K102140 · Edan Instruments, Inc. · HGM · Jan 13, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K102140
Device Name
FETAL MONITORS, MODELS F2 AND F3
Applicant
Edan Instruments, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Jan 13, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The F2 and F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
Device Story
Fetal monitor provides non-invasive and invasive monitoring of fetal heart rate (FHR) and uterine activity. Inputs: ultrasound Doppler signals (via external transducer), TOCO transducer signals (external uterine activity), direct fetal ECG (DECG) signals, and intrauterine pressure catheter (IUPC) signals. Device processes these inputs to display real-time fetal and maternal data. Used in antepartum examination rooms, labor, and delivery rooms by trained clinical personnel. Connectivity: RJ45 interface for central monitoring system; DB9 interface for wireless network module. Output: visual monitoring data for clinical assessment of fetal well-being during labor and delivery. Benefits: enables continuous fetal surveillance to support clinical decision-making.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including software, hardware, safety, and environmental testing, alongside risk analysis and final validation.
Technological Characteristics
Non-invasive ultrasound Doppler, external TOCO, and direct fetal ECG (DECG) sensing. Connectivity via RJ45 (wired) and DB9 (wireless module). Form factor: Fetal monitor console. Software-based signal processing. Sterilization: Not specified.
Indications for Use
Indicated for non-invasive and invasive fetal monitoring during antepartum examination, labor, and delivery in pregnant women from the 28th week of gestation. Contraindicated for use in intensive care units, operating rooms, or home use.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
CADENCE, CADENCE DUAL, CADENCE II FETAL MONITOR (K082369)
Submission Summary (Full Text)
{0}------------------------------------------------
F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION
Section 1
# 510(k) Summary of Safety and Effectiveness
JAN 1 3 201
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
Submitter: Edan Instruments, Inc 3/F - B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, shekou, Nanshan Shenzhen, 518067 P.R. China Tel: 86-755-26882220. . . . . . . . . . . . . Fax:86-755-26882223 Contact person: Yue Qiuhong Official correspondent: William Stern Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 X27 Fax: 914 376 6111 Date of Preparation: 2010-7-22 Proprietary Name: F2, F3 Fetal Monitor (Models: F2, F3) Classification Name: System, Monitoring, Perinatal Product code: HGM, HGL Predicate Devices:
| Predicate devices | CADENCE, CADENCE DUAL,<br>CADENCE II FETAL MONITOR |
|-------------------|----------------------------------------------------|
| Manufacturer | Edan Instruments, Inc |
| K # | K082369 |
Device Description:
With non-invasive ultrasound Doppler, external TOCO and direct fetal ECG technique, Fetal Monitor provides Non-Stress testing for pregnant women from the 28th week of gestation. You can externally monitor the fetal heart rate using ultrasound and uterine activity via a TOCO transducer. Alternatively, you can internally monitor one of the fetal heart rate with direct fetal ECG technique and uterine activity with an Intrauterine Pressure Catheter.
{1}------------------------------------------------
## F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION ·
Section 1
The Fetal Monitor has six models: Cadence, Cadence DUAL, Cadence PRO, Cadence II, F2 and F3. The former four models Cadence, Cadence DUAL, Cadence PRO, Cadence II have already been cleared under K082369, while the other two models F2, F3 are new models of Fetal Monitor in this Traditional 510(K) application, which have the same intended use and constructions to the existing models.
The Fetal monitor can be connected with Central Monitoring System via RJ45 interface. Also it can be connected to wireless network module via a DB9 interface, and the wireless network module will complete the data switch of the monitor and the Obstetrical Central Monitoring System.
### Comparison with predicate device .
The F2, F3 fetal monitors have the same device characteristics as the predicate approved device cleared under K082369. Both models use the same technology and circuitry as the already approved device models Cadence, Cadence DUAL, Cadence PRO, Cadence II cleared under K082369. Hence the models F2, F3 above are substantially equivalent to the predicate devices cited.
#### Intended Use
The F2, F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
Contraindications:
It is not intended for use in intensive care units, operating rooms or for home use.
Test Summary:
The following quality assurance measures were applied to the development of the F2, F3 Fetal Monitors.
- · Software testing
- · Hardware testing
- · Safety testing
- · Environment test
- · Risk analysis · · ..
- · Final validation
#### Conclusion:
Verification and validation testing was done on the Fetal Monitor models F2, F3. This premarket notification submission demonstrates
5/22
{2}------------------------------------------------
, KIDZL 40 , કર્ભ રે page
## F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION
Section 1
that Fetal Monitor models F2, F3 is substantially equivalent to the predicate device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with its wings spread, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the border of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Yue Qiuhong Certification Engineer Edan Instruments, Inc. 3/F-B, Nanshan Medical Equipment Park Nanhai Road 1019# Shenzhen, Guangdong, 518067 CHINA
APR 2 6 2012
K102140 Re:
Trade/Device Name: Fetal Monitor model F2 and F3 Regulation Number: 21 CFR § 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM, HGL Dated: December 20, 2010 Received: December 27, 2010
Dear Mr. Oiuhong:
This letter corrects our substantially equivalent letter of January 13, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Fetal Monitor model F2 and F3, as described in your premarket notification:
## MS3-109301
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found
{4}------------------------------------------------
Page 2 - Mr. Yue Oiuhong
in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Ms. Kathryn Daws- Kopp at (301) 796-6535. は
Sincerely yours,
t
Herbert Hoover
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosures
{5}------------------------------------------------
F2, F3 Fetal Monitors TRADITIONAL 510(K) SUBMISSION
Section 1
## Indication for Use
510(k) Number (if.known):
K102140
Device Name: F2, F3 Fetal Monitors (Models: F2, F3)
The F2 and F3 Fetal Monitor are intended for non-invasive and invasive monitoring of fetus during antepartum examination, labor and delivery. They are intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
They both provide Non-stress testing for pregnant women from the 28th week of gestation. They can externally monitor the FHRs using ultrasound and uterine activity via a TOCO transducer. Alternatively, they can internally monitor one of the FHRs with DECG and uterine activity with an IUPC.
× Prescription Use (21 CFR Part 801 Subpart D)
And/Or Over the Counter Use _ (21 CFR Part 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ivision Sign
of Reproductive. Gastro-R
{6}------------------------------------------------
## Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 1MHz PW fetal probe
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | Mode Operation | | | | | | | | | | |
|----------------------------------|----------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | P | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheraln Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
ndication; P=previously cleared by FDA; e=ADDED UNDER appendix E Additional Comments: _________________________________________________________________________________________________________________________________________________________
The above is a 1MHz PW transducer for the fetal heart rate detection
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I PNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Tracy
r. 13, 2012
Section 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.