K101923 · Depuy Spine · MAX · Dec 1, 2010 · Orthopedic
Device Facts
Record ID
K101923
Device Name
CONCORDE CURVE SYSTEM
Applicant
Depuy Spine
Product Code
MAX · Orthopedic
Decision Date
Dec 1, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CONCORDE, CONCORDE Bullet, CONCORDE Curve, COUGAR, DEVEX, and LEOPARD Systems are indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF (CONCORDE, CONCORDE Bullet), TLIF (CONCORDE, CONCORDE Bullet, CONCORDE Curve, DEVEX, LEOPARD) or anterior (COUGAR) approach using autogenous bone. When used as interbody fusion devices these implants are intended for use with DePuy Spine supplemental internal fixation products. The CONCORDE, CONCORDE Bullet, CONCORDE Curve, COUGAR, DEVEX, and LEOPARD Systems are indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. These systems are also indicated for treating fractures of the thoracic and lumbar spine. These systems are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device these systems are intended for use with DePuy Spine supplemental internal fixation products.
Device Story
CONCORDE Curve System consists of carbon-fiber reinforced polymer cages with tantalum wires; used as intervertebral body fusion or vertebral body replacement devices. Implants placed via PLIF or TLIF approach in lumbar spine or thoracolumbar spine; requires use of DePuy Spine supplemental internal fixation. Facilitates fusion using autogenous bone; restores biomechanical integrity of spinal column. Used in clinical settings by surgeons; output is physical stabilization of spinal segments. Benefits include pain relief, spinal decompression, and structural support for diseased or fractured vertebrae.
Clinical Evidence
Bench testing only. Performance evaluated via static and dynamic compression, compression shear, and torsion per ASTM F2077, and subsidence per ASTM F2267. Results demonstrated performance equivalent to predicate devices.
Technological Characteristics
Carbon-fiber reinforced polymer with tantalum wires. Intervertebral body fusion/vertebral body replacement cage form factor. Mechanical stabilization principle. Tested per ASTM F2077 (compression, shear, torsion) and ASTM F2267 (subsidence).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (discogenic back pain, confirmed by history/radiography) at 1-2 contiguous lumbar levels (L2-S1) after 6 months of failed non-operative treatment; also indicated for thoracolumbar (T1-L5) vertebral body replacement due to tumors, decompression of spinal cord/neural tissues, height restoration of collapsed vertebral bodies, and treatment of thoracic/lumbar fractures.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# KI01923
#### IX. 510(k) Summary
SUBMITTER:
DePuy Spine, Inc. 325 Paramount Drive Raynham, MA 02780
DEC 1 2010
CONTACT PERSON: Eugene Bang Regulatory Affairs Associate Tel: 508-977-3966 Fax: 508-828-3797 Email: ebang@its.jnj.com
DATE PREPARED: . July 8, 2010
CLASSIFICATION NAME: Intervertebral Body Fusion Device Spinal Intervertebral Body Fixation Orthosis
- REGULATION NUMBER: 888.3060, 888.3080
- PRODUCT CODE: MAX, MQP
DEVICE CLASS: Class II
PROPRIETARY NAME: CONCORDE® Curve System
PREDICATE DEVICES: LEOPARD® System (K081917) CONCORDE® Bullet System (K081917)
DEVICE DESCRIPTION: CONCORDE® Curve System consists of cages of various sizes.
> This system also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.
INTENDED USE: The CONCORDE. CONCORDE Bullet. CONCORDE Curve. COUGAR, DEVEX, and LEOPARD Systems are indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient historv and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF (CONCORDE, CONCORDE Bullet), TLIF (CONCORDE, CONCORDE
Ps 1 of 3
{1}------------------------------------------------
## K101923
Bullet, CONCORDE Curve, DEVEX, LEOPARD) or anterior (COUGAR) approach using autogenous bone. When used as interbody fusion devices these implants are intended for use with DePuy Spine supplemental internal fixation products.
The CONCORDE, CONCORDE Bullet, CONCORDE Curve, COUGAR, DEVEX, and LEOPARD Systems are indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. These systems are also indicated for treating fractures of the thoracic and lumbar spine. These systems are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device these systems are intended for use with DePuy Spine supplemental internal fixation products.
MATERIALS: The CONCORDE® Curve System is made from carbon-fiber reinforced polymer with tantalum wires.
### TECHNOLOGICAL CHARACTERISTICS:
The purpose of this submission is to obtain market clearance for the proposed components to the CONCORDE® Curve System. The key difference between the subject and predicate devices is the addition of various sizes of the implants. These proposed components have the same design characteristics, performance, intended use, and packaging as the predicate devices.
## PERFORMANCE DATA:
The testing of the CONCORDE® Curve System was performed in accordance with the following,
- Static and dynamic compression per ASTM F2077 .
- Static and dynamic compression shear per ASTM F2077 .
- . Static and dynamic torsion per ASTM F2077
- Subsidence per ASTM F2267 .
The testing results demonstrated that CONCORDE® Curve implants performed as well as the predicate devices.
!
i
Pg 2 of 3
{2}------------------------------------------------
K101923
CONCLUSION:
Based on the predicate comparison and testing, the subject . device CONCORDE® Curve System is substantially equivalent to the predicate device.
Pg 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized image of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple, using black lines on a white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
DePuy Spine, Inc. % Mr. Eugene Bang Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02780
DEC 1 2010
Re: K101923
Trade/Device Name: CONCORDE® Curve System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX. MOP Dated: November 12, 2010 Received: November 15, 2010
Dear Mr. Bang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
#### Page 2 - Mr. Eugene Bang
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Molkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### III. Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
CONCORDE® Curve System Device Name:
#### Indications For Use:
The CONCORDE, CONCORDE Bullet, CONCORDE Curve, COUGAR, DEVEX, and LEOPARD Systems are indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF (CONCORDE, CONCORDE Bullet), TLIF (CONCORDE, CONCORDE Bullet, CONCORDE Curve, DEVEX, LEOPARD) or anterior (COUGAR) approach using autogenous bone. When used as interbody fusion devices these implants are intended for use with DePuy Spine supplemental internal fixation products.
The CONCORDE, CONCORDE Bullet, CONCORDE Curve, COUGAR, DEVEX, and LEOPARD Systems are indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of turnors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. These systems are also indicated for treating fractures of the thoracic and lumbar spine. These systems are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device these systems are intended for use with DePuy Spine supplemental internal fixation products.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
K101923 510(k) Number_
Pg 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.