NUVUE COLORMARK9TM) NEEDLE

K101312 · Nuvue Therapeutics, Inc. · IYN · Feb 4, 2011 · Radiology

Device Facts

Record IDK101312
Device NameNUVUE COLORMARK9TM) NEEDLE
ApplicantNuvue Therapeutics, Inc.
Product CodeIYN · Radiology
Decision DateFeb 4, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2

Indications for Use

The NuVue ColorMark™ Needle is intended for use when one desires visualization of an interventional needle (e.g. biopsy needle) on B-scan ultrasound screens which have color flow Doppler imaging capability. The NuVue ColorMark™Needle is indicated for use in clinical applications and at anatomical sites for which the biopsy needle and the ultrasound system have received marketing clearance by the FDA.

Device Story

Hand-held, battery-operated, single-use electro-mechanical device; integrates biopsy needle into driving handpiece. Handpiece uses two electromagnets to induce orbital-like vibration (<1 kHz) in needle via sleeve coupling. Vibration mimics fluid motion, triggering Doppler motion-sensing function of standard ultrasound systems to display needle as colored line on B-scan screen. Enables 360-degree visualization of needle orientation. Used by clinicians during interventional procedures (e.g., biopsies) to improve needle tracking. If device fails, needle remains visible via conventional ultrasound techniques. Benefits include enhanced needle visibility and improved ergonomics compared to linear-vibration predecessors.

Clinical Evidence

Bench testing only. Energy and imaging performance testing conducted to verify needle visualization effectiveness. Studies demonstrate device is effective for needle visualization and easy to use. No clinical trial data presented.

Technological Characteristics

Electro-mechanical handpiece with integrated needle sleeve. Uses dual electromagnets for orbital vibration (<1 kHz). Battery-operated. Single-use. Sterilized with Ethylene Oxide. Compatible with biopsy needles (19-27 gauge, 10-20 cm length). No software/firmware; purely mechanical/electromagnetic signal generation.

Indications for Use

Indicated for visualization of interventional needles (e.g., biopsy) during ultrasound-guided procedures. Applicable to clinical sites and anatomical regions where the specific biopsy needle and ultrasound system are FDA-cleared for use.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510 (K) Submission Summary # FEB - 4 2011 # NuVue ColorMark™ Needle (K101312) ## Common/Classification Name: Ultrasonic Locator of Interventional Needles, 21CFR 892.1550 ### Contact: Roger S. Kolasinski, Prepared: February 3, 2011 #### Legally marketed Predicate Devices 1. The NuVue ColorMark™ Needle is substantially equivalent to the EchoTip Needle Guidewire Introducer Needle (K031173) for ultrasound imaging visibility and to the ColorMark™ Visualization System (K926351) for Color flow Doppler Ultrasound imaging visibility. ### II. Intended Use Indications for Use: The NuVue ColorMark™ Needle is intended for use when one desires visualization of an interventional needle (e.g. biopsy needle) on B-scan ultrasound screens which have color flow Doppler imaging capability. The NuVue ColorMark™Needle is indicated for use in clinical applications and at anatomical sites for which the biopsy needle and the ultrasound system have received marketing clearance by the FDA. ### III. Substantial equivalence Summary The technology used in the NuVue ColorMark™Needle is substantially equivalent to that used in the ColorMark™ Visualization System that was previously approved under K926351. The NuVue ColorMark™Needle has improved ergonomics with similar technology and performance as its predecessor in that it presents the image of the needle as a colored line on the B-scan screen, by making the needle vibrate at low (sonic) frequencies. The Needles purchased by NuVue therapeutics have the same characteristics as similar type, gauge, and length needles for similar applications. These new characteristics do not raise new types of safety or effectiveness questions. # IV. Device description The NuVue ColorMark™Needle combines a hand-piece and echogenic needles like its predicates. The NuVue ColorMark™Needle is intended to allow the visualization by ultrasound of a needle- such as biopsy needle- inserted into the body. The principle is that the needle is made to vibrate at low sonic frequencies (<1 kHz), which makes it visible on a Doppler Color-flow system, since the vibration of the needle appears as motion to the Doppler system. As Stipulated, NuVue's ColorMark™ Needle, a hand-held needle device, consists of the driving handpiece and integrated needle. The system is a self-contained, battery-operated, and single-use device that integrates and vibrates interventional biopsy needle(s). The hand-piece accommodates a range of NuVue Therapeutics compatible biopsy needles that fit within its integrated sleeve. The sleeve is the guide and vibration transmitter for the needle. {1}------------------------------------------------ The NuVue ColorMark™Needle is an electro-mechanical needle device. The mechanical coupling of the needle with the hand-piece exciter is mediated by two electromagnets that make the hand-piece sleeve to vibrate in an orbital-like motion, through a phasing of the coils +- 90 degrees apart. This motion is transferred to the interventional needle that is coaxially catheterizing the hand-piece sleeve. This allows for the representation of color Doppler imaging in 360 degrees of rotation of the needle while it is in NuVue's new hand held interventional driver device and in the on position. The needle motion stimulates the Doppler motion sensing function of the imager to display colors corresponding to turbulent fluid motion pattern in close vicinity with the biopsy needle. NuVue Therapeutics will purchase in bulk needles designated by application (biopsy, aspiration, other), needle type (Chiba, Franseen, spinal, other), needle gauge (19 to 27). The needle device, as stipulated, is a single use, hand-piece that integrates and guides the needle for lengths of tissue insertion from 10 to 20 cm. The needle device will be repackaged as a single use device, sterilized with Ethylene Oxide, and distributed commercially under the NuVue Therapeutics, Inc. Name, with the stipulation on the packaging, do not reuse or re-sterilize. #### V. Testing NuVue Therapeutics has carried out energy and imaging performance testing on the NuVue ColorMark™ Needles. The changes implemented into the NuVue ColorMark™ Needle do not introduce any added safety risk nor do they lessen any imaging performance compared to the predicate devices. NuVue has the ability of Quadriture, which allows visualization of any of its needles in a 360 degree visualized capability, compared to the cleared ColorMark™ Visualization System, who's vibratory signal was designed in a linear responsibility, demanding the operator of the device to always be assured of the orientation of the needle to the head of the Doppler Transducer, per the manufactures mandates. A detailed analysis of the safety of NuVue's device shows no addition risk issues. In case of failure of the device, Color Doppler visualization may be lost, but the needle is still visible and can be followed by conventional techniques. Our imaging studies show that NuVue's ColorMark™ Needle is effective in the visualization of image of the biopsy needle, and can be easily used by the operator. #### VI. Conclusion This premarket submission has demonstrated Substantial equivalence as defined by the Federal Food and Drug and Cosmetic Act. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines forming its body and wing. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 NuVue Therapeutics, Inc. % Mr. Roger S. Kolasinski, CEO 11135 Sedgefield Road Fairfax, Virginia 22030 FEB - 4 201 Re: K101312 Trade/Device Name: NuVue Colormark™ Needle Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN Dated: December 15, 2010 Received: December 17, 2010 Dear Mr. Kolasinski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indice) ons for use stated in the enclosure) to legally marketed predicate devices marketed in internate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, insting of devices, good manufacturing practice, labeling, and prohibitions against misbaratif insand adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not migleding. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Mr. Roger S. Kolasinski, CEO comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm. Sincerely yours. Ais B. R. for Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number: K101312 Device Name: NUVUE COLORMARK™ NEEDLE Indications for Use: The NuVue ColorMark™ Needle is intended for use when one desires the visualization of an interventional needle (e.g. biopsy needle) on B-scan ultrasound screens which have color flow Doppler imaging capability. The NuVue ColorMark™Needle is indicated for use in clinical applications and at anatomical sites for which the biopsy needle and the ultrasound system have received marketing clearance by the FDA. Prescription Use X X Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K101312
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...