← Product Code [LLZ](/productcode/LLZ) · K102148

# CURIEPACS (DICOM PACS SOFTWARE) (K102148)

_Moksha Digital Software Pvt Limited · LLZ · Aug 17, 2010 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K102148

## Device Facts

- **Applicant:** Moksha Digital Software Pvt Limited
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.

## Device Story

CuriePACS is a scalable PACS solution for DICOM imaging; consists of MD Athena (server for storage/connectivity) and MD Vision (viewer for retrieval/workflow). Inputs include DICOM-compliant images from CR, CT, DX, MG, MR, NM, RF, SC, US, XA, and ES modalities. Operates on network-connected PCs; used by medical professionals for electronic viewing, storage, and communication. Provides image processing tools: window/level, zoom, pan, magnifying glass, probe, spatial locator, shutter, and various ROI/measurement tools. Output displayed on FDA-approved medical-grade monitors. Facilitates clinical decision-making by enabling remote access and standardized review of diagnostic images across distributed networks.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence consists of software validation, risk analysis, and demonstration of DICOM 3 compliance.

## Technological Characteristics

Software-based PACS; runs on Windows XP/Vista; requires .NET Framework 3.5 SP1, Intel Core 2 Duo, 2GB RAM, 320GB HDD, and Gigabit Ethernet. Supports DICOM, JPEG, and JPEG 2000. Connectivity via network-connected server. Requires FDA-approved self-calibrating medical-grade monitor with DVI input for diagnostic interpretation.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Voyager PACS System ([K062062](/device/K062062.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(K) Summary K102148

Moksha Digital Software PVT Limited #1611, Janardhana Towers, 7th Cross, 19th Main, Sector 1, HSR Layout, Bangalore 560034, India Phone : +91 (80) 4110 1208 Fax : +91 (80) 4110 1824 E-mail : sales@mokshadigital.com Contact: Deepak Sharma, Managing Director Date prepared: August 12, 2010

AUG 1 7 2010

- 1 . Trade Name: CuriePACS Dicom PACS Software Common Name: PACS Software Classification Name: System, image processing, radiological
- 2. Regulation Description Picture archiving and communications system, product code LLZ, Regulation: 892.2050 Class of device: Class II.
- 3. The legally marketed device to which we are claiming equivalence Voyager PACS System, Voyager Imaging, K062062.
- 4. Description of device: CuriePACS is a comprehensive solution for Dicom imaging needs. This scalable PACS solution provides electronic viewing, storage and communication in a secure environment. The device consists of: 1. Dicom Server (MD Athena) - For storage and connectivity with all the modalities and 2. Image Viewer (MD Vision) - To connect to the server and retrieve images based on search criteria and other workflow requirements. Image modalities supported: CR. CT. DX. MG. MR. NM. RF. SC. US. XA (x-ray), and ES.
- న. Indications for use: CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
- Technological characteristics: The software is installed on a network connected PACS server 6. with the following minimum requirements:
	- Microsoft® Windows® XP Service Pack 2 or above .
	- Microsoft® Windows® Vista Service Pack 1 or above .
	- . Microsoft .Net Framework 3.5 SP1
	- Microsoft Internet Explorer 7 or above .
	- Intel Core 2 Duo Processor .
	- 320 GB HDD .

{1}------------------------------------------------

- 2 GB RAM (Min) •
- Dedicated or Onboard Graphics card with minimum 1440 x 900 resolution and DVI output. •
- . Gigabit Ethernet Adaptor
- . Display: The display used for diagnosis purposes should be FDA approved self calibrating medical grade monitor which is driven by a digital input (DVI output from graphics card).

| Characteristic                                           | Predicate Device<br>Voyager PACS System, Voyager<br>Imaging, K062062                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | CuriePACS (Dicom PACS Software)<br>(This submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                                   | Voyager PACS System is a software based device that receives digital images and data from<br>various sources (i.e. CT scanners, MR scanners,<br>ultrasound systems, R/F Units,<br>computed & direct radiographic devices.<br>secondary capture dev ices, scanners, imaging<br>gateways, etc.). Images and data can be<br>captured, stored, communicated, processed and<br>displayed within the system and or across<br>computer networks at distributed locations.<br>Lossy compressed mammographic images and<br>digitized film screen images must not be<br>reviewed for primary image interpretation.<br>Mammographic images may only be interpreted<br>using an FDA approved monitor that offers at<br>least 5 Mpixel resolution and meets other<br>technical specifications reviewed and accepted<br>by FDA. | CuriePACS (Dicom PACS Software) Software<br>is a software device that receives digital<br>images and data from various sources<br>(Computed Radiography, Magnetic Resonance<br>Imaging, Ultrasound, Endoscopy, Computed<br>Tomography, Digital X-Ray Mammography,<br>Nuclear Medicine Imaging, Secondary<br>Capture, Radio Fluoroscopy, X-Ray or other<br>Angiograms such as CT Angio). Images and<br>data can be captured, stored, communicated,<br>processed and displayed within the system and<br>or across computer networks at distributed<br>locations. Lossy compressed mammographic<br>images and digitized film screen images must<br>not be reviewed for primary image<br>interpretation. Mammographic images may<br>only be interpreted using an FDA cleared<br>monitor that offers at least 5 Megapixel<br>resolution and meets other technical<br>specifications reviewed and accepted by FDA. |
| Use, key feature<br>Connection                           | Medical professionals, WEB based<br>WEB and Ethernet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Multimodality<br>Connectivity,<br>DICOM<br>compatibility | Voyager PACS can receive with ease<br>any DICOM compliant image,<br>irrespective of the source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Connect all DICOM Modalities CR,<br>CT, DX, MG, MR, NM, RF, SC,<br>US, XA (x-ray), and ES.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Image Sources                                            | Dicom, JPEG and JPEG 2000 compliant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Target hardware                                          | PC compatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Image Processing<br>Tools                                | • Brightness/ Contrast<br>• Mouse button operation for quick<br>image manipulation • Selectable<br>multiple image view• Magnify, zoom<br>and pan• Flip (left/ right, top/bottom)<br>and rotate• Image inversion<br>• Measurement and angles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | All processing tools Stack,<br>Window/Level, Zoom, Pan,<br>Magnifying Glass, Probe, Spatial<br>Locator, Draw Shutter, Ruler,<br>Elliptical ROI, Rectangular ROI,<br>Polygonal ROI, Text Area, Text<br>Callout, Protractor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Security                                                 | SSL Encryption                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

#### 7. Comparison Table

Performance: The results of nonclinical tests submitted showing Dicom 3 compliance, 8. along with bench testing (software validation and risk analysis) shows that this new device has equivalent indications and performs in an equivalent fashion to the named predicate. Therefore this device poses no new issues of safety or effectiveness, and is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Moksha Digital Software PVT Limited % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25" Street NW BUFFALO MN 55313

AUG 1 7 2010

Re: K102148

Trade/Device Name: CuriePACS (Dicom PACS Software) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and Communications System Regulatory Class: II Product Code: LLZ Dated: July 29, 2010 Received: July 30, 2010

### Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

Enclosure

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{4}------------------------------------------------

# Indications for Use

K102148

510(k) Number (if known): K102148.

Device Name: CuriePACS (Dicom PACS Software)

Indications For Use:

CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of 1

510K K102148

---

**Source:** [https://fda-staging.innolitics.com/device/K102148](https://fda-staging.innolitics.com/device/K102148)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).