The Suspension™ Acromioclavicular (AC) Repair System can be used for adult patients. The AC Repair System is indicated for securing suture to bone for acromioclavicular separations and coracoclavicular displacement.
Device Story
Internal fixation system for acromioclavicular (AC) repair; utilizes non-absorbable suture anchor, pre-threaded suture, and metallic accessories to eliminate knot-tying. Components include 316L stainless steel coracoid anchor, clavicle set screw, clavicle sleeve, and washer. Suture is USP size 5 co-braided polyethylene. Used by surgeons in clinical settings for orthopedic stabilization. System includes manual instruments: torque-limiting driver, hex driver, drill bits, and suture tensioner. Surgeon implants anchor into bone and secures suture to stabilize AC joint; provides mechanical fixation to support healing of separations and displacement.
Clinical Evidence
No clinical data. Bench testing only, including driving torque, pull-out strength, lock screw torque, anchor torque, and suture testing.
Technological Characteristics
Materials: 316L stainless steel (implantable components). Suture: USP size 5 co-braided polyethylene. Principle: Mechanical bone fixation via suture anchor and set screw. Form factor: Implantable anchor system with manual surgical drivers and drill bits. Single-use, sterile. No software or electronic components.
Indications for Use
Indicated for adult patients requiring suture-to-bone fixation for acromioclavicular separations and coracoclavicular displacement.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
# 510(k) Summary
p5 1/2
## Suspension TM Acromioclavicular (AC) Repair System (ref .: K102143)
DEC - 8 2010
December 3, 2010
Submitter/Regulatory Contact: Curtis Raymond Orchid Design - 80 Shelton Technology Center - Shelton, CT 06484
## Sponsor/Manufacturer:
Suspension Orthopaedic Solutions, LLC 2635 Riva Road, Suite 100 - Annapolis, MD 21401
## FDA Establishment Registration Number: Pending
## Trade Name, Common Name, Classification:
Device Trade Name: Suspension™ Acromioclavicular (AC) Repair System
Device Common or Usual Names: Non-absorbable Suture Anchor & Accessories
Classification: Class II, 21 CFR 888.3030, JDR (Suture Anchor Accessories) & MBI (Suture Anchor)
## Predicate Device:
The substantial equivalence of the AC Repair System is based on similarities in indications for use, design, and material composition to the following currently marketed predicate device:
- AxyaLoop™ Non-absorbable Bone Anchor (K060970) �
- 0 AxyaLoop™ Self Tapping Bone Anchor (K052491)
- 0 TeleFlex Force Fiber Blue Co-Braided Polyethylene Suture (K063778)
- � LM Anchor (K960825)
## Description of the Device:
The Suspension" Acromioclavicular (AC) Repair System is an internal fixation system consisting of a non-absorbable suture anchor, pre-threaded suture and other attachment accessories that eliminate the need for tying a knot in the suture. The metallic suture anchor and metallic accessories are fabricated from 316L stainless steel. The suture is a USP size 5, co-braided, white/blue polypropylene, non-absorbable suture.
The system consists of the following implantable components that are included in the sterile blister package:
{1}------------------------------------------------
4102143
pg 2/2
- a USP size 5, co-braided, white/blue polyethylene suture (pre-threaded through the Coracoid Anchor)
- Coracoid Anchor �
- Clavicle Set Screw a
The following instruments are included in the sterile pack:
- ള The Torque-Limiting Driver
- The Coracoid Anchor Driver (a 3.0mm hex driver). o
The system includes non-sterile implantables included in a carier tray:
- Clavicle Sleeve Extension Washer e
- a Clavicle Sleeve
The following instruments are included non-sterile in a carier tray:
- � 5.5mm Drill bit
- ﮟ 4.0mm Drill Bit
- o Suture Tensioner Device
#### Intended Use:
The Suspension " Acromioclavicular (AC) Repair System can be used for adult patients. The AC Repair System is indicated for securing suture to bone for acromioclavicular separations and coracoclavicular displacement. .
#### Technological Characteristics:
The subject devices are substantially equivalent to the predicate devices. Both the subject device and predicate devices are provided sterile and are single use. Both devices utilized braided polyethylene suture and manual surgical instruments. The devices are composed medical grade metals and have similar diameters. The axial pull test data was compared and analyzed.
#### Performance:
The material (316L Stainless Steel) selected is commonly used for orthopedic implants and has a long history of biocompatibility.
The devices have been subjected to recognized consensus standards for these types of devices and perform in a manner equivalent to the predicate devices. The device has been subjected to non-clinical testing including Driving Torque, Pull-Out Strength, Lock Screw Torque, Anchor Torque, and Sulure Testing.
#### Conclusion:
We believe that based on the predicate device comparison and the non-clinical testing performed the subject devices are substantially equivalent to the predicate devices and conclude that the subject devices are as safe and effective as the predicate devices.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of a human figure in profile, with three flowing lines representing the hair or head. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Suspension Orthopaedic Solutions, LLC % Mr. Joseph Azary, Senior Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484
[DEC --- 18 (200
Re: K102143
Trade/Device Name: Suspension Acromioclavicular Repair System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR, MBI Dated: November 17, 2010 Received: November 19, 2010
Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
#### Page 2 - Mr. Joseph Azary
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
A. B. Mr.
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
(revised)
510(k) Number (if known): K102143
Device Name: Suspension Acromioclavicular (AC) Repair System
Indications for Use:
ික්රමිණ දිස්ත්රී ලංකාව
The Suspension™ Acromioclavicular (AC) Repair System can be used for adult The AC Repair System is indicated for securing suture to bone for acromioclavicular separations and coracoclavicular displacement.
Prescription Use _X___ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEE IF (PLEASE DO NOT WRITE BELOW THIS LINE-NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melleson
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_KO2143
(Page 1 of 1)
DEC - 8 2010
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.