K100981 · United Orthopedic Corp. · JWH · Jun 14, 2010 · Orthopedic
Device Facts
Record ID
K100981
Device Name
UNITED STEM, PSA TYPE
Applicant
United Orthopedic Corp.
Product Code
JWH · Orthopedic
Decision Date
Jun 14, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Stem - PSA Type is included in U2 Total Knee System-PSA type. U2 Total Knee System-PSA type is indicated for use in skeletally mature patients with severe knee pain and disability undergoing surgery for knee arthroplasty due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue imbalance. U2 Total Knee System-PSA type is designed for cemented use only.
Device Story
Stem-PSA Type is an extension of the U2 Total Knee System (K082424) providing additional stem lengths (150, 200, 240mm) for femoral/tibial revision or augmentation. Device consists of Ti-6Al-4V alloy stems in straight or curved configurations; features spline and flute design for canal filling, rigid fixation, and torsional resistance; distal taper addresses proximal stress shielding. Used by orthopedic surgeons in clinical settings during total knee arthroplasty. Output is a mechanical implant component providing structural support and stabilization for the tibiofemoral joint. Benefits include improved fixation and stability in patients with compromised bone stock or soft tissue imbalance.
Clinical Evidence
Bench testing only. Locking strength test of the Morse Taper was performed as part of the design assurance process to demonstrate safety and effectiveness.
Technological Characteristics
Material: Ti-6Al-4V alloy. Design: Straight and curved stems with spline and flute geometry for canal filling and distal taper for stress shielding reduction. Dimensions: 10mm to 24mm diameter; 150mm, 200mm, and 240mm lengths. Fixation: Cemented. Mechanical interface: Morse Taper.
Indications for Use
Indicated for skeletally mature patients with severe knee pain and disability requiring knee arthroplasty due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, or posttraumatic joint loss. Includes patients requiring augmentation/stem extensions for inadequate bone stock or constrained stabilization for soft tissue imbalance. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
DEPUY Anatomic Modular Knee (AMK) System (K864671)
SMITH & NEPHEW LEGION Revision Knee System (K043440)
Submission Summary (Full Text)
{0}------------------------------------------------
K100981(1/3)
JUN 1 4 2010
(0) Stem, PSA Type : Additional Size
510(k) Summary
## 510(k) Summary of Safety and Effectiveness
### Submission Information
| Company: | United Orthopedic Corporation |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan |
| Contact Person: | Fang-Yuan Ho, |
| | Regulatory Affairs Manager |
| | Phone:+886-3-5773351 ext. 212 |
| | Fax:+886-3-5777156 |
| Date Prepared: | March 15, 2010 |
| Device Identification | |
| Device Name: | Stem, PSA Type |
| Common Name: | Semi-constrained total knee prostheses |
| Classification Name and Reference : | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis per 21CFR 888.3560.<br>This falls under the Orthopedics panel. |
| Predicate Device: | 1. “UNITED” U2 Total Knee system - PSA Type (K082424)<br>2. “DEPUY” Anatomic Modular Knee (AMK) System (K864671)<br>3. “SMITH & NEPHEW” LEGION Revision Knee System (K043440) |
### Device Description:
"UNITED Stem-PSA Type is an additional size extension of stem to the existing, previously cleared "UNITED" U2 Total Knee system product line (K082424). The
Page: Summary-1/3
ﮩﯿ -
.
{1}------------------------------------------------
## (9) Stem, PSA Type : Additional Size
510(k) Summary
materials, design, safety and effectiveness of this subject are identical to the previously cleared U2 Total Knee system. The only difference from the existing products is the length of stem. This subject stem made of Ti-6A1-4V alloy and will be offered in two types which include straight stem and curved stem. The length of extended straight stem distribute to 150, 200 and 240mm, while the curved stem has 240mm only. Each size of stem has fifteen options in diameter which ranging from 10mm to 24mm. Canal filling stems with spline and flute design provide immediate rigid fixation and resistance to torsional movements. Distal taper is designed to address the problems associated with proximal stress shielding, resulting from distal fixation.
### Intended Use
Stem - PSA Type is included in U2 Total Knee System-PSA type. U2 Total Knee System-PSA type is indicated for use in skeletally mature patients with severe knee pain and disability undergoing surgery for knee arthroplasty due to rheumatoid arthritis. osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue imbalance. U2 Total Knee System-PSA type is designed for cemented use only.
Basis for Substantial Equivalence:
Page: Summary-2/3
{2}------------------------------------------------
# (0) Stem, PSA Type : Additional Size
### 510(k) Summary
The safety and effectiveness of Stem - PSA Type are substantially equivalent to the legal on-market stem components in the U2 Total Knee System - PSA Type (straight stem: available in length 30, 75, and 100 mm; curved stem: 150, 200 mm), except for its larger length (straight stem: available in length 150, 200, and 240 mm; curved stem: 240mm). The diameters for the stem components in Stem - PSA Type and U2 Total Knee System - PSA Type both range from 10 mm to 24mm. In addition, the subject straight stem is also substantial equivalence to the "DEPUY" Anatomic Modular Knee (AMK) System (K864671), while the curved stem is also substantial equivalence to the "SMITH & NEPHEW" LEGION Revision Knee System (K043440). The materials for the subject and predicate devices are identical.
### Performance Data:
The indications for use, materials, design, manufacturing process, sterilization of the Stem - PSA Type are all identical to the legally on-market stem components in the U2 Total Knee System - PSA Type. Besides, locking strength test of Morse Taper. completed as part of the design assurance process demonstrated that this device is safe and effective and is substantially equivalent to the predicate device.
Page: Summary-3/3
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, represented by three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
United Orthopedic Corporation % Ms. Fang-Yuan Ho Regulatory Affairs Manager 57 Park Avenue 2, Science Park Hsinchu China (Taiwan) 300
JUN 1 4 2010
Re: K100981
Trade/Device Name: Stem, PSA Type Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: May 14, 2010
Received: May 17, 2010
Dear Ms. Ho:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Fang-Yuan Ho
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark A. Millersen
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indication for Use
510 (k) Number (if known): K10098 |
Device Name: Stem, PSA Type
### Indications for Use:
Stem - PSA Type is included in U2 Total Knee System-PSA type. U2 Total Knee System-PSA type is indicated for use in skeletally mature patients with severe knee pain and disability undergoing surgery for knee arthroplasty due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue imbalance. U2 Total Knee System-PSA type is designed for cemented use only.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE: BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Juta for mxm
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number
Page of — 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.