K101782 · Staar Japan · MSS · Dec 22, 2010 · Ophthalmic
Device Facts
Record ID
K101782
Device Name
MICROSTAAR MSI INJECTOR SYSTEM CARTRIDGE
Applicant
Staar Japan
Product Code
MSS · Ophthalmic
Decision Date
Dec 22, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The MicroSTAAR MSI Injector System is intended to fold and insert STAAR Surgical Company Collamer phakic and aphakic 1-piece and 3-piece Collamer® intraocular lenses for surgical placement in the human eye. The MSI Cartridges in this submission are just one component of the MicroSTAAR MSI Injector System.
Device Story
MicroSTAAR MSI Injector System Cartridges (models CQ, SFC 25, SFC 45) facilitate loading, folding, and delivery of Collamer intraocular lenses (IOLs) into the human eye through small incisions. Cartridges feature a loading area tapering to a tip; CQ model uses a hinged loading area, while SFC-45 and SFC-25 use flat loading areas. Device is a single-use component used with an injector system. Operated by surgeons in clinical/OR settings. IOL is placed in the cartridge, folded/pressed, and delivered via the injector. Benefits include controlled, minimally invasive IOL insertion, reducing surgical trauma. Device is supplied sterile.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Cartridges fabricated from polypropylene mixed with Glycerol Monostearate (GMS). Three models: CQ (hinged loading area), SFC-25, and SFC-45 (flat loading areas). Sterile, single-use, ethylene oxide (EtO) sterilized. Mechanical delivery system; no software or electronic components.
Indications for Use
Indicated for folding and inserting STAAR Surgical Company Collamer phakic and aphakic 1-piece and 3-piece intraocular lenses into the human eye during surgery.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
{0}------------------------------------------------
MicroSTAAR™ (MSI) INJECTOR SYSTEM CARTRIDGES
K101782
# 510(k) SUMMARY
| Submitter Information | STAAR Surgical Company<br>Contact: Jack Coggan<br>Director of Regulatory Affairs<br>1911 Walker Avenue<br>Monrovia, California 91016<br>Phone: 626.303.7902 ext 2616<br>Fax: 626.303.2962 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Registration Number | 2023826 |
| Submission Correspondent | STAAR Surgical Company<br>Contact: Jack Coggan<br>Director of Regulatory Affairs<br>1911 Walker Avenue<br>Monrovia, California 91016<br>Phone: 626.303.7902 ext 2616<br>Fax: 626.303.2962 |
| Date Summary Prepared | January 4, 2011 |
| Classification Name | Lens, Guide, Intraocular<br>(Class I) - KYB 21 CFR 886.4300<br>Folders and Injectors, Intraocular Lens (IOL)<br>(Class I) - MSS 21 CFR 886.4300 |
| Common/Usual Name | Intraocular Lens Injector Cartridge |
| Device Trade Name | MicroSTAAR™ MSI Injector System<br>Models: CQ, SFC 25, SFC 45 |
The primary devices used for comparison in this summary are: Epiphany Injector System (K090161) and MicroSTAAR™ Indigo Injector System (K073586).
#### Intended Use
The MicroSTAAR MSI Injector System is intended to fold and insert STAAR Surgical Company Collamer phakic and aphakic 1-piece and 3-piece Collamer® intraocular lenses for surgical placement in the human eye. The MSI Cartridges in this submission are just one component of the MicroSTAAR MSI Injector System.
The intended use of the MSI Cartridges in this submission are substantially equivalent to the intended use of the predicate devices K940593 (CQ cartridge FP), K980696 (SFC-45 FP), and K073586 (SFC-25 FP). Please refer to Attachment 1 for a copy of the Directions for Use for the predicate devices which will be the same for the MSI Cartridges.
{1}------------------------------------------------
#### 1990085840 M 11252 JUTO HVA H8M 473 Home Ho
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ារ
::
# '557' 등' 2402 문제 12
. ಅಂದಿಭಾರತದಿಗಳು ಇವರಿಗಳು ಸ
មានប្រទេសនាក់នោះ និងប្រទេសស្រី ប្រពន្ធមានប្រ
n abdury, on 2007 and 14:00 p
breage for asteradosmil
on Holly Produced is
orenić fora Tarmardio O
ara 21 aberti server :
унеото 3 врімай УЗАДУ -ா ஏர்தித்தும், மேற்கும் ation recinion 14 lociolorio trafi ported in to 1,30 | 11 | 1 6 lot Replace":[1]2] [stabuse 4 6181.128 1997 2005.01 61 611511000 1700 107 510 113
0232011
villa 30 Jan 2017 SAATR स्विमा को मिलते ही बात के ridio () voorhali/" he modern 9119 8 255, 157 177 Amar Christill Onliner M of of 145 for 1. Ha rollers 1 1 1971. 08:30 14 1
1991 de resulted.
ക്കാരങ്ങിനു പ്രധാന 用具体的角度(C) LE 器内液 - (Smart) a Kazama leritoorial ,216 meint hao ze bloži 요 11 310 300 2009 11 28M = 4 38000 1
์ ที่ยังสามารถเลือก และไทเป็นต้อมาใหม
n te vel prisojni le M M An | Potar | 21.012 21.0 02.00 : 0.0 : Malais
makayi takayi yandigitý li na gumunias zím zi pokinájace něklávana v Libben mag odl (882E502) aparea2 marajid andan2 mintal2 handaris is to (1846) . 2.)
145 between to
þaðgrófu á 1.5.7.0.1.2017 hann bast er húsinari af merevit 10:41 Hr.hljóð 11.1.1.12.10.11.201 ാറിക്കുന്നത് നിന്നും അത്തിന് അമിന്റെ പ്രിമ്പിച്ച് കന്സിന്റെ മലയാളിക്കുന്നത്. എന്നും വാഹിത്യ പ്രവ eno salipas egisa arina aidi di sempluraD (715) will neks minumbed, di new contributorio pasterji sabajal 1914 11:44 11 14 12 14 11 10 10 1000 1000
diff of terdesciapS gibilitassegon son occusarity of that call letter and Arcan edi to a min the high a bhaile a than and the should was the same on the soule of the security of ത്തിക്കി ഉപയോഗ്യിന്റെ ഇരാഹരി 1 അക്കാപക്ഷിച്ചു 1- 21,4 വരെ വേഴ് ැනුම් සිට විවිධිය වේ. මේ හෝ කොට පෙන් හිට අතර විය. මෙය විය. අධිරි වූ
{2}------------------------------------------------
## Description
The injector system is designed to facilitate the loading, folding and delivery of an intraocular lens (IOL) into the human eye through a small incision. The MicroSTAAR™ MSI injector System Cartridge utilizes three different cartridges for the insertion of an IOL lens. The three cartridges; CQ cartridge, SFC-45 cartridge and the SFC-25 cartridge are designed to facilitate a folded IOL lens for delivery. Each cartridge has a loading area which tapers down to a small tip, for the SFC-45 and SFC-25 cartridges, the lens loading area is flat and for the CQ cartridge the loading area is hinged. It is a component that is fabricated from polypropylene mixed with Glycerol Monostearate (GMS). Currently, the cartridge is cleared for fabrication from polypropylene mixed with GMS and is supplied sterile.
## Technological Characteristics
The MicroSTAAR™ MSI injector System Cartridge has substantially equivalent technological characteristics to the predicate device. Refer to Table 1 in the following section, entitled Comparison Analysis, for a summation of technological characteristics such as design, specifications and materials.
| | MicroSTAAR Injector System<br>K940593, K980696, K073586 | Epiphany Injector System<br>K090161<br>Epiphany | MicroSTAAR MSI Injector<br>System Cartridges | Substantially<br>Equivalent |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Product<br>Description | System is designed to facilitate<br>the loading, folding and<br>delivery of an intraocular lens<br>(IOL) into the human eye<br>through a small incision. | System is designed to facilitate<br>the loading, folding and<br>delivery of an intraocular lens<br>(IOL) into the human eye<br>through a small incision. | System is designed to facilitate<br>the loading, folding and<br>delivery of an intraocular lens<br>(IOL) into the human eye<br>through a small incision. | Same |
| Intended Use | Intended to fold and insert<br>STAAR Surgical Company 1-<br>piece or 3-piece Collamer<br>intraocular lenses for surgical<br>placement in the eye. | Intended to fold and insert<br>STAAR Surgical Company 3-<br>piece Silicone intraocular<br>lenses for surgical placement in<br>the eye. | Intended to fold and insert<br>STAAR Surgical Company 1-<br>piece and 3-piece Collamer<br>intraocular lenses for surgical<br>placement in the eye. | Same |
| Design | Single cartridge to be used in<br>conjunction with appropriate<br>injector system and 1-piece or<br>3-piece Collamer lenses for<br>surgical placement of lens into<br>eye. | A single component medical<br>device used in conjunction with<br>3-piece silicone lenses for<br>surgical placement in the eye. | Single cartridge to be used in<br>conjunction with appropriate<br>injector system and 1-piece or<br>3-piece Collamer lenses for<br>surgical placement of lens into<br>eye. | Same |
| Materials | Cartridge is manufactured from<br>Polypropylene purchased from<br>US Supplier. | Cartridge is manufactured from<br>Polypropylene MG05 and<br>Polypropylene MG03<br>purchased from Japan Supplier. | Cartridge is manufactured from<br>Polypropylene MG05 and<br>Polypropylene MG03<br>purchased from Japan Supplier. | Same |
| Mechanical<br>Safety | | | | |
| Manufacturing | Utilization of contract<br>manufacturers and in-house<br>processing at Monrovia facility | Utilization of contract<br>manufacturers and in-house<br>processing at the STAAR Japan<br>facility | Utilization of contract<br>manufacturers and in-house<br>processing at the STAAR Japan<br>facility, while molding process<br>is still done by a US contractor. | Similar |
# Comparison Analysis
{3}------------------------------------------------
### MicroSTAAR™ (MSI) INJECTOR SYSTEM CARTRIDGES
| 510(k) | MicroSTAAR Injector System<br>K940593, K980696, K073586 | Epiphany Injector System<br>K090161<br>Epiphany | MicroSTAAR MSI Injector System Cartridges | Substantially Equivalent |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Operating<br>Principle | The IOL is placed into the<br>loading area and folded shut or<br>gently pressed towards tip.<br>Then cartridge is loaded onto<br>the injector. | The cartridge is part of the<br>Epiphany Injector. The lens is<br>loaded into the cartridge,<br>snapped shut, and pushed into<br>pre-operative position. | Depending on the cartridge<br>type, the IOL is either placed<br>into the loading area and folded<br>shut, or it is pressed forward<br>towards the tip. Then cartridge<br>is loaded onto the injector. | Same |
| Packaging | Cartridge is placed in tray,<br>sealed with Tyvek lid and<br>labeled. Tray is placed in Kraft<br>paper pouch and then a foil<br>pouch, and labeled again. US<br>Packaging Supplier. | The Epiphany injector is placed<br>into a tray, which is labeled.<br>This unit is placed inside a<br>pouch which is sealed. The<br>sealed pouch is then put into a<br>box with a package insert.<br>Japanese Packaging Supplier. | Cartridge is placed in tray,<br>sealed with Tyvek lid and<br>labeled. Tray is placed in Kraft<br>paper pouch. Japanese<br>Packaging supplier. | Similar |
| Sterility | Sterile Ethylene Oxide (EtO)<br>by Subcontractor | Sterile Ethylene Oxide (EtO)<br>by STAAR Japan | Sterile Ethylene Oxide (EtO)<br>by STAAR Japan | Same |
| Manufacturer | STAAR Surgical | STAAR Japan | STAAR Japan | Same |
8 .
Table 1 – Summary of Design comparison for 21 CFR § 807.87(f)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
FEB - 7 2011
STAAR Surgical Company c/o Mr. Jack Coggan Director of Regulatory Affairs 1911 Walker Avenue Monrovia, CA 91016
Re: K101782
Trade/Device Name: MicroSTAAR Injector System Cartridges Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I, reserve Product Code: MSS Dated: December 2, 2010 Received: December 6, 2010
Dear Mr. Coggan:
This letter corrects our substantially equivalent letter of December 22, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bocasined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to commerce provided to May 20, 1977, the oncordance with the provisions of the Federal Food, Drug, devices that have occh require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, marker the devices subject to eno gements for annual registration, listing of devices, controls provisions of the frect merade song prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (000 as ) Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA s issuated or wouldevice complies with other requirements of the Act that I DA has made a acterimansions administered by other Federal agencies. You must of any I cuthal saltutes and regulations and limited to: registration and listing (21 Comply with an the For 8 requirements (01); medical device reporting (reporting of medical
{5}------------------------------------------------
#### Page 2 - Mr. Jack Coggan
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/5/Picture/7 description: The image shows a handwritten word, "for", in cursive script. The letter "f" is elongated with a large loop extending below the baseline, and the letters "o" and "r" are connected in a fluid, continuous stroke. The writing appears to be done with a pen or marker, and the ink is dark and bold.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K101782
Device Name: MicroSTAAR™ Injector System Cartridge
Indications For Use:
The MicrSTAAR™ MSI Injector System is a device intended to fold and insert STAAR™ Surgical Company Collamer phakic and aphakic one-piece and three-piece Collamer® intraocular lenses for surgical placement in the eye.
Prescription Use > (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jennifer Brown
Page 1 of
(Division Sign Off) (Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K101782
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.