K100991 · International Medsurg Connection · KKX · Oct 5, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100991
Device Name
IMC SURGICAL DRAPE
Applicant
International Medsurg Connection
Product Code
KKX · General, Plastic Surgery
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
These devices are intended to be used as protective coverings used to isolate incision sites and protect against contamination during surgical procedures with the additional feature of a finger cot which is latex-free rubber. This submission includes drapes that will be sold both sterile and non-sterile. Non-sterile drapes are to be sold to OEMs for EO sterilization according to ANSI/AMMI/ISO 11135. Sterile drapes are to be sold directly to users after EO sterilization validation to ANSI/AMMI/ISO 11135
Device Story
IMC Surgical Drape consists of patient protective coverings designed to isolate surgical incision sites and prevent contamination. Devices feature an integrated latex-free rubber finger cot and a fluid collection bag. Available in multiple configurations (Bladder Suspension, GYN, TUR) with varying dimensions and aperture sizes. Used in clinical surgical environments by healthcare professionals. Drapes are provided either sterile or non-sterile (for OEM EO sterilization). Functionality relies on physical barrier properties to maintain a sterile field during procedures.
Clinical Evidence
Bench testing only. Performance characteristics evaluated against ISO and ASTM standards: Flammability (16 CFR Part 1610, Class I), Hydrostatic Pressure (AATCC 127-2008), Impact Penetration (AATCC 42-2007), Linting (ISO 9073-10), and Tensile Testing (ASTM D5034-2009). All results met intended use requirements and were comparable to predicate devices.
Technological Characteristics
Materials tested per ISO 10993-5, 10993-10, and 10993-11. Features include non-latex rubber finger cot and fluid collection bag. Sterilization via Ethylene Oxide (EO) per ANSI/AMMI/ISO 11135. Mechanical properties verified via ASTM D5034-2009 and AATCC standards.
Indications for Use
Indicated for use as protective surgical drapes to isolate incision sites and prevent contamination during surgical procedures, including bladder suspension, GYN, and TUR procedures. Suitable for use in clinical surgical settings.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Medline Blockade and Resistat Non-Sterile Surgical Gowns and Surgical Drapes (K003755)
Submission Summary (Full Text)
{0}------------------------------------------------
#### K100991 510(K) SUMMARY
| Trade Name: | IMC Surgical Drape | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Common Name: | Surgical Drape | |
| Classification Name: | Surgical Drape and Drape Accessories<br>(21 CFR subpart E §878.4370) | |
| Submitter Information: | International Medsurg Connection<br>935 N Plum Grove Rd, STE F<br>Schaumburg, Illinois 60173 | OCT 5 2010 |
| Summary Prepared By: | Peter Kim<br>Quality Manager<br>International Medsurg Connection<br>935 N Plum Grove Rd, STE F<br>Schaumburg, Illinois 60173<br>Telephone: 847-619-9926<br>Fax: 847-619-9927<br>e-mail: peterkim@intlmedsurg.com | |
| Date Prepared: | June 7, 2010 | |
| Predicate Devices: | • 3M Steri-Drape™ (K031287)<br>• Medline (K003755) | |
### Device Name(s):
IMC Surgical Drape
#### Classification Panel:
General and Plastic Surgery
#### Legally Marketed Device Under Which Substantial Equivalence is Being Claimed:
International Medsurg Connections, Inc is claiming substantial equivalence of the IMC Surgical Drape with the currently marketed:
| Description | 510(k) Number | Clearance Date |
|-------------------------------------------------------------------------------------|---------------|----------------|
| 3M Steri-DrapeTM | K031287 | 9/10/2003 |
| Medline BlockadeTM and ResistatTM Non-Sterile<br>Surgical Gowns and Surgical Drapes | K003755 | 2/16/2001 |
#### Device Description
The IMC Surgical Drape devices are patient protective coverings used to isolate incision sites and protect against contamination during surgical procedures with the additional feature of a finger cot which is made of non-latex rubber.
{1}------------------------------------------------
# K 100991
The Bladder Suspension Drape, the GYN Drape and the TUR drape have the same features and same materials but have different dimension of drape and aperture.
#### Statement of Intended Use
These devices are intended to be used as protective coverings used to isolate incision sites and protect against contamination during surgical procedures with the additional feature of a finger cot which is latex-free rubber. This submission includes drapes that will be sold both sterile and non-sterile.
Non-sterile drapes are to be sold to OEMs for EO sterilization according to ANSI/AMMI/ISO 11135. Sterile drapes are to be sold directly to users after EO sterilization validation to ANSI/AMMI/ISO 11135
| Catalog Number | Description |
|----------------|--------------------------|
| 31-0051 | Bladder Suspension Drape |
| 31-0051NS | Bladder Suspension Drape |
| 61-0091 | GYN Drape |
| 61-0091NS | G YN Drape |
| 81-0011 | TUR Drape |
| 81-0011NS | TUR Drape |
#### New Devices as Compared to Marketed Device(s)
The IMC Surgical Drape and the predicate devices (3M Steri Drape TM, and Medline Blockade™ and Resistat™ Non-Sterile Surgical Gowns and Surgical Drapes) to be used as protective coverings used to isolate incision sites and protect against contamination during surgical procedures with the additional feature of a finger cot.
| Feature/<br>Characteristic | IMC Surgical Drape K100991 | 3M Steri-DrapeTM<br>K031287<br>(Predicate) | Medline<br>K003755<br>(Predicate) |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------------------|
| Material Composition<br>Aperture for surgical<br>site | These materials were tested in<br>accordance with ISO 10993-<br>5:2009, 10993-<br>10:2002/Amd.1:2006, and<br>10993-11:2006(E) test<br>methods and were found to be<br>acceptable for the intended use | Similar | Similar |
| Absorbent Prevention<br>Fabric | These materials were tested in<br>accordance with ISO 10993-<br>5:2009, 10993-<br>10:2002/Amd.1:2006, and<br>10993-11:2006(E) test<br>methods and were found to be | Same | Same |
{2}------------------------------------------------
510(h) Summery K100991
| Feature/<br>Characteristic | IMC Surgical Drape K100991 | 3M Steri-DrapeTM<br>K031287<br>(Predicate) | Medline<br>K003755<br>(Predicate) |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-----------------------------------|
| Finger Cot | acceptable for the intended use<br>These materials were tested in<br>accordance with ISO 10993-<br>5:2009, 10993-<br>10:2002/Amd.1:2006, and<br>10993-11:2006(E) test<br>methods and were found to be<br>acceptable for the intended use | Same | same |
| Color Additive | These materials were tested in<br>accordance with ISO 10993-<br>5:2009, 10993-<br>10:2002/Amd.1:2006, and<br>10993-11:2006(E) test<br>methods and were found to be<br>acceptable for the intended use | Same | Same |
| Design Feature<br>Finger Cot | Bladder Suspension Drape,<br>GYN Drape,<br>TUR Drape | Same | Same |
| Fluid Collection<br>Bag | Bladder Suspension Drape,<br>GYN Drape,<br>TUR Drape | Same | Same |
| Aperture<br>Bladder<br>Suspension | 40cm (15.7") x 41cm (16.1") | Similar | Similar |
| GYN | 25cm (9.8") x 25 cm (9.8") | Similar | Similar |
| TUR | 12.5cm (4.9") x 7.5cm (2.95") | Similar | Similar |
## Performance Data:
10 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
ាច្រើន ប្រ
100 million in the state of the states of the states of the states of the states of the states of the states of the states of the states of the states of the states of the st
| Performance<br>Characteristics | Test Method | IMC Surgical<br>Drape K100991 | 3M Steri-DrapeTM<br>K031287<br>(Predicate) | Medline<br>K003755<br>(Predicate) |
|-----------------------------------------------------|------------------|-------------------------------|--------------------------------------------|-----------------------------------|
| Flammability of Clothing<br>Textiles<br>Section 11 | 16 CFR Part 1610 | Class I | Same | Same |
| Hydrostatic Pressure<br>Section 12 | AATCC 127-2008 | Meets Intended<br>use | Same | Same |
| Impact Penetration<br>Section 13 | AATCC 42-2007 | Meets Intended<br>use | Same | Same |
| Determine Lint and Other<br>particles<br>Section 14 | ISO 9073-10 | Meets Intended<br>use | Same | Same |
| General Tensile Testing of<br>Fabric<br>Section 15 | ASTM D5034-2009 | Meets Intended<br>use | Same | Same |
{3}------------------------------------------------
### Conclusions:
The indications for use, technology, specification, safety of the IMC Surgical Drapes and the two predicate devices K031287 and K003755 are essentially the same. The differences between the drapes are minor and do not raise new issues of safety or effectiveness. Hence, the IMC Surgical Drapes are substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three abstract shapes resembling arms or limbs extending upwards. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# DCT 5 2010
International Medsurg Connection C/O Mr. Peter Kim 935 N. Plum Grove Road, Suite F Schaumberg, Illinois 60173
Re: K100991
Trade/Device Name: IMC Surgical Drape Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: KKX Dated: September 16, 2010 Received: September 20, 2010
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Kim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh fer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K100991
Device Name: IMC Surgical Drape
Indications for Use:
These devices are intended to be used as protective coverings used to isolate incision sites and protect against contamination during surgical procedures with the additional feature of a finger cot which is latex-free rubber. This submission includes drapes that will be sold both sterile and non-sterile.
Non-sterile drapes are to be sold to OEMs for EO sterilization according to ANSIIAMMIIISO 11135. Sterile drapes are to be sold directly to users after EO sterilization validation to ANSIIAMMIIISO 11135
| Catalog Number | Description |
|----------------|--------------------------|
| 31-0051 | Bladder Suspension Drape |
| 31-0051NS | Bladder Suspension Drape |
| 61-0091 | GYN Procedure Drape |
| 61-0091NS | GYN Procedure Drape |
| 81-0011 | TUR Procedure Drape |
| 81-0011NS | TUR Procedure Drape |
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Elizabeth F. (Laurie) Weller
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.