The Apex Knee ™ Modular Tibia System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate: • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis • Rheumatoid arthritis • Correction of functional deformity • Revision procedures where other treatments or devices have failed. The porous coated femoral component may be used cemented or uncemented (biological fixation). The porous coated tibial baseplate component may be used uncemented (biological fixation). All other femoral, tibial baseplate and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibia System Tibial Augments are intended to be bolted to the Tibia baseplate and cemented to the prepared tibia.
Device Story
Apex Knee Modular Tibia System is a modular orthopedic implant for total knee replacement. Components include a symmetrical tibial baseplate (sizes 1-6), modular stems (9-17mm diameter; 75-150mm length), and stackable tibial augments (4mm thick). Components attach via Morse taper connections and locking bolts. The system provides rotational stability through modular stems or pegs. Used in surgical settings by orthopedic surgeons to replace damaged knee joints. The device is implanted to restore joint function and alleviate pain associated with degenerative or inflammatory joint disease. The modular design allows surgeons to customize the tibial construct based on patient anatomy and bone loss requirements.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including augment attachment strength (ASTM F1814-97), fretting analysis (ASTM F1800-07), and stem/tray attachment strength (ASTM F1814-97). All samples met acceptance criteria.
Technological Characteristics
Materials: ASTM F75 Cobalt-chromium (baseplate, augments), ASTM F136 Ti-6Al-4V ELI titanium alloy (bolts, caps, stems, keels, pegs). Design: Modular tibial baseplate with Morse taper stem/keel connection and locking bolt. Symmetrical, stackable augments. Connectivity: None. Energy: None. Sterilization: Not specified.
Indications for Use
Indicated for primary or revision total knee replacement in patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or failed previous treatments/devices.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
# K101994 (1/3)
### 510(k) Summary
## SEP 2 8 2010
.
| Apex Knee ™ Modular Tibia System | | 22 September, 2010 | |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-----------------------------------------------------------------------------------|
| Submitter | OMNI life science, Inc.<br>50 O'Connell Way<br>E. Taunton MA 02718 | Contact | Radhika Pondicherry<br>Regulatory Affairs<br>774-226-1852<br>(508) 822-6030 (fax) |
| Preparation Date | 22 September 2010 | | |
| Device Nam<br>Trade Name | Apex Knee ™ Modular Tibia System | | |
| Common/Classification<br>Name | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented<br>prosthesis | | |
| Regulatory Class | Class II per 21 CFR §888.3560 | | |
| Product Code | JWH, MBH | | |
| Legally Marketed<br>Predicate Device(s) | K060192- Apex Knee System, cleared 15Jul2006<br>K094017- Apex Knee System Tibial Baseplate Augment, cleared 05Mar2010<br>K080361- Biomet, Regenerex™ Tibial Component, cleared 21April2008 | | |
| Device Description | The Modular Tibial Baseplate is offered in sizes 1 thru 6 and is a symmetrical design. It is 5mm<br>thick, and accepts Tibial Stems and Augment Blocks on its inferior surface.<br>Modular Tibia Stems are available from 9-17mm diameter in lengths of 75,100 or 150mm.<br>The stem mates to the Modular Tibial Baseplate via a Morse taper connection, secured by a<br>Locking Bolt.<br>Modular Tibia Augments are available in, size 1 thru 6. Each Augment is 4mm thick, and may<br>be placed on either the medial or lateral side of the baseplate. They may be stacked up to<br>three high in equal or descending sizes to create either a uniform or stepped lateral profile.<br>Bone cement should not be used between stacked augments. Once stacked the augments<br>must be secured to the tibial tray using a locking bolt of appropriate length to match the<br>height of the augment stack. The Tibial tray (tibial baseplate) - augment assembly is then<br>cemented to the prepared tibia. | | |
| Indications for Use | The Apex Knee ™ Modular Tibia System is intended for use as a primary or revision total knee<br>replacement.<br>This prosthesis may be used for the following conditions, as appropriate: | | |
| | • Non-inflammatory degenerative joint disease, including osteoarthritis and avascular<br>necrosis<br>• Rheumatoid arthritis<br>• Correction of functional deformity<br>• Revision procedures where other treatments or devices have failed | | |
| | The porous coated femoral component may be used cemented or uncemented (biological<br>fixation). The porous coated tibial baseplate component may be used uncemented (biological<br>fixation). All other femoral, tibial baseplate and patellar components are indicated for<br>cemented use only.<br>The Apex Knee™ Modular Tibia System Tibial Augments are intended to be bolted to the Tibia<br>baseplate and cemented to the prepared tibia. | | |
:
·
{1}------------------------------------------------
#### Predicate Device Comparison
.
| | Apex Knee Modular<br>Tibial<br>(subject device) | Apex Tibial Baseplate<br>Augment<br>(K094017) | Biomet- Regenerex<br>Tibial Component<br>(K080361) |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | | | |
| Intended Use | Primary and revision<br>total knee<br>replacement | Primary and revision<br>total knee<br>replacement | Painful and disabled<br>knee joint resulting<br>from osteoarthritis,<br>rheumatoid arthritis,<br>traumatic arthritis<br>where one or more<br>compartments are<br>involved. Correction<br>of varus, valgus, or<br>posttraumatic<br>deformity. Correction<br>or revision of<br>unsuccessful<br>osteotomy,<br>arthrodesis, or failure<br>of previous joint<br>replacement<br>procedure. |
| Design | | | |
| Modular Tibial<br>Baseplate<br>insert locking<br>mechanism | Dovetail rails for<br>engaging the<br>UHMWPE tibial<br>insert with a central<br>threaded hole for<br>receiving the locking<br>bolt | Dovetail rails for<br>engaging the<br>UHMWPE tibial insert<br>with a central<br>threaded hole for<br>receiving the locking<br>bolt | N/A |
| Modular Tibia<br>Stem, Keel and<br>Cap | Modular Stem, Keel<br>and Cap- Attaches to<br>the Tibial Baseplate<br>via a Morse taper<br>connection, secured<br>by a Locking Bolt.<br>Provides rotational<br>stability and a point<br>of fixation for the<br>locking bolt to thread<br>onto. | Monoblock Tibial<br>Baseplate and keel. | Modular press fit stem<br>- taper junction and<br>screw. |
| Modular Tibia<br>Augments | Symmetrical<br>Augments are<br>stackable and<br>flipable and attach to<br>the bottom of the<br>Tibial Baseplate. | Designed to mate with<br>and be cemented to<br>the Apex Knee System<br>Tibial Baseplate. | N/A |
| Modular Tibia<br>Augment Bolt | Locking bolt used to<br>secure the<br>attachment of the<br>Tibial Augments to<br>the Tibial Baseplate. | Locking bolt used to<br>attach Tibial<br>Augments,<br>polyethylene insert<br>and Tibial Baseplate. | N/A |
| Insert | Locking bolt used to | secure the attachment | |
| Retaining Bolt<br>(UHMWPE) | secure the<br>attachment of the<br>Insert to the Tibial<br>Baseplate | of the Insert to the<br>Tibial Baseplate | |
| Modular Tibia<br>Pegs | Attaches to the<br>threaded holes in the<br>bottom of the tibial<br>Baseplate. Modular<br>Peg provides<br>additional rotational<br>stability to the tibial<br>component when the<br>Tibial Augments are<br>not being used. | NA | Modular Tibial Pegs<br>used to stabilize the<br>Tibial Plate on the<br>tibial plateau. |
| Materials and Standards | | | |
| Knee<br>components | ASTM F75-Cobalt<br>chromium- Tibial<br>Baseplate, Tibial<br>Augments | ASTM F1537-<br>Wrought cobalt<br>chromium- Tibial<br>Baseplate | CoCrMo alloy-<br>Modular Pegs |
| | ASTM 136- Ti-6Al-4V<br>ELI titanium alloy-<br>Inset locking bolt,<br>Augment locking<br>bolt, Cap, Stem, Keel<br>and Pegs | ASTM 136-<br>Ti-6Al-4V ELI titanium<br>alloy- Tibial Baseplate<br>Augment | Titanium Alloy-<br>Modular press fit stem |
.
,
{2}------------------------------------------------
Augment Attachment Strength per ASTM F1814-97AR03 .
- Fretting Analysis per ASTM F1800-07 .
- Stem Attachment and Tray/Augment Attachment Strength ASTM F1814-97 .
All samples tested met the acceptance criteria.
No clinical studies were performed. Clinical Test Summary
#### Conclusions
Non-Clinical Test
Summary
The Apex Knee ™ Modular Tibịa System is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OMNI Life Science, Inc. c/o Ms. Radhika Pondicherry 50 O'Connell Way Suite 10 East Taunton, MA 02718
#### Re: K101994
SEP 2 8 2010
Trade/Device Name: Apex Knee Modular Tibia System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBH
Dated: July 13, 2010 Received: July 15, 2010
Dear Ms. Pondicherry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Radhika Pondicherry
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For Pete D
Rumm
Rumm
Dep D.R
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
#### 510(k) Number: K101994
K101999
SEP 28 2010
Device Name: Apex Knee ™ Modular Tibia System
The Apex Knee ™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate:
- . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular Necrosis;
- . Rheumatoid arthritis;
- Correction of functional deformity; .
- . Revision procedures where other treatments or devices have failed.
The porous coated femoral component may be used cemented or uncemented (biological fixation). The porous coated tibial baseplate component may be used uncemented (biological fixation). All other femoral, tibial baseplate and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibia System Tibial Augments are intended to be bolted to the Tibia baseplate and cemented to the prepared tibia.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Smuta for mxn
(Division Sign-Off)
Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K101994
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.