K101062 · Trireme Medical, Inc. · LIT · May 12, 2010 · Cardiovascular
Device Facts
Record ID
K101062
Device Name
GLIDER PTA BALLOON CATHETER
Applicant
Trireme Medical, Inc.
Product Code
LIT · Cardiovascular
Decision Date
May 12, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The GliderXtreme™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. NOT for use in the coronary or cerebral vasculature.
Device Story
Standard PTA balloon dilatation catheter; braided shaft; atraumatic tapered tip; semi-compliant balloon; radiopaque markers for positioning; hydrophilic coating. Available in Rapid Exchange or Over-The-Wire configurations. Used in clinical settings by physicians for peripheral artery angioplasty. Device inserted over guidewire (0.014" or 0.018") via introducer sheath; balloon inflated via luer connector to dilate stenotic lesions. Benefits include restoration of blood flow in peripheral arteries. Single-use, sterile device.
Clinical Evidence
Bench testing only. In vitro tests included balloon rated burst pressure, inflation/deflation, fatigue, bond strength, torsional strength, catheter diameter, balloon profile, tip configuration, balloon compliance, trackability, pushability, kink resistance, and device interface compatibility.
Technological Characteristics
Braided shaft; semi-compliant balloon; hydrophilic coating; radiopaque markers; hypotube proximal end with plastic hub/strain relief. Compatible with 0.014"/0.018" guidewires and 5F/6F introducers. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for balloon dilatation of peripheral vascular lesions in iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. Contraindicated for coronary or cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K101062
#### 510(k) Summary [As required by 21 CFR 807.92(c)]
- 1. Submitter's Name / Contact Person
| Submitter: | TriReme Medical, Inc.<br>7060 Koll Center Parkway, Suite 300<br>Pleasanton, CA 94566 |
|-----------------|--------------------------------------------------------------------------------------|
| Contact Person: | Shiva Ardakani<br>VP of RA/QA<br>Phone: 925-931-1300 Ext 209<br>Fax: 925-931-1361 |
| Date Prepared: | May 4, 2010 |
General Information
| Trade Name: | GliderXtreme™ PTA Balloon Catheter |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Angioplasty Catheter |
| Classification Name: | Percutaneous Catheter (21 CFR 870.1250) |
| Product Code: | DQY / LIT |
| 510(k) number: | |
| Predicate Devices: | NanoCross PTA Dilatation Catheter (K090849)<br>Sterling ES MR and OTW PTA Balloon Dilatation Catheters |
MAY 12 2010Balloon Dilatation Catheters (K080982)
EverCross .035" OTW PTA Dilatation Catheter (K082579)
#### 3. Intended Use
2.
The GliderXtreme™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries. NOT for use in the coronary or cerebral vasculature.
- 4. Device Description
The GliderXtreme™ PTA Balloon Catheter is a standard balloon dilatation catheter with a braided shaft and an atraumatic tapered tip. The device is compatible with commonly used accessories, including standard 0.014" and 0.018" guidewires and 5F introducer sheaths (or 6F guide catheters). Overall catheter length is approximately 135 cm.
The distal end of the catheter has a semi-compliant balloon that expands to known diameters at specific pressures. The balloon is available in multiple sizes and contains one or two radiopaque markers to assist with positioning. The shaft is braid reinforced and has a lubricious hydrophilic coating. The proximal end of the device is a common PTA catheter design of a hypotube connected to a plastic hub and strain relief. The hub is used to inflate the balloon; the luer connector is compatible with standard inflation
{1}------------------------------------------------
devices. A second lumen intended for guidewire use is contained within the catheter. The balloon catheters are available in Rapid Exchange or Over-The-Wire configuration.
The GliderXtreme™ PTA Balloon Catheters are supplied sterile and intended for single use only.
- 5. Performance Data
Bench testing was performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. The following in vitro tests were performed:
- Balloon Rated Burst Pressure .
- . Balloon Inflation and Deflation
- . Balloon Fatigue
- Catheter Body Strength (Bond Strength) .
- . Torsional Strength
- Catheter Diameter, Balloon Profile and Tip Configuration
- Balloon Compliance ●
- . Trackability, Pushability
- Kink Resistance .
- Device Interface Compatibility ◆
#### ર્ભ . Substantial Equivalence Comparison and Conclusion
All GliderXtreme™ PTA Balloon Catheters are substantially equivalent to the predicate devices in design, materials, packaging, fundamental scientific technology, manufacturing, sterilization and intended use. Performance testing demonstrated that the GliderXtreme™ PTA Balloon Catheter reliably achieved the desired effect and is safe for its intended use. No new questions of safety or effectiveness were identified during device testing. Therefore, All GliderXtreme™ PTA Balloon Catheters are considered substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized image of a bird-like figure, possibly representing a symbol of health or service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
TriReme Medical, Inc. c/o Chris Lavanchy Primary Reviewer CITECH 5200 Butler Pike Plymouth Meeting, PA 19462-1298
MAY 1 2 2010
Re: K101062
Trade/Device Name: Glider PTA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 10, 2010 Received: May 11, 2010
Dear Mr. Lavanchy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 -- Mr. Chris Lavanchy
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
er R. lahmen
Bram D. Zuckerman, M.D.
Director. Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and · Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (K094049): K i U (06 2
Device Name: Glider™ PTA Balloon Catheter
Indications for Use:
The Glider™ PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, NOT for use in the coronary or cerebral vasculature.
The indication for use is same for all Glider PTA Balloon Catheter Family.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
unner R. Vc. Anner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.