K102117 · Shippert Medical Technologies Corp. · MUU · Aug 6, 2010 · General, Plastic Surgery
Device Facts
Record ID
K102117
Device Name
TISSU TRANS FILTRON SYRINGE FILL
Applicant
Shippert Medical Technologies Corp.
Product Code
MUU · General, Plastic Surgery
Decision Date
Aug 6, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5040
Device Class
Class 2
Indications for Use
Tissu Trans Filtron Syringe Fill is intended to be used with house vacuum and/or cleared pumps, tubing and cannulas, for the collection of aspirated fat, for aesthetic body contouring. If the fat is untreated, it may be reinjected via a cleared injection apparatus.
Device Story
Sterile, single-use disposable canister accessory for Tissu Trans Filtron (k092482); contains six 60cc syringes for collection of aspirated fat. Connects to vacuum source and primary filtration device via silicone tubing. Physician-operated in OR/clinical settings; vacuum draws aspirated tissue into syringes sequentially. Once filled, syringes are detached, fitted with physician-supplied luer-lock injection cannulas, and used for immediate autologous fat reinjection. Simplifies tissue harvesting, filtering, and transfer process; enables direct reinjection without additional handling.
Clinical Evidence
Clinical testing conducted by Ronald D. Shippert, M.D. to validate design specifications and performance. Testing confirmed the device functions as intended, meeting all design requirements and demonstrating safety and effectiveness equivalent to the predicate device.
Technological Characteristics
Sterile, single-use disposable canister with six 60cc syringes. Materials tested per ISO 10993 for external communication (tissue contact < 24 hours). Operates via external vacuum source (15" Hg). Sterilized by Gamma Radiation. Accessory to existing suction lipoplasty system.
Indications for Use
Indicated for patients undergoing aspiration of soft tissue in plastic, reconstructive, neuro, gastrointestinal, urological, general, orthopedic, gynecological, thoracic, or laparoscopic surgery. Used for collection of aspirated fat for aesthetic body contouring and potential autologous reinjection.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
Tissu Trans Filtron (k092482)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) SUMMARY Section 19:
## k# 102117
As required by 807.97
#### Prepared By:
Shippert Medical Technologies . 6248 South Troy Circle, Unit A Centennial, Colorado 80111 Tele: 303.754.0044 Fax: 303.754.0318 Contact: Sarah Lake Shippert Email: sarah@shippertmedical.com Date: July 20, 2010
SPECIAL 510(k) APPLICATION
AUG - 6 2010
Telephone ( 303) 888.4965
Submitted by: Sarah Lake Shippert
FDA Establishment # 1718903
## Proposed Device:
| Tradename: | Tissu Trans Filtron Syringe Fill |
|-----------------------|----------------------------------|
| Model/Product Number: | 3-TT-Filtron Syringe Fill |
| Common Name: | Suction Lipoplasty System |
| Classification name: | Suction Lipoplasty System |
| Product Code: | MUU |
| Regulation Number: | 21 CFR 878.5040 |
| Device: Class II, | Sterile |
| Submission Purpose: | Modification to: |
|---------------------|-------------------------------------------------|
| | Shippert Medical's Tissu Trans Filtron, k092482 |
| Predicate Device: | Shippert Medical's Tissu Trans Filtron, k092482 |
## Device Classification and Product Code:
As shown in 21 CFR 878.5040, Suction Lipoplasty Systems are defined as devices consisting of collection bottles, cannulas and connecting tubing for use in aesthetic body contouring or collection battles, calinoplasty Systems are classified as Class II. They have been assigned Product Code MUU.
## Indications For Use
Tissu Trans Filtron Syringe Fill is intended to be used with house vacuum and/or cleared pumps, tubing and cannulas, for the collection of aspirated fat, for aesthetic cleared purilips, tubing and ournalias, for may be reinjected via a cleared injection apparatus.
{1}------------------------------------------------
## con't. Section 19: 510(k) SUMMARY:
#### k102117
Intended Use: Shippert Medical's Tissu Trans Filtron Syringe Fill is intended to be used with house vacuum and/or cleared pumps, tubing and cannulas, for the collection of aspirated fat, for aesthetic body contouring. If the fat is untreated, it may be reinjected via a cleared injection apparatus.
Tissu Trans Filtron Syringe Fill is used in the aspiration, harvesting, filtering and transferring of autologous tissue.
Tissue Trans Filtron is intended for use in the following surgical specialties when the aspiration of soft tissue is desired:
- Plastic and Reconstructive Surgery . .
- Neurosurgery .
- Gastrointestinal and Affiliated Organ Surgery ●
- Urological Surgery .
- General Surgery ◆
- Orthopedic Surgery .
- Gynecological Surgery .
- Thoracic Surgery .
- Laparoscopic Surgery
The Indications For Use and the Intended Use are identical to the Predicate Device, Tissu Trans Filtron k0924.82.
#### Design Characteristics:
Tissu Trans Filtron Svringe Fill is an optional accessory to Shippert Medical's Tissu Trans Filtron, k092482. Shippert Medical's Tissu Trans Filtron Syringe Fill is a sterile, single use, disposable device used in the aspiration, harvest, filtering and transferring of autologous tissue. Used by physicians in the harvesting of body tissues, this device is a single use, disposable sterile canister with six enclosed 60cc syringes used for collecting the harvested tissue after harvesting/aspiration.. The harvested tissue is stored in the Syringe Fill syringes and is intended for immediate reinjection of the harvested tissue back into the donor patient.
## Device Components:
The Tissu Trans Filtron Syringe Fill connects to the inverted Tissu Trans Filtron, k092482, with an accessory silicone tubing piece which is included in the Tissu Trans Filtron Syringe Fill packaging. This tubing is connected onto the the Tissu Trans Filtron Svringe Fill's lid at the appropriate in-take port. This same transfer tubing is then connected onto the inverted Tissu Trans Filtron at its out-take port. The vacuum tubing from the Tissu Trans Filtron is repositioned from the Tissu Trans Filtron's vacuum port to the Tissue Trans Filtron Syringe Fill's vacuum port. When the vacuum pump is turned
{2}------------------------------------------------
#### con't. Section 19: 510(k) SUMMARY:
on the Tissu Trans Filtron Syringe Fill's six syringes begin to fill up as the physician releases the pinch clamp. As one syringe is filled, the tissue continues to fill the subsequent svringes. A pinch clamp is used to close off the silicone transfer tubing after the final sixth syringe is filled and the vacuum pump is turned off. The physician or OR staff will remove the filled syringes and place a physician supplied FDA approved re-injection cannula onto the syringe's luer lock. The syringe's plunger will be inserted at the large open end of the syringe and re-injection of the autologous tissue can proceed.
Material Composition: The components of the Tissu Trans Filtron Syringe Fill do not have direct patient contact. The necessary FDA cleared liposuction cannula (direct patient contact) is supplied by the physician.
All components of the Tissu Trans Filtron Svringe Fill have been tested and have passed the ISO 10993 testing regimen for External Communication Devices. Tissue contact, of less than 24 hours. All necessary biocompatibility testing was performed on the sterile, finished device.
#### Sterility:
The Tissu Trans Filtron Syringe Fill is sterilized by Gamma Radiation.
#### In Vitro Testing :
Clinical testing was conducted by Ronald D. Shippert, M.D. to validate and verify that the proposed device met all design specifications. Validation tests were conducted to validate the design control activity. Performance of the Tissu Trans Filtron Syringe Fill satisfies all design specifications. Mechanical testing of the Tissu Trans Filtron Syringe Fill demonstrates that the device is substantially equivalent to the predicate device. Device Performance was satisfied by clinical testing performed by Ronald Shippert, M.D. All testing proved to be safe and effective. The device performed as desired and was as safe and as effective as the predicate device.
Equivalence To Marketed Device:_Shippert Medical's Tissu Trans Filtron Syringe Fill is substantially equivalent to the Tissu Trans Filtron device under K092482. This new device is an optional accessory to the Shippert Medical Technologies Tissu Trans Filtron device.
Shippert Medical Technologies Tissu Trans Filtron k092482 Classification Name: Suction Lipoplasty System Class 2 Device, Product Code MUU Regulation Number 878.5040, General Hospital
{3}------------------------------------------------
#### con't. Section 19: 510(k) SUMMARY:
As stated in 21 CFR 878.5040 , (product code MUU), Suction Lipoplasty Systems are defined as devices consisting of collection bottles, cannulas, and connecting tubing for use in aesthetic body contouring procedures. Suction Lipoplasty Systems are classified as Class 2. They have been assigned Product Code MUU.
As 21 CFR 878.5040 states the existence of a "collection bottle and detachable tube" in the description of Product Code MUU, Shippert Medical is now modifying the original Tissu Trans Filtron, k092482 to offer an additional canister (collection bottle) and a detachable tube as an optional accessory.
Technological Characteristics: Shippert Tissu Trans Filtron Syringe Fill is used in conjunction with a hospital or facility supplied vacuum-powered body fluid suction apparatus which is capable of a low pressure setting of 15" of mercury. The Tissu Trans Filtron Syringe Fill is an optional accessory to the Tissu Trans Filtron, k 092482. The technological characteristics of the Tissu Trans Filtron Syringe Fill follow those of its predicate, Tissu Trans Filtron, k092482.
This device is sterilized by Gamma Radiaiton, same as the predicate device, k092482.
Summary: The Shippert Medical Tissu Trans Filtron Syringe Fill device described in this submission is a modification to the previously 510(k) cleared device. Tissu Trans Filtron k092482. This modification is substantially equivalent to the Tissu Trans Filtron k092482 predicate device and is safe and effective.
The modified device has the same classification, the same indications, the same intended use, the same fundamental technology, the same design structure as described in 878.5040, the same sterility protocol and the same safety and effectiveness performance as the predicate device.
Given the "Intended Use, Indications For Use, device technology, materials and basic design structure of Product Code MUU, Shippert Medical Technologies claim Substantial Equivalence to Tissu Trans Filtron, k092482.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shippert Medical Technologies % Ms. Sarah Lake Shippert 6248 South Troy Circle, Unit A Centennial, Colorado 80111
Re: K102117
Trade/Device Name: Tissu Trans Filtron Syringe Fill Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: July 20, 2010 Received: July 28, 2010
Dear Ms. Shippert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
AUG - 6 2010
{5}------------------------------------------------
Page 2 - Ms. Sarah Lake Shippert
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm:
Sincerely yours.
cerely yours,
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K102117
# Indications for Use
510(k) Number (if known): k102117
AUG - 6 2010
Tissu Trans Filtron Syringe Fill Device Name:
Tissu Trans Filtron Syringe Fill is intended to be used with house vacuum and/or cleared pumps, tubing and cannulas, for the collection of aspirated fat, for aesthetic body contouring. If the fat is untreated, it may be reinjected via a cleared injection apparatus.
X Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krone for MXM
---
(Division Sign Off)
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102117
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.