The DSC/ALF Spinal System is a vertebral body replacement system intended for use in the thoracolumbar spine (TI-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The DSC/ALF Spinal System is to be used with supplemental fixation. Specifically, the DSC/ALF spinal system is to be used with the anterior lateral plate that is included in the system, and may also be used with the Sintea Biotech PLS Spinal System. Additionally, the DSC/ALF spinal system is intended to be used with autograft.
Device Story
Vertebral body replacement (VBR) system for thoracolumbar spine; replaces collapsed/damaged/unstable vertebrae due to tumor or trauma. System comprises hollow internal module, external module, and anterior lateral fixation (ALF) plate. Internal module slides within external module; locked via two screws into tooth rails on internal cage. Endplates optimize contact with vertebral bodies for primary stability. ALF plate connects to anterior side of vertebra via two screws; plates connected by rod. Used by surgeons in clinical settings. Provides structural support; requires supplemental fixation and autograft. Benefits patient by restoring spinal stability.
Clinical Evidence
Bench testing only. Testing performed per ASTM F2077 and ASTM F2267. Results: Static compression, static torsion, subsidence, and fatigue compression showed better results than the predicate; fatigue torsion results were comparable to the predicate.
Technological Characteristics
Materials: Medical grade titanium alloy (Ti-6Al-4V) per ASTM F136. Design: Modular VBR with sliding internal/external modules and anterior lateral fixation plate. Mechanical locking via screws into tooth rails. Standards: ASTM F2077, ASTM F2267.
Indications for Use
Indicated for patients with collapsed, damaged, or unstable vertebral bodies in the thoracolumbar spine (T1-L5) resulting from tumor or trauma (fracture). Requires use with supplemental fixation (included anterior lateral plate or Sintea Biotech PLS Spinal System) and autograft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
K101886
JUL 21 2011
sintea PLUSTEK
DSC/ALF Spinal System 510(k) Summary June 2010
Sintea Plustek, LLC. Company: 407 Lincoln Rd. Suite 10L Miami Beach, FL 33139 (305) 673-6226
Proprietary Trade Name: DSC/ALF System II.
> 888.3060 Regulation Number: Spinal Intervertebral Body Fixation Orthosis Regulation Name: Product Code: MQP, KWQ
### III. Product Description
Sintea Plustek, LLC. 407 Lincoln Rd, Suite 10L
Miami Beach, FL 33139
P. 305-673-6226 F. 305-673-3312
l.
This 510(k) consists on the submission to add additional lengths and cross-sections, an improved locking mechanism, cage body redesign, and end-plate charges to a previously cleared system. The DSC/ALF Spinal System provides two basic components: A VBR (DSC) and an anterior lateral fixation plate (ALF). The DSC has a hollow internal module that can slide in relation to an external module, which enables the former to be locked in the most suitable position by means of two screws. These screws are inserted into two tooth rails on the internal cage, on both the anterior and posterior part, and mechanically lock the two modules. Both modules have endplates that optimize the contact with the vertebral bodies between which the device is inserted, thus ensuring better primary stability. The cages present a symmetrical geometry in order to make device placement simpler for the surgeon. The ALF is composed of two plates that connect to the anterior side of a vertebra by two screws. The two plates are connected by a rod. All components of the DSC/ALF Spinal System are made of medical grade titanium alloy (Ti-6Al-4V) as described by ASTM Standard F136.
## IV. Indications
The DSC/ALF Spinal System is a vertebral body replacement system intended for use in the thoracolumbar spine (TI-LS) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The DSC/ALF Spinal System is to be used with supplemental fixation. Specifically, the DSC/ALF spinal system is to be used with the anterior lateral plate that is included in the system, and may also be used with the Sintea Biotech PLS Spinal System. Additionally, the DSC/ALF spinal system is intended to be used with autograft.
## Performance Data - Overview V.
1. Static compression - This test on the new device achieved better results than those of the predicate device.
2. Static torsion - This test on the new device achieved better results than those of the predicate device.
{1}------------------------------------------------
Sintea Plustek, LLC. 407 Lincoln Rd, Suite 10L Miami Beach, FL 33139 P. 305-673-6226 F. 305-673-3312
Image /page/1/Picture/1 description: The image shows the logo for sinteaPLUSTEK. The logo consists of the word "sinteaPLUSTEK" in a simple, sans-serif font. The word is enclosed in a curved shape that resembles a parenthesis or a crescent moon. The logo is simple and modern, and the use of a single color makes it easy to reproduce.
3. Subsidence – This test on the new device achieved better results than those of the predicate device.
4. Expulsion - This test on the new device achieved comparable results to those of the predicate device.
5. Fatigue compression - This test on the new device achieved better results than those of the predicate device.
6. Fatigue torsion - This test on the new device achieved comparable results to those of the predicate device.
Testing standards followed: ASTM F2077 ASTM F2267
## VI. Substantial Equivalence
Sintea Plustek, LLC. believes that the additions to the DSC/ALF Spinal System are substantially equivalent to the Sintea Biotech's DSC/ALF Spinal System (K070181) with respect to functional design, indications for use, principles of operation, performance, and materials.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an emblem of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Sintea Plustek, LLC % Ms. Danielle Wernikowski 407 Lincoln Road, Suite 10L Miami Beach, FL 33139
JUL 2 1 2011
Re: K101886
Trade/Device Name: DSC/ALF Spinal System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP, KWQ Dated: July 06, 2011 Received: July 11, 2011
Dear Ms. Wernikowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Danielle Wernikowski
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
E. I. Keith
o Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Sintea Plustek, LLC. 407 Lincoln Rd, Suite 10L Miami Beach, FL 33139 P. 305-673-6226 F. 305-673-3312
Image /page/4/Picture/1 description: The image shows the logo for "sintea PLUSTEK". The logo consists of the company name written in a simple, sans-serif font. The entire name is enclosed within a double-lined oval shape, adding a visual element to the design. The text and oval are black.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# Indications for Use
K101886 510(k) Number (if known):
Device Name: DSC/ALF_Spinal System
Indications for Use:
The DSC/ALF Spinal System is a vertebral body replacement system intended for use in the thoracolumbar spine (TI-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The DSC/ALF Spinal System is to be used with supplemental fixation. Specifically, the DSC/ALF spinal system is to be used with the anterior lateral plate that is included in the system, and may also be used with Sintea Biotech's PLS Spinal System. Additionally, the DSC/ALF spinal system is intended to be used with autograft.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use _
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
signature
(Division Sign-Off) Dision of Surgical, Orthopedic. and Restorative Devices
510(k) Number K101886
sheet 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.