MICROLIFE UPPER ARM AUTOMATIC DIGITAL BLOOD PRESSURE MONITOR, WATCHBP OFFICE TARGET (BP3MD1-4).
Applicant
Microlife Intellectual Property GmbH
Product Code
DXN · Cardiovascular
Decision Date
Jul 1, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
Device Story
Device measures systolic/diastolic blood pressure, pulse rate, and pulse pressure using oscillometric method; inflatable cuff wraps around upper arm; resistive pressure sensor converts cuff pressure changes to electrical signals; software algorithm processes signals to determine BP and pulse. Device features target BP setting (140/90 mmHg or 130/80 mmHg for patients with diabetes, coronary heart disease, or chronic kidney disease); visual feedback via red/green backlight based on target comparison; pulse pressure calculation. Used in clinical or home settings; operated by patient or clinician. Output displayed on screen; assists in monitoring hypertensive patients and those with specific comorbidities.
Clinical Evidence
No clinical data provided. Substantial equivalence based on technical identity to predicate BP3BT0-1 regarding blood pressure measurement technology. Safety and effectiveness supported by bench testing including reliability (storage, operating, vibration, drop, life) and EMC testing in accordance with FDA guidance and ANSI/AAMI SP10:2008.
Technological Characteristics
Non-invasive oscillometric blood pressure monitor. Uses resistive pressure sensor, automated pump, and factory-set exhaust valve. Features include target BP setting, red/green backlight, and pulse pressure calculation. Software-based signal processing. Class II device.
Indications for Use
Indicated for adult individuals to measure systolic/diastolic blood pressure, pulse rate, and pulse pressure via non-invasive oscillometric technique using an upper arm cuff.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3BT0-K013485
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (Twin200), K082357
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K101572
#### 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland
JUL -1 2010
Date Summary Prepared: May 25, 2010 .
Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch
#### 2. Name of the Device:
Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4)
Requlation Number: 21 CFR Part 870.1130 Requlation Name: Non-Invasive Blood Pressure Measurement System Requiatory Class: II Product Code: DXN
#### 3. Information for the 510(k) Cleared Device (Predicate Device):
- a. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3BT0-K013485. Microlife Corporation.
- b. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (Twin200), K082357, Microlife Intellectual Property GmbH.
#### 4. Device Description:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Tarqet (BP3MD1-4) is designed to measure systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an individual by using a noninvasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses a resistive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".
{1}------------------------------------------------
The target blood pressure value 140/90 mmHg is used for hypertensive patient. The target blood pressure value 130/80 mmHq is used for hypertension patients with diabetes, coronary heart disease or chronic kidney disease. If both the measured systolic and diastolic blood pressure values are below the set target values (140/90 or 130/80), the screen will light in green with two <<OK>> signs and show the pulse pressure (PP). If either the measured systolic result is above the set target value (140/90 or 130/80), the screen will light in red with the difference to the goal value shown in display as number.
#### 5. Intended Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
#### Comparison to the 510(k) Cleared Devices (Predicate Devices): 6.
a) Subject Modified Device Compared to BP3BT0-1 Predicate, K013485:
The subject modified device, Model WatchBP Office Target (BP3MD1-4) and the predicate device model BP3BT0-1, use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. Upper arm cuff is inflated automatically, deflation rate is controlled by one factory set exhaust valve and the deflation pressures are transferred to one sensor.
The differences between the two models are the sensor type and additional features such as blood pressure targets setting function, red/green backlight function and PP automatically calculation function. However, the differences do not affect the accuracy and normal use of this device based on the internal clinical tests comparing different sensors and cuff bladder materials.
- b) Subject Modified Device Compared to WatchBP Office (Twin200) Predicate, K082357:
The subject modified device. WatchBP Office Target (BP3MD1-4) uses the same oscillometric method as the predicate WatchBP Office (Twin200) to determine the svstolic and diastolic blood pressure, pulse rate and pulse pressure (PP). Upper arm cuff is inflated automatically by pump. the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units.
They differ by the intended use, measuring location, measuring mode, power source type. MAP automatically calculation function etc., however, the differences do not affect the accuracy and normal use of this device because they use the same fundamental scientific technology. Based upon the aforementioned information, the two devices are substantially equivalent.
The intended use of the subject modified device, the WatchBP Office Target (BP3MD1-4), has not changed from the 510(k) cleared predicate device labeling
{2}------------------------------------------------
since it is a combination of the two (2) predicate devices stated in this submission. All comparison parameters stated in our comparison chart (included in the submission) come from either one of the predicate devices cited.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.
The following testing was conducted:
- a. Reliability Test Storage test
- b. Reliability Test Operating test
- c. Reliability Test Vibration test
- d. Reliability Test Drop test
- e. Reliability Test Life test
- EMC Test نه
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) tested met all relevant requirements of the aforementioned tests.
#### 8. Discussion of Clinical Tests Performed:
The subject modified device Model WatchBP Office Target (BP3MD1-4) is from the technical point of view, identical to the blood pressure monitor Model BP3BT0-1. The differences between them do not related to blood pressure measurement technology; therefore the clinical accuracy in terms of blood pressure detection will not be affected. Based on Microlife's risk analysis, repeated clinical testing in accordance with ANSI/AAMI SP10 is therefore not required.
#### 9. Software information:
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### 10. Conclusions:
We have demonstrated that there are no significant differences between the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) and the predicate devices, in terms of safety and effectiveness based on electrical, mechanical and environmental test results per the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and the ANSI/AAMI Voluntary Standard, SP10: 2008.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings, rendered in black lines. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Microlife Intellectual Property GmbH Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021
**JUL - 1 2010**
Re: K101512
Device Name: Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (Two) Product Code: DXN Dated: May 26, 2010 Received: June 1, 2010
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Exhibit B
# Indications for Use
510(k) Number (if known):
K101512
Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4)
### Indications For Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Target (BP3MD1-4) is a device intended to measure the systolic and diastolic blood pressure, pulse rate and pulse pressure (PP) of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
J.W.C.
**Page** 1 of 1
ivision Sign-C diovascular Devices Division of Car
510(k) Number K101502
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.