COOK WORD BARTHOLIN GLAND CATHETER (FINAL TRADE NAME NOT YET DETERMINED)
K102141 · Cook Urological, Inc. · KNA · May 6, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K102141
Device Name
COOK WORD BARTHOLIN GLAND CATHETER (FINAL TRADE NAME NOT YET DETERMINED)
Applicant
Cook Urological, Inc.
Product Code
KNA · Obstetrics/Gynecology
Decision Date
May 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Used for the treatment of abscesses and cysts of the Bartholin gland.
Device Story
Manual surgical instrument set for Bartholin gland cyst/abscess drainage; includes 15 FR silicone balloon catheter (3 ml capacity, 5.5 cm length), 1" 21 gauge needle, and scalpel with #1 blade. Used by clinicians in clinical settings. Procedure involves lancing cyst/abscess with scalpel; inserting catheter; inflating balloon via syringe/needle to secure position. Catheter remains in situ for up to 28 days to facilitate duct re-epithelialization and permanent drainage. Provides immediate relief and prevents recurrence by maintaining patency of accessory duct.
Clinical Evidence
Bench testing only. Testing included balloon integrity, volume, and leak testing per BS EN 1616:1997/ASTM F623-99; balloon burst testing; and biocompatibility testing per ISO 10993 series. No clinical data presented.
Technological Characteristics
Silicone balloon catheter (15 FR, 5.5 cm length, 3cc balloon). Includes stainless steel scalpel and needle. Biocompatibility per ISO 10993. Performance testing per BS EN 1616:1997 and ASTM F623-99. Sterile, single-use device.
Indications for Use
Indicated for patients requiring treatment of Bartholin gland abscesses or cysts.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Word Bartholin Gland Catheter for Cyst and Abscess Drainage (K861385)
{0}------------------------------------------------
K102141/
pg. 1 of 3
## MAY - 6 2011
| Date of Summary: | June 30, 2010 |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitted by<br>and Primary Contact: | Cindy Foote<br>Regulatory Affairs Specialist<br>Cook Urological, Incorporated<br>1100 West Morgan Street<br>Spencer, IN 47460<br>Tel: 812-829-4891 x7281<br>cindy.foote@cookmedical.com |
| Device Name: | Word Catheter-Silicone Bartholin Gland Balloon Set<br>(final trade name yet to be determined) |
| Classification Name: | Instrument, Manual, Specialized Obstetric-Gynecologic |
| Panel: | Obstetrics/Gynecology |
| Regulation Number: | 21 CFR 884.4530 |
| Regulation Description: | Obstetric-gynecologic specialized manual instrument |
| Device Class: | Class II |
| Product Code: | KNA |
| Indication for Use: | |
Used for the treatment of abscesses and cysts of the Bartholin gland.
### Device Description:
510(k) SUMMARY
The Word Catheter-Silicone Bartholin Gland Balloon Set is used for treatment of abscesses and cysts of the Bartholin gland. The Word Catheter-Silicone Bartholin Gland Balloon Set is a 3 ml balloon catheter which is 5.5 cm in length and constructed of silicone. The catheter is sold with a 1" 21 gauge needle and a scalpel with a # 1 blade is used to lance the Bartholin gland cystlabscess, the syringe with needle is used to inflate the balloon catheter, and the catheter is placed in the cyst or abscess remaining for up to 28 days to allow the duct to re-epithelialize. The devices are provided sterile and are intended for one time use. Tim
## Predicate Devices:
510(k) number: Current Trade Name:
K861385 Word Bartholin Gland Catheter for Cyst and Abscess Drainage
{1}------------------------------------------------
K102/41
page 2 of 3
Distributed by: 510(k) Applicant: CooperSurgical Incorporated Milex Products Incorporated
510(k) number: Current Trade Name: Distributed by: 510(k) Applicant:
K880497 Word Bartholin Catheter Rusch Medical (A Teleflex Medical Company) Hoosier Specialties Incorporated
The Cook Word Bartholin Gland Catheter Set is equivalent in technology, materials, and intended use as the Word Bartholin Catheter for Cyst and Abscess Drainage manufactured by CooperSurgical Incorporated, The Cook Word Bartholin Gland Catheter Set is equivalent in technology and intended use as the Word Bartholin Catheter, 10 French manufactured by Rusch (Teleflex Medical) Medical.
| Device | Word Catheter-<br>Silicone Bartholin<br>Gland Balloon Set | Word Bartholin Catheter for Cyst<br>and Abscess Drainage | Word Bartholin<br>Catheter, 10 French |
|--------------|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Manufacturer | Cook Urological,<br>Incorporated | CooperSurgical, Incorporated | Rusch Medical (Teleflex<br>Medical) |
| Intended Use | Used for treatment of<br>abscesses and cysts<br>of the Bartholin Gland | Used following an incision and drainage of a<br>Bartholin cyst or abscess providing<br>immediate relief. However, Bartholin<br>cysts/abscesses have a high rate of<br>recurrence unless the accessory duct is<br>provided to permanently drain the obstructed<br>acini. This can be accomplished by use of<br>the Word Catheter. | Specifically developed<br>for the treatment of<br>Bartholin gland cysts. |
| Materials | Silicone balloon<br>Silicone Catheter | Silicone balloon<br>Silicone catheter | Latex balloon<br>Silicone Catheter |
| Dimensions | 15 FR with a 3cc<br>balloon and 5.5 cm<br>length | 15 FR with a 3cc balloon and 5.5 cm<br>length | 10 Fr, 5cm with a 5cc<br>balloon |
| Components | Syringe with Needle<br>and Scalpel | Syringe with Needle | None known |
#### Device Comparison
There are few differences between the Word Catheter-Silicone Bartholin Gland Set offered by Cook Urological, Incorporated, the Word Bartholin Catheter for Cyst and Abscess Drainage offered by CooperSurgical, Incorporated, and the Word Bartholin Catheter, 10 French offered by Rusch Medical. The material used to construct the balloon of the Rusch Balloon is different as it is Latex as opposed to silicone, which is used for the Cook and CooperSurgical Word Balloon. The dimensions of the Rusch balloon catheter are slightly different from the Cook and CooperSurgical balloon catheters, which are identical. The Word Catheter-Silicone Bartholin Gland Set is sold with a syringe with a needle and a scalpel, the Word Bartholin Catheter for Cyst and Abscess Drainage is sold with a needle by CooperSurgical, Incorporated. The Word Bartholin Catheter is sold as a stand alone device by Rusch Medical.
{2}------------------------------------------------
K102141
pg. 3 of 3
## Summary of Testing:
The Word Catheter-Silicone Bartholin Gland Balloon Set was tested by the following non-clinical methods to demonstrate that the device is substantially equivalent to the predicate devices in functionality, safety and effectiveness
- . Balloon Integrity, Volume, Leak Testing, using modified recommendations from BS EN 1616; 1997 (which is nearly identical to ASTM F623-99)
- Balloon Burst Testing, using modified recommendations from BS EN 1616:1997 (which is nearly . identical to ASTM F623-99)
- Biocompatibility testing methods as outlined in the ISO 10993 series, conducted using Good . Laboratory Practices (GLP).
Biocompatibility, sterility and performance testing were performed in accordance to Food and Drug Administration guidance's and recognized international standards.
## Comparison with Predicate Devices:
The results of non-clinical and bench testing indicate that the Word Catheter-Silicone Bartholin Gland Balloon Set is as safe and effective as the predicate devices. The Word Cathefer-Silicone Bartholin Gland Balloon Set is similar to the predicate devices in terms of technical characteristics, design, Indications for Use, patient population, performance, and size. The data that was presented for the Word Catheter-Silicone Bartholin Gland Balloon Set prove substantial equivalence to the product devices, prove that the products are safe and effective for their intended use and do not raise any new questions regarding safety and effectiveness.
The Word Catheter-Silicone Bartholin Gland Balloon Set is comparable with respect to intended use to the published predicate device descriptions and meets the requirements for 510(k) substantial equivalence.
1:7 +
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of an eagle with three lines extending from its head, symbolizing health, services, and people.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G( Silver Spring, MD 20993-0002
Ms. Cindy Foote Regulatory Affairs Specialist Cook Urological, Incorporated Cook Women's Health 1100 West Morgan. Street SPENCER IN 47460
MAY - 6 2011
Re: K102141
Trade Name: Word Catheter-Silicone Bartholin Gland Balloon Set Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: KNA Dated: April 18, 2011 Received: April 19, 2011
Dear Ms. Foote:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate informations related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleding.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898: In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Patt 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutliDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliante. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Leuner MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
:
| 510(k) Number (if known): | K102141 |
|---------------------------|-----------------------------------------------------------------------|
| Device Name: | Word Catheter-Silicone Bartholin Gland Balloon Set |
| Indications for Use: | Used for the treatment of abscesses and cysts of the Bartholin gland. |
Prescription Use? X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
| 510(k) Number: | K102141 |
|----------------|---------|
|----------------|---------|
ﭘﯿﻨﮯ ﮐﮯ ﻧﺎﻡ
.
.
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.