K101338 · Technologia Y Diseno Industrial, S.A DE C.V · KTT · Mar 14, 2011 · Orthopedic
Device Facts
Record ID
K101338
Device Name
NEWFIX EXTERNAL FIXATION SYSTEM
Applicant
Technologia Y Diseno Industrial, S.A DE C.V
Product Code
KTT · Orthopedic
Decision Date
Mar 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Newfix® External Fixation System, consisting of axial fixators and frame components, is indicated for stabilization of open and/or unstable fractures where soft tissue injury precludes the use of other fracture treatments such as IM rodding or casting. NEWFIX® External Fixation System is intended for the fixation and for temporary or extended stabilization in cases such as: open fractures, closed fractures, poly-trauma fractures, supplement in the stabilization of minimal fixations in intra and extra articular fractures.
Device Story
Modular external fixation system; components include clamps, distractors, pins, carbon fiber rods, stainless steel tubes. Used in orthopedic/trauma settings for fracture stabilization. Physician assembles frame constructs using clamps (rod-to-rod, pin-to-rod) and fixation pins (smooth or self-drilling/self-tapping with diamond tip). System provides temporary or extended stabilization for fractures where internal fixation or casting is contraindicated. Benefits include modularity, frame construction flexibility, and stabilization of complex fractures. Supplied non-sterile for single-patient use.
Clinical Evidence
Bench testing only. Testing included validation of locking mechanisms per ASTM F 1541 and engineering analysis comparing Newfix components to Orthofix predicate devices. Materials and mechanical performance were verified using force verification machines and torque wrenches.
Indicated for stabilization of open/unstable fractures in pediatric and adult patients where soft tissue injury precludes IM rodding or casting; includes open, closed, and poly-trauma fractures, and stabilization of intra/extra-articular fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Orthofix Modulsystem Dynamic Axial Fixation System (K831576, k955848)
Synthes ® External Fixation System (K961350, k011034, k090658)
Extrafix External fixation System by QFX Technologies (k091258)
Tecnología y Diseño Industrial (TDI) Newfix Screw, Wire, and Fixation System (K101254)
Treu-Instrumente GmbH Bone Fixation Screws and Pins (K083912)
Submission Summary (Full Text)
{0}------------------------------------------------
Tecnología y Diseño Industrial. S.A.P.I. de C.V Newfix® External Fixation System k101338 510(k) Summary
#### MAR 1 4 2011 SUMMARY OF SAFETY AND EFFECTIVENESS
March 11, 2010
Trade Name: Newfix® External Fixation System
Common Name: External Fixation System
Applicant:
Tecnología y Diseño Industrial, S.A.P.I. de C.V. José Adán Ortega Torres Operation Manager
José Guadalupe Zuno 48 Col. Parque Industrial los Belenes Zapopan, Jalisco Mexico, C.P. 45101
Tel: +52 (33) 38 36 45 00 Ext. 2960 +52 (33) 36245132 Fax: +52 (33) 38 36 45 02
Classification Name: Single/multiple component metallic bone fixation appliances and accessories
Classification Panel: Orthopedic
All questions and/or comments concerning this document should be made to:
Robert J. Bard, Esq. Managing Director
HealthCare Technologies Consultants PO Box 506 South Lyon, MI 48178 Tel: 248-573-5040 Cell: 734-330-5990 email: rbard@reglaw.net
{1}------------------------------------------------
K101138
Tecnología y Diseño Industrial, S.A..P.I. de C.V. Newfix® External Fixation System k101338 510(k) Summary
#### 1.0 DEVICE SUMMARY
The Tecnología y Diseño Industrial, S.A. de C.V. Newfix® External Fixation System consists of various components including fixators, clamps, bars, distractors, carbon fiber rod, and stainless steel rods for use in orthopedic and trauma treatments.
#### 1.1 Classification Information
### Table SE1: Device Classification
| Classification or descriptor | Name or designation |
|------------------------------|--------------------------------------------------------------------------------|
| Common Name | External Fixation System |
| Device Trade Name | Newfix® External Fixation System |
| Device Classification Name | Single/multiple component metallic bone fixation<br>appliances and accessories |
| Device Classification | Class II |
| Reviewing Panel | Orthopedic |
| Regulation Number | 21 CFR 880.3030 |
| Product Code | KTT |
#### 2.0 PREDICATE DEVICES
The Newfix® External Fixation System is substantially equivalent to the following predicate products.
# Table SE2: Predicate Products
| Manufacturer and Product | Cleared Predicate<br>Product |
|---------------------------------------------------------------------------------|------------------------------|
| Orthofix Modulsystem Dynamic Axial Fixation System | K831576, k955848 |
| Synthes ® External Fixation System | K961350, k011034,<br>k090658 |
| Extrafix External fixation System by QFX Technologies | k091258 |
| Smith & Nephew External Fixation System | K994143 |
| Hoffmann II Compact External Fixation System | K971755 |
| Zimmer Wristore Distal Radius Fracture Fixator | k012294 |
| Tecnología y Diseño Industrial (TDI) Newfix Screw, Wire, and<br>Fixation System | K101254 |
| Treu-Instrumente GmbH Bone Fixation Screws and Pins | K083912 |
{2}------------------------------------------------
16101338
Tecnología y Diseño Industrial, S.A..P.I. de C.V. Newfix® External Fixation System k101338 510(k) Summary
#### 3.0 INTENDED USE AND INDICATIONS FOR USE
The Newfix® External Fixation System, consisting of axial fixators and frame components, is indicated for stabilization of open and/or unstable fractures where soft tissue injury precludes the use of other fracture treatments such as IM rodding or casting. NEWFIX® External Fixation System is intended for the fixation and for temporary or extended stabilization in cases such as: open fractures, closed fractures, poly-trauma fractures, supplement in the stabilization of minimal fixations in intra and extra articular fractures.
#### 4.0 DEVICE DESCRIPTIONS
- 4.1 The Newfix External Fixation System is a modular system that consists of various components including clamps, distractors, pins, carbon fiber rod, and stainless steel tubes for use in orthopedic and trauma treatment.
- 4.2 The system is a modular design to provide options in frame construction, simplicity in frame components, and ease of use. The system is comprised of aluminum and stainless steel clamps (rod-to-rod and pin-to-rod), stainless steel fixation pins and stainless steel and carbon fiber connector rods. Pins include smooth and self drilling/self-tapping type threaded pin with diamond tip and cutting edges.
- 4.3 The Newfix® External Fixation System is supplied non sterile and is intended for single patient use. None of the components of the Newfix System are reusable. The Newfix® External Fixation System includes instruments and accessory devices necessary for its use.
{3}------------------------------------------------
Tecnologia y Diseño Industrial, S.A..P.I. de C.V.
COMPARISON OF DEVICE UNDER REVIEW AND ITS PREDICATI 5.0
| Comparison<br>Element | TDI Newfix® External Fixation<br>System K101338 | Orthofix Modulsystem Dynamic<br>Axial Fixation System K955848 | Synthes® External Fixation<br>System K961350, k011034, k090658 | XtraFix (Extrafix) External fixation<br>System k091258 | Comparison<br>Element | Materials | Duration of Use | Reuse Capability |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------|------------------|
| Manufacturer | Tecnología y Diseño Industrial<br>S.A.P.I. de C.V. | Orthofix | Synthes (USA) | ExtraOrtho (QFX Technologies) | TDI Newfix® External Fixation<br>System K101338 | Carbon fiber<br>Stainless Steel 17-4PH (AISI 630)<br>Stainless steel 304<br>Titanium alloys<br>Aluminum 6061-T6<br>Stainless steel 316LS | Greater than 30 days | Single use |
| Indication for/<br>Intended Use | The Newfix® External Fixation<br>System, consisting of axial fixators<br>and frame components, is indicated<br>for stabilization of open and/or<br>unstable fractures where soft tissue<br>injury precludes the use of other<br>fracture treatments such as IM<br>rodding or casting. NEWFIX®<br>External Fixation System is<br>intended for the fixation and for<br>temporary or extended stabilization<br>in cases such as: open fractures,<br>closed fractures, poly-trauma<br>fractures, supplement in the<br>stabilization of minimal fixations in<br>intra and extra articular fractures. | Orthofix Dynamic Axial Fixation<br>System is a unilateral external<br>fixation device, which is intended<br>for use in the treatment of bone<br>conditions including leg<br>lengthening, osteotomies,<br>arthrodesis, fracture fixation and<br>other bone conditions amenable to<br>treatment by use of the external<br>fixation modality." | The Synthes (USA) External Fixation<br>Systems (K961350, k011034,<br>k090658) are intended for use in the<br>construction of an external fixator<br>frame for the treatment of pediatric<br>and adult fractures. | The Extrafix External Fixation System<br>(k091258) includes various elements<br>designed to build a fixator construct. The<br>system includes clamps, posts, bars, and<br>fixation pins.<br>The Extrafix External Fixation System is<br>indicated for use in construction of an<br>external fixation frame for treatment of<br>long bone (foot, femur, and tibia) and<br>pelvic fractures that require external<br>fixation. Specifically, the system is<br>intended for:<br>o Temporary stabilization of open or<br>closed acute fractures with soft tissue<br>injuries;<br>o Definitive stabilization of open or closed<br>fractures where open or alternative<br>closed<br>treatment is undesirable or otherwise<br>contraindicated;<br>o Stabilization of fractures in the context<br>of polytrauma;<br>o Temporary or definitive stabilization of<br>certain pelvic fractures or pelvic ring<br>injuries;<br>o Arthrodesis and osteotomies with<br>associated soft tissue problems;<br>o Stabilization of limbs after removal of<br>total joint (knee and ankle) arthroplasty<br>for infection or other failure;<br>o Neutralization of fractures stabilized<br>with limited internal fixation;<br>o Stabilization of non-unions; and<br>o Intraoperative temporary stabilization<br>tool to assist with indirect reduction. | Orthofix Modulsystem Dynamic<br>Axial Fixation System K955848 | Aluminum, stainless steel,<br>titanium, and composite materials | Greater than 30 days | Single use |
| Synthes®. External Fixation<br>System K961350, k011034, k090658 | Clamps -Stainless steel and titanium<br>alloy<br>Dynamization Clip - Stainless steel<br>Rod Attachment - Stainless steel and<br>titanium alloy<br>Rods - Carbon fiber reinforced epoxy<br>(CFRE)<br>Screws - stainless, titanium alloys | Greater than 30 days | Single use | | | | | |
| XtraFix (Extrafix) External fixation<br>System k091258 | Aluminum, stainless steel, titanium, and<br>composite materials | Greater than 30 days | Single use | | | | | |
上101338
Page 4 of 6
{4}------------------------------------------------
Tecnologia y Diseño Industrial, S.A..P.I. de C.V.
Page 5 of 6
{5}------------------------------------------------
-101336
Tecnología y Diseño Industrial, S.A..P.I. de C.V. Newfix® External Fixation System k101338 510(k) Summary
#### 5.1 Performance Tests
- 5.1.1 Performance tests were conducted to validate the locking mechanism of the Newfix External Fixation System against various components of the Orthofix predicate devices.
# TDI Products tested:
- · E0300104 Lateral Cylinder for Pins (Straight Clamp).
- · E0600201 T-Shaped Clamp.
- . E0600301 Angled Clamp.
- E0600401 Ankle Clamp. I
- · E0900104/5 RadioLucent Wrist Axial Fixator Clamp.
# Predicate Devices tested:
- · Orthofix ProCallus Fixator Straight Clamp
- · Orthofix T-Shaped Clamp
- Orthofix Ankle Clamp .
- · Orthofix RadioLucent Wrist Fixator Clamp
# Materials used:
- · Ø6mm (0.236in) and Ø4mm (0.157in) series 316 stainless steel rods.
- Ø6mm (0.236in) and Ø4mm (0.157in) series 316 stainless steel rods with . a hexagonal nut for torque tests.
- . Socket Set Screws Flat Point ISO 4026 M8x1.25.
# Measuring tools:
- · TDI.QC.023 PTS Analog push pull gauge SKN-5 S/N3409090760
- · TDI.QC.027 Dial torque wrench 6178A S/N 0207801708
### Referenced Standards
- · ASTM F 1541 Standard Specification and Test Methods for External Skeletal Fixation Devices.
- · ASTM E 4 Practices for Force Verification of Testing Machines
#### 6.0 CONCLUSION
- Interconnection testing per ASTM F 1541 and engineering analysis comparing lock 6.1 mechanisms demonstrated equivalence to predicates.
- The Newfix External Fixation System is substantially equivalent to the identified predicate 6.2 systems based on the substantial equivalence of indication for use, design features, operating principles, performance tests and material of composition.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Technologia y Diseño Industrial, S.A.P.I. de C.V. % Healthcare Technologies Consultants Mr. Robert Bard PO Box 506 South Lyon, Michigan 48178
MAR 1 4 2011
Re: K101338
Trade/Device Name: Newfix® External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: February 25, 2011 Received: March 01, 2011
Dear Mr. Bard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Robert Bard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Alvy B. Benson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indication for Use
510(k): k101338
Device Name: Newfix® External Fixation System
The Newfix® External Fixation System, consisting of axial fixators and frame components, is indicated for stabilization of open and/or unstable fractures where soft tissue injury precludes the use of other fracture treatments such as IM rodding or casting. NEWFIX® External Fixation System is intended for the fixation and for temporary or extended stabilization in cases such as: open fractures, closed fractures, poly-trauma fractures, supplement in the stabilization of minimal fixations in intra and extra articular fractures.
Prescription Use X (21 CFR Part 801 Subpart D) and/or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Melkerson
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k)_
Number K101338
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.