K101311 · Philips Medical Systems North America Co. · LLZ · Sep 30, 2010 · Radiology
Device Facts
Record ID
K101311
Device Name
EP NAVIGATOR R3
Applicant
Philips Medical Systems North America Co.
Product Code
LLZ · Radiology
Decision Date
Sep 30, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
EP navigator is intended to enable users to segment previously acquired 3D CT or other datasets and overlay and register these 3D segmented data sets with live fluoroscopy X-ray images of the same anatomy in order to support catheter/device navigation. The 3D segmented data set can be displayed with a color map annotation received from an external source.
Device Story
EP navigator is PC-based software used in cathlabs by radiologists or cardiologists to assist in interventional heart rhythm procedures. It takes previously acquired 3D DICOM CT or other image data as input. The software segments this 3D data and registers/overlays it onto live fluoroscopic X-ray images from a Philips Allura Xper FD system. It allows for visual marker placement on 3D volumes and provides an 'EndoView' for internal 3D volume visualization. The system integrates with EP Logix, receiving color map annotations and transmitting visual marker positions. By providing a real-time anatomical overlay, the device aids clinicians in navigating catheters and guidewires, potentially improving procedural accuracy and patient outcomes during cardiac interventions.
Clinical Evidence
Clinical evaluation was performed to demonstrate safety and effectiveness in the intended clinical environment. Non-clinical verification and validation testing included software verification, validation, and DICOM conformance testing against requirement specifications and risk management results.
Technological Characteristics
PC-based software platform. Integrates with Philips Allura Xper FD angiography X-ray systems. Supports DICOM 3D CT and other image datasets. Connectivity includes data exchange with external sources (e.g., EP Logix).
Indications for Use
Indicated for patients with heart rhythm disorders undergoing X-ray fluoroscopy-guided interventional procedures. Provides navigation support for intra-cardiac instruments (catheters, guidewires) by overlaying segmented 3D anatomical data onto live fluoroscopic images. No contraindications.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "PHILIPS" in large, bold, black letters. Above the word "PHILIPS" is the handwritten text "KL01311". The text is smaller than the word "PHILIPS" and is written in black ink.
# 510(k) Summary
, ৮
The following information is submitted in accordance with the requirements of 21CFR 807.92.
## ldentification of manufacturer
| Company: | Philips Medical Systems Nederland B.V |
|----------------------|--------------------------------------------------|
| Address: | Veenpluis 4-6,<br>5684-PC, Best, The Netherlands |
| Registration number: | 3003768277 |
# ldentification of submitter
| Name: | Lynn Harmer |
|----------------|------------------------------------|
| Position: | Senior Manager, Regulatory Affairs |
| Telephone: | (425) 487-7312\ |
| Fax: | (425) 487-8666 |
| Date prepared: | May 7, 2010 |
## Device identification
| Trade name: | Philips |
|-------------------------|---------------------------------------------|
| Device name: | EP navigator |
| Regulation description: | Picture archiving and communications system |
| Regulation number: | 21CFR 892.2050 |
| Class: | II |
| Product code: | 90L--LZ |
## Legally marketed devices
| Trade names: | Philips EP navigator, Philips Brilliance CT,<br>Philips Allura 3D-CA, Philips Integris 3D-RA,<br>Philips Xper CT, GE Innova Vision Applications |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips and GE |
| 510(k) numbers: | K062650, K042293, K042334, K040254, K060749<br>K092639 |
{1}------------------------------------------------
#### Device description
EP navigator image software processing algorithms are executed on a PC based hardware platform, which can perform the following functions:
- segment previously acquired DICOM 3D CT or other image data.(the acquisition of . the image data from a rotational angiogram is known as 3D atriography (3D ATG))
- superimpose the segmented 3D CT or other dataset on a live fluoroscopic X-ray image of the same anatomy, obtained on a Philips Allura Xper FD angiography X-ray system,
- . register the segmented 3D CT or other data with live fluoroscopic X-ray images obtained on a Philips Allura Xper FD angiography X-ray system for specified procedures.
- The 3D segmented data set can be displayed with a color map annotation received ● from an external source.
- position visual markers on the 3D volume ●
- visualize the inside of the 3D volume (EndoView) .
- certain buttons on the user interface control EP Logix functions; ●
- visual marker positions are transmitted to EP Logix; .
- color map information is received from EP Logix. .
#### Intended use
EP navigator is intended to enable users to segment previously acquired 3D CT or other datasets and overlay and register these 3D segmented data sets with live fluoroscopy X-ray images of the same anatomy in order to support catheter/device navigation. The 3D segmented data set can be displayed with a color map annotation received from an external source.
#### Indications for Use
Medical purpose EP navigator is intended to provide navigation support for intra-cardiac instruments, such as catheters and guidewires, during the interventional treatment of heart rhythm disorders, by overlaying acquired and segmented 3D anatomical image data over live fluoroscopic X-rav images of the same anatomy.
Patient population EP navigator is intended to be used with patients who suffer from heart rhythm disorders and who are capable of undergoing a procedure guided by X-ray fluoroscopy.
Contact with body part / tissue type The heart and surrounding tissue are illuminated by X-ray fluoroscopy (by the Allura Xper system).
Operator profile The Operator is a Radiologist or Cardiologist who is fully skilled and responsible for sound clinical judgment and for applying the best clinical procedure. The Operator may also be a nurse assisting who is authorized by the Radiologist or Cardiologist. Application EP navigator is to be used in a controlled environment in the context of a control room of a Cathlab. The system is connected to an Allura Xper system.
Contra-indications Within the mentioned patient population, there are no contraindications.
510(k) Summary
{2}------------------------------------------------
#### Technological characteristics
EP navigator image software processing algorithms are executed on a PC based hardware platform
#### Summary of testing
EP navigator 3 complies with standards as detailed in annex 009 of this premarket submission. Clinical evaluation was performed to show safety and effectiveness to of EP-Navigator in the intended clinical environment. Non-clinical verification and validation tests were performed relative to the requirement specifications and risk management results, specifically including software verification, validation and DICOM conformance testing. Corresponding clinical evaluation report and test results are included in this submission.
Conclusion:...................................EP-Navigator is substantially equivalent to the currently legally marketed devices.
This opinion is based on the following:
- EP navigator does not introduce new indications for . use,
- . EP navigator has the same technological characteristics as the predicate devices,
- EP navigator does not introduce new potential haz-. ards or safety risks.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Lynn Harmer Manager, Regulatory Submissions Philips Medical Systems 22100 Bothell Evert Highway BOTHELL WA 98041-3003
SEP 3 O 2010
Re: K101311
Trade/Device Name: EP Navigator Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 10, 2010 Received: May 11, 2010
Dear Ms. Harmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kl01311
Image /page/5/Picture/1 description: The image shows the word "Philips" in a stylized, bold font. The letters are black and have a slightly distressed or textured appearance. The word is presented in a single line and is centered in the image.
Appendix 005
Indications for Use Statement
Device Name
EP navigator
#### Indications for Use
Medical purpose EP navigator is intended to provide navigation support for intra-cardiac instruments, such as calheters and guidewires, during the interventional treatment of heart rhythm dlsorders, by overlaying acquired and segmented 3D anatomical image data over live fluoroscopic Xray images of the same anatomy.
Patient population EP navigator is intended to be used with panents who suffer from heart rhything disorders and who are capable of undergoing a procedure guided by X-ray fluoroscopy.
Contact with body part / tissue type The heart and surrounding tissue are illuminated by X-ray fluoroscopy (by the Allura Xper system).
Operator profile The Operator is a Radiologist or Cardiologist who is fully skilled and responsible for sound clinical judgment and for applying the best clinical procedure. The Operator may also be a nurse assisting who is authorized by the Radiologist or Cardiologist
Application EP navigator is to be used in a controlled environment in the context of a control room of a Cathlab. The system is connected to an Allura Xper system.
Contra-Indications Within the mentioned patient population, there are no contra-indications.
#### Intended use
EP navigator is intended to enable users to segment previously acquired 3D CT or other datasets and overlay and register these 3D segmented data sets with live fluoroscopy X ray images of the same anatomy in order to support catheter/device navigation. The 3D segmented data set can be displayed with a color map annotation received from an external source.
Prescription Use ___yes_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Indications for Use Statement.docx
1/1
(Division Sign-Off) (Division of Radiological Devices Division of Radiological Device Evaluation and Safety
Office of In Vitro Diagnostic Device Evaluation and Safety
K101311
510K
# SEP 30 2010
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.