← Product Code [LPH](/productcode/LPH) · K101761

# EXACTECH NOVATION EMPIRE ACETABULAR AUGMENT WITH INTEGRIP MODEL 186-01-08/11/13, 186-02-08/11/13, 186-03-08/11/13 (K101761)

_Exactech, Inc. · LPH · Sep 21, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K101761

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Sep 21, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Empire Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. The Empire Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with Empire Acetabular Augments. The assembled construct is intended for press-fit fixation. The construct may also be used with bone cement at the discretion of the surgeon.

## Device Story

Empire Acetabular Augments are titanium alloy implants manufactured via additive manufacturing to create a porous structure for mechanical and biological fixation. Used in primary or revision hip replacement surgery to address segmental acetabular deficiencies; serves as an alternative to structural allograft. Augments are affixed to Novation Crown Cup acetabular shells using PMMA bone cement; construct is intended for press-fit fixation, though bone cement may be used at surgeon discretion. Augments feature holes for 4.5mm bone screws for adjunct fixation. Available in three heights (8mm, 11mm, 13mm) for shell diameters 48-58mm. Operated by orthopedic surgeons in clinical/OR settings. Provides structural support to the acetabulum, potentially improving implant stability and patient outcomes in complex revision cases.

## Clinical Evidence

No clinical trials performed. Evidence includes cadaver lab validation of design features, mechanical testing per FDA guidance for modified metallic surfaces, animal study comparing biological fixation of porous Ti-6Al-4V to historical data, and clinical literature review of predicate devices.

## Technological Characteristics

Material: Ti-6Al-4V titanium alloy. Manufacturing: Additive manufacturing (porous structure). Dimensions: 8mm, 11mm, 13mm heights; 48-58mm diameters. Fixation: PMMA bone cement to shell; press-fit or cement to bone. Sterilization: Provided sterile for single use.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Zimmer Hedrocel Acetabular Augment ([K001471](/device/K001471.md))

## Reference Devices

- Novation Crown Cup acetabular shells ([K070479](/device/K070479.md))
- Exactech Bone Screws ([K896601](/device/K896601.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Exactech® Novation® Empire™ Acetabular Augments Traditional 510(k) – 510(k) Summary

## 510(k) Summary

| Company:             | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville FL, 32653                                                                                                                                                                                                 | SEP 21, 2010 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Date:                | September 17, 2010                                                                                                                                                                                                                                            |              |
| Contact Person:      | Darrell Kassner<br>Director, Regulatory Affairs<br>Telephone: (352) 377-1140<br>Fax: (352) 378-2617                                                                                                                                                           |              |
| Proprietary Name:    | Exactech® Novation® Empire™ Acetabular Augments<br>with InteGrip™                                                                                                                                                                                             |              |
| Common Name:         | Acetabular Augment                                                                                                                                                                                                                                            |              |
| Classification Name: | 21 CFR 888.3358 - Hip joint metal/polymer/metal<br>semi-constrained porous-coated uncemented prosthesis.<br>Device Class: II, Classification Panel: Orthopedic,<br>Product Code: LPH - prosthesis, hip, semi-constrained,<br>metal/polymer, porous uncemented |              |

## Legally Marketed Devices to Which Substantial Equivalence Is Claimed

Zimmer Hedrocel Acetabular Augment, 510(k) K001471 .

### Device Description

The proposed acetabular augments are manufactured by an additive manufacturing process from titanium alloy to create a porous structure designed to provide initial mechanical fixation in addition to the potential for long-term biological fixation. The augments feature holes that allow for the use of 4.5mm bone screws for adjunct fixation of the acetabular augment and the surrounding bone. The acetabular augments are available in 3 heights (8mm, 11mm, and 13mm) and cover acetabular shell outer diameter sizes 48-58mm.

The proposed augments mate with the following devices:

- Novation Crown Cup acetabular shells (K070479) .
- Exactech Bone Screws (K896601). .

ﺎ ، ﻭﻳ

{1}------------------------------------------------

## Exactech® Novation® Empire™ Acetabular Augments Traditional 510(k) - 510(k) Summarv

### Indications for Use

The Empire Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.

The Empire Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with Empire Acetabular Augments. The assembled construct is intended for press-fit fixation. The construct may also be used with bone cement at the discretion of the surgeon.

#### Summary of Technological Characteristics

The rationale for substantial equivalence is based on consideration of the following characteristics:

- . Intended Use. The proposed Empire Acetabular Augments and predicate Zimmer Hedrocel Acetabular Augments are intended for use as a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies and have similar indications for use statements.
- . Materials. The proposed Empire Acetabular Augments and predicate Zimmer Hedrocel Acetabular Augments are composed of materials conforming to recognized industry standards for permanent implants.
- Sterilization processes. The proposed Empire Acetabular Augments and . predicate Zimmer Hedrocel Acetabular Augments are provided sterile for single use and conform to recognized industry standards.
- Design Features. The proposed Empire Acetabular Augments and predicate . Zimmer Hedrocel Acetabular Augments incorporate similar design features.
- Performance specifications. The proposed Empire Acetabular Augments and . predicate Zimmer Hedrocel Acetabular Augments conform to recognized performance standards for total hip replacement devices.

#### Substantial Equivalence Conclusion

Results from mechanical tests, animal studies, simulated-use tests, and engineering analyses demonstrate the proposed Empire Acetabular Augments are substantially equivalent to the predicate Zimmer Hedrocel Acetabular Augments. A summary of these tests and analyses are as follows:

- Cadaver lab validation demonstrating the design features (including outside . geometry and overall scope).
- Test summary and reports of the material based on FDA Guidance Document for . Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement.
- Animal study report comparing the biological fixation characteristics of the . porous Ti-6Al-4V to historical data.
- Clinical literature review of predicate acetabular augment devices. .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Exactech, Inc. % Mr. Darrell J. Kassner Director, Regulatory Affairs 2320 N.W. 66th Court Gainesville, Florida 32653

EP 2 1 2010

Re: K101761

Trade/Device Name: Exactech® Novation® Empire™ Acetabular Augments with InteGrip™ Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Codes: LPH, JDI, LWJ, LZO Dated: June 21, 2010 Received: June 23, 2010

Dear Mr. Kassner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Ваввадвник

Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Exactech® Novation® Empire™ Acetabular Augments Traditional 510(k) - Indications for Use Statement

# Indications for Use Statement

510(k) Number:

K101761

Device Name: Exactech® Novation® Empire™ Acetabular Augments with InteGrip™

## INDICATIONS FOR USE:

The Empire Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies.

The Empire Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with Empire Acetabular Augments. The assembled construct is intended for press-fit fixation. The construct may also be used with bone cement at the discretion of the surgeon.

Prescription Use_ and/or X (Part 21 CFR 801 Subpart D)

**Over-T**
(21 CFR

Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Souther
(Division Sign-Off)
for mxm

Division of Surgical, Orthopedic. and Restorative Devices

510(k) Number K101761

---

**Source:** [https://fda-staging.innolitics.com/device/K101761](https://fda-staging.innolitics.com/device/K101761)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).