The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) or adult patients (> 66 lbs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.
Device Story
BiPAP Synchrony 2 is a microprocessor-controlled, blower-based, bi-level positive pressure system. It delivers two pressure levels (IPAP/EPAP) to provide non-invasive ventilation. Input data includes real-time flow and pressure measurements from sensors in the patient airway. A microprocessor controller processes these inputs to regulate the blower assembly. The device features a user interface for clinicians to adjust parameters and view clinical data. Used in hospitals or homes, it is operated by clinicians or patients/caregivers. The system improves patient comfort by providing pressure support therapy. The current submission adds two small-sized pediatric full-face masks (AF531 EE Leak 1 and Leak 2) to the existing system, supported by anthropometric analysis and bench testing for pressure performance, triggering, cycling, and alarm functionality.
Clinical Evidence
Bench testing only. Testing included pressure performance, waveform analysis, triggering, cycling, and alarm functionality to verify compatibility of new pediatric mask sizes with the existing BiPAP Synchrony 2 platform. Anthropometric analysis confirmed mask fit for the pediatric population (>7 years, >40 lbs).
Technological Characteristics
Microprocessor-controlled blower-based bi-level positive pressure system. Masks constructed of silicone cushions, polycarbonate faceplates, and polycarbonate/silicone entrainment valves. Connectivity includes a user interface for clinical parameter adjustment. System utilizes flow and redundant pressure sensors for closed-loop control.
Indications for Use
Indicated for pediatric patients ≥7 years (>40 lbs) and adults (>66 lbs) with respiratory insufficiency or obstructive sleep apnea requiring non-invasive ventilation in hospital or home settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Respironics BiPAP Synchrony 2 Ventilatory Support System (K092043)
{0}------------------------------------------------
・
.
------
. "
:
. .
TAB 3
510(K) SUMMARY OF SAFETY & EFFECTIVENESS
| Original Date of Submission | April 21, 2010 |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | BiPAP Synchrony 2 SEP 2 2010 |
| Common/Usual Name | Ventilator, continuous, non-life supporting |
| Establishment Registration # | 2518422 |
| Official Contact / Address of<br>Mfr. Facility | Zita A. Yurko<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>(724) 387-4120 FAX (724)-387-7490 CELL (724) 882-4120<br>Zita.Yurko@Respironics.com |
| Classification | Class II device |
| Classification Panel | Anesthesiology Devices |
| Classification Reference | 21 CFR 868.5895 |
| Product Code | MNS - Continuous ventilator |
| Predicate Device(s) | Respironics BiPAP Synchrony 2 Ventilatory Support System (K092043)<br>Respironics Performax Total Face Mask (K072592)<br>Respironics Spectrum 2 Full Face Mask (K002465) |
| Labeling | Draft Labeling can be found in Tab 5. |
| Intended Use | The Respironics BiPAP Synchrony 2 is intended to provide non-<br>invasive ventilation for pediatric patients 7 years or older (> 40 lbs) or<br>adult patients (> 66 lbs) with respiratory insufficiency or obstructive<br>sleep apnea, in the hospital or home. |
| Reason for Submission | Include two additional Respironics pediatric full face masks to be used<br>with this device. Device is unchanged as a result of the addition of<br>either of these masks. |
-
{1}------------------------------------------------
# Substantial Equivalence
The BiPAP Synchrony 2 system has the following similarities to the previously cleared predicate device:
- D Same intended use.
- O Same operating principle.
- O Same technology.
- Same manufacturing process. ០
There is no change to the intended use, operating principle, technology or manufacturing process for the BiPAP Synchrony. Furthermore, there is no modification required to the electromechanical platform of the BiPAP Synchrony 2 as a result of adding additional Respironics pediatric masks to the existing cleared device (K092043). To demonstrate compatibility of the small size Respironics AF531 EE Leak 1 Full Face Mask (no built-in exhalation) and the small size Respironics AF531 EE Leak 2 Full Face Mask (with built-in exhalation) with the BiPAP Synchrony 2, mask compatibility testing was performed with each mask. This testing is provided in Tab 8 of this submission. This testing included pressure performance, waveform nee, triggering, cycling and alarm functionality testing. All tests were verified to meet the required acceptance criteria. Results of this testing concluded that the verification testing raises no new issues of safety or effectiveness by adding the small size of either of these full face masks to the cleared BiPAP Synchrony 2.
Respironics has followed the FDA's Guidance for Industry and FDA Staff document "pre-market assessment of pediatric medical devices" and applied the principle of FDA's Least Burdensome Approach to demonstrate the Substantial Equivalence of the BiPAP Synchrony 2 system. As a result, we conclude that the existing indications for use can be safely and effectively applied to this device with the small size of either the Respironics AF531 EE Leak 1 Full Face Mask or the Respironics AF531 EE Leak 2 Full Face Mask.
### Intended Use
The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) or adult patients (>66 lbs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.
{2}------------------------------------------------
## Device Description
The Respironics BiPAP Synchrony 2 device is unchanged from K092043. The BiPAP Synchrony 2 is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.
The only change to the Respironics BiPAP Synchrony 2 device is to include the small size Respironics AF531 EE Leak 1 Full Face Mask (no built-in exhalation) and the small size Respironics AF531 EE Leak 2 Full Face Mask (with built-in exhalation) as options for use by its pediatric users. These two masks consist of a silicon cushion, polycarbonate faceplate, and an integrated entrainment valve elbow made of polycarbonate with a silicone flapper. All of the materials used within these masks are predicate materials. The small size Respironics AF531 EE Leak 1 Full Face Mask has an entrainment elbow with no additional exhalation and will require a separate exhalation device. Whereas, the small size Respironics AF531 EE Leak 2 Full Face Mask has an entrainment elbow with built-in exhalation, negating the need for a separate exhalation device. Both masks are designed to be used with either the four point style headgear or the capstrap style headgear, allowing the patient an option.
Anthropometric analysis was performed on the small size of the Respironics AF531 EE Leak 1 Full Face Mask and the small size Respironics AF531 EE Leak 2 Full Face Mask. This analysis has determined that these two small masks will appropriately fit the pediatric population (> 7 years of age, > 40 lbs).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three lines representing its wings and body. The eagle faces to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Respironics, Incorporated C/O Ms. Zita A. Yurko Director, Regulatory Affairs Sleep & Home Respiratory Group 1740 Golden Mile Highway Monroeville, Pennsylvania 15146
SEP 2 2010
Re: K101130
Trade/Device Name: BiPAP Synchrony 2 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: August 5, 2010 Received: August 10, 2010
Dear Ms. Yurko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 will ), it thay of color of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Yurko
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K101130
Page _ 1 of 1
#### Indications for Use
510(k) Number (if known): _ .
Device Name: _________________________________________________________________________________________________________________________________________________________________
The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) and adult patients (> 66 lbs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) a Olli (Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anesthesion Devices Division of Antool, Dental Devices 1101130 510(k) Number: __
000037
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.