K102207 · Synca Marketing, Inc. · EBI · Nov 23, 2010 · Dental
Device Facts
Record ID
K102207
Device Name
FIBER FORCE
Applicant
Synca Marketing, Inc.
Product Code
EBI · Dental
Decision Date
Nov 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
· As structural reinforcement for fabricating and/or repairing removable prostheses (dentures) and removable implant supported dentures (ISDs) · As reinforcement for temporary and/or permanent plastic/composite partial and full crowns and bridges. · As reinforcement for customized splints used to immobilize teeth which may be required for orthodontia, periodontia, traumatology and any dental activity requiring bonding of natural teeth.
Device Story
Fiber Force is a system of light-cured resin pre-impregnated E-glass fibers; used by dental professionals for structural reinforcement of dental prostheses, crowns, bridges, and splints. The device consists of mesh, fiber, and rope formats impregnated with resin via an industrialized process to ensure homogeneity with acrylic or composite interface materials. Operation involves manual placement of the fiber material into the dental restoration, followed by light-curing using a standard dental photopolymerization lamp. The output is a reinforced, hardened dental structure. The device provides mechanical strength to dental appliances, aiding in the immobilization of teeth or the repair of dentures. Clinical benefit includes improved durability and stability of dental restorations.
Clinical Evidence
Bench testing only. Biocompatibility evaluation performed per ISO 10993-5 and ISO 7405, including cytotoxicity, sensitization, irritation, and toxicity testing. In vitro cytotoxicity testing confirmed no cell lysis or toxicity (reduction of cellular density <25%). No clinical trials were conducted; safety is supported by the use of materials with established histories in dental applications (e.g., UDMA, TEGDMA, E-glass).
Technological Characteristics
Pre-impregnated E-glass fibers in light-cured resin. Materials include UDMA, TEGDMA, camphorquinone, polymerization accelerators, and pigments (titanium dioxide, iron oxide, ultramarine blue). Form factors include mesh, UD fiber, and braided rope. Polymerization via dental photopolymerization lamp. Biocompatibility tested per ISO 10993-5 and ISO 7405.
Indications for Use
Indicated for patients requiring structural reinforcement of removable dentures, implant-supported dentures, temporary or permanent crowns and bridges, and customized dental splints for orthodontia, periodontia, or traumatology.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
# 510(k) Summary
# NOV 2 3 2010
This revised summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.
K10 2207
Date
November 11, 2010
#### Repackager / Relabeler
Synca Marketing Inc. 337 Marion Street Le Gardeur, Quebec, J5Z 4W8 Canada
Tel: 800-667-9622 Fax: 800-554-1021 Contact person: Ms. Rolande Villemure
## United States Official Correspondent (U.S. Designated agent)
Mtech Group 12946 Kimberley Ln Houston TX 77079
Tel: +713-467-2607 Fax: +713-464-8880 Contact person: Mr. Dave Kim Email: davekim@mtech-inc.net
## Trade/Proprietary Name:
# Fiber Force products (repackaged product of Fast Splint):
Pink/white Mesh (118mmx53mm), Pink/white UD Fiber (2mmx150mm), Pink/white Braided Rope (1mmx150mm), Pink/white Braided Rope (2mmx150mm), Light Cure Pink/white Resin (3ml + 4tips)
{1}------------------------------------------------
## Common or Classification Name
Glass fiber reinforcement material for Denture relining, repairing, or rebasing resin (21CFR 872.3760, Product code EBI, Class2)
# Description:
Fiber Force is a system of light-cured resin pre-impregnated E-glass fibers used for reinforcement in a variety of applications in dentistry. All formats of Fiber Force use specially treated E-glass fibers that impregnated in resin using an industrialized process that ensures homogeneity with the acrylic or composite resin interface material.
# Indication for use:
· As structural reinforcement for fabricating and/or repairing removable prostheses (dentures) and removable implant supported dentures (ISDs)
· As reinforcement for temporary and/or permanent plastic/composite partial and full crowns and bridges.
· As reinforcement for customized splints used to immobilize teeth which may be required for orthodontia, periodontia, traumatology and any dental activity requiring bonding of natural teeth.
### Predicate Device:
| Manufacturer | : Stick Tech., Ltd. |
|---------------|--------------------------------------------------|
| Device | : everStick/ everStick Net |
| 510(k) Number | : K011788 (Nov. 05, 2001)/K011799 (Nov.05, 2001) |
#### Substantial Equivalence:
Fiber Force described in this 510(k) has the same intended use and similar technical characteristics as everStick / everStick Net of Stick Tech Ltd.
The indications for use, material, form factor, performance, and safety characteristics between Fiber Force and the predicate device are the same. The primary difference is size, packaging,
{2}------------------------------------------------
structure only. Accordingly we can claim the substantially equivalence of Fiber Force to the predicate device.
## Technological Characteristics
The following tests were performed according to standard ISO 10993-5 and ISO 7405 and the product is tested to be safe and has the same technological characteristics as the predicate device since both have same indications for use, similar polymers composition and same working technique:
### 1. Biological Evaluation
According to the standards applicable to biocompatibility evaluation. PRP's belong to the category of devices in contact with a surface (mucous membranes) in the case of Clinical application and to that of devices without contact in the case of Laboratory application (in this latter case, biological development is not requested). The medical device and its many applications have been described, particularly the components of the formulation, the impurities, the leachable products and degradation. The initial evaluation tests taken into consideration for the Clinical application were: cytotoxicity, sensitization, intradermal irritation or reaction, subchronic and subacute toxicity and genotoxicity. We have also decided to perform additional tests, such as carcinogenicity and toxicity on reproduction and development, in order to have an in-depth evaluation of our product. The study of these toxicological risks showed that it is improbable that they cause unacceptable health risk. The formulation, optimized to achieve effective polymerization, minimizes the leaching of soluble products into the water in the oral environment. The cytotoxicity test performed by an outside agency proved the total absence of cytotoxicity of the reticulated product.
However, the following risks were able to be identified:
- . irritation or sensitization (allergy) during contact with the non-polymerized resin with the fingers of certain predisposed practitioners.
- . mechanical irritation due to airborne glass fibers during cutting of the product.
{3}------------------------------------------------
These risks can be eliminated by taking precautions such as wearing gloves, protective goggles, and a mask, and by grinding under water or with surgical aspiration during cutting or elimination of the product. Also, the existence of these risks was mentioned in the instructions for use, as well as precautions to take to avoid them. To the best of our knowledge, we have not identified any major or unacceptable risks associated with the PRP product, by ensuring that the instructions for use are scrupulously respected.
# 2. Vitro Cytotoxicity
An in vitro biocompatibility test on the "Fiber Force" removable denture reinforcement, RES reference prp L018A was performed with the goal of determining the cytotoxic potential. This study was conducted according to the requirements of the standard NF EN ISO 10993: Biological Evaluation of Medical Devices, Part 5: Tests concerning in vitro cytotoxicity.
Under the conditions of this study, the extract of the tested product did not provoke any sign of cell lysis or toxicity (reduction of the cellular density lower than 25%). The extract of the tested product satisfied the test requirements. The positive and negative controls were compliant with the expected results.
Chemical Component Data: Chemical component data for Fiber Force including CAS number is described in the product description.
# Non-clinical Considerations: Safety and Effectiveness of the Subject Device
The fiber used is fiberglass E, the same composition as the fiberglass E used for manufacturing of Postec and Precipost posts by BCM since 2001. UDMA and TEGDMA monomers are very common in dentistry and have been used for more than 30 years particularly for filling products, adhesives, etc. The camphorquinone is the photocatalyst best adapted to the spectra of dental photopolymerization lamps and housing. It is found in the photopolymerizable dental materials for more than 15 Ethyl-4-(dimethylamino)-benzoate is a polymerization accelerator which is years.
{4}------------------------------------------------
#### 510(k) Submission- Fiber Force
found in several other dental products, particularly Dyract, an adhesive of De Trey Dentsply, which is not cytotoxic (67) and has never been the subject of a health alert (35).
The blue pigment (Ultramarine blue) is used in the coloring of cosmetic products (make-up and mascara) and is universally accepted for the coloring of objects intended for contact with food and for the manufacturing of toys.
The white pigment (titanium dioxide) is used for food coloring (E171) in cosmetics. medications and in the European and US pharmacopoeia.
The red pigment (iron oxide (red M)) is accepted for manufacturing of toys (in Europe according to EN71/3 and in the USA according to the ASTM), and for the coloring of objects intended for contact with food (Germany, Austria, and Australia). The red pigment S is iron oxide III. This latter is a food additive of which the ADI (acceptable daily intake) is 0.5mg/kg (18).
The four pigments, the silica, the camphorquinone and the TEGDMA are already in use since 1996 by BCM for the manufacturing of the product Vectris. It is also a photopolymerizable pre-impregnated fiber intended for the manufacturing of dental crowns and bridges: it is recovered and buried in various reconstitution resins. The chemical formulation, whose photocatalysts are used for the photopolymerization of PRP, is identical to that used in composite filling products, for example TETRIC EVO CERAM from the IVOCLAR Company. (79) For this product, the thickness of the product, fully hardenable using a photopolymerizing office lamp is greater than 2 mm (enamel shade transparent) and 1.5 mm for strongly colored products (dentine shade). A study by Prof. Antonio CERUTTI over six and twelve months showed excellent stability of the ingredients used for TETRIC EVO CERAM when they were polymerized for 10 seconds with a high intensity lamp (1200mw) and 20 seconds with an average lamp(700mW). (80)
The study is also based on the product VARIOLINK II (86) or ultra from the IVOCLAR Company (90): it is a photo- and auto-polymerizable composite (dual cure), radio-opaque for adhering restorations. It is found in the non-polymerized state and is thus in contact with I mineral and gingival or pulpal tissues. (86)
{5}------------------------------------------------
The formulation contains compounds similar to our PRP product (UDMA, TEGDMA).
# Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Synca Marketing Ltd. concludes that Fiber Force is safe and effective and substantially equivalent to predicate device as described herein.
END
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings and three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Synca Marketing, Incorporated C/O Mr. Dave Kim MTech Group 12946 Kimberley Lane Houston, Texas 77079
NOV 2 3 2010
Re: K102207
Trade/Device Name: Fiber Force products (repackaged product of Fast Splint) Regulation Number: 21 CFR 872.3670 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: November 15, 2010 Received: November 15, 2010
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
# Page 2- Mr. Kim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Th for
Anthony D. Watson, B.S.,M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
NOV 2 3 2010
510(K) Number (if known): K10 2207
Fiber Force products (repackaged product of Fast Splint): Device Name: Pink/white (118mmx53mm), UD Fiber Pink/white Mesh (2mmx150mm), Pink/white Braided Rope (1mmx150mm), Pink/white Braided Rope (2mmx150mm), Light Cure Pink/white Resin (3ml + 4tips)
Glass fiber reinforcement material for Denture relining, repairing, or Classification: rebasing resin
#### Indications for Use:
· As structural reinforcement for fabricating and/or repairing removable prostheses (dentures) and removable implant supported dentures (ISDs)
· As reinforcement for temporary and/or permanent plastic/composite partial and full crowns and bridges.
· As reinforcement for customized splints used to immobilize teeth which may be required for orthodontia, periodontia, traumatology and any dental activity requiring bonding of natural teeth.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
uson
(Division Sign-Off) (Division of Anesthesiology, Geegral Hospital Infection Control, Dental Devices
510(k) Number: K100267
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.