K101641 · Ag Digital Technology Corp. · LHQ · Jun 16, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K101641
Device Name
AG THERMOGRAPHIC CAMERA, MODEL ATIR-M301
Applicant
Ag Digital Technology Corp.
Product Code
LHQ · Obstetrics/Gynecology
Decision Date
Jun 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2980
Device Class
Class 1
Indications for Use
The AG Thermographic Camera (Model no.: ATIR-M301) is intended to use as an adjunct to other clinical diagnostic procedures by quantifying differences in skin surface temperature changes. It can visualize and digitalize body surface temperature distribution and temperature changes. The device is intended for use by qualified healthcare personnel trained in its use.
Device Story
Non-contact, non-invasive infrared system; captures passive infrared emission from human body surface via uncooled microbolometer FPA; processes thermal data through proprietary PC-based software; displays temperature distribution patterns as images. Used by qualified healthcare personnel in hospitals, clinics, nursing homes, or home settings. Output consists of digital thermal images and temperature statistics; assists clinicians in visualizing surface temperature changes as an adjunct to other diagnostic procedures. Benefits include non-radiating, non-contact assessment of skin temperature distribution.
Clinical Evidence
Bench testing only. Performance validated via EN 60601-1 and EN 60601-1-2 electrical safety standards. Functional performance testing included temperature accuracy, environmental temperature factor, focus, distanced object temperature, thermal image handling, small area temperature statistics, two-area comparison, and patient history record handling.
Technological Characteristics
Uncooled microbolometer FPA (Amorphous Silicon); spectral response 8-14um; 320x240 array resolution; thermal sensitivity <0.1°C; electric motor focus (0.5M-infinity); USB 2.0 data output; PC-based processing and GUI display; AC power supply (110/240VAC).
Indications for Use
Indicated for use as an adjunct to clinical diagnostic procedures to visualize and quantify skin surface temperature changes and distribution in adult human patients. Intended for use by trained healthcare personnel in clinical, hospital, or home settings.
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
TELETHERMOGRAPHIC CAMERA, Model No. SERIES A, E, S AND P (K033967)
TELESIS DIGITAL INFRARED THERMAL IMAGE SYSTEM, Model No. SPECTRUM 9000MB (K020783)
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(K) SUMMARY
1 :: :
K101641
This summary of 510(k) safety and effectiveness information is being submitted in دccordance with the requirements of SMDA and 21 CFR §807.92
JUN 1 6 2011
REV.【C】
| 1. Submitter's Name: | AG Digital Technology Corp. |
|----------------------|---------------------------------------------------------|
| Address: | 11F, No.15, Sec.1, Chi-Nan Road Taipei, TAIWAN (CHINA) |
| | 10051 |
| Phone: | +886-2- 23575809 |
| Fax: | +886-2-23575845 |
| Contact: | Mr. Telvin Ju / Genral Manager |
| 2. Device Name : | |
| Trade Name: | AG Thermographic Camera, |
| | Model no.: ATIR-M301 |
| Common Name: | Telethermographic system |
| Classification name | Telethermographic system (adjunctive use) |
| 3. DEVICE CLASS | The AG Thermographic Camera (Model no.: |
| | ATIR-M301) has been classified as |
| | Regulatory Class: I |
| | Product Code: LHQ |
| | Regulation Number: 21 CFR 884.2980 |
| 4. Predicate Device: | The predicate device is the |
| | TELETHERMOGRAPHIC CAMERA , Model No. SERIES |
| | A, E, S AND P (K033967) marketed by FLIR SYSTEMS |
| | TELESIS DIGITAL INFRARED THERMAL IMAGE |
| | SYSTEM , Model No. SPECTRUM 9000MB (K020783) |
| | marketed by TELESIS TECHNOLOGIES INC. |
| Intended Use: | The AG Thermographic Camera (Model no.: ATIR |
| | Product: AG Thermographic Camera (Model no.: ATIR-M301) |
Section 4 - 510(k) Summary
Page 1 of 5
{1}------------------------------------------------
M301) is intended to use as an adjunct to other clinical diagnostic procedures by quantifying differences in skin surface temperature changes. It can visualize and digitalize body surface temperature distribution and temperature changes. The device is intended for use by qualified healthcare personnel trained in its use.
#### 6. Device Description: The AG Thermographic Camera (Model no.:
ATIR-M301) consists of an infrared camera, accessories, and PC visualization/analysis software. It is a non-contact, non-invasive, and non-radiating infrared system capable of imaging and storing thermal patterns generated by the human body. It employs passive infrared emission sensing technology to capture the thermal data and uses proprietary software to display the temperature distribution pattern as an image. It is suitable for imaging adult human targets and can be used in hospitals, acute and sub-acute healthcare settings, nursing homes, clinics, homes, public areas such as airports, and any environment where healthcare is provided by a healthcare professional.
### 7. Comparison to 510(k) Predicate Devices
The AG Thermographic Camera (Model no.: ATIR-M301) have the same intended use, principles of operation, and similar technological characteristics વડ predicate devices. Please find Technological Characteristics as follows
| Item | AG Digital<br>ATIR-M301 | FLIR A20M | TELESIS<br>SPECTRUM<br>9000MB |
|---------------------------|------------------------------------------|--------------------------------------|----------------------------------------|
| Technology | FPA uncooled<br>Mircrobolometer | FPA uncooled<br>Mircrobolometer | FPA uncooled<br>Mircrobolometer |
| Material | Amorphous<br>Silicon | Amorphous<br>Silicon | VOx |
| Spectral<br>Response | 8 ~ 14um | 7.5 ~ 13 um | 7 ~ 14 um |
| Array<br>Resolution | 320x240 | 160x120 | 320x240 |
| Thermal<br>Sensitivity | < 0.1 °C | 0.09~0.12°C | 0.07°C |
| Optics | 50mm F=0.7<br>FOV= 18.1°(H)<br>×13.7°(V) | F=1.2<br>FOV= 24°(H)<br>×18°(V) | 50mm F=0.83<br>FOV=<br>18.7°(H)×14°(V) |
| Focus | Electric motor<br>driving 0.5M -∞ | Manual, 0.3M - ∞ | Manual, 0.6M - ∞ |
| Spatial<br>Resolution | ≤ 1 mrad | 2.7mrad | 1 mrad |
| Temperature<br>Range | 30°C ~40°C | -20-+250 °C<br>(-4-+482 °F) | 10°C to 40°C |
| Temperature<br>Resolution | 0.1°C | 0.1°C | 0.1°C |
| Accuracy | < ±1°C | ± 2 °C or ± 2 %<br>of reading | - |
| Refresh Rate | 30 Frames/sec | 60 Frames/sec | 60 Frames/sec |
| Working<br>temperature | 20 - 26°C | -15°C ~ 50°C | 18°C~28°C |
| Operating<br>Temperature | 24°C ± 2°C | -15°C ~ 50°C | 20-22.2°C |
| Data Output | USB2.0 | RS-232,<br>IEEE-1394,<br>TCP/IP | RS-170 |
| Data<br>processing | personal<br>computer | Camera and<br>personal<br>compute | personal<br>compute |
| Display | GUI on PC | GUI on PC | GUI on PC |
| User<br>interface | Software on PC | Software on PC | Software on PC |
| Power Supply | AC adapter<br>110/240VAC,50/<br>60Hz | AC adapter<br>110/220VAC,50<br>/60Hz | 115~230 VAC<br>50~60 Hz |
| Weight | 1.3 Kgs | 1.7 LB | 3 LB |
Product: AG Thermographic Camera (Model no.: ATIR-M301) Section 4 - 510(k) Summary REV.【C】 Page 2 of 5
{2}------------------------------------------------
Product: AG Thermographic Camera (Model no.: ATIR-M301) Section 4 – 510(k) Summary Page 3 of 5 REV.【C】
ﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﺘﻲ ﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1999 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﺘﻲ ﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1999 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﺘﻲ ﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1999 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ
{3}------------------------------------------------
- 8. Substantial Equivalence Discussion
AG Thermographic Camera (Model no.: ATIR-M301) has the same general design with the predicate devices
Since the Applicant has selected Legally Marketed Devices
- TELETHERMOGRAPHIC CAMERA , Model No. SERIES . ● A, E, S AND P (K033967) marketed by FLIR SYSTEMS
- TELESIS DIGITAL INFRARED THERMAL IMAGE ● SYSTEM , Model No. SPECTRUM 9000MB (K020783) marketed by TELESIS TECHNOLOGIES INC.
The has the following similarities to the predicate devices in:
- · having the same intended use.
- · using similar operating principle..
- using similar technological characteristics
In summary, the AG Thermographic Camera (Model no.: ATIR-M301) described in this submission are, in our opinion, substantially equivalent to the predicate devices.
## 9. Performance AG Thermographic Camera (Model no.: ATIR-M301) meets EN 60601-1 (IEC 60601-1), EN 60601-1-2 (IEC Data 60601-1-2), and Performance tests including · Temperature accuracy • Environment temperature factor test · Focus test · Distanced object temperature test · Thermal image handling · Small area temperature statistics and two-area comparison
- · Patient history record handling
Product: AG Thermographic Camera (Model no.: ATIR-M301) Page 4 of 5 Section 4 - 510(k) Summary REV. [C)
{4}------------------------------------------------
### 10. Safety and
#### Effectiveness
ATIR-M301) meet electrical Safety standard such as EN 60601-1 (IEC 60601-1), EN 60601-1-2 (IEC 60601-1-2). so it is safe for human body. Additionally, Performance tests including
- · Temperature accuracy
- Environment temperature factor test
The AG Thermographic Camera (Model no .:
- Focus test
- Distanced object temperature test
- · Thermal image handling
- · Small area temperature statistics and two-area comparison
- · Patient history record handling
were conducted to ensure the device can work as the intended performance.
Therefore, the device is of safety and effectiveness.
#### Conclusion
AG Thermographic Camera (Model no.: ATIR-M301) have the same intended use, principles of operation, and similar technological characteristics as predicate devices. Moreover, bench testing contained in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the AG Thermographic Camera (Model no.: ATIR-M301) are substantially equivalent to the predicate devices.
Product: AG Thermographic Camera (Model no.: ATIR-M301) Section 4 - 510(k) Summary REV.【C】 Page 5 of 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AG Digital Technology Corp. % Ms. Jennifer Reich Senior Consultant Harvest Consulting Corp. 2904 N. Boldt Drive FLAGSTAFF AZ 86001
JUN 1 6 201
Re: K101641
Trade/Device Name: AG Thermographic Cmaera, Model no: ATIR-M301 Regulation Number: 21 CFR 884.2980 Regulation Name: Telethermographic system Regulatory Class: I Product Code: LHQ Dated: January 26, 2011 Received: January 7, 2011
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{6}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K101641
Device Name: AG Thermographic Camera, Model no.: ATIR-M301 AG Digital Technology Corp.
Indications For Use:
The AG Thermographic Camera(Model no.: ATIR M301) is intended to use as an adjunct to other clinical diagnostic procedures by quantifying differences in skin surface temperature changes. It can visualize and digitalize body surface temperature distribution and temperature changes. The device is intended for use by qualified healthcare personnel trained in its use.
Prescription Use AND/OR ﺳﮯ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael D'Orazio
(Division Sign Off)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K101 641
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.