NEOWIRE PTA/PTCA GUIDEWIRE

K101729 · Neometrics, Inc. · DQX · Sep 23, 2010 · Cardiovascular

Device Facts

Record IDK101729
Device NameNEOWIRE PTA/PTCA GUIDEWIRE
ApplicantNeometrics, Inc.
Product CodeDQX · Cardiovascular
Decision DateSep 23, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The NeoWire Guidewire is intended to facilitate the introduction and placement of catheters and interventional devices within the coronary and/or peripheral vasculature. Models with Stiff tips (> 9g) are intended for peripheral use only. Models with Floppy, Soft, and Medium tips (≤ 9g) are intended for peripheral or coronary use. The guidewire is not intended for use in the cerebral vasculature.

Device Story

NeoWire is a steerable guidewire used to navigate catheters and interventional devices through coronary and peripheral vasculature. Device consists of nitinol alloy core with distal radiopaque coil; may include PTFE, silicone, or hydrophilic lubricious coatings. Distal tip is shapeable to assist navigation. Operated by physicians in clinical settings (e.g., cath lab) to provide a rail for catheter advancement. Output is mechanical support and guidance for interventional tools. Benefits include improved access to target vascular sites. Device is not for cerebral use.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including guidewire dimensions, torqueability, torque strength, coating adherence, tip flexibility, sterile package integrity (ASTM F1929-98, ASTM F88-09), reverse bend, tensile testing, fracture resistance, corrosion resistance (ISO 11070), tip load, and rail support. Biocompatibility testing (ISO 10993) performed on NeoWire or leveraged from previously cleared NeoMetrics products.

Technological Characteristics

Nitinol alloy core wire; distal radiopaque coils; PTFE, silicone, or hydrophilic lubricious coatings. Joined via adhesive and welds. Diameters: 0.014" and 0.018". Lengths: 80, 180, 190, 300cm. Flexible, shapeable distal tip. Sterilized. Mechanical device; no software or electronics.

Indications for Use

Indicated for facilitating introduction and placement of catheters and interventional devices in coronary and peripheral vasculature. Contraindicated for use in cerebral vasculature. Stiff tip models (>9g) restricted to peripheral use; Floppy, Soft, and Medium tip models (≤9g) for peripheral or coronary use.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KCO1729 ### Section 5.0: 510(k) Summary ## SEP 2 3 2010 #### 5.1 Administrative Information | Name: | NeoMetrics, Inc. | |--------------------------|--------------------------------------------------------| | Address: | 2605 Fernbrook Lane North, Suite<br>Plymouth, MN 55447 | | Contact Person: | Gene Champeau | | Title: | President | | Phone: | (763)-559-4440 | | Fax: | (763) 559-7676 | | FDA Registration Number: | 2135342 | | Date: | April 2, 2010 | #### 5.2 Device Information | Name of Device: | NeoWire PTA/PTCA Guidewire | |----------------------|------------------------------------------| | Common Name: | Guides, Guidewires, or spring Guidewires | | Classification Name: | Catheter Guide Wire (870.1330) | | Product Code: | DQX | #### 5.3 Predicate Device Information The following commercially available guidewires are predicate devices for NeoWire. | 510(k) Number | Trade or Model Name | Manufacturer | |---------------|------------------------------|------------------| | K070150 | VascuPuncture PICC Guidewire | NeoMetrics, Inc. | | K983927 | Zinger Guidewire | Medtronic | | K013024 | Selectiva | NeoMetrics, Inc. | | K082304 | Roll-X | St. Jude Medical | | K081337 | Approach CTO | Cook | #### 5.4 Device Description NeoWire is a steerable guidewire constructed of a nitinol alloy core with a coiled segment on the distal end. The core wire is either PTFE coated or uncoated and the wire may also contain silicone or hydrophilic lubricious coatings. This distal end of the wire can be shapeable and is radiopaque. #### 5.5 Intended Use The NeoWire Guidewire is intended to facilitate the introduction and placement of catheters and interventional devices within the coronary and/or peripheral vasculature. Models with Stiff tips (> 9g) are intended for peripheral use only. Models with Floppy, Soft, and Medium tips (≤ 9g) are intended for peripheral or coronary use. The guidewire is not intended for use in the cerebral vasculature. {1}------------------------------------------------ #### 5.6 Technological Characteristics A comparison of the characteristics of the proposed device to the predicate devices show the proposed device to have the following same or similar technological characteristics to the devices which have received 510(k) clearance: - . Same intended use; - . Same operating principle; - . Same shelf life and sterilization process; - Similarities in Design, Material Types, and Technology include: . - Nominal diameters: 0.014" and 0.018" o - Similar Lengths: 80, 180 190, and 300cm o - Nitinol alloy core wires o - Distal radiopaque coils o - 0 Flexible, shapeable distal tip segments - o Joined by a combination of adhesive and welds - o Lubricious coatings To ascertain similarity, the following performance testing was conducted on both the proposed device and the predicate devices: - . Guidewire Dimensions (overall length, radiopaque tip length, and diameter) per FDA Coronary and Cerebrovascular Guidewire Guidance January 1995 - . Torqueability and Torque Strength per FDA Coronary and Cerebrovascular Guidewire Guidance January 1995 - . Coating Adherence per FDA Coronary and Cerebrovascular Guidewire Guidance January 1 ਰੇਰੇਟ - Tip Flexibility per FDA Coronary and Cerebrovascular Guidewire Guidance January 1995 . - . Sterile Package Integrity per ASTM F 1929-98 - Sterile Package Integrity per ASTM F 88-09 . - Reverse Bend (Fatigue Failure) per ISO 11070 . - Tensile testing per ISO 11070 - . Fracture Resistance per ISO 11070 - Corrosion Resistance per ISO 11070 . - . Tip Load per NeoMetrics' test method TM-3010 - . Rail Support per NeoMetrics' test method TM-3002 Due to the destructive nature of the tests and the availability of predicate devices not all testing was conducted on each predicate device, however competitive devices were tested along with the proposed device in each test. #### 5.7 Summary of Non-Clinical Testing Non-clinical testing of the NeoWire Guidewire includes bench testing, biocompatibility testing, shelflife and packaging testing and sterilization evaluation. Bench testing, shelf-life testing, and sterilization testing were conducted on NeoWire. Biocompatibility testing was conducted on NeoWire and leveraged from other previously cleared NeoMetrics products. A summary table of the test data source is shown: {2}------------------------------------------------ | | Tested and Leveraged Biocomaptibility Data | | | | | | | | |-----------------------------------------|--------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------|-----------------------------------|-----------------------------------------------------|--------------------------|-------------------------------------------------| | Test Guidance<br>Document | ISO 10993-4 | ISO 10993-4<br>(ASTM<br>Guideline<br>F-756-00) | ISO 10993-5 | ISO 10993-11 | ISO-10993-10 | ISO 10993-10 | | | | Material | Partial<br>Thromboplastin<br>Time (PTT) | ASTM<br>Hemolysis Assay<br>Direct Contact<br>Method | MEM Elution<br>Assay with L-<br>929 Mouse<br>Fibroblast Cells<br>(Cytotoxicity) | Acute Systemic<br>Injection Test | Intracutaneous<br>Reactivity Test | Guinea Pig<br>Maximization<br>Sensitization<br>Test | Complement<br>Activation | In Vivo<br>Thrombosis<br>(Thrombo-<br>genicity) | | Core Wire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | | Hydrophilic<br>Coating | Leveraged | Leveraged | Leveraged | Leveraged | Leveraged | Leveraged | Tested on<br>NeoWire | Tested on<br>NeoWire | | Silicone<br>Coating | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Leveraged | Leveraged | | Proximal Coll<br>(dual coll<br>designs) | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Leveraged | Leveraged | | Radiopaque<br>Distal Coll<br>(Option 1) | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | | Radiopaque<br>Distal Coll<br>(Option 2) | Leveraged<br>Tested on<br>NeoWire | Leveraged<br>Tested on<br>NeoWire | Leveraged<br>Tested on<br>NeoWire | Leveraged<br>Tested on<br>NeoWire | Leveraged<br>Tested on<br>NeoWire | Leveraged<br>Tested on<br>NeoWire | Leveraged | Leveraged | | Adhesive | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Leveraged | Leveraged | | PTFE Coating | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Tested on<br>NeoWire | Results of this testing demonstrate that the guidewire design meets the product specifications and intended uses. #### 5.8 Substantial Equivalence Conclusion The NeoWire Guidewire described in this 510(k) is substantially equivalent to the devices listed in section 5.3. The intended use, design, material types, technology, and performance of the NeoWire Guidewire are equivalent to the predicate devices. There are no differences between devices which would raise issues of safety or effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 NeoMetrics, Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25" Street NW Buffalo, MN 55313 SEP 2 3 2010 Re: K101729 Trade/Device Name: NeoWire PTA/PTCA Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (two) Product Code: DQX Dated: September 7, 2010 Received: September 8, 2010 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Mark Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. .z Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K101729 Section 4: Indications for Use Statement SEP 2 3 2010 # Indications for Use 510(k) Number (if known): unavailable Device Name: NeoWire PTA/PTCA Guidewire Indications for Use: The NeoWire Guidewire is intended to facilitate the introduction and placement of catheters and interventional devices within the coronary and/or peripheral vasculature. Models with Stiff tips (> 9g) are intended for peripheral use only. Models with Floppy, Soft, and Medium tips (S 9g) are intended for peripheral or coronary use. The guldewire is not intended for use in the cerebral vasculature. Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use: (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) \$\mu PDZ ivision Sign-Off) Division of Cardiovascular Devices 510(k) Number: K101729
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...