A powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable powdered natural latex examination glove; worn on examiner's hand/finger; acts as physical barrier to prevent cross-contamination between patient and examiner; used in clinical/medical settings; non-sterile; single-use.
Clinical Evidence
No clinical data; bench testing only. Performance verified against ASTM D 3578-01a standards for dimensions, physical properties, pinhole freedom, powder content, and water-soluble protein content. Biocompatibility testing (primary skin irritation and dermal sensitization) confirmed no adverse reactions.
Technological Characteristics
Natural rubber latex; powdered; non-sterile. Dimensions and physical properties meet ASTM D 3578-01a. Powder content <10 mg/dm² (ASTM D 6124-05). Water-soluble protein content <200 ug/dm² (ASTM D 5712-99). Biocompatibility: pass (no primary skin irritation or dermal sensitization).
Indications for Use
Indicated for use by medical personnel as a disposable barrier worn on the hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the letters BPG in a bold, sans-serif font. The letters are large and take up most of the frame. The letters are black, and the background is white. The letters are slightly distressed, with some small imperfections and rough edges.
### I PUTRA GLUVES SUN.BHD. (580515-T)
& 1358, Jin Kg. Mohd Taib, Kawas 3000 Rawang, Selangor Da +03-6092 1042 +03-6092 1142 Fax: +0 E-mail : bog@streamyx.com Website
Image /page/0/Picture/3 description: The image contains two sets of logos. The first set includes the "nqa. ISO 9001 Registered" logo and the "UKAS ENVIRONMENTAL MANAGEMENT 015" logo. The second set includes the "nqa. ISO 14001 Registered" logo and the "UKAS ENVIRONMENTAL MANAGEMENT 015" logo.
# <101106
### 510 (K) SUMMARY
#### 1.0 Submitter:
NOV 1 0 2010
0143
| Name | : Ratnam S.Vythilingam |
|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | : Best Putra Gloves Sdn Bhd, Jln Kg Mohd Taib, Kawasan Perindustrian Sg Choh, 48000, Rawang, Selangor Darul Ehsan. Malaysia |
| Phone No. | : +03-60921042 |
| Fax No. | : +03-60912820 |
Date of Summary Prepared: 20 August 2010
#### 2.0 Contact Person:
| Name | Dr. Effendi Tenang |
|-----------|--------------------|
| Phone No. | +03-60921042 |
| Fax No. | +03-60912820 |
#### 3.0 Name of the device:
| Trade Name | : 1) BPG Latex Examination Gloves<br>2) Multiple or Customers' Trade Name |
|---------------------|---------------------------------------------------------------------------|
| Device Name | : Powdered, Natural Color, Latex Examination Gloves,<br>Non-Sterile |
| Common Name | : Examination Gloves |
| Classification Name | : Patient Examination Gloves (Class 1) |
#### 4.0 Identification of the legally marketed device:
Class I patient examination gloves, powdered, that meets all the requirements of
ASTM standard D 3578-01a22 and FDA 1000 ml Water Leak Test.
#### 5.0 Description of the Device:
The Powdered, Natural Color, Latex Examination Gloves, Non-Sterile meets all the requirements of ASTM standard D 3578-01a22 and FDA 1000 ml Water Leak Test.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the letters 'BPG' in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance, giving them a rough or worn look. The letters are large and take up most of the frame.
(580515-1)
Lot 1357 & 1358, Jln Kg. Mohd Taib, Kawasan Perindustrian Sg. Choh
48000 Rawang, Selangor Darul Ehsan, Malaysia.
Tel: +03-6092 1042, +03-6092 1142 Fax: +03-6091 2820
E-mail : bpg@streamyx.com Website : www.bpgloves.com
Image /page/1/Picture/2 description: The image contains two sets of logos related to quality and environmental management systems. The first set includes the "nqa. ISO 9001 Registered" logo and the "UKAS" logo with the number "015" below it. The second set includes the "nqa. ISO 14001 Registered" logo and the "UKAS" logo with the number "015" below it. The logos suggest that the organization has been certified for both ISO 9001 (quality management) and ISO 14001 (environmental management) standards by an accredited body.
K101106
#### 6.0 Intended Use of the Device:
A powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### 7.0 Summary of the technological characteristic of the devices:
The Powdered, Natural Color, Latex Examination Gloves, Non-Sterile are summarized with following technological characteristics compared to ASTM or equivalent standards.
| CHARACTERISTICS | STANDARDS | DEVICE<br>PERFORMANCE |
|-----------------------|-----------------------|------------------------------|
| Dimensions | D 3578-01at | Meets |
| Physical Properties | D 3578-01a€ | Meets |
| Freedom from Pinhole | D 3578-01a84 | Meets |
| | FDA CFR 800.20 | |
| Powder Amount | D 3578-01a64 | <10 mg/dm- |
| | D 6124 - 05 | |
| Water Soluble Protein | D 3578-01a84 | <200 ug/dm² |
| Content | D 5712-99 | |
| Biocompatability | Primary Skin | Pass |
| | Irritation in Rabbits | (No primary skin irritation) |
| | Dermal Sensitization | Pass |
| | | (No contact sensitizer) |
#### 8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data.
The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.
#### 9.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data.
Clinical data is not needed for gloves or for most devices cleared by the 510 (k) process.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the letters 'BPG' in a bold, sans-serif font. The letters are black and have a distressed or grunge texture, giving them a worn or aged appearance. The letters are large and take up most of the frame.
## BESI PUIKA GLOVES SDN.BHD. (580515-T)
Lot 1357 & 1358, Jin Kg. Mohd Taib, Kawasan Perindustrian Sg. Choh 48000 Rawang, Selangor Darui Ehsan, Malaysia. Tel: +03-6092 1042, +03-6092 1142 Fax: +03-6091, 2820 E-mail : bpg@streamyx.com Website : www.bpgloves.com
Image /page/2/Picture/3 description: The image contains two sets of logos related to quality and environmental management systems. The first set includes the "nqa. ISO 9001 Registered" logo alongside the UKAS (United Kingdom Accreditation Service) environmental management logo with the number 015. The second set features the "nqa. ISO 14001 Registered" logo, also accompanied by the UKAS environmental management logo with the same number 015, indicating certification or registration under these standards.
K101106
p3df
#### 10.0 Conclusion
It can be concluded that the Powdered , Natural Color, Latex Examination Gloves, Non-Sterile will perform according to the glove performance standards referenced in Section (7) above and meet ASTM standards, and FDA requirements for Water Leak Test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.
{3}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Dr. Effendi Tenang Managing Director Best Putra Gloves SDN BHD Lot 1357-1358, JLN Kg. Mohd Taib Kawasan Perindustrian Sg. Choh 48000 Rawang, Selangor Darul Ehsan, Malaysia
NOV 1 0 2010
Re: K101106
Trade/Device Name: Powdered, Latex Examination Gloyes, Natural Color, Non-Sterile Model: MEP1 Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: October 4, 2010 Received: October 21, 2010
Dear Dr. Tenang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, , that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and appears to be a scanned image.
{4}------------------------------------------------
Page 2- Dr. Tenang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Seren Luures
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the letters "BPG" in a bold, sans-serif font. The letters are large and take up most of the frame. The letters are solid black, and the background is white.
T PUTRA GLOVES SDN.BF
Lot 1357 & 1358, Jln Kg. Mohd Taib, Kawasan Perindustrian Sg. Choh,
48000 Rawang, Selangor Darul Ehsan, Malaysia.
Tel: +03-6092 1042, +03-6092 1142 Fax: +03-6091 2820
E-mail: bpg@streamyx.com Website: www.bpgloves.com
Image /page/5/Picture/3 description: The image shows two sets of logos. The first set includes a logo with the text "nga. ISO 9001 Registered" and another logo with a checkmark and the number "015". The second set includes a logo with the text "ISO 14001 Registered" and another logo with a checkmark and the number "015".
1 0 2019
## INDICATIONS FOR USE
510( k ) Number: K101106
Device Name: Powdered, Latex Examination Glove, Natural Color, Non-Sterile, Model: MEP1
Indications For Use:
A powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Sincerely
Dr. Effendi Tenang
(Managing Director)
Prescription Use_ (Part 21 CFR 801 Subpart D)
OR
Over - The - Counter Use _XXX_ (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _
Elaine S. Mayhall
Offision Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Seneral Section of Anesthesionogy, General Devices Division of Anesthoolor
510(k) Number: K101106
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.