The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
Device Story
Sharps Container Lns-T1 is a single-use, injection-molded polypropylene container designed for safe disposal of used needles and syringes. Used in hospitals, clinics, labs, and emergency vehicles by healthcare professionals. Features include biohazard warning labels, puncture-resistant walls, and a locking lid mechanism to prevent accidental needle sticks and leakage during storage, handling, or transport. Device sits on floors or desks; provides physical containment of medical waste to reduce sharps-related injury risk.
Clinical Evidence
Bench testing only. Device performance validated against BS7230:1990, ASTM F2132:2008, and OSHA 29CFR1910 standards for mechanical and environmental safety. No clinical data provided.
Indicated for healthcare professionals and small quantity waste generators in facilities including nursing stations, labs, dental/veterinarian offices, and emergency vehicles for safe disposal of hazardous sharps.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
K 101567 2 2010 SEP
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: June 7, 2009
1. Company and Correspondent making the submission:
Name - Cixi Leinuo Plastic Co., Ltd
Address - No. 56, Luoming South Road, Guanhaiwei Town, Cixi , Ningbo,
Zhejiang, China,315316
Telephone - +86-574-63673001
Fax - +86-574-63671001
Contact - Mr. Leiping Wang
Email – cxleinuo@163.com
2. Device :
Trade/proprietary name: Sharps Container Lns-T1
Common Name : Sharps Container Classification Name : Container, Sharps
Predicate Devices:
| Model | Manufacturer | K Number | Submitted Device |
|-----------------------------------------|-----------------|----------|------------------|
| Sharps Away<br>Disposable<br>Containers | Solutions, Inc. | K072667 | Lns-T1 |
3. Classifications Names & Citations :
21CFR 884.6120, MMK, Sharps Container, Class II
15 of 58
{1}------------------------------------------------
#### 4. Description :
4.1 General
The Sharps Container is a piece of equipment used in hospitals, physicians' offices, dental offices, laboratories, home health areas, patient rooms and other locations of medical waste. The sharps container is designed to safely contain used needles or syringes and prevent accidental needle sticks from used syringes or needles.
## Structure
- The device is made of injected molded polypropylene material. The material . is made of molded plastic into a specific shape. The actual shape is designed to sit on the floor or a desk.
- · On the outer casing of all the sharps containers is the Biohazard Warning label in Orange, noticeably displayed.
- The molded plastic container is designed and manufactured such that there . will be no leakage during storage, handling or transport.
for the daving thinker
| Please refer to the table below for the device thickness. | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Comments of the comments of the comments of the comments of | - A L L L L L L L L L L L L L L L L B L B R L B R B R B R B R B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L B L | | |
| | | | | |
| Model Name | Size (cm) | Volume (L) | Empty<br>Weight<br>(grams) | Thickness of material |
|------------|-----------|------------|----------------------------|-----------------------|
| LNS-T1 | 10.5*8*18 | 1.0 | 80g | body:1.2mm lid:1.3mm |
5. Indication for use : Device Name:
Lns-T1
Indications for Use:
The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
{2}------------------------------------------------
6. Comparison with predicate device :- Please see the next page
:
. . . . . .
.
. . . .
and the same of the same of the same
:
:
:
.
1. September 2017
and the comments of the comments of the comments of
{3}------------------------------------------------
| Element of comparison | Subject Device | Claimed SE Device |
|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Company | Cixi Leinuo Plastic Co., Ltd. | Solutions Inc. |
| FDA510(K) Number | N/A | K072667 |
| Device Name | Sharps Container | Sharps Away Disposable Container |
| Intended use(s) | Leinuo sharps containers are<br>intended to be used for the safe<br>disposal of hazardous sharps. | Sharps Away Disposable Containers are<br>intended to be used for the safe disposal<br>of hazardous sharps. |
| Target Population | Healthcare professional | Healthcare professional |
| where used: Hospital, Physicians<br>Offices, Dental offices, Laboratories,<br>Home health, Patient room, etc. | Yes | Yes |
| Material | Polypropylene | Polypropylene |
| Sharps closure | Flaps are closed and locked in place<br>for removal | Flaps are closed and locked in place for<br>removal |
| Impact Resistance | Yes | Yes |
| Puncture Resistance | Yes | Yes |
| Leak Resistance | Leak-proof on the sides and bottom | Leak-proof on the sides and bottom |
| Single Use | Yes | Yes |
18 of 58
:
{4}------------------------------------------------
- 7. Performance Testing :
Mechanical, environmental safety and performance testing have been accomplished according to BS7230:1990, ASTM F2132:2008 and OSHA 29CFR1910. The product passed all tests. These tests demonstrate this product is substantially equivalent to the predicate device.
## 8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification The Cixi Leinuo Plastic Co., Ltd. Sharps Container, Model Lns-T1 is substantially equivalent to the predicate devices as described herein.
- 9. Cixi Leinuo Plastic Co., Ltd will update and include in a summary any other information deemed Yeasonably necessary by the FDA.
END
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cixi Leinuo Plastic Company, Limited C/O Mr. Charles Mack IRC 77325 Joyce Way Echo, Oregon 97826
S.. 2 2010
Re: K101567
Trade/Device Name: Sharps Container, Models Lns-T1 Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: March 25, 2010 Received: June 4, 2010
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Mr. Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
Device Name: Sharps Container, Models Lns-T1 Indications For Use:
The sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
Prescription Use ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth T. Clavie-Will
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number: KI01567
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.