The SONOACE R3 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Neonatal Cephalic, Trans-rectal, trans-vaginal, Muscular-Skeletal (conventional, Superficial), Cardiac Adult and Peripheral-vessel.
Device Story
SONOACE R3 is a portable, software-controlled diagnostic ultrasound system. It acquires ultrasound data via transducers and displays it in B-mode, M-mode, color/power Doppler, PW/CW spectral Doppler, harmonic imaging, and freehand 3D. Operated by healthcare professionals in clinical settings, the system provides anatomical measurements and analysis packages to assist in diagnosis. It features real-time acoustic output display (mechanical and thermal indices). The system processes received echoes to generate on-screen images of structures and fluid flow. Output is used by clinicians to guide clinical decision-making, such as biopsy procedures or monitoring follicle development. Benefits include non-invasive visualization of internal structures and blood flow.
Clinical Evidence
Bench testing only. The device complies with safety standards including UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD-2, NEMA UD-3, IEC 61157, and ISO 10993-1 for biocompatibility.
Technological Characteristics
Portable, software-controlled ultrasound system. Transducers: C2-4/20, EC4-9, CN2-8, L5-12/60, CN4-9, LN5-12/40. Imaging modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic, 3D. Safety standards: UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD-2/UD-3, IEC 61157, ISO 10993-1. Connectivity: Standard ultrasound interface.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid analysis in fetal, abdominal, pediatric, small organ (thyroid, parathyroid, breast, scrotum, penis), neonatal cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional, superficial), adult cardiac, and peripheral vessel applications. Includes biopsy guidance.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
SONOACE PICO Diagnostic Ultrasound System (K061213)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 16 2010
6101829
510(k) Premarket Notification
SONOACE R3 Diagnostic Ultrasound System
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
# 1. Submitter's Information: 21 FR 807.92(a)(1)
MEDISON CO., LTD. 1003, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea
Contact Person: Mr. Kyung-Am, Shim Regulatory Affairs Manager
Telephone: 82.2.2194.1381 Facsimile: 82.2.2194.1399
Data Prepared: April 6, 2010
### 2. Name of the device:
| Common/Usual Name: | | |
|----------------------------------------------|-----------|--------------|
| Diagnostic Ultrasound System and Accessories | | |
| Proprietary Name: | | |
| SONOACE R3 Diagnostic Ultrasound System | | |
| Classification Names: | FR Number | Product Code |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasound Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
# 3. Identification of the predicate or legally marketed device:
K081676, SONOACE X6 Diagnostic Ultrasound System K061213, SONOACE PICO Diagnostic Ultrasound System
510(k) Summary / Statement Certification
#### ATTACHMENT 1
{1}------------------------------------------------
### 4. Device Description:
The SONOACE R3 is a general purpose, portable, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, color Doppler imaging, power Doppler imaging, PW/CW spectral Doppler mode, Harmonic imaging, Freehand 3D imaging mode or as a combination of these modes. The SONOACE R3 also gives the operator the ability to measure annomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The SONOACE R3 has real time acoustic output display with two basic indices, a mechanical index and a thermised index, which are both automatically displayed.
The SONOACE R3 has been designed to meet the following product safety standards:
- UL 60601-1, Safety requirements for Medical Equipment
- CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
- IEC60601-2-37, Diagnostic Ultrasound Safety Standards
- EN/IEC60601-1, Safety requirements for Medical Equipment
- EN/IEC60601-1-2, EMC requirements for Medical Equipment
- NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- IEC 61157, Declaration of the acoustic output
- ISO10993-1. Biocompatibility
#### 5. Intended Uses:
The SONOACE R3 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Neonatal Cephalic, Trans-rectal, trans-vaginal, Muscular-Skeletal (conventional, Superficial), Cardiac Adult and Peripheral-vessel.
#### Technological Characteristics: 6.
The SONOACE R3 is substantially equivalent to the SONOACE X6 Diagnostic Ultrasound System, cleared via K081676, and the SONOACE PICO Diagnostic Ultrasound System, cleared via K061213. All systems transmit ultrasonic encry into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.
### END of 510(K) Summary
510(k) Summary / Statement Certification
ATTACHMENT 1
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
# JUL 1 3 2010
Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K101829
Trade/Device Name: SONOACE R3 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: June 30, 2010 Received: July 1, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SQNOACE R3 Diagnostic Ultrasound System, as described in your premarket notification:
### Transducer Model Number
| C2-4/20 | EC4-9 |
|---------|-----------|
| CN2-8 | L5-12/60 |
| CN4-9 | LN5-12/40 |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542.
Sincerely vours.
Maha D. O'Brien for
Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
510(k) Premarket Notification
SONOACE R3 Diagnostic Ultrasound System
# Indications for Use
510(k) Number (if known):
Device Name: SONOACE R3 Diagnostic Ultrasound System
Indications for Use:
The SONOACE R3 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal, Abdominal, Pediatric, Small Organs, Neonatal Cephalic, Trans-rectal, trans-vaginal, Muscular-Skeletal (conventional, superficial), Cardiac Adult and Peripheral-vessel.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Division of Radiological Device Office of In Vitro Diagnostic Device Evaluation and Safety
510K K101829
Page 1 of 12
Indications For Use
.
{5}------------------------------------------------
510(k) Premarket Notification
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
### 510(k) No .:
Device Name: SONOACE R3 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | N | N | N | N | N | Note 1 | Notes 2,7,8 |
| | Abdominal | N | N | N | N | N | Note 1 | Notes 2,7,8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | N | N | Note 1 | Notes 2,5,6,7,8 |
| | Small Organ (See Note 5) | N | N | N | N | N | Note 1 | Note 2,5,6 |
| | Neonatal Cephalic | N | N | N | N | N | Note 1 | Notes 2,8 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | N | N | Note 1 | Note 2,7,8 |
| | Trans-vaginal | N | N | N | N | N | Note 1 | Note 2,7,8 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | N | N | Note 1 | Note 2,5,6 |
| | Musculo-skel. (Superfic.) | N | N | N | N | N | Note 1 | Note 2,5,6 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | Note 1 | Notes 4 |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | N | N | Note 1 | Notes 2,5,6,8 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+Color, B+PD, B+Color+PW, B+PD+PW, B+Color+Color M
510
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
*(Division Sign-Off)*
Indications For Use
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
{6}------------------------------------------------
### 510(k) No.:
Device Name: C2-4/20 for use with SONOACE R3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | N | N | N | N | N | Note 1 | Notes 2,7,8 |
| | Abdominal | N | N | N | N | N | Note 1 | Notes 2,7,8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | N | N | Note 1 | Notes 2,5,7,8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | N | N | N | N | N | Note 1 | Notes 4 |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+Color, B+PD, B+Color+PW, B+PD+PW, B+Color+Color M
510k
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Indications For Use .
(Division Sign-Off) Division of Radiological Devices Division of Radiological Bovice Evaluation and Safety
**Section 1.3, Page 3**
{7}------------------------------------------------
### 510(k) No.:
Device Name: CN2-8 for use with SONOACE R3 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2,7,8 |
| | Abdominal | N | N | N | | N | Note 1 | Notes 2,7,8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Notes 2,7,8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Steel Bopper morades Fower (Ampinade) Doppler
Note B+M, B+PW, B+Color, B+PD, B+Color+PW, B+PD+PW, B+Color+Color M
510
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
Mh. S. D.K.
(Division Sign-Off)
Indications For Use
Division of Radiological Devices Office of Ir fitro Diagnostic Device Evaluation and Safety
**Section 1.3, Page 4**
{8}------------------------------------------------
### 510(k) No.:
Device Name: CN4-9 for use with SONOACE R3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal (See Note 3) | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Notes 2,8 | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | N | N | N | | N | Note 1 | Notes 2,8 | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Notes 2,8 | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+Color, B+PD, B+Color+PW, B+PD+PW, B+Color+Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Indications For Use
Division of Radiological Devices Office of Ir Evaluation and Safety
510K K101829
{9}------------------------------------------------
SONOACE R3 Diagnostic Ultrasound System
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
#### 510(k) No.:
.
Device Name: EC4-9 for use with SONOACE R3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| General<br>(Track I only) | Clinical Application<br>Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
|---------------------------|------------------------------------------------------|---|---|-----|-----|-------------------|----------------------|------------------|
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | | N | Note 1 | Note 2,7,8 |
| | Trans-vaginal | N | N | N | | N | Note 1 | Note 2,7,8 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+Color, B+PD, B+PD, B+PD+PW, B+Color+Color M
510
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
*(Division Sign-Off)*
Indications For Use
Division of Radiological Devices Office of iagnostic Device Evaluation and Safety
**Section 1.3, Page 6**
{10}------------------------------------------------
### 510(k) No.:
Device Name: L5-12/60 for use with SONOACE R3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Small Organ (See Note 5) | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent) | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Musculo-skel. (Superfic.) | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Note 5,6 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note I : B+M, B+PW, B+Color, B+PD, B+Color+PW, B+PD+PW, B+Color+Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
5101
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
Indications For Use
h. D'H
(Division Sign-Off)
Division of Radiological Devices Vitro Diagnostic Device Evaluation and Safety Office of In
{11}------------------------------------------------
### 510(k) No.:
Device Name: LN5-12/40 for use with SONOACE R3
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|--------------------------------|--|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | (Tracks I & III)<br>Ophthalmic | | | | | | | | |
| | Fetal (See Note 3) | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Small Organ (See Note 5) | | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Musculo-skel. (Superfic.) | | N | N | N | | N | Note 1 | Note 2,5,6 |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | N | N | N | | N | Note 1 | Note 5,6 |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note I : B+M, B+PW, B+Color, B+PD, B+Color+PW, B+Color+PW, B+Color+Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
510
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients.
Note 6: Abdominal congress and registers I vocal
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)
**Indications For Use**
Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
**Section 1.3, Page 8**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.