← Product Code [NKB](/productcode/NKB) · K102204

# CHOICE SPINE STARFIRE PEDICLE SCREW SYSTEM (K102204)

_Choicespine, LP · NKB · Nov 15, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102204

## Device Facts

- **Applicant:** Choicespine, LP
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Nov 15, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Choice Spine STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

## Device Story

STARFIRE™ Pedicle Screw System is a top-loading, polyaxial spinal fixation system. Components include polyaxial screws, rods, and connectors. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Operates via a set screw mechanism to secure rods to screws. Provides mechanical support to the spine to facilitate fusion and treat instabilities or deformities. Benefits patients by stabilizing the spinal column, potentially reducing pain and correcting deformities.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717-09. Static compression bending tests confirmed equivalence to non-cannulated predicate systems. Compression bending fatigue tests showed equivalent or greater endurance strength. Static torsional tests confirmed resistance exceeding requirements for the human lumbar spine.

## Technological Characteristics

Polyaxial screw, rod, and connector system. Materials: Titanium alloy (ASTM F136), commercially pure titanium (ASTM F67), and stainless steel (ASTM F138). Mechanical fixation via set screw mechanism. Non-sterile, clean delivery.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- OPTIMA™ Spinal System ([K024096](/device/K024096.md), [K031585](/device/K031585.md))
- Firebird™ Spinal Fixation System ([K081684](/device/K081684.md))
- Coral™ Spinal Fixation System ([K041592](/device/K041592.md))
- Expedium™ 6.35 Spinal Fixation System ([K062174](/device/K062174.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 8. 510(k) Summary

Preparation Date

November 11, 2010

## Sponsor

Choice Spine, LP 306 Erin Drive Knoxville, TN 37919 p: 865.246.3333 f: 865.588.4045

#### Contact

G. Todd Hawkins Director of Regulatory Affairs/Quality Assurance

### Trade Name

Choice Spine STARFIRE™ Pedicle Screw System

### Common Name

Pedicle Screw Spinal System; Pedicle Screws

#### Regulatory Classification & Device Product Codes

888.3070

MNI - Orthosis, Spinal Pedicle Fixation MNH - Orthosis, Spondylolisthesis Spinal Fixation NKB - Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease

#### Predicate Devices

OPTIMA™ Spinal System (K024096, K031585) Firebird™ Spinal Fixation System (K081684) Coral™ Spinal Fixation System (K041592) Expedium™ 6.35 Spinal Fixation System (K062174)

## Device Description

The Choice Spine STARFIRE™ Pedicle Screw System is a top-loading, polyaxial screw spinal fixation system including polyaxial screws, rods, and connectors. The components are provided clean and non-sterile. Various sizes of the implants are provided.

#### Intended Use

The Choice Spine STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities

NOV 1 5 2010

{1}------------------------------------------------

or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

#### Materials

The Choice Spine STARFIRE™ Pedicle Screw System implants are manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136, commercially pure titanium (G2) as described by ASTM F67 and stainless steel (316 LVM) as described by ASTM F138.

## Technological Characteristics

The Choice Spine STARFIRE™ Pedicle Screw System consists of a range of polyaxial screws, rods, and connectors identical to the identified predicate systems.

The Intended Use of the STARFIRE™ System is identical to the predicate systems. The STARFIRE™ System materials are similar to the predicate systems.

The STARFIRE™ Pedicle Screw System utilizes a set screw mechanism to secure the rod in the same manner as the predicate systems.

## Substantial Equivalence

Documentation was provided that demonstrates the Choice Spine STARFIRE™ Pedicle Screw System to be substantially equivalent to previously cleared device systems. The substantial equivalence is based upon equivalence in intended use, indications, anatomic location, materials, and performance.

Mechanical testing was performed according to the guidelines outlined in ASTM F1717-09 "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model."

Static compression bending tests showed that cannulated STARFIRE pedicle screw constructs were equivalent, with respect to bending loads and stiffness, to noncannulated predicate pedicle fixation systems.

Compression bending fatigue tests demonstrated that the cannulated STARFIRE pedicle screw constructs had equivalent or greater endurance strength compared to non-cannulated predicate pedicle fixation systems.

In static torsional tests the cannulated STARFIRE pedicle screw constructs, without cross-connectors, were able to resist loads greater than the range of resistance required for the human lumbar spine.

Test results indicate that the STARFIRE™ System is equivalent to the predicates in each of the test modes. Documentation of the testing is available in Section 16.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three lines extending from its head, representing growth and development. To the left of the figure, the text "DEPARTMENT OF HEALTH &" is arranged vertically, followed by "HUMAN SERVICES" in a curved formation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Choice Spine, LP % Mr. G. Todd Hawkins Director of Regulatory Affairs/Quality Assurance 306 Erin Drive Knoxville, Tennessee 37919

NOV 1 5 2010

Re: K102204

Trade/Device Name: Choice Spine STARFIRE™ Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: October 21, 2010 Received: October 22, 2010

Dear Mr. Hawkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Mr. G. Todd Hawkins

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Barbara Muchup

Mark N. Melker Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 7. Indication for Use Statement

510(k) Number (if known): KLU2204

NOV 1 5 2010

Device Name: Choice Spine STARFIRE" Pedicle Screw System

Indications for Use:

The STARFIRE™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

- degenerative disc disease (DDD; defined as back pain of discogenic origin . with degeneration of the disc confirmed by history and radiographic studies);
- spondylolisthesis; ●
- trauma (i.e., fracture or dislocation); ●
- . spinal stenosis;
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); .
- . tumor;
- pseudoarthrosis; and .
- . failed previous fusion

Prescription Use _ _ _ _ X _ Over-The-Counter Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number_________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K102204](https://fda-staging.innolitics.com/device/K102204)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
