← Product Code [KGN](/productcode/KGN) · K101546

# ENDOFORM DENTAL TEMPLATE (K101546)

_Mesynthes, Ltd. · KGN · Jun 23, 2010 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K101546

## Device Facts

- **Applicant:** Mesynthes, Ltd.
- **Product Code:** [KGN](/productcode/KGN.md)
- **Decision Date:** Jun 23, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

Endoform™ Dermal Template is supplied sterile and is intended for single use in the treatment of the following wounds: partial and full-thickness wounds . pressure ulcers . venous ulcers . diabetic ulcers . chronic vascular ulcers . tunneled/undermined wounds . surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, . podiatric, wound dehiscence) trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) . . draining wounds

## Device Story

Endoform™ Dermal Template is a sterile, single-use wound care product composed of one or two layers of ovine extracellular matrix. Supplied as a perforated or non-perforated sheet up to 400 cm2, it is applied topically to the wound site. The device acts as a scaffold to support wound healing. It is intended for use by healthcare professionals in clinical settings for various wound types, including chronic ulcers, surgical wounds, and trauma-related injuries. The device provides a physical matrix to facilitate tissue repair, potentially benefiting patients by promoting healing in complex or chronic wound environments.

## Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility and performance were established via non-clinical testing.

## Technological Characteristics

Material: Ovine extracellular matrix. Form factor: Sterile perforated or non-perforated sheets up to 400 cm2. Single-use. No electronic components or software.

## Predicate Devices

- Endoform™ Dermal Template ([K092096](/device/K092096.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1/2

JUN 2 3 2010

## 510(k) K101546 Summary . for the Endoform™ Dermal Template

#### 1. SUBMITTER/510(K) HOLDER

Mesynthes Ltd 69 Gracefield Road Lower Hutt Wellington New Zealand 5010

| Contact Person | Brian Ward       |
|----------------|------------------|
| Telephone:     | (+64 4) 931-3275 |

Date Prepared: June 17, 2010

### 2. DEVICE NAME

| Proprietary Name:  | Endoform™ Dermal Template               |
|--------------------|-----------------------------------------|
| Common/Usual Name: | Extracellular matrix wound care product |
| Classification:    | Unclassified                            |

#### 3. PREDICATE DEVICES

- Endoform™ Dermal Template (Mesynthes) (K092096) .
#### 4. DEVICE DESCRIPTION

The Endoform™ Dermal Template is a wound care device comprised of one or two layers of ovine extracellular matrix and is supplied as a sterile perforated or non perforated sheet ranging in size up to 400 cm2.

### 5. INTENDED USE

Endoform™ Dermal Template is supplied sterile and is intended for single use in the treatment of the following wounds:

- partial and full-thickness wounds .
- pressure ulcers .
- . venous ulcers

{1}------------------------------------------------

K101546 page 2/2

- diabetic ulcers .
- chronic vascular ulcers .
- tunneled/undermined wounds .
- surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, . wound dehiscence)
- trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) .
- . draining wounds

#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The substantial equivalence of this device was established based on a declaration of conformity to design control requirements and on an assessment of the effects of the changes to the predicate device based on a risk assessment. As such, the Endoform™ Dermal Template (K101546) is substantially equivalent to the predicate device with respect to material composition, device characteristics and intended use.

#### 7. PERFORMANCE TESTING

The Endoform™ Dermal Template (K092096) has been subjected to extensive non-clinical testing to assess the biocompatibility and the performance of the device. Endoform™ Dermal Template (K092096) was shown to be safe and effective as a wound care product.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 2 3 2010

Mesynthes % Medical Device Consultants, Inc. Ms. Mary McNamara 49 Plain Street North Attleboro, Massachusetts 02760-4153

Re: K101546 · Trade/Device Name: Endoform™ Dermal Template Product Code: KGN Dated: June 2, 2010 Received: June 3, 2010

Dear Ms. McNamara:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Mary McNamara

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

incerely yours,

Mark N. Melkerson

Mark N. Melkers Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

K101546 510(k) Number:

#### Endoform™ Dermal Template Device Name:

#### Indications for Use:

Endoform™ Dermal Template is supplied sterile and is intended for single use in the treatment of the following wounds:

- partial and full-thickness wounds .
- pressure ulcers .
- venous ulcers .
- diabetic ulcers �
- chronic vascular ulcers .
- tunneled/undermined wounds ●
- surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, . podiatric, wound dehiscence)
- trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) .
- draining wounds .

Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Keene for MXM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101546

---

**Source:** [https://fda-staging.innolitics.com/device/K101546](https://fda-staging.innolitics.com/device/K101546)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).