K101874 · GE Healthcare · IYN · Oct 5, 2010 · Radiology
Device Facts
Record ID
K101874
Device Name
GE LOGIQ P6/P6 PRO
Applicant
GE Healthcare
Product Code
IYN · Radiology
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is a general-purpose ultrasound system. Specific clinical applications and exam types include: Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal (TE); Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, vascular and neuro).
Device Story
Mobile ultrasound console with keyboard, controls, and color LCD display; uses electronic-array transducers to acquire, process, and display real-time digital ultrasound data. Operates in various modes including B, M, PW/CW Doppler, Color/Power Doppler, and Elastography. Used in clinical settings (radiology, cardiology, vascular) by trained healthcare professionals. Provides real-time imaging and analysis for diagnostic evaluation. Output displayed on integrated monitor for clinician interpretation to support clinical decision-making and patient diagnosis.
Clinical Evidence
No clinical studies were required to support substantial equivalence. The device was evaluated via bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing.
Technological Characteristics
Mobile console with color LCD; electronic-array transducers; supports B, M, Doppler (PW/CW/Color/Power), and Elastography modes. Materials are biocompatible. System conforms to medical device safety standards for electrical, thermal, and electromagnetic safety. Connectivity includes digital acquisition and display.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, urological, transesophageal, transrectal, transvaginal, and intraoperative applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' in a stylized font. The letters are intertwined and appear to be in a bold, sans-serif typeface. The logo is surrounded by a slightly jagged or pixelated edge, suggesting it may be a low-resolution image or a scan of an older print.
SIO(k) Premarket Notification Submission
# OCT - 5 2010 K101874
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| <b>Date:</b> | June 28, 2010 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Submitter:</b> | GE Healthcare [GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC] 9900 Innovation Dr Wauwatosa, WI 53226 |
| <b>Primary Contact Person:</b> | Bryan Behn Regulatory Affairs Manager GE Healthcare, GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC.] T:(414)721-4214 F:(414)918-8275 |
| <b>Secondary Contact Person:</b> | Jim Turner Regulatory Affairs Manager America's Service GE Healthcare. GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC T:(262) 544-3359 F:(414)908-9225 |
| <b>Device: Trade Name:</b> | LOGIQ P6/P6 Pro BT11 Ultrasound System |
| <b>Common/Usual Name:</b> | LOGIQ P6/P6 Pro BT11 |
| <b>Classification Names:</b> | Class II |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| <b>Predicate Device(s):</b> | K073297 GE LOGIQ P6 Ultrasound System K092271 GE LOGIQ E9 Diagnostic Ultrasound System |
| <b>Device Description:</b> | The subject device consists of a mobile console with keyboard, specialized controls, a color video LCD display with electronic- array transducers. It has the same general appearance, dimensions and weight as the unmodified device, it is a Track 3 general- purpose imaging and analysis system providing real-time digital acquisition, processing and display capability intended for general radiology imaging and evaluation with some cardiology and vascular applications. |
| <b>Intended Use:</b> | The device is a general-purpose ultrasound system. Specific clinical applications and exam types include: Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are depicted in black, creating a strong contrast against the white background.
SIO(k) Premarket Notification Submission
thyroid); Neonatal Cephalic; Adult Cophalic; Cardiac (adult and Peripheral Vascular pediatric): (PV); Musculo-skeletal Conventional and Superficial; Urology (including prossate); Transesophageal (TE); Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracie, vascular and neuro.
#### The LOGIQ P6 BT11 employs the same fundamental scientific Technology technology as its predicate devices.
#### Determination of Substantial Equivalence:
### Summary of Non-Clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIO P6/P6 Pro and its applications comply with voluntary standards as detailed in Section 9, 11 and
17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- . Requirements Reviews
- Design Reviews .
- Testing on unit level (Module verification) .
- Integration testing (System verification) ●
- Final Acceptance Testing (Validation) .
- Performance testing (Verifieation)
- Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission. LOGIQ P6/P6 Pro, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the LOGIQ P6/P6 Pro to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle or bird with three curved lines representing its wings or body. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters. The text is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
OCT - 5 2010
Mr. Bryan Behn Regulatory Affairs Manager GE Healthcare GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC 9900 Innovation Dr. WAUWATOSA WI 53226
Re: K101874
Trade/Device Name: LOGIQ P6/P6 Pro BT11 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: July 26, 2010 Received: September 10, 2010
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the LOGIQ P6/P6 Pro BT11 Ultrasound, System as described in your premarket notification:
# Transducer Model Number
| i12L | 3Sp |
|--------|-------|
| ML6-15 | 5Sp |
| 4D8C | 6Tc |
| 4D5C-L | 11L |
| 3CRF | BE9CS |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 4
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (301) 796-6881.
Sincerely yours,
Mahal O'Hara Lin
David G. Brown. Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a black and white drawing of a circular object. The object has a swirling pattern on its surface, with darker areas contrasting against lighter ones. The overall impression is that of a stylized or abstract representation of a sphere or similar rounded form.
K101874
0CT - 5 2010
510(k) Number (if known): K101874
Device Name: LOGIQ P6/P6 Pro BT11
Indications for Use:
The device is a general-purpose ultrasound system. Specific elinical applications and exam types include: Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal (TE); Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracie, vascular and neuro).
Prescription Use X AND/OR (Part 2) CFR 801 Subpart D)
Over-The-Counter Use_NA (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Coneurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Michael D. O'Rear for David b. Brown
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K101874
Page 1 of 1
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' in a stylized font. The letters are intertwined and appear to be in a sans-serif typeface. The logo is surrounded by a circular border.
# Diagnostic Ultrasound Indications for Use Form
#### GE LOGIQ P6/P6 Pro Ultrasound System
# Inlended Use: Diagnostic ultrasound maging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | P |
| Abdominal" | P | P | P | P | P | P | P | P | P | P | P,[6] |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P,[6] |
| Small Organ 2 | P | P | P | | P | P | P | P | P | P | [6] |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | |
| Cardiac5 | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal<br>Conventional | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal<br>Superficial | P | P | P | | P | P | P | P | P | P | [6] |
| Other4 | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type. Means of Access | | | | | | | | | | | |
| Transesophageal | P | P | P | P | P | P | P | P | P | P | |
| Transrectal | P | P | P | | P | | P | P | P | P | P |
| Transvaginal | P | P | P | | P | | P | P | P | P | P |
| Transuretheral | | | | | | | | | | | |
| Intraoperative1 | P | P | P | | P | | P | P | P | P | [6] |
| Intraoperative<br>Neurological | P | P | P | | P | P | P | P | P | P | [6] |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| | | | | | | | | | | | |
* new indication. P = previously cleared by FDA: E = added under Appendix E Notes.
[1] Abdominal indudes renal, GYN/Pelvic
(2) Small organ includes breast, testes, and thyoid.
[3] Cardiac is Adult and Pediatric
[4] Other use includes Urology/Prostate
[5] Intraoperative indudes abdominal, thoracic (cardiac), and vascular (PV).
[6] Elastography Imaging
[1] Combined modes are BiM, BiColor M, BiPWD or CWD, BlColoriPWD or CWD, BiPower/PWD.
(""| Other mode is 4D / Realtime 3D
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IE NEEDED)
Concurrance of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 21 C(FR 801.109)
Mark A. O'Dea for David 6. Brown
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K101874
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo of General Electric (GE). The logo is a circular shape with the letters "GE" intertwined in the center. The logo is black and white, with the letters "GE" being white and the background being black. The logo is a well-known symbol of the General Electric company.
510(k) Premarket Notification Submission
# Diagnostic Uitrasound Indications for Use Form GE LOGIQ P6 with i121, Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | | | | | P | | P | P | P | P | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ[2] | P | P | P | | P | | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | P | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative[5] | P | P | P | | P | | P | P | P | P | |
| Intraoperative<br>Neurological | E | E | E | | E | | E | E | E | E | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication: P = previously desred by FDA (Logiq V K030934); E = added under Approadix H
Notes. [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, lesles, and thyroid.
[3] Cardiac is Adult and Pediatric
[4] Other use includes Urology(Prostate
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV),
1") Combined modes are BIM, B/Color M, B/PWD or CWO. B/Color/PWD or CWD, B/Power/PWD.
[™] Other mode is 4D / Reall:me 3D
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IE MEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 21 CFR 801,109)
Michael D'Them for David G Brown
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K101874/
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white, with the letters and border appearing in black against a white background.
Diagnostic Ultrasound Indications for Use Form
# GE LOGIO P6 with ML6-15 Transducar
Intended Use: Diagnostic with as und imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|-------------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal1 | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] |
| Small Organ2 | P | P | P | | P | P | P | P | P | P | [6] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac3 | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other4 | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative5 | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Noles: [1] Abdominal includes renal. GYN/Pelvic
[7] Small organ includes breast, testes, and thyroid.
(J) Cardiac Is Adult and Pediatac
[4] Other use includes Uralogy: Prosiste
[5] intraoperative includes abdominal. thoracic (canthac), and vascular (PV)
[6] Elastography unaging
[0] Elibstograph| This and
[ 7] Ovner mode is 4D / Reallime JD
ISTEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOIRER FAGE IE NEEDEDI Cancurrence of CORH. Office of Device Evaluation (DDE)
Prescription User (Per 21 CFR 801.109)
Michael D'Ollone for David G. Brown
(Division Sign-Off) (Division of Radiological Devices
510K. K161874
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' in a stylized, intertwined font. The letters are surrounded by a textured or dotted pattern, giving the logo a vintage or slightly distressed appearance. The overall design is simple yet recognizable.
#### Diagnostic Ultrasound Indications for Use Form
#### GE LOGIO P6 with 4D8C Transducer
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|--------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other- |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominall | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | P |
| Small Organl | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Otherl | | | | | | | | | | | |
| Exam Type Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperativeli | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| N = new indication; P - previously cleared by FDA (LOGIQ 9 Ku61129); E = added under Appendix F<br>Notes (1) Abdominal includes renal, GYN/Pelvic | | | | | | | | | | | |
(2) Small organ includes breast, testes, and thyrou
(3) Cardian is Adult and Pediatine
[4] Other use includes Urology/Prostate
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
( 1 Combined modes are 8/M, B/Color M, B/PND ar CWD, B/Color/PWD or CWD, B/Power/PWD
[*] Other mode is 4D / Reallime 3D
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANDIHER PAGE IE NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 21 CFR 801.109)
Murdock D'Ochler for David 6 Brown
(District Sign Off)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K101874
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with the letters 'GE' inside a circle. The letters are stylized and intertwined. The logo is black and white and appears to be slightly distressed or aged.
# Diagnostic Ultrasound Indications for Use Form
# GE LOGIO P6 with 4D5C-L Transducer
#### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|-----------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>** |
| Ophthalmic | E | E | E | | E | E | E | E | E | E | E |
| Fetal / Obstetrics | E | E | E | | E | E | E | E | E | E | E |
| Abdominal(1) | E | E | E | | E | E | E | E | E | E | E |
| Pediatric | E | E | E | | E | E | E | E | E | E | E |
| Small Organ(2) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac(3) | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (4) | E | E | E | | E | E | E | E | E | E | E |
| Exam Type. Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative(5) | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic<br>N = new indication, P = previously cleared by FDA; E = added under Appendix E | | | | | | | | | | | |
[1] Abdominal includes renal, GYN!Petric Noles.
[2] Small organ includes breast, testes, and thyroid.
[3] Cardiac is Adult and Pediatric
[4] Other use includes Urology:Proslate
15) intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
(v) miraoperative linduss Blashinal, the book (conode). 6-x) - Institution (r 7).
["] Combined modes are B/M, B/Color M. B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD
[**] Other mode is 40 / Realtime 3D
(BLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Preseription User (Per 21 CFR 801.109)
Michael D. O'Hern for David G. Brown
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K101874
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a black and white logo. The logo is a circular shape with the letters 'GE' intertwined in the center. The letters are stylized and appear to be in a cursive font. The logo is surrounded by a dashed line, suggesting it may be part of a larger image or document.
5 10(k) Premarket Notification Submission
#### Diagnostic Ultrasound Indications for Use Form
#### GE LOGIQ P6 with 3CRF Transducer
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | |
| Abdominal(1) | P | P | P | | P | P | P | P | P | P | |
| Pediatric | | | | | | | | | | | |
| Small Organ(2) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac(3) | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other(4) | P | P | P | | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative(5) | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| N = new indication; P - previously cleared by FDA(LOGIQ F9 KU92271); E = added under Appendix E<br>Notes: (1) Abdominal includes renal, GYN/Pelvic | | | | | | | | | | | |
1. Abdominal includes renal, GYN/pelvic
(2) Small organ includes breast, lestes, and thyroid.
(3) Cardiac is Adull and Pediatric
14] Other use includes UrologyProsiate
15) Intraoperative includes abdominal. thorecic (cardiac), and vascular (PV)
[1] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
1™| Other mode is 4D / Realtime 3D
(P) EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of Device Evakuation (ODE)
Preseription User (Per 2) CFR 801.109)
Michael D. O'Hern for David C. Brown
Division of Rediological Devices
510K K101874
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized manner. The logo is enclosed within a dotted circle.
# 510(k) Premarket Notification Submission
# Diagnostic Ultrasound Indications for Use Form GE LOGIO P8 with 3Sp Transducer
Intended Use: Diagnostic ultrasound imaging or fluid thow analysis of the human body as lollows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal1 | E | E | E | E | E | E | E | E | E | E | |
| Pediatric | | | | | | | | | | | |
| Small Organ2 | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | E | E | E | E | E | E | E | E | E | E | |
| Cardiac3 | E | E | E | E | E | E | E | E | E | E | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other4 | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative5 | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
previously clear cd by ( ation; P DA: I:
Notes: (1) Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, and thyroid.
[3] Cardrac is Adult and Pediatric.
[4] Other use includes UrologyProstate
[5] Intracperative includes abdominal, thoracic (cardiac), and vascular (PV),
[1] Combined modes are BiM. BIColor M. BiPWD or CWD, B/Color/PWD or CWD, B/Power/PWD
| " | Other mode is 40 / Realume 30
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON PHOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 21 L'FR 801.109)
Fletcher D. Hannon G- David b. Brown
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K161874
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized manner. The logo is surrounded by a circular pattern, giving it a distinctive and recognizable appearance.
510(k) Premarket Notification Submission
# Diagnostic Ultrasound Indications for Use Form
#### GE LOGIO P6 with 5Sp Transducer
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|-------------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal1 | E | E | E | E | E | E | E | E | E | E | E |
| Pediatric | E | E | E | E | E | E | E | E | E | E | |
| Small OrganPl | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | E | E | E | E | E | E | E | E | E | E | |
| Cardiac3 | E | E | E | E | E | E | E | E | E | E | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other14 | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative5 | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| N - new indication: P - previously cleared by FDA: E = added under Appendix E | | | | | | | | | | | |
Notes. { } | Abdominal includes renal. GYN/Pelvic
{2] Small organ includes breast. testes, and thyroid.
(3) Cardiac is Adult and Pedialiic
[4] Other use includes Urology/Prossate
(4) Carlin LES Visions Orders abdominal, theracic (cardiac), and vascular (PV).
[1] Combined modes are B/M, B/Color M, B/PWD or CWD, B/ColoriPWD or CWD, B/PoweriPWD.
(**) Other mode is 4D / Reallime 3D
IPLE ASE DO NOT MRITE BELOW THIS LINE - CONTINUE ON PARTINER PAGE IE MEEDEDI Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 2) CFR 801.109)
Michael O'Dell for David 6. Brown
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
s10K K101874
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows a black and white logo. The logo is circular and contains a stylized monogram of the letters 'GE'. The monogram is composed of curved lines and swirls, creating an abstract design. The logo is surrounded by a thin, slightly jagged border.
#### Diagnostic Ultrasound Indications for Use Form
#### GE LOGIQ P6 with 6Tc Transducer
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as tollows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------|
| Anatomy/Region of<br>Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes' | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.