 0.010% | Low Negative >0.010 to 0.020% | Near Cut-off Negative (>0.020 to <0.040%) | Near Cut-off Positive (0.040 to 0.060%) | High Positive (>0.060) |
| --- | --- | --- | --- | --- | --- |
| Positive | 0 | 0 | 10 | 13 | 21 |
| Negative | 21 | 7 | 4 | 0 | 0 |
{14}
15
Instructions for use survey results:
| Question: | Yes | No | Strongly Disagree | Agree | Don’t Know | Disagree | Strongly Disagree |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1. Did you have any problems performing any of the tests? | 3 | 27 | N/A | N/A | N/A | N/A | N/A |
| 2. The CheckPoint instructions are easy to understand. | N/A | N/A | 19 | 10 | 0 | 1 | 0 |
| 3. The CheckPoint results are easy to interpret. | N/A | N/A | 17 | 12 | 0 | 1 | 0 |
| 4. The Alco-Breath Tube instructions are easy to understand. | N/A | N/A | 18 | 10 | 0 | 2 | 0 |
| 5. The Alco-Breath Tube results are easy to interpret. | N/A | N/A | 16 | 14 | 0 | 0 | 0 |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{15}
5. Expected values/Reference range:
Using these types of devices, alcohol is not detectable in the breath of persons who have not ingested alcohol.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.