URGENT PC STIMULATOR; URGENT PC LEAD SET

K101847 · Uroplasty, Inc. · NAM · Oct 21, 2010 · Gastroenterology, Urology

Device Facts

Record IDK101847
Device NameURGENT PC STIMULATOR; URGENT PC LEAD SET
ApplicantUroplasty, Inc.
Product CodeNAM · Gastroenterology, Urology
Decision DateOct 21, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5310
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Urgent PC Neuromodulation System is intended to treat patients with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence.

Device Story

Urgent PC Neuromodulation System is a minimally invasive device for Percutaneous Tibial Nerve Stimulation (PTNS); delivers retrograde access to sacral nerve via percutaneous electrical stimulation of tibial nerve. System comprises battery-operated external pulse generator (Urgent PC Stimulator) and single-use needle electrode lead set (Urgent PC Stimulation Lead Set). Used in clinical settings; operated by healthcare professionals. Stimulator transfers electrical current through needle electrode to tibial nerve. Treatment aims to alleviate OAB symptoms; provides non-surgical neuromodulation alternative for patients.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use identical to the predicate device.

Technological Characteristics

Percutaneous tibial nerve stimulator; consists of battery-operated external pulse generator and single-use needle electrode lead set. Delivers electrical stimulation to tibial nerve for retrograde sacral nerve access. Class II device (21 CFR 8876.5310, Product Code NAM).

Indications for Use

Indicated for patients with Overactive Bladder (OAB) experiencing urinary urgency, urinary frequency, and urge incontinence.

Regulatory Classification

Identification

A nonimplanted, peripheral electrical continence device is a device that consists of an electrode that is connected by an electrical cable to a battery-powered pulse source. The electrode is placed onto or inserted into the body at a peripheral location and used to stimulate the nerves associated with pelvic floor function to maintain urinary continence. When necessary, the electrode may be removed by the user.

Special Controls

*Classification.* Class II, subject to the following special controls:(1) That sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter. (2) That the labeling must bear all information required for the safe and effective use of the device as outlined in § 801.109(c) of this chapter, including a detailed summary of the clinical information upon which the instructions are based.

In addition to the general controls of the act, the Percutaneous SANS Device is subject to the following special controls: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) Clinical data are necessary. (3) The labeling must include (a) warnings indicating the need to evaluate the patient during treatment for pain or skin irritation/inflammation; (b) precautions indicating the need to read and understand all directions before using this device; (c) specific instructions to describe proper patients set-up prior to the start of treatment, including proper placement of the electrodes; (d) stimulation parameters and duration and frequency of treatments in accordance with the protocol used during clinical trial; and (e) a summary of the clinical data collected including a listing of adverse events reported.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Uroplasty, Inc. Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System #### Page 7 of 36 K101847 ## Section 5: 510(k) Summary | Date Prepared | October 8, 2010 | | |------------------|--------------------------------------------------------------------------------------------------------|--------------| | New Device Name | Urgent® PC Neuromodulation System | OCT 2 1 2010 | | Predicate Device | Urgent® PC Neuromodulation System (K071822) | | | Contact | Uroplasty, Inc.<br>5420 Feltl Road<br>Minnetonka, MN 55343 USA<br>Tel: 952.426.6140; Fax: 952.426.6199 | | #### Intended Use i The Urgent PC Neuromodulation System is intended to treat patients with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence. ## Device Description The Urgent® PC Neuromodulation System is a minimally invasive neuromodulation system designed to deliver retrograde access to the sacral nerve through percutaneous electrical stimulation of the tibial nerve. The method of treatment is referred to as Percutaneous Tibial Nerve Stimulation (PTNS). The Urgent PC Neuromodulation System is a combination of the Urgent PC Stimulator and the Urgent PC Stimulation Lead Set. The Urgent PC Stimulator is a battery-operated external pulse generator and is designed, constructed, and manufactured for multiple use, only in conjunction with the Urgent PC Stimulation Lead Set. The Urgent PC Stimulation Lead Set transfers the electrical current from the Urgent PC Stimulator to the tibial nerve via the Needle Electrode. The entire Stimulation Lead Set is intended for single use only and is not to be reused. #### Indications Statement The primary difference between the new and predicate devices is the wording of indications statement. The new device indications statement incorporates the overactive bladder terminology listed in the intended use. This difference in indication statement represents an update in wording rather than a change in the device patient population. #### Technological Characteristics The new and predicate devices are technologically the same; they are percutaneous tibial nerve stimulator devices with lead sets intended to deliver retrograde access to the sacral nerve for the overactive bladder symptoms of urinary urgency, urinary frequency, and urge incontinence. Both devices have the same intended use and principles of action. Only the wording of the indications statement differs between this device and the predicate. {1}------------------------------------------------ Uroplasty, Inc. Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System # Performance The Urgent PC Neuromodulation System and Stimulation Lead Set allows for the successful performance of the product's intended use. ## Conclusion The subject device of the 510(k) submission is substantially equivalent to the previously cleared Urgent PC Neuromodulation System by Uroplasty, Inc. (K071822). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized representation of a human figure, composed of three curved lines, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN" is arranged vertically along the left side of the logo, with "SERVICES" completing the circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002 Mr. Michael Morrell, RAC Director of Regulatory Affairs and Quality Assurance Uroplasty. Inc., 5420 Feltl Road MINNETONKA MN 55343 # OCT 2 1 2010 Re: K101847 Trade/Device Name: Urgent® PC Neuromodulation System Regulation Number: 21 CFR 8876,5310 Regulation Name: Non-implanted, peripheral electrical continence device Regulatory Class: II Product Code: NAM Dated: October 8, 2010 Received: October 14, 2010 Dear Mr. Morrell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate . commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {3}------------------------------------------------ Page 2 adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Hubert Lemmer MD Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Uroplasty, Inc. Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System Page 6 of 36 K101847 ## Section 4: Indications for Use Statement 510(k) Number (if known): K101847 OCT 2 1 2010 Device Name: Urgent® PC Neuromodulation System The Urgent PC Neuromodulation System is intended to treat patients Indications for Use: with Overactive Bladder (OAB) and associated symptoms of urinary urgency, urinary frequency, and urge incontinence. Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) HReinu (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K101847 CONFIDENTIAL
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...