# JOULE MULTI-PLATFORM SYSTEM (K101916)

_Sciton, Inc · GEX · Mar 18, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K101916

## Device Facts

- **Applicant:** Sciton, Inc
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Mar 18, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The JOULE Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology. The integral thermo-electric cooler is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to: Reduce pain during and/or associated with light or laser treatment (via partial anesthesia from cooling); Reduce discomfort during and/or associated with light or laser treatment; Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light or laser treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation; Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions); and Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).

## Device Story

JOULE Multi-Platform System is a modular laser/light device for dermatology and surgery. System comprises control console, power supply, cooling system, and various delivery accessories (fiber optics, articulated arms, handpieces, scanners). Device delivers thermal energy via selective photothermolysis or direct tissue ablation/coagulation. Operated by physicians in clinical settings. Input is user-selected laser/light parameters; output is controlled energy delivery to target tissue. Integral thermo-electric cooler protects epidermis during treatment. Clinical benefits include precise tissue interaction, reduced thermal injury, and pain management via cooling. Healthcare providers use device to perform procedures like hair removal, resurfacing, and vascular lesion treatment, directly impacting clinical outcomes through controlled tissue modification.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and indications for use identical to previously cleared predicate devices.

## Technological Characteristics

Modular multi-wavelength laser/light system. Wavelengths include 755nm, 1064nm, 1319nm, 1470nm, 1550nm, 2940nm, and BBL (300-1400nm). Delivery via fiber optics, articulated arms, handpieces, and scanners. Includes integral thermo-electric cooling system. Power supply and control console integrated. System is Class II.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Profile Multi-Platform System ([K081352](/device/K081352.md))
- Candela Family of Q-Switched Alexandrite ([K081324](/device/K081324.md))
- Cynosure Accolade II Laser ([K072868](/device/K072868.md))
- Profile Multi-Platform System ([K070388](/device/K070388.md))
- Profile Multi-Platform System ([K060033](/device/K060033.md))
- Profile 1064 Laser System ([K023881](/device/K023881.md))
- Hoya ConBio RevLite Q-Switched Nd:YAG Laser System ([K083899](/device/K083899.md))
- Alma Harmony XL Multi-Platform Q-Switched Nd:YAG ([K072564](/device/K072564.md))
- Cynosure Affinity QS Q-Switched Nd:YAG Laser System ([K050382](/device/K050382.md))
- Quanta System Polysurge Diode Laser Family ([K083613](/device/K083613.md))
- Xintec Vectra Family of Laser Systems and Accessories ([K082230](/device/K082230.md))
- Biolitec Ceralas D 1470 nm Diode Laser ([K082225](/device/K082225.md))
- Palomar Lux 1440 ([K101506](/device/K101506.md))
- Palomar Lux1540 ([K090195](/device/K090195.md))
- Fraxel SR1500 Laser System ([K070284](/device/K070284.md))
- Palomar LUX2940 ([K100270](/device/K100270.md))
- Cutera Titan ([K042165](/device/K042165.md))
- RevLight Skin Care System ([K042630](/device/K042630.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 长101916

MAR 1 8 2011

.

# Attachment IV

# 510(k) Summary

| Submitter:                            | Sciton, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                              | 925 Commercial Street, Palo Alto, CA 94303                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Phone:                                | (650) 493-9155                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Fax:                                  | (650) 493-9146                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact Person:                       | Jay M. Patel, VP of Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Date Prepared:                        | March 16, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Trade Name:                    | JOULE Multi-Platform System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name:                          | Laser/Light Powered Surgical Device (and Accessories)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name:                  | Laser Surgical Instrument, 21 CFR 878.4810.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Legally Marketed<br>Predicate Device: | 755 nm Laser System<br>K081352: Profile Multi-Platform System<br>Q-Switched 755 nm Alex<br>K081324: Candela Family of Q-Switched Alexandrite<br>K072868: Cynosure Accolade II Laser<br>1064 nm Laser System<br>K070388: Profile Multi-Platform System<br>K060033: Profile Multi-Platform System<br>K023881: Profile 1064 Laser System<br>Q-Switched 1064/532 nm Nd:YAG<br>K083899: Hoya ConBio RevLite Q-Switched Nd:YAG Laser System<br>K072564: Alma Harmony XL Multi-Platform Q-Switched Nd:YAG<br>K050382: Cynosure Affinity QS Q-Switched Nd:YAG Laser System<br>1319 nm Laser System<br>K081352: Sciton Profile Multi-Platform System<br>K070388: Sciton Profile Multi-Platform System<br>K060033: Sciton Profile Multi-Platform System<br>1470 nm Laser System<br>K083613: Quanta System Polysurge Diode Laser Family<br>K082230: Xintec Vectra Family of Laser Systems and Accessories<br>K082225: Biolitec Ceralas D 1470 nm Diode Laser<br>1470 nm Laser System (Fractional)<br>K101506: Palomar Lux 1440<br>K090195: Palomar Lux1540<br>K070284: Reliant Technologies Fraxel SR1500 Laser System<br>1550 nm Laser System (Fractional)<br>K090195: Palomar Lux1540<br>K070284: Reliant Technologies Fraxel SR1500 Laser System<br>2940 nm Laser System (Fractional)<br>K081352: Sciton Profile Multi-Platform System<br>K070388: Profile Multi-Platform System<br>K060033: Profile Multi-Platform System<br>K100270: Palomar LUX2940<br>BBL System<br>K070388: Sciton Profile Multi-Platform System<br>K060033: Sciton Profile Multi-Platform System<br>K042165: Cutera Titan<br>K042630: RevLight Skin Care System |

·

{1}------------------------------------------------

Description of The JOULE Multi-Platform System is a modular, multi-wavelength laser/light system. JOULE Multi-Platform The system uses scanning and focusing optics to deliver a pattern of thermal System: energy to the treatment site. The system consists of control console which houses the power supply, cooling system, fiber optic delivery system and/or articulated arm delivery system with handpiece and/or scanner. Intended Use: The 755 nm Alexandrite system is designed for use in: The ClearScan ALX 755nm Alexandrite laser system with its accessories is indicated for stable long-term, or permanent hair reduction for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions and wrinkles. Q-Switched 755 nm Alexandrite Indications for Use: The JOULE Q-Switched 755 nm Alex Multi-Platform Systems with accessories are indicated for the following uses: Pigmented Lesions Tattoos The 1064 nm Nd:YAG Indications for Use: Coagulation and hemostasis of benign vascular lesions such as, but not limited to, telangiectasias and rosacea. Removal of unwanted hair (for stable, long term or permanent hair reduction) through selective targeting of melanin in hair follicles and for the treatment of Pseudofolliculitis Barbae (PFB). Treatment of facial wrinkles. The JOULE 1064 nm Nd:YAG with fiber delivery is indicated for laser assisted lipolysis. Q-Switched 1064/532 nm Nd: YAG Indications for Use: The JOULE Q-Switched 1064/532 nm Nd:YAG Multi-Platform Systems with accessories are indicated for the following uses; Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis. Specific Indications: 1064 nm wavelength Tattoo Removal (dark ink: blue and black) Nevus of Ota Removal or lightening of hair with or without adjuvant preparation Skin Resurfacing for Acne Scars and Wrinkles Benign Cutaneous Lesions, such as, but not limited to: striae and scars Reduction of Red Pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. 532 nm wavelength Tattoo Removal (light ink: yellow, red, light blue, green) Vascular Lesions including but not limited to: port wine stains/birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi Epidermal Pigmented Lesions including but not limited to: Cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus Other Pigmented Cutaneous Lesions including verrucae, skin tags, keratoses and plaques Skin Resurfacing for Acne Scars and Wrinkles Benign Cutaneous Lesions, such as, but not limited to: striae and scars

{2}------------------------------------------------

Reduction of Red Pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

#### The 1319 nm Nd:YAG Indications for Use:

Treatment of fine lines and wrinkles.

Treatment of atrophic acne scars.

Treatment of mild to moderate inflammatory acne vulgaris.

The JOULE 1319 nm Nd:YAG with fiber delivery is indicated for laser assisted lipolysis.

The JOULE 1319 nm Nd:YAG with fiber delivery is indicated for the treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities, and for treatment of incompetence and reflux of superficial veins in the lower extremity.

#### 1470 nm Indications for Use:

The JOULE 1470 nm Multi-Platform Systems and delivery accessories are intended for delivery of laser light to soft tissue for use in incision, excision, vaporization, ablation, hemostasis or coaqulation of soft tissue. The device is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

The JOULE 1470 nm Multi-Platform Systems with ProFractional handpiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.

#### 1550 nm Indications for Use:

The JOULE 1550 nm Multi-Platform Systems with ProFractional handpiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.

#### 2940 nm Indications for Use:

The JOULE 2940 nm Multi-Platform Systems with delivery accessories are designed for use in surgical applications requiring the excision, incision, ablation, vaporization, and coagulation of soft tissue, and for skin resurfacing. Aesthetic Surgery: Skin resurfacing and treatment of wrinkles. Dermatology & Plastic Surgery: Indications include epidermal nevi, actinic cheilitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision, debulking benign tumors, and decubitis ulcers. It is also used for laser assisted site preparation for hair transplantation.

The JOULE 2940 nm Multi-Platform Systems with ProFractional handpiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing, and ablation and coagulation of soft tissue.

#### BBL (300-1400nm) Indications for Use:

The Joule Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dematology.

It is intended for use for:

- · Mild to moderate inflammatory and pustular inflammatory acne vulgaris.
	- (420nm BP filter, 515nm LP filter, 560 LP filter)
- · The treatment of benign pigmented lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles) (515nm LP filter, 560nm LP filter)
- · The treatment of cutaneous lesions including warts, scars and striae; (515nm LP filter, 560nm LP filter)
- · The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias,

{3}------------------------------------------------

rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations; (560nm LP filter, 590nm LP filter)

- · The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction (590nm LP filter, 640nm LP filter, 695 LP filter)
- Topical heating for the purpose of elevating tissue temperature for the . temporary relief of minor muscle and ioint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, it may also help muscle spasms, minor sprains and strains, and minor muscular back pain. (800 nm LP filter)

The integral thermo-electric cooler is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to: Reduce pain during and/or associated with light or laser treatment (via partial anesthesia from cooling);

Reduce discomfort during and/or associated with light or laser treatment:

Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light or laser treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation;

Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions); and

Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).

The JOULE Multi-Platform System shares the same indications for use, similar design features (including wavelength, laser/light medium and delivery systems, power supply, cooling and control system), functional features (including power output, repetition rate, energy, spot size and fluence), and is therefore substantially equivalent to the above legally marketed predicate devices.

The indications for use are based upon the indications for use for predicate systems. Technologically, the JOULE Multi-Platform System is substantially equivalent to the listed predicate devices. Therefore, the risks and benefits for the JOULE Multi-Platform System are comparable to the predicate devices.

Conclusion:

Safety and

Effectiveness:

Technological

Characteristics:

The JOULE Multi-Platform System shares similar indications for use, design features, and similar functional features as, and therefore is substantially equivalent to, the currently marketed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Sciton. Inc. % Mr. Jay M. Patel VP of Regulatory Affairs 925 Commercial Street Palo Alto, California 94303

MAR 1 8 2011

Re: K101916

Trade/Device Name: JOULE Multi-Platform System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ONG, ORK, ONF Dated: March 17, 2011

Received: March 17, 2011

Dear Mr. Patel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 – Mr. Jay M. Patel

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Aty B. nh
for

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Attachment III

# Statement of Indications for Use

JOULE Multi-Platform System

510(k) Number (if known): K101916

Indications for Use:

Device Name:

#### The 755 nm Alexandrite system is designed for use in:

The ClearScan ALX 755nm Alexandrite laser system with its accessories is indicated for stable long-term, or permanent hair reduction for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions and wrinkles.

# Q-Switched 755 nm Alexandrite Indications for Use:

The JOULE Q-Switched 755 nm Alex Multi-Platform Systems with accessories are indicated for the following uses: Pigmented Lesions Tattoos

# The 1064 nm Nd:YAG Indications for Use:

Coagulation and hemostasis of benign vascular lesions such as, but not limited to, telangiectasias and rosacea.

Removal of unwanted hair (for stable, long term or permanent hair reduction) through selective targeting of melanin in hair follicles and for the treatment of Pseudofolliculitis Barbae (PFB).

Treatment of facial wrinkles.

The JOULE 1064 nm Nd:YAG with fiber delivery is indicated for laser assisted lipolysis.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ 
(Per 21CFR801)

OR Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Udelon for maxim

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 5

:10(k) Number K101916

{7}------------------------------------------------

#### Q-Switched 1064/532 nm Nd: YAG Indications for Use:

The JOULE Q-Switched 1064/532 nm Nd:YAG Multi-Platform Systems with accessories are indicated for the following uses:

Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis.

Specific Indications:

1064 nm wavelength

Tattoo Removal (dark ink: blue and black)

Nevus of Ota

Removal or lightening of hair with or without adjuvant preparation Skin Resurfacing for Acne Scars and Wrinkles

Benign Cutaneous Lesions, such as, but not limited to: striae and scars Reduction of Red Pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

#### 532 nm wavelength

Tattoo Removal (light ink: yellow, red, light blue, green)

Vascular Lesions including but not limited to: port wine stains/birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi

Epidermal Pigmented Lesions including but not limited to: Cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus

Other Pigmented Cutaneous Lesions including verrucae, skin tags, keratoses and plaques

Skin Resurfacing for Acne Scars and Wrinkles

Benian Cutaneous Lesions, such as, but not limited to: striae and scars Reduction of Red Pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

# The 1319 nm Nd:YAG Indications for Use:

Treatment of fine lines and wrinkles.

Treatment of atrophic acne scars.

1

Treatment of mild to moderate inflammatory acne vulgaris.

The JOULE 1319 nm Nd:YAG with fiber delivery is indicated for laser assisted lipolysis.

The JOULE 1319 nm Nd:YAG with fiber delivery is indicated for the treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities, and for treatment of incompetence and reflux of superficial veins in the lower extremity.

| Prescription Use (Per 21CFR801) | X | OR Over-The-Counter Use |  |
|---------------------------------|---|-------------------------|--|
|---------------------------------|---|-------------------------|--|

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) | <div style="text-align:left">Nekeda for nxm</div> |
|--------------------------------------------------------|---------------------------------------------------|
|                                                        | (Division Sign-Off)                               |
|                                                        | Division of Surgical, Orthopedic,                 |
| Page 2 of 5                                            | and Restorative Devices                           |

· 510(k) Number

{8}------------------------------------------------

#### 1470 nm Indications for Use:

The JOULE 1470 nm Multi-Platform Systems and delivery accessories are intended for delivery of laser light to soft tissue for use in incision, excision, vaporization, ablation, hemostasis or coaqulation of soft tissue. The device is indicated for the treatment of reflux of the saphenous veins associated with varicose veins and varicosities.

The JOULE 1470 nm Multi-Platform Systems with ProFractional handbiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.

#### 1550 nm Indications for Use:

The JOULE 1550 nm Multi-Platform Systems with ProFractional handbiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.

#### 2940 nm Indications for Use:

The JOULE 2940 nm Multi-Platform Systems with delivery accessories are designed for use in surgical applications requiring the excision, incision, ablation, vaporization, and coagulation of soft tissue, and for skin resurfacing. Aesthetic Surgery: Skin resurfacing and treatment of wrinkles. Dermatology & Plastic Surgery: Indications include epidermal nevi, actinic cheilitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision, debulking benign tumors, and decubitis ulcers. It is also used for laser assisted site preparation for hair transplantation.

The JOULE 2940 nm Multi-Platform Systems with ProFractional handpiece and delivery system is intended for use in dermatological procedures requiring skin resurfacing, and ablation and coagulation of soft tissue.

Nifredde formxn

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101916

Prescription Use × (Per 21CFR801)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

{9}------------------------------------------------

### BBL (300-1400nm) Indications for Use:

The JOULE Multi-Platform Systems and Accessories are indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology.

It is intended for use for:

- · Mild to moderate inflammatory and pustular inflammatory acne vulgaris.
(420nm BP filter, 515nm LP filter, 560 LP filter)

- · The treatment of benign pigmented lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles) (515nm LP filter, 560nm LP filter)
- · The treatment of cutaneous lesions including warts, scars and striae:

(515nm LP filter, 560nm LP filter)

- · The treatment of benign cutaneous vascular lesions, including port wine stains, hemanqiomas, facial, truncal and leg telangjectasias. rosacea, erythema of rosacea, angiomas and spider angiomas. poikiloderma of Civatte, leg veins and venous malformations; (560nm LP filter, 590nm LP filter)
- · The removal of unwanted hair from all skin types, and to effect stable long-term, or permanent, hair reduction (590nm LP filter, 640nm LP filter, 695 LP filter)
- · Topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, it may also help muscle spasms, minor sprains and strains, and minor muscular back pain. (800 nm LP Filter)

The integral thermo-electric cooler is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to:

Reduce pain during and/or associated with light or laser treatment (via partial anesthesia from cooling);

| Prescription Use<br>(Per 21CFR801) | X | OR |
|------------------------------------|---|----|
|------------------------------------|---|----|

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|                                   | <span style="text-decoration: overline;">Nuk K. Ogles</span> Concurrence of CDRH, Office of Device Evaluation (ODE) |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------|
| (Division Sign-Off)               |                                                                                                                     |
| Division of Surgical, Orthopedic, |                                                                                                                     |
| and Restorative Devices           |                                                                                                                     |
|                                   | Page 4 of 5                                                                                                         |
| 510(k) Number                     | <span style="text-decoration: overline;">K101916</span>                                                             |

{10}------------------------------------------------

Reduce discomfort during and/or associated with light or laser treatment;

Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light or laser treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation;

Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions); and

Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).

Neil R.P. Isle for mkr
Division of 88

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101916

Prescription Use X (Per 21CFR801)

OR Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda-staging.innolitics.com/device/K101916](https://fda-staging.innolitics.com/device/K101916)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).