K101643 · Care Fusion · CBK · Jan 26, 2011 · Anesthesiology
Device Facts
Record ID
K101643
Device Name
MODEL LTV 1100 VENTILATOR
Applicant
Care Fusion
Product Code
CBK · Anesthesiology
Decision Date
Jan 26, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LTV® 1100 ventilator is intended to provide continuous or intermittent ventilatory support for the care of the individuals who require mechanical ventilation. The ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5kg (11 lbs.), who require the following types of ventilatory support: Positive Pressure Ventilation, delivered invasively (via endotrach tube or trach tube) or non-invasively (via mask or nasal prongs). Assist/Control, SIMV, CPAP, and NPPV modes of ventilation. The ventilator is suitable for use in institutional, home, or transport settings.
Device Story
LTV 1100 is a portable, turbine-powered ventilator (10.5" x 13.5" x 3.25", 14.5 lbs) providing invasive and non-invasive positive pressure ventilation. It operates without external compressed gas. Features include CPAP, SIMV, Control, Assist/Control, Apnea Backup, and NPPV modes; volume control and pressure support; and spontaneous breathing trials. Input includes patient circuit pressure and flow (via optional flow sensor). The device processes these inputs to manage breath delivery, leak compensation, and alarm monitoring (e.g., high/low pressure, apnea). Operated by trained personnel in clinical, home, or transport environments. Output includes real-time pressure/flow data and alarm notifications on a front panel display. Healthcare providers use these outputs to adjust ventilation parameters, supporting patient weaning and respiratory stability.
Clinical Evidence
Bench testing only. Performance testing verified that the LTV 1100 ventilator meets its performance requirements and is substantially equivalent to legally marketed predicate devices.
Technological Characteristics
Turbine-based ventilation; no external compressed gas required. Dimensions: 10.5" x 13.5" x 3.25"; Weight: 14.5 lbs. Power: AC, internal battery, or external DC. Features: Flow sensor/no-flow sensor modes, leak compensation, lockable controls, real-time pressure display. Software-controlled ventilation modes (CPAP, SIMV, Control, Assist/Control, NPPV).
Indications for Use
Indicated for adult and pediatric patients weighing ≥5kg requiring continuous or intermittent mechanical ventilatory support (invasive or non-invasive). Suitable for institutional, home, or transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon contains a stylized letter "C" inside of a circle. The text is in a bold, sans-serif font.
K101643
## 510 (K) Summary as required by 21 CFR 807.92
| 510 (K) Submitter: | CareFusion<br>Yorba Linda, CA 92887<br>(714) 283 – 2228<br>JAN 26 2011 |
|------------------------------------|-------------------------------------------------------------------------------|
| Contact Person: | Monther Abushaban<br>(714) 919 – 3660<br>Monther.Abushaban@CardinalHealth.com |
| Establishment Registration Number: | 2050001 |
| Date Prepared: | January 11,2011 |
| Name of the Device: | LTV 1100 Ventilator |
| Common/ usual name: | ventilator |
Classification:
The LTV 1100 ventilator is classified as a class fl device under the following classification code
| Product code | CFR Section | Panel |
|--------------|-----------------|----------------|
| 73 CBK | 21 CFR 868.5895 | Anesthesiology |
Predicate Device: LTV 1200 ventilator (K060647), Vela (K093094), LTV800 (K981371)
Device Description: The LTV® 1100 ventilator is designed to provide the functionality in a small package. The LTV® 1100 ventilator provides the following features:
- ventilator is (10.5" x 13.5" x 3.25", 14.5 lbs). A
- Turbine technology allows the ventilator to operate without an external compressed gas source. A
- CPAP2, SIMV3, Control, Assist/Control and Apnea Backup ventilation modes. A
- NPPV4 mode ventilation, providing an alarm package suitable for mask ventilation of patients A that do not require life support ventilation.
- Volume Control and Pressure Support ventilation. A
- Spontaneous Breathing Trial (SBT) to assist with weaning and discontinuation of ventilatory A support.
- Variable alarm settings including High Peak Pressure, Low Peak Pressure, Low Minute Volume, A Apnea, High Breath Rate, High PEEP and Low PEEP.
- Low-Pressure Oxygen Bieed-in A
- Lockable front panel controls. A
- > Eookubio from panel Control Continute Ventilation (VE), PEEP, PIP and Tidal Volume (Vte).
- Real-time patient circuit pressure display with Peak Inspiratory Pressure indicator. A
- > Variable termination conditions for Pressure Support breaths, including maximum inspiratory time termination and percentage of peak flow.
- Leak Compensation to improve triggering when a circuit leak is present. a
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon features a stylized letter "C" within a circle, and the text is in a bold, sans-serif font.
- A Single or dual tone output capabilities.
- A Operation from a variety of power sources including AC power, internal battery and external DC power sources.
ﺮ ﻧﺴﺒﺔ
・・・
- > Flow Sensor and No Flow Sensor configuration modes allow for a flow sensing patient circuit or a non-flow sensing patient circuit to be used.
Intended Use: The LTV® 1100 ventilator is intended to provide continuous or intermittent ventilatory support for the care of the individuals who require mechanical ventilation. The ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5kg (11 lbs.), who require the following types of ventilatory support:
- > Positive Pressure Ventilation, delivered invasively (via endotrach tube or trach tube) or non-invasively (via mask or nasal prongs).
- A Assist/Control, SIMV, CPAP, and NPPV modes of ventilation.
The ventilator is suitable for use in institutional, home, or transport settings.
- Substantial Equivalence: The LTV 1100 ventilator is the same device as the predicate device LTV 1200. LTV800 and Vela ventilator. Modifications to the software by adding more features did not affect the intended use. The modifications to the software associated with this submittal are as follows:
- > Translations for LEAK COMP On & OFF
- > Sigh
- > Patient Circuit Selection
In summary, the LTV 1100 ventilator described in this submission is, in our opinion substantially equivalent to the predicate device.
- Summary of Testing and Validation: Performance testing verified that the LTV 1100 ventilator meets it's performance requirements that this device is substantially equivalent to medical devices currently legally marketed in the United States.
- Reason for the submission: This Traditional 510 (k) Premarket Notification is a submission for minor software and hardware changes to a current device, LTV 1200 510(k) No. K060647. These changes are features that have been added such as Sigh Breaths, Patient Circuit Mode and Adjustable rate in NIV.
The modified device also includes minor changes of appearance, operator controls and displays. There are no changes to the fundamental technology as compared to the predicate device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Monther Abushaban Regulatory Affairs Manager Carefusion 22745 Savi Ranch Parkway Yorba Linda, California 92887
JAN 2 6 250
Re: K101643
Trade/Device Name: Model LTV 1100 Ventilator Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: November 9, 2010 Received: January 6, 2011
> : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
Dear Mr. Abushaban:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Abushaban
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon contains a stylized letter "C" inside a circle. The text "CareFusion" is in a bold, sans-serif font.
510 (K) Number (if Known): ___________________________________________________________________________________________________________________________________________________ Device Name: LTV 1100 Ventilator
## Indications for Use
Device Name: LTV 1100
Indications for Use:
The LTV® 1100 ventilator is intended to provide continuous or intermittent ventilatory support for the care of the individuals who require mechanical ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5kg (11 lbs.), who require the following types of ventilatory support:
졌 Positive Pressure Ventilation, delivered invasively (via endotrach tube or trach tube) or non-invasively (via mask or nasal prongs).
X Assist/Control, SIMV, CPAP, and NPPV modes of ventilation.
The ventilator is suitable for use in institutional, home, or transport settings.
Prescription Use X (Part 21CFR 801 subpart D) AND/ OR
Over the Counter Use: (Part 21CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schult
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: *k101643*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.