M800 Handheld Pulse Oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.
Device Story
Handheld pulse oximeter; inputs SpO2 and pulse rate via LED/detector sensor; processes signals via internal CPU/driving circuit; outputs data to OLED display (waveform, %SpO2, pulse rate, pulse amplitude) and speaker (audible alarms/indicators). Used by clinicians or patients in hospitals, clinics, or home environments. Provides real-time monitoring and alarm functionality for physical parameters (user-settable limits) and technical status. Assists healthcare providers in assessing patient oxygenation status for clinical decision-making.
Clinical Evidence
Bench testing only. Verification tests performed per risk management protocols demonstrated that risks associated with design changes (OLED display, alarm functionality, speaker) were reduced to acceptable levels.
Technological Characteristics
Handheld device; OLED display; speaker for audio alerts; LED emitter/detector sensor; CPU-based processing; battery-powered. Features include physical parameter and technical alarming. No specific material standards or software architecture details provided.
Indications for Use
Indicated for spot checking functional arterial oxygen saturation (SpO2) and pulse rate in adult, pediatric, and neonatal patients in hospital, hospital-type, and home care settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Premarket Notification Special Section 510(k) Submission Report SN: JN00520100609FDA Submission Date: 11 JUN 2010
Exhibit #3 510(k) Summary
Image /page/0/Picture/3 description: The image shows the number K101694 at the top. Below the number is a logo with the letters BLT and the word BIOLIGHT. The logo also contains a graphic of a plant with three leaves.
Exhibit #3 510(k) Summary
SEP 0 8 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## The assigned 510(k) Number is: K101694
| Date of Preparation | 11 JUN 2010 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Guangdong Biolight Meditech Co., Ltd [Reg #:3007305624]<br>Innovation First Road, Technology Innovation Coast<br>Zhuhai, Guangdong, 519085, China<br>Contact Person: Mr. Tianbao Li, Chief Engineer<br>Tel: +86-756-3399963 Fax: +86-756-3399989 E-mail: li_tb@blt.com.cn |
| Submission<br>Correspondent | Ms. Diana Hong / Mr. Lee Fu<br>Shanghai Mid-Link Business Consulting Co., Ltd<br>Suite 5D, No.19, Lane 999, Zhongshan Road (S-2), Shanghai, 200030, China<br>T: +86-21-64264467 F: 240-238-7587 E: info@mid-link.net |
| Proposed Device | M800 Handheld Pulse Oximeter<br>Modification to: M700 Handheld Pulse Oximeter (K081712) as predicate device<br>21 CFR 870.2700 DQA Class II |
| Intended Use | M800 Handheld Pulse Oximeter is indicated for spot checking of functional arterial<br>oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in<br>hospital, hospital type facilities as well as in the home care environment. |
| Device Description | The proposed device is a handheld device, which can display %SpO2, waveform,<br>pulse rate value, pulse amplitude bar indication. It consists of detector and emitter<br>LED, OLED display module, CPU, driving circuit, power supply circuit and battery.<br>In addition, it has alarming function including physical parameter alarming and<br>technical alarming. The physical alarming limit can be set by user. The modifications<br>to M700 Handheld Pulse Oximeter are presented in Table 3-1 Modifications List on<br>the next page. |
| Test Summary | Per the risk management during the design change control, the verification tests<br>performed demonstrated that risks of each hazard are reduced to acceptable region. |
| Conclusion | The information in this 510(k) Summary demonstrate that the proposed device,<br>M800 Handheld Pulse Oximeter, is Substantially Equiavlent (SE) to the predicate<br>device, M700 Handheld Pulse Oximeter (K081712), with respect of effectiveness<br>and safety. |
3-1
{1}------------------------------------------------
Premarket Notification Special Section 510(k) Submission
.
.
・ ,
.. -."
Report SN: JN00520100609FDA Submission Date: 11 JUN 2010
Exhibit #3
510(k) Summary
.
.
Image /page/1/Picture/3 description: The image shows a logo with the letters "BLT" in bold, black font on the left side. Above the letters, there is a stylized graphic of a person with their arms raised. To the right of "BLT", there are Chinese characters, followed by the word "BIOLIGHT" in bold, black font. The logo appears to be for a company or organization called "Biolight".
| Table 5-4 Modifications List | | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | M700 Handheld Pulse Oximeter<br>(Predicate/Existed) | M800 Handheld Pulse Oximeter<br>(Proposed / Modified) |
| Display | LED Display, which can display the following information:<br>Pulse Rate<br>SpO2<br>Pulse Strength<br>System Information<br>Status Information. | OLED Display, which can display the following information:<br>Pulse Rate<br>SpO2<br>Pulse Strength<br>System Information<br>Status Information.<br><br>In addition, it can display:<br>SpO2 waveform, Alarming Limit and<br>Alarming Information. |
| Audio | Beeper, which can audio indicate the followings:<br>Sensor Detachment Indicating;<br>Pulse Indicating;<br>Key Pressing Indicating;<br>Low Voltage Indicating; | Speaker, which can audio indicate the followings:<br>Sensor Detachment Indicating;<br>Pulse Indicating;<br>Key Pressing Indicating;<br>Low Voltage Indicating;<br><br>In addition, it can indicate: Starting<br>and Alarming. |
| Alarming | Without Alarming Function | With Alarming Function |
Table 3-1 Modifications List
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and is evenly spaced around the circular border.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Guangdong Biolight Meditech Company, Limited C/O Ms. Diana Hong Shanghai Midlink Business Consulting Company, Limited Suite 5D, No. 19, Lane 999 Zhongshan Road (S-2) Shanghai China 200030
SEP 0 8 2010
Re: K101694
Trade/Device Name: M800 Handheld Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 11, 2010 Received: August 11, 2010
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Ms. Hong
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Jawed for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Premarket Notification Special Section 510(k) Submission Report SN: JN00520100609FDA Submission Date: 11 JUN 2010
Section III Indication for Use Form
K101694
**BLT. BIOLIGHT** 廣萊特
SEP 0 8 2010
Section III Indication for Use Form
510(k) NUMBER (if known):
DEVICE NAME: M800 Handheld Pulse Oximeter
## INDICATION FOR USE:
M800 Handheld Pulse Oximeter is indicated for spot checking of functional arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric and neonatal patients in hospital, hospital type facilities as well as in the home care environment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Z. Schulten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K101694
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.