K101670 · United Orthopedic Corp. · KWY · Oct 8, 2010 · Orthopedic
Device Facts
Record ID
K101670
Device Name
UNITED U2 BIPOLAR IMPLANT
Applicant
United Orthopedic Corp.
Product Code
KWY · Orthopedic
Decision Date
Oct 8, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
U2 Bipolar Implant is intended for use in combination with UNITED Femoral System for cemented or cementless hip replacement in skeletally mature patients with the following conditions: non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; inflammatory degenerative joint disease such as rheumatoid arthritis; correction of function deformity; revision procedures where other treatments or devices have failed; treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Device Story
U2 Bipolar Implant is a hemiarthroplasty device for hip replacement; consists of outer metal shell with permanently assembled bearing insert; features factory-assembled stopper ring and metal wire. Primary articulation occurs at femoral head/inner polyethylene bearing interface; secondary articulation at outer shell/acetabulum interface. Available in 41-73 mm outer diameters; compatible with 26mm, 28mm, or 36 mm femoral heads. Used by orthopedic surgeons in clinical settings for hip reconstruction. Device provides dual-mobility articulation to restore joint function and reduce wear. Highly polished outer metal surfaces articulate with patient's acetabulum.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included evaluation of locking mechanism integrity and range of motion.
Technological Characteristics
Bipolar endoprosthesis; metal/polymer construction. Features a locking ring mechanism for the polyethylene liner. Available in 41-73 mm outer diameters. Designed for use with 26mm, 28mm, or 36 mm femoral heads. Minimum polyethylene thickness of 5mm at load-bearing area. Highly polished outer metal shell.
Indications for Use
Indicated for skeletally mature patients requiring cemented or cementless hip replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, revision of failed treatments, or proximal femur fractures (nonunion, femoral neck, trochanteric) with head involvement.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
{0}------------------------------------------------
K101670
OCT - 8 2010
510(k) Summary
# 510(k) Summary of Safety and Effectiveness
## Submission Information
| Company: | United Orthopedic Corporation |
|-----------------|------------------------------------------------------|
| Address: | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan |
| Contact Person: | Fang-Yuan Ho, |
| | Regulatory Affairs Manager |
| | Phone:+886-3-5773351 ext. 212 |
| | Fax:+886-3-5777156 |
| Date Prepared: | June 11, 2010 |
## Device Identification
| Device Name: | U2 Bipolar Implant |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bipolar endoprosthesis |
| Classification Name<br>and Reference : | Hip Joint femoral (Hemi-Hip) Metal/Polymer Cemented or<br>Uncemented Prosthesis per 21CFR 888.3390. This falls under<br>the Orthopedics panel. |
| Predicate Device: | 1. "UNITED" U1 Hip System – Bipolar (K050269)<br>2. "Wright" GLADIATOR Bipolar System (K062693) |
#### Device Description:
"UNITED" U2 Bipolar Implant, designed with a simple one-step assembly interface between femoral head and U2 Bipolar Implant, is a hemiarthroplasty design. U2 Bipolar Implant consists of an outer shell into which a bearing insert has been permanently assembled. The bearing insert has a factory assembled stopper ring and metal wire. The assembled prosthesis provides for primary articulation at the femoral
UOC-FDA-014
Page: Summary-1/3
{1}------------------------------------------------
# 0 U2 Bipolar Implant (K 101670)
head/inner polyethylene bearing interface and secondary articulation at the outer shell/acetabulum interface. The internal aspect of the shell is designed to lock the polyethylene liner. The outer metal surfaces of the bipolar hip prosthesis are highly polished for articulation with the patient's acetabulum. Components are available in a range of sizes to fit varying anatomical requirements.
U2 Bipolar Implant are offered in 41-73 mm outer diameters and is designed to be used with either 26mm, 28mm, or 36 mm femoral heads, depending on the inner diameter liner. The minimum thickness of the polyethylene at its thinnest point at a load bearing area of the liner is at lest 5.mm in U2 Bipolar Implant.
### Intended Use
U2 Bipolar Implant is intended for use in combination with UNITED Femoral System for cemented or cementless hip replacement in skeletally mature patients with the following conditions: non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; inflammatory degenerative joint disease such as rheumatoid arthritis; correction of function deformity; revision procedures where other treatments or devices have failed; treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
#### Basis for Substantial Equivalence:
The safety and effectiveness of U2 Bipolar Implant are substantially equivalent to the previously cleared U1 Hip System -- Bipolar (K050269), except for an extension in the size distribution and increasing stopper ring design. In addition, the subject device is also substantial equivalence to the "Wright" GLADIATOR Bipolar System (K062693).
UOC-FDA-014
Page: Summary-2/3
{2}------------------------------------------------
These two devices both use a locking ring mechanism.
## Performance Test - Bench:
The following tests were performed:
- 1. The integrity of locking mechanism
- Range of motion evaluation 2.
**UOC-FDA-014**
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a black and white logo. The logo features a stylized bird in flight, with three parallel lines forming its body and wings. The bird is oriented towards the right. Encircling the bird is text that appears to follow the curve of the circle. The text is not clearly legible due to the image quality.
#### Public Health Scrvice
OCT 0 8 2010
Food and Drug Administration 10903 New Hampshire Avenuc Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
United Orthopedic Corporation % Ms. Fang-Yuan Ho 57 Park Ave Science Park Hsinchu, TW 300
Re: K101670
Trade/Device Name: U2 Bipolar Implant Regulation Number: 21 CFR 888.3390 Regulation Name: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented Regulatory Class: Class II Product Code: KWY Dated: October 4, 2010 Received: October 7, 2010
Dear Ms. Ho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Pederal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, invince of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and warranties. We remind you, however, that device labeling must be nuth to book not midlers ing.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that PDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registroution and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of maniti device-rolated adverse events) (21 CFR 803); good manufacturing practice requirements as et
{4}------------------------------------------------
Page 2 – Ms. Fang-Yuan Hol.
ು ಸ
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 53)-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please the Courter for gover DA/CentersOffices/CDRH/CDRHOffices/ucm/15800, bttp://www.fov the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation ontitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDK regulation (2) CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollillibe
1800) 638-2041 - 12011-7011 706-700 (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcessforYou/Industry/default.htm.
Sincerely_yours.
Mark N. Melkerson
Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indication for Use
510 (k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: _U2 Bipolar Implant
Indications for Use:
: 11 11 11
This device is intended for use in combination with UNITED Femoral System for comented or cementless hip replacement in skeletally mature patients with the following conditions: non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; inflammatory degenerative joint disease such as theumatoid arthritis; correction of function deformity; revision procedures where other treatments or devices have failed; treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Prescription Use __x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Dand Kame for MXM*
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
Page 1 of 1
510(k) Number. K101670
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.