K101535 · Utah Medical Products, Inc. · OQY · Jan 11, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K101535
Device Name
BT-CATH
Applicant
Utah Medical Products, Inc.
Product Code
OQY · Obstetrics/Gynecology
Decision Date
Jan 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K101535 · Jan 11, 2011
BT-CATH
Utah Medical Products, Inc.
Clinical feedback from routine clinical use
Clinical feedback from approximately 840 uses of the BT-Cath by U.S. clinicians was used to demonstrate that the device consistently achieved its intended purpose without reported patient injury.
Clinical feedback; Postmarket experience; Routine clinical use
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical feedback of BT-Cath use; Retrospective review of clinical feedback/experience
Patients undergoing treatment for postpartum hemorrhage; Sample Size: Approximately 840 uses
Not applicable for this study
Achievement of intended purpose and absence of patient injury
Indications for Use
BT-Cath is intended to provide temporary control or reduction of uterine bleeding during postpartum hemorrhage that is unresponsive to standard conservative therapy including uterine massage and oxytocin administration.
Device Story
BT-Cath is an intrauterine balloon tamponade catheter used to manage postpartum hemorrhage. The device consists of an extruded silicone dual-lumen catheter with a molded silicone tamponade balloon at the distal end. A clinician inserts the device into the uterus; saline is infused through one lumen to inflate the balloon, which provides mechanical compression against bleeding sites (torn arteries, ruptured sinuses, lacerations). The second lumen acts as a drainage port for blood and fluids. A proximal stopcock retains saline, and a check valve facilitates filling. The device is used in clinical settings by healthcare providers. By maintaining uterine compression, the device allows the uterus to contract and helps control bleeding, potentially preventing further patient injury.
Clinical Evidence
No formal clinical trial data provided. Evidence consists of mechanical bench testing (structural integrity, burst volume, leak tests, tensile strength) and retrospective clinical feedback from approximately 840 uses by U.S. clinicians, which indicated the device consistently achieved its intended purpose without reported patient injury.
Technological Characteristics
Extruded silicone dual-lumen catheter with molded silicone tamponade balloon. Latex-free. Sterile. Mechanical compression principle. Proximal stopcock and check valve for saline inflation and drainage. Dimensions and configuration similar to predicate.
Indications for Use
Indicated for temporary control or reduction of postpartum uterine hemorrhage in patients unresponsive to standard conservative therapy, including uterine massage and oxytocin administration.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
K101535...
Page 1/2
Image /page/0/Picture/1 description: The image shows the logo for Utah Medical Products Inc. The logo consists of the text "UTAH MEDICAL PRODUCTS INC." in a sans-serif font, with a line underneath the text. Below the text is a stylized symbol that resembles the letter "U" or a horseshoe shape, with a vertical line extending upwards from the center of the "U" shape.
JAN 1 1 2011
## SECTION 9: 510(k) SUMMARY
and the state of the states and
| 510(k) Owner:<br>Address: | Utah Medical Products, Inc.<br>7043 South 300 West<br>Midvale, UT 84047 USA |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone No .:<br>Fax No .: | (801) 566-1200<br>(801) 566-2062 |
| Contact: | Kevin L. Cornwell, CEO |
| Date: | 3 January 2011 |
| Trade Name: | BT-Cath |
| Common Name: | Intrauterine Balloon Tamponade Catheter |
| Classification: | Obstetric-gynecologic specialized manual instrument,<br>21 CFR 884.4530; Product Code OQY |
| Predicate Devices: | Bakri Uterine Balloon Tamponade, (K013597, K062438)<br>Cook Ob/Gyn |
| Indications for Use: | BT-Cath is intended to provide temporary control or reduction of<br>uterine bleeding during postpartum hemorrhage that is unresponsive<br>to standard conservative therapy including uterine massage and<br>oxytocin administration. |
{1}------------------------------------------------
# 7080 2
#### Device Description:
Comparison to predicate devices:
Performance Data:
The BT-Cath consists of an extruded silicone dual lumen catheter with an inflatable molded silicone tamponade balloon attached on the distal end. One lumen of the catheter serves as the conduit through which saline is infused into the balloon after insertion into the bleeding uterus. When inflated properly to press against the uterine wall, the balloon provides compression against multiple sites of torn arteries, ruptured sinuses, lacerations and other tissue evoking hemorrhage. The second lumen, which is open at the distal end of the catheter, provides a drainage port for blood and other bodily fluids. At the proximal end of the catheter, a stopcock retains the saline in the balloon until a clinician gradually drains it allowing the uterus to contract. For ease of filling the balloon with multiple syringes, a check valve is also provided at the proximal end. The device is latex-free and provided sterile.
There are insignificant differences in the physical dimensions and physical configuration of the balloon tamponade catheters. The technological characteristics, methods of use, materials, intended use, indications for use and contraindications are the same as the predicate devices. Biocompatibility and shelf-life testing were also assessed to demonstrate substantial equivalence.
Substantial equivalence is also based on an assessment of mechanical tests where the structural integrity of the balloon, including the relationship of volume of saline infused into the balloon and resulting balloon diameter, the burst volume of the balloon, and the ability of the balloon to hold its shape and not leak during repeated and prolonged inflations; the structural integrity of the catheter using leak tests; and the tensile strength in each joint in the devices were evaluated. Both the BT-Cath and the Bakri device were tested and found to achieve structural integrity with a reasonable safety factor. Clinical feedback was obtained after about 840 uses of BT-Cath by U.S. clinicians that indicates the device consistently achieved its intended purpose without any reported likelihood of patient injury.
K. Howell
Kevin L. Cornwell
Kevin L. Cornwell Chairman & CEO Utah Medical Products, Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN USA" around the edge. Inside the circle is a stylized eagle symbol. The eagle is depicted with three thick, curved lines representing its body and wings. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Kevin L. Cornwell Chairman & CEO Utah Medical Products, Inc. 7043 South 300 West MIDVALE UT 84047
JAN 1 1 201
Re: K101535 Trade Name: BT-Cath® Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: OQY Dated: January 3, 2011 Received: January 3, 2011
Dear Mr. Cornwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Herbert Semler MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### SECTION 3: INTENDED USE / INDICATIONS for USE
| 510(k) Number: | K101535 | JAN 1 1 2011 |
|----------------|---------|--------------|
|----------------|---------|--------------|
Device Name: UTMD BT-Cath - Balloon Tamponade Catheter for Postpartum Uterine Hemorrhage (BTC-100)
#### Indications for Use:
BT-Cath (BTC-100) is intended to provide temporary control or reduction of uterine bleeding during postpartum hemorrhage that is unresponsive to standard therapy including massage and oxytocin administration.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _ X (Per 21 CFR 801.109) OR
Lori K. Wertz
Visible Sign Off
10
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.