JETSTREAM G3 SE SYSTEM, JETSTREAM G3 SF SYSTEM MODEL PV31230 (JETSTREAM G3 SE CATHETER AND CONTROL POD); PVCN100 (PV CON
K101334 · Pathway Medical Technologies, Inc. · MCW · Jun 11, 2010 · Cardiovascular
Device Facts
Record ID
K101334
Device Name
JETSTREAM G3 SE SYSTEM, JETSTREAM G3 SF SYSTEM MODEL PV31230 (JETSTREAM G3 SE CATHETER AND CONTROL POD); PVCN100 (PV CON
Applicant
Pathway Medical Technologies, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Jun 11, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Story
Jetstream G3 SE and SF systems are atherectomy catheters for debulking and treating peripheral vascular disease. System components include a sterile, single-use catheter with a differentially cutting tip and a reusable console. The catheter features integrated lumens for continuous aspiration of excised tissue/thrombus and infusion of sterile saline. The console houses peristaltic pumps, a system controller, and a keypad interface. During procedures, the catheter is inserted into the peripheral vasculature; the console remains outside the sterile field. The device uses mechanical cutting to break apart thrombus and plaque, which are then aspirated into a collection bag. The G3 SE model includes expandable cutting blades, while the G3 SF model uses a fixed-diameter cutter. The system provides clinicians with real-time operational status via LED indicators. By removing obstructive material from peripheral arteries, the device aims to restore blood flow and improve patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional verification, catheter burst/pull strength, system reliability, infusion/aspiration flow rates, torque to failure, trackability, aspiration efficiency, crossing time, heat generation, rotational speed, guidewire/device wear, sterilization validation, control pod logic verification, and electrical safety.
Technological Characteristics
System consists of an electrically driven catheter with a differentially cutting tip and a reusable console with two peristaltic pumps. Materials include standard medical-grade components. Connectivity includes electrical connectors for the control pod and console. Sterilization is performed via Ethylene Oxide (EO). Software includes control pod logic for device operation. Dimensions modified for smaller crossing profile compared to predicates.
Indications for Use
Indicated for atherectomy and thrombus removal in upper and lower extremity peripheral arteries. Contraindicated for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
K101334
Image /page/0/Picture/1 description: The image shows the logo for Pathway Medical Technologies. The logo consists of a black circle with a white, stylized mountain range inside of it. To the right of the circle, the word "PATHWAY" is written in large, sans-serif font, and below that, the words "MEDICAL TECHNOLOGIES" are written in a smaller, sans-serif font.
JUN 1 1 2010
## 510(k) SUMMARY
General Information: Date of Summary Preparation:
June 4, 2010
Name and Address of Manufacturer:
Pathway Medical Technologies, Inc. 10801 120th Ave NE Kirkland, Washington 98033
Contact Person:
Trade Name:
Common Name:
Regulation Number:
Regulation Name:
Regulatory Class:
Classification Panel:
Product Code:
Cyndy Adams Regulatory Affairs Sr. Manager Phone: 425-636-4047 Fax: 425-636-4001
Jetstream G3TM SE System Jetstream G3TM SF System
Peripheral Atherectomy Catheter
21 CFR 870.4875
Intraluminal Artery Stripper
Class II
Cardiovascular
MCW
{1}------------------------------------------------
Performance Standards: Performance standards do not currently exist for these devices. None are established under Section 514.
Device Description: The Jetstream G3 SE and Jetstream G3 SF Systems are atherectomy catheter devices intended for use in debulking and treating vascular disease in the peripheral vasculature. The Jetstream G3 SE System is designed with expandable cutting blades; the Jetstream G3 SF System has only a fixed diameter cutter; that is, it does not incorporate the expandable blades. Separate lumens within the Catheters allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.
The Jetstream G3 SE and G3 SF Systems each consist of two primary components: (1) a Catheter with Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows:
- Catheter with Control Pod: A sterile, single-use unit consisting of an . electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
- . PV Console: A reusable compact Console, with two (2) peristaltic pumps for aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.
This 510(k) is for the same device most recently cleared under 510(k) K093456, but with a slightly increased catheter length and decreased catheter and tip diameter to achieve a slightly smaller crossing profile. The motor gear ratio is changed and the expandable blades are eliminated from one of the device models.
Indications for Use: The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Substantially Equivalent Devices: Pathway Medical cites the following as substantially equivalent predicate devices:
- · Jetstream G3TM System (K100462)
- · Jetstream G3TM System (K093456)
- · Jetstream G3TM System (K092332)
- · Jetstream G2TM NXT System (K091509)
{2}------------------------------------------------
- · Jetstream Pathway PVTM Atherectomy System (K082186)
- · Pathway PVTM Atherectomy System (K081328)
Testing Summary: To demonstrate substantial equivalence of the Jetstream G3 SE and Jetstream G3 SF Systems with the predicate devices, the technological and performance characteristics were evaluated using in vitro testing as outlined below:
- . Dimensional Verification
- Catheter Burst .
- . Catheter Pull
- System Reliability/Life .
- Speed Drop Flexibility .
- Accessory Compatibility .
- Infusion and Aspiration Flow Rates .
- . Torque to Failure
- Catheter Trackability and Pushability .
- Aspiration Efficiency and Crossing Time ◆
- Heat Generation .
- Rotational Speed .
- Guidewire and Device Wear .
- Sterilization (EO) Validation .
- Control Pod Logic Verification .
- Electrical Safety .
The results from these tests:
- demonstrate that the technological and performance characteristics of the . ﺮ Jetstream G3 SE and Jetstream G3 SF Systems are comparable to those of the predicate devices.
- . support the safety and effectiveness of the modifications that are the subject of this 510(k), and
- ensure the modified devices can perform equivalently to the listed predicate . devices, which have the identical intended use.
Conclusion (Statement of Equivalence): The data and information presented within this submission and the similarities between the modified and predicate devices, support the substantial equivalence of the Jetstream G3 SE and Jetstream G3 SF Systems to the specified predicate devices. The devices have identical indications for use and equivalent technological characteristics. Bench testing supports the safety and effectiveness of the modifications that were the subject of this 510(k).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure, which is the department's official emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
JUN 1 1 2010
Pathway Medical Technologies, Inc. c/o Mr. Brian Cleary Vice President of Clinical and Regulatory Affairs 10801 120th Ave NE Kirkland, WA 98033
Re: K101334
Trade/Device Name: Jetstream G3 SE & SF Systems Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II (two) Product Code: MCW Dated: May 11, 2010 Received: May 12, 2010 Amended: June 4, 2010
Dear Mr. Cleary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. Brian Cleary
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W. Ward
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K101334
Device Names: Jetstream G3TM SE System Jetstream G3TM SF System
## Indications for Use:
.
The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, illiac or renal vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W. Mac.
(Division Sign-Off) Division of Cardiovascular Devi
510(k) Number
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.