POWDER FREE VINYL PATIENT EXAMINATION GLOVES

K102097 · Shijiazhuang Great Vision Plastic Products Co., LT · LYZ · Nov 17, 2010 · General Hospital

Device Facts

Record IDK102097
Device NamePOWDER FREE VINYL PATIENT EXAMINATION GLOVES
ApplicantShijiazhuang Great Vision Plastic Products Co., LT
Product CodeLYZ · General Hospital
Decision DateNov 17, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable vinyl patient examination gloves; worn on examiner's hand or finger; intended to prevent cross-contamination between patient and examiner. Used in clinical or general medical settings. Device is non-sterile. Provides barrier protection during patient examinations. Benefits include reduced risk of contamination transmission.

Clinical Evidence

No clinical data. Bench testing only. Testing included physical and dimension measurements (ASTM D-5250-06e1), FDA 1000 ml water-fill test (AQL 2.5), primary skin irritation and sensitization testing, and residual powder testing (ASTM D-6124-06) to confirm powder-free status (<2 mg/glove).

Technological Characteristics

Vinyl patient examination glove. Meets ASTM D-5250-06e1 standards. Powder-free (residual powder <2 mg/glove). Available in yellow, white, blue, and pink. Non-sterile. Class I device, product code LYZ.

Indications for Use

Indicated for use as a disposable, non-sterile patient examination glove worn by examiners to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # EXHIBIT #1 Page 1 of 2 ### 510(K) SUMMARY This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR \$807.92. The assigned 510(K) number is: 《102097 #### 1. Owner's Identification: Mr. Jianping Wang Shijiazhuang Great Vision Plastic Products Co., Ltd. Cangshi South Rd., East Dist. Jinzhou City, Hebei Province, CHINA 052260 Date Summary Prepared: June 7, 2010 ### 2. Name of the Device: Shijiazhuang Great Vision Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) ### 3. Predicate Device Information: Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) ### 4. Device Description: Classified by FDA's General Hospital and Personal Use Device panel as Class I. 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1. # 5. Intended Use: A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. # 6. Comparison to Predicate Devices: Shijiazhuang Great Vision Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) is substantially equivalent in safety and effectiveness to the Better Care Plastic Products Co., Ltd's Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink). {1}------------------------------------------------ # K102097 # EXHIBIT #1 Page 2 of 2 - # 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows: The standards used for Shijiazhuang Great Vision Plastic Products Co., Ltd.'s glove production are based on ASTM D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove). ### 8. Labeling: There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. ## 9. Discussion of Clinical Tests Performed: Not Applicable - There is no hypoallergenic Claim. ### 10. Conclusions: Shijiazhuang Great Vision Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three arms or lines extending from the body. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002 Shijiazhuang Great Vision Plastic Products Company, Limited C/O Ms. Kathy Liu Surprotect Incorporated 3973 Schaefer Avenue Chino, California 91710 000 77 2000 Re: K102097 Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: October 23, 2010 Received: October 27, 2010 Dear Ms. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Ms. Liu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersQffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATION FOR USE 0102 77 200 510 (k) NUMBER (if KNOW): _ KIO 2,097 APPLICANT: DEVICE NAME: Shijiazhuang Great Vision Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) INDICATIONS FOR USE: A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR Over-The-Counter-Use √ (21CFR 801Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Elaine S. Mayhall Dergisfi Exgaluation (ODE) Concurrent of CDRH, Office of De (Division of Anesthesiology, General Hospital Division Control, Dental Devices 510(k) Number: _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...