NanoGen is indicated for use in the following ways; by itself in bone regenerative techniques; mixed with other suitable bone filling agents to prevent particle migration in an osseous defect; and, to provide a resorbable barrier over other bone graft materials.
Device Story
NanoGen is a synthetic bone filling augmentation material composed of nanocrystalline calcium sulfate particles (400-850 microns) and calcium sulfate hemihydrate powder. Clinicians mix the powder with normal saline to create a putty-like paste for application into bone defects. The device sets in vivo and undergoes controlled, slower degradation (10-12 weeks) compared to traditional calcium sulfate (4-6 weeks). It functions as a scaffold or barrier to support bone regeneration and prevent migration of other graft materials. The calcium content contributes to mineralization of new bone. Used by surgeons in bone regenerative procedures; the device is applied directly to the defect site.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including chemical composition analysis (USP NSF 18), in vitro dissolution studies in simulated body fluid, and literature review documenting the use of calcium sulfate particles as bone augmentation material. Biocompatibility is supported by literature and adherence to recognized standards.
Technological Characteristics
Material: Nanocrystalline calcium sulfate and calcium sulfate hemihydrate. Dimensions: 400-850 micron particles. Energy: None (mechanical/chemical). Connectivity: None. Sterilization: Not specified (references ISO 11137-1). Standards: ASTM F2224-03. Form: Powder/particle mixture reconstituted with saline into a putty.
Indications for Use
Indicated for patients requiring bone regenerative procedures, including use as a standalone bone filler, as a carrier/stabilizer for other bone graft materials in osseous defects, or as a resorbable barrier over bone grafts.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K102208
# SECTION 1
## 510 (K) SUMMARY
Submitted by:
| Company Name: | Orthogen LLC |
|---------------|-----------------------------------------------------|
| Address: | 505 Morris Avenue, Ste 104<br>Springfield, NJ 07081 |
| Telephone: | 973-467-2404 |
| Fax: | 973-467-1218 |
CONTACT PERSON:
Harold Alexander, Ph.D. 505 Morris Avenue, Suite 104 Springfield, NJ, 07081 Ph: 973-467-2404 Fax: 973-467-1218 Email : halexander(@orthogencorp.com
DATE PREPARED:
08/03/2010
TRADE NAME:
NanoGen (Nanocrystalline calcium sulfate particles)
COMMON NAME: Bone filling augmentation material
#### SUBSTANTIALLY EQUIVALENT TO:
SurgiPlaster K 011403, BoneGen-TR K 060285
DESCRIPTION of the DEVICE: NanoGen is a mixture of particles composed of densely packed grains of nanocrystalline calcium sulfate and medical grade calcium sulfate hemihydrate powder. The particles range from 400 - 850 microns in size. The particles of nanocrystalline calcium sulfate are purely synthetic and are manufactured from medical grade calcium sulfate hemihydrate through a proprietary process. Because of its unique structure, NanoGen undergoes controlled, slower degradation as compared to traditional calcium sulfate.
NanoGen is designed to set-up in vivo and may be used either alone or mixed with DFDBA (demineralized freeze-dried bone allograft) or autogenous bone in bone in bone regenerative procedures. It is mixed with small amounts of normal saline to produce a putty-like paste and is then applied to the bone defects. Set NanoGen dissolves with time. The calcium present in NanoGen is believed to contribute to the mineralization of newly regenerated bone. In vitro dissolution studies conducted by incubating NanoGen particles in simulated body fluid confirmed that NanoGen undergoes degradation in 12 weeks as opposed to the 4 - 6 week degradation time found for medical grade calcium sulfate powder. This finding has been confirmed by in vivo studies.
{1}------------------------------------------------
#### 510(k) Summary-Continued
### INDICATIONS FOR USE:
NanoGen is indicated for use in the following ways; by itself in bone regenerative techniques; mixed with other suitable bone filling agents to prevent particle migration in an osseous defect; and, to provide a resorbable barrier over other bone graft materials.
#### COMPARISON WITH PREDICATE DEVICES:
#### 1. Comparison of NanoGen and SurgiPlaster:
NanoGen is substantially equivalent to:
SurgiPlaster Calcium Sulfate hemihydrate Bone Graft Plaster K011403
Equivalency is determined by:
- NanoGen has the same indication for use and labeling claims. >
- √ NanoGen has similar material composition.
| Device Description | Device (NanoGen) | Predicate (SurgiPlaster) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | 1. By itself in bone<br>regeneration<br>2. Mixed with other<br>bone grafts<br>3. Providing a<br>resorbable barrier<br>over other bone<br>grafts | 1. By itself in bone<br>regeneration<br>2. Mixed with other bone<br>grafts<br>3. Providing a resorbable<br>barrier over other bone<br>grafts |
| Device Design | 0.65 grams medical grade<br>calcium sulfate particles<br>400 - 850 microns in size<br>mixed with 0.35 grams<br>medical grade calcium<br>sulfate powder; regular set<br>solution (saline) provided | Medical grade calcium sulfate<br>hemihydrate powder, supplied<br>with regular set (saline) and fast<br>set (potassium sulfate) liquids |
| Composition of<br>Materials | Medical grade calcium<br>sulfate, normal saline<br>(Regular Set) | Medical grade calcium sulfate,<br>normal saline (Regular Set) and<br>4% Potassium sulfate solution<br>(Fast Set) |
| Physical Properties | Completely resorbs in 10-<br>12 weeks, mix of powder<br>and particles | Resorbs in 4-6 weeks, provided<br>in powder form |
{2}------------------------------------------------
#### 510(k) Summary-Continued
| FDA-Recognized<br>Standards | ASTM F2224-03 for<br>properties of calcium<br>sulfate<br>510(k) Sterility Review | ASTM F04.13.15 for properties<br>of calcium sulfate<br>EN 552, 554 and 556 for<br>sterilization standard |
|-----------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| | Guidance K90-1<br>ISO 11137-1:2006 | |
### 2. Comparison of NanoGen and BoneGen - TR:
NanoGen is substantially equivalent to:
BoneGen-TR K060285
Equivalency is determined by:
- √ NanoGen has the same indication for use and labeling claims.
- ✔ NanoGen is manufactured using same technology.
- √ NanoGen is sterilized in the same manner.
| Device Description | Device (NanoGen) | Device (BoneGen – TR) |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | 1. By itself in bone regeneration<br>2. Mixed with other bone grafts<br>3. Providing a resorbable barrier over other bone grafts | 1. Oral Surgery: Post-extraction<br>2. Periodontics: Infra-osseous defects.<br>3. Endodontics: Apicoectomy, Root perforations, Open apices.<br>4. Implantology: Dehiscences, fenestrations, sinus lifts. |
| Device Design | 0.65 grams of medical grade calcium sulfate particles 400-850 microns in size mixed with 0.35 grams of medical grade calcium sulfate powder; regular set solution (saline) also provided. | Composite particles (425-850 micron) of medical grade calcium sulfate and PLLA (ratio of 96:4) supplied in 1.5 gram packs |
| Composition of Materials | Medical grade calcium sulfate, normal saline (Regular Set) | Medical grade calcium sulfate and PLLA |
{3}------------------------------------------------
#### 510(k) Summary-Continued
| Physical Properties | Completely resorbs in 10-12<br>weeks, mix of powder and<br>particles | Completely resorbs in 16 weeks,<br>particle size range is 425 to 850<br>micron |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| FDA-Recognized<br>Standards | ASTM F2224-03 for<br>properties of calcium sulfate<br>510(k) Sterility Review<br>Guidance K90-1<br>ISO 11137-1:2006 | ASTM F2224-03 for properties of<br>calcium sulfate<br>ANSI/AAMI/ISO standard<br>11737-1995 for sterilization |
## SUMMARY of TESTING:
Orthogen LLC has 1) tested the chemical composition of Calcium Sulfate Hemihydrate used to make NanoGen by an independent laboratory to a recognized test method (USP NSF 18) and reported the results in this submission; and, 2) has provided extensive literature articles documenting the use of calcium sulfate particles as a bone augmentation material. Therefore this material has been qualified as biocompatible and safe for its intended use. This device is not marketed as non-pyrogenic.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing strokes, representing service to the people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY - 6 2011
Harlod Alexander, Ph.D. Chief Executive Officer Orthogen, LLC 505 Morris Avenue, Suite 104 Springfield, New Jersey 07081-1033
Re: K102208
Trade/Device Name: NanoGen (nanocrystanlline Calcium Sulfate Particles) Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: April 26, 2011 Received: April 27, 2011
Dear Dr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Alexander
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
K102208
## Indications for Use
510(k) Number K102208
Device Name: NanoGen (nanocrystalline calcium sulfate particles)
Indications For Use:
- 1. By itself in bone regenerative techniques;
- 2. Mixed with other suitable bone filling agents to prevent particle migration in an osseous defect; and,
- 3. To provide a resorbable barrier over other bone graft materials.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Rent P.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Susta
510(k) Number: k102208
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.