# CARDO MEDICAL FEMORAL CEMENT RESTRICTOR, MODEL 503-MCRX (K101032)

_Cardo Medical, Inc. · JWH · Jul 8, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K101032

## Device Facts

- **Applicant:** Cardo Medical, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Jul 8, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Cardo Medical Total Knee System is indicated for use in individuals undergoing surgery for: Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis; Post-traumatic loss of knee joint configuration and function; Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; Revisions of previous unsuccessful knee replacement or other procedures. Additional indications for posteriorly stabilized components: Ligamentous instability requiring implant bearing surfaces with increased constraint; Absent or non-functioning posterior cruciate ligament. These devices are single use only and are intended for implantation with bone cement.

## Device Story

The Cardo Medical Femoral Cement Restrictor is an optional accessory for the Cardo Medical PS femoral components. It is a non-load-bearing plug manufactured from UHMWPE. During total knee arthroplasty, the surgeon implants the device to occlude the opening in the femoral component, preventing bone cement migration. The device is for single use only and is implanted with bone cement.

## Clinical Evidence

Bench testing only. The device is a non-load-bearing plug that does not alter the worst-case product configuration for testing; previous testing for the system applies.

## Technological Characteristics

Material: UHMWPE. Form factor: Non-load-bearing plug. Intended for use with bone cement. Single-use.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Howmedica Osteonics Tibial Tray Screw Hole Plugs ([K032479](/device/K032479.md))

## Reference Devices

- Cardo Medical Inc. Total Knee System ([K081127](/device/K081127.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101032

Cardo Medical Femoral Cement Restrictor Premarket Notification Submission - 510(k)

Image /page/0/Picture/2 description: The image shows the logo for Cardo Medical. The logo consists of a stylized graphic to the left of the words "Cardo Medical" stacked on top of each other. Below the logo is the address "10 Clifton Blvd. Suite B1 Clifton, NJ 07011".

JUL--8 2010

# Summary of Safety and Effectiveness

Cardo Medical, Inc. Submitter/Owner: 10 Clifton Blvd., Suite B1, Clifton, NJ 07011

Dina L. Weissman, J.D. Contact Person: Director, Quality Assurance, Regulatory Affairs and Government Compliance 973-777-8832 Phone: Fax: 973-777-8851 dweissman@cardomedical.com

July 2, 2010 Date Prepared:

Cardo Medical Femoral Cement Obturator Trade Name:

Email:

Cement Obturator Common Name:

Surgical Mesh, (21 CFR 878.3300, Product Code: LZN, Class II) and Classification Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, (21 CFR 888.3560, Product Code: JWH, Class II)

Howmedica Osteonics Tibial Tray Screw Hole Plugs (K032479) Predicate Device: Cardo Medical Inc. Total Knee System (K081127, cleared as ACCIN Total Knee System)

The Cardo Medical Femoral Cement Restrictor is only intended for use with Device Description: Cardo Medical PS femoral components (previously cleared in K081127). The device is an optional accessory and is manufactured from UHMWPE.

The Cardo Medical Cement Restrictor is for use at the option of the surgeon in Intended Use: total knee arthroplasty to occlude the opening in the femoral component to help prevent bone cement migration. Because the device may only be used in total knee arthroplasty, it carries the indications for use for the overall system:

> The Cardo Medical Total Knee System is indicated for use in individuals undergoing surgery for:

· Painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis;

· Post-traumatic loss of knee joint configuration and function

· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;

· Revisions of previous unsuccessful knee replacement or other procedures.

Additional indications for posteriorly stabilized components:

· Ligamentous instability requiring implant bearing surfaces with increased constraint;

· Absent or non-functioning posterior cruciate ligament.

These devices are single use only and are intended for implantation with bone cement.

{1}------------------------------------------------

Cardo Medical Femoral Cement Restrictor Premarket Notification Submission - 510(k)

#### Comparison to Predicates:

The Cardo Medical Femoral Cement Restrictor is an ultrahigh molecular weight polyethylene component for use in total knee arthroplasty. The device is substantially equivalent to the Howmedica Osteonics 7 for use in total kneed annoplaty. 176 02479), having the same technological characteristics. The Tibler Tray Done Oceew The same material, are used in total knee arthroplasty and serve device and its predioute are opening to help prevent bone cement from migrating.

Cardo Medical, Inc. has determined that any differences in the proposed device will not impact the safety or effectiveness of the total knee system for its intended use.

### Synopsis of Test Methods and Results:

Oynopsis or rese mothed and non-load-bearing plug that does not change the worst case products for testing purposes, therefore the previous testing applies to these components.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines. The overall design is simple and conveys a sense of national identity and governmental authority.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Cardo Medical, Inc. % Ms. Dina L. Weissman. J.D. Director, Quality Assurance Regulatory Affairs and Government Compliance 10 Clifton Boulevard, Suite B1 Clifton, New Jersey 07011

Re: K101032

Trade/Device Name: Cardo Medical Total Knee System - Femoral Cement Restrictor Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis

JUL - 8 2010

Regulatory Class: II Product Codes: JWH: LZN Dated: April 13, 2010 Received: April 14, 2010

Dear Ms. Weissman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Dina L. Weissman, J.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kaibau Buchur

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

0103

510(k) Number (if known):

#### Device Name: Cardo Medical Total Knee System - Femoral Cement Restrictor

#### INDICATIONS FOR USE

The Cardo Medical Total Knee System is indicated for use in individuals undergoing surgery for:

- Painful, disabling joint disease of the knee resulting from degenerative arthritis, . rheumatoid arthritis or post-traumatic arthritis;
- Post-traumatic loss of knee joint configuration and function ●
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be . returned to adequate function and stability;
- . Revisions of previous unsuccessful knee replacement or other procedures.

Additional indications for posteriorly stabilized components:

- Ligamentous instability requiring implant bearing surfaces with increased constraint; �
- Absent or non-functioning posterior cruciate ligament. .

These devices are single use only and are intended for implantation with bone cement.

Prescription Use XXX (Per 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

barbare buehm.

(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K101632

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K101032](https://fda-staging.innolitics.com/device/K101032)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).