The Powered PD-IO provides intraosseous access in the distal tibia of pediatric patients as an alternative to IV access in emergent, urgent, or medically necessary cases for up to 24 hours. The Humeral Head EZ-IO provides vascular access when standard venous access is not possible. Humeral head IO access is indicated in pediatrics in emergent, urgent, or medically necessary cases when intravenous access is difficult or impossible to obtain for up to 24 hours. The Powered PD-IO Intraosseous Infusion System provides intraosseous access in the proximal tibia as an alternative to IV access in emergent, urgent, or medically necessary cases for up to 24 hours. The device is for use in pediatric patients (approximate weight range: 3kg-39kg).
Device Story
Powered PD-IO and EZ-IO Humeral Head systems provide intraosseous (IO) vascular access for drug and fluid delivery. Device utilizes a powered driver to insert a needle into the bone marrow space (proximal tibia, distal tibia, or humeral head). Used in emergent/urgent clinical settings when IV access is difficult or impossible. Operated by healthcare professionals. Output is established vascular access, allowing clinicians to administer life-saving medications and fluids directly into the systemic circulation. Benefits include rapid, reliable access in pediatric patients where peripheral venous access is challenging.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and design to previously cleared predicate devices.
Technological Characteristics
Powered intraosseous infusion system consisting of a driver and needle set. Materials, design, and components are identical to previously cleared predicates. Intended for manual or powered insertion into bone marrow. Sterility and biocompatibility specifications remain unchanged from predicate devices.
Indications for Use
Indicated for pediatric patients (approx. 3kg-39kg) requiring emergent, urgent, or medically necessary vascular access when IV access is difficult or impossible. Provides intraosseous access in the proximal tibia, distal tibia, or humeral head for up to 24 hours.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
JUL 2 7, 2010
ﮩ ﺗﮭﺎ ﺗ
Image /page/0/Figure/1 description: The image shows the text "5. 510(k) Summary" above the word "vidacare" in a stylized font. The number 5 is followed by a period. The word summary is on the same line as 510(k).
4350 Lockhill Selma Road Shavano Park, TX 78249+2095 210-375-8500 www.Vidacare.com
### SUMMARY
Submitter's name: Address:
Phone: Fax number: Vidacare Corporation 4350 Lockhill Selma Road Shavano Park, TX 78249 210-375-8500 210-375-8537
Name of contact person:
Grace Holland Regulatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821
Date the summary was prepared: April 9, 2010
Name of the devices: Powered PD-IO Intraosseous Infusion System; EZ-IO Humeral Head; Powered PD-IO.
| Common or usual name: | Intraosseous Infusion System |
|-----------------------|--------------------------------|
| Classification name: | Hypodermic single lumen needle |
| Class: | Class II |
| CFR Reference number: | 880.5570 |
| Product Code: | FMI |
### The legally marketed devices to which we are claiming equivalence [807.92(a) (3)]:
| | 510(k)<br>Number | | Trade or Proprietary or Model Name | | Manufacturer |
|---|------------------|---|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---|----------------|
| 1 | K051992 | 1 | Powered PD-IO Infusion System | 1 | Vidacare Corp. |
| 2 | K052408 | 2 | EZ-IO Humeral Head | 2 | Vidacare Corp. |
| 3 | K063142 | 3 | Powered PD-IO | 3 | Vidacare Corp. |
| 4 | K091140 | 4 | Vidaport Intraosseous Infusion System<br>(formerly K032885);<br>EZ-IO Humeral Head (formerly K052408;<br>EZ-MIO, EZ-IO, Distal Tibia (formerly<br>K062956). | 4 | Vidacare Corp. |
{1}------------------------------------------------
Summary of the technological characteristics of our device compared to the predicate devices:
This submission seeks to extend the indications for use for the Powered PD-IO Intraosseous Infusion System (K051992); EZ-IO Humeral Head (K052408) and Powered PD-IO (K063142) for emergent, urgent and medically necessary intraosseous vascular access for the delivery of drugs and fluids to pediatric patients. Vidacare manufactures all of these products. The expansion of indications for these products requires no new technology to facilitate the safe application of the product. There have been no changes to the design or components of the devices cleared under 510(k) K051992, K052408, K063142 and therefore the comparison of technological characteristics listed below are identical.
Target Population Driver Design Features Needle Design Technique Sterility Biocompatibility Where Used
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Vidacare Corporation C/O Ms. Grace Holland Regulatory Specialist, Incorporated 3722 Ave. Sausalito Irvine, California 92606
JUL 27 2010
Re: K101026
Trade/Device Name: Powered PD-IO Intraosseous Infusion System; EZ-IO Humeral Head; Powered PD-IO Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: July 9, 2010 Received: July 13, 2010
Dear Ms. Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Ms. Holland
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Turner
Anthony D. Watson, B.S., M.
Anthony D atson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{4}------------------------------------------------
# 4. Indications for Use Statement
Indications for Use
510(k) Number (if known): ***_*****_**
Device Name: The Powered PD-IO
The Powered PD-IO provides intraosseous access in the distal tibia of pediatric patients as an alternative to IV access in emergent, urgent, or medically necessary cases for up to 24 hours.
1 27 2018
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ==============================================================================================================================================================================
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Rld C. 7/28/10
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Regulatory Specialists, Inc.
{5}------------------------------------------------
## 4. Indications for Use Statement
## Indications for Use
510(k) Number (if known): _
Device Name: _________________________________________________________________________________________________________________________________________________________________ EZ-10 Humeral Head
> The Humeral Head EZ-IO provides vascular access when standard venous access is not possible. Humeral head IO access is indicated in pediatrics in emergent, urgent, or medically necessary cases when intravenous access is difficult or impossible to obtain for up to 24 hours.
× AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ali C. A 7/28/10
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Contr
Regulatory Specialists, Inc.
510(k) Number:
{6}------------------------------------------------
## 4. Indications for Use Statement
Indications for Use
JUI 2 7 2010
510(k) Number (if known): _
Device Name: Powered PD-IO Intraosseous Infusion System
The Powered PD-IO Intraosseous Infusion System provides intraosseous access in the proximal tibia as an alternative to IV access in emergent, urgent, or medically necessary cases for up to 24 hours. The device is for use in pediatric patients (approximate weight range: 3kg-39kg).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rid C. Chape 7/27/10
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page
ge 1 of 1
510(k) Number: *_*_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.