K102172 · Global Manufacturing Technology · MEH · Mar 1, 2011 · Orthopedic
Device Facts
Record ID
K102172
Device Name
MSA HIP SYSTEM
Applicant
Global Manufacturing Technology
Product Code
MEH · Orthopedic
Decision Date
Mar 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MSA Hip System is indicated for cementless use and is intended to replace a hip joint. The device is intended for: skeletally mature individuals undergoing primary surgery for total hip replacement patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur. patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis patients suffering from disability due to previous fusion patients with acute femoral neck fractures
Device Story
Modular hip prosthesis system; consists of short stem and modular neck. Stem features plasma spray and hydroxyapatite coatings. Designed for cementless implantation; mates with previously cleared femoral heads and acetabular components. Used by orthopedic surgeons in clinical settings for total hip replacement. Provides structural replacement of hip joint to alleviate pain and restore function in patients with degenerative or traumatic hip conditions. Clinical benefit derived from mechanical stabilization of the hip joint.
Clinical Evidence
Bench testing only. Performance data included neck endurance, stem fatigue, interconnection strength of modular connections, and ROM testing. No clinical data provided.
Technological Characteristics
Modular hip system. Stem: Ti-6Al-4V (ASTM F136) with commercially pure titanium plasma spray coating (ASTM F1580). Neck: CoCr alloy (ASTM F1537). Cementless fixation. Mechanical assembly.
Indications for Use
Indicated for skeletally mature patients requiring primary total hip replacement due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, femoral head nonunion, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, previous fusion, or acute femoral neck fractures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
## 510(k) Summary
MAR - 1 2011 Manufacturer: Global Manufacturing Technology 8-10 Resolution Drive Unanderra NSW 2526 AUSTRALIA Phone: (011) 61 (2) 4272 4493 Device Trade Name: MSA Hip System Contact: Mr. Ricardo Lucchini Director Of Regulatory Affairs Prepared By: Musculoskeletal Clinical Regulatory Advisers, LLC 1331 H Street, NW. 12th Floor Washington, DC 20005 Phone: (202) 552-5800 Fax: (202) 552-5798 Date Prepared: February 10, 2011 Classification: 21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Class II MEH Product Codes: The MSA Hip System is indicated for cementless use and is Indications For Use: intended to replace a hip joint. The device is intended for: skeletally mature individuals undergoing primary � surgery for total hip replacement patients suffering from severe hip pain and disability . due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur. patients with congenital hip dysplasia, protrusio . acetabuli, or slipped capital femoral epiphysis patients suffering from disability due to previous fusion . patients with acute femoral neck fractures ●
{1}------------------------------------------------
Device Description: The MSA Hip System is a modular hip system consisting of a short stem which mates with a modular neck. The proximal portion of the stem has a plasma spray coating and a hydroxyapatite coating. The MSA Hip System can be mated with a number of previously cleared femoral heads and acetabular components. The hip system is intended for cementless implantation.
> The hip stems are made of Ti-6AI-4V (ASTM F136) with a commercially pure titanium plasma spray coating (ASTM F1580) and the modular necks are made of CoCr alloy (ASTM F1537).
> > Comparative information presented in the 510(k) supports the substantial equivalence of the MSA Hip System with respect to its indications for use, design, materials, and This 510(k) demonstrates the method of fixation. substantial equivalence of the MSA Hip System to the following predicate devices: Aesculap Implant System's Metha Short Stem Hip System (K071916 and K080584); Zimmer's Mayo Conservative Hip Prosthesis (K030733 K943230): Wright Technology's and PROFEMUR Hip System (K081090 and K091423); and OMNI life science. Inc.'s Apex ARCTM Hip System (K090845).
The MSA Hip Stem shares the same indications for use for which the predicate systems were cleared. The geometry and design of the subject hip stem matches that of the predicate hips. The materials of the stem and neck are the same and the predicate stems and necks. Like the MSA Hip, the majority of the predicate stems are intended for cementless fixation.
The non-clinical tests performed by the company include Performance Data: neck endurance, stem fatigue, interconnection strength of modular connections, and ROM testing. The test results demonstrate that the MSA Hip System is substantially equivalent to legally marketed predicate devices.
Materials:
Predicate Devices:
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Global Manufacturing Technology % Musculoskeltal Clinincal Regulatory Advisers, LLC Hollace S. Rhodes Director, Orthopedic Regulatory Affairs 1331 H St NW, 12th Floor Washington DC, 20005
MAR - 1 2011
Re: K102172
Trade/Device Name: MSA Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: MEH Dated: February 24, 2011 Received: February 25, 2011
Dear Ms. Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drinas intact a and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page - 2 - Ms. Hollace S. Rhodes
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aliz. B. R k
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use 4.
510(k) Number (if known): _KIO2172 (gg 1/1)
Device Name: MSA Hip System
The MSA Hip System is indicated for cementless use and is intended to replace a hip joint.
The device is intended for:
- . skeletally mature individuals undergoing primary surgery for total hip replacement
- patients suffering from severe hip pain and disability due to rheumatoid arthritis, . osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fracture of the femur.
- patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral . epiphysis
- patients suffering from disability due to previous fusion .
- . patients with acute femoral neck fractures
Prescription Use (Part 29 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use no (29 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melbern
(Division/Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102172
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.