K100946 · Terumo Europe N.V. · FPA · Aug 2, 2010 · General Hospital
Device Facts
Record ID
K100946
Device Name
INTRAVSCULAR ADMINISTRATION SET
Applicant
Terumo Europe N.V.
Product Code
FPA · General Hospital
Decision Date
Aug 2, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Surflo Winged Infusion Sets with Needle Protection (Surshield) are intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices, Additionally, after withdrawal of the needle from the patient's vein. the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.
Device Story
Sterile, single-use winged infusion set; 23G and 25G needle sizes with 350mm microbore tubing. Device accesses peripheral vascular system for fluid infusion or blood withdrawal via syringe or luer adapter. Features hinged, transparent shield cover attached to wing hub; manually activated by user post-withdrawal to cover needle; audible click confirms latching. Designed to keep user fingers behind needle point to minimize accidental stick risk. Used in clinical settings by healthcare professionals. Output is physical access to vascular system; safety mechanism reduces sharps injury risk.
Clinical Evidence
Bench testing only. No clinical data. Verification activities included flow rate, dead space volume, air/liquid leakage, fitting strength, bonding strength, needle penetration resistance, and safety feature activation force. All tests met acceptance criteria.
Technological Characteristics
Materials: standard medical-grade components. Design: winged hub, microbore tubing, luer adapter, hinged shield cover. Sterilization: Ethylene oxide (EN ISO 11135-1:2007) to SAL 10^-6. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring peripheral vascular access for intravenous fluid administration or blood specimen withdrawal. No specific age or gender restrictions. Contraindications not specified.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887)
Surflo Winged Infusion Set with Needle Protection (Surshield) (K072894)
Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547)
Submission Summary (Full Text)
{0}------------------------------------------------
K100946
### 510(k) SUMMARY as required by 807.92 Summary of Safety & Effectiveness Information
AUS 0 2 2010
Submitter information
| Prepared for: | TERUMO EUROPE N.V.<br>Interleuvenlaan 40,<br>3001 Leuven,<br>BELGIUM |
|-----------------------------|-------------------------------------------------------------------------------------------------|
| Prepared by/Contact person: | Mrs. M.J. Aerts - Manager Regulatory Affairs<br>Tel. (+32) 16 38 13 53<br>Fax (+32) 16 40 02 49 |
| Date prepared: | June 2010 |
### .1. Device Name
### Proprietary Name
SURFLO® Winged Infusion Set with Needle Protection (Surshield™)
Classification Name
Intravascular Administration Set (-FPA)
21 CFR, Section 880.5440
Classification: Class II
#### 2. Reason for Submission
This 510k is being submitted to extend Terumo Europe's cleared Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887 and K072894) product line to include 23G and 25G needle sizes with microbore tubing taking into consideration the potential issues of safety and effectiveness specific for a thinner tubing.
This 510k will provide supporting information that the 23G Surflo Winged Infusion Sets with Needle Protection (Surshield) and microbore tubing design are safe and effective and an acceptable extension of the current cleared Surflo Winged Infusion Set with Needle Protection (Surshield) product line (K052887 and K072894).
- ે. Intended Use
The Surflo Winged Infusion Sets with Needle Protection (Surshield) are intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices, Additionally, after withdrawal of the needle from the patient's vein. the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.
Note: This is the same intended use as the predicate device. Surflo Winged Infusion Set with Needle Protection (Surshield), cleared under KU52887 and KU72894.
Page 10
{1}------------------------------------------------
### 4. Description
The 23G and the 25G Terumo Surflo Winged Infusion Sets with Needle Protection (Surshield) are sterile, single use devices consisting of a needle attached to a winged hub, microbore tubing, adapter and adapter cap, und a hinged shield cover that attaches to the wing just below the needle-to-wing junction.
The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a one- or twohanded technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it.
The device possesses a 350 min length microbore tubing,
#### Substantjal Equivalence 5.
The 23G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) with microbore tubing, manufactured by Terumo Europe N:V., submitted in this 510(k) file are substantially equivalent in intended use, descriptions, technology/principles of operation, materials and performance to the following cleared devices:
- 1. Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887), manufactured by Teruno Europe N.V.
- 2. Surflo Winged Infusion Set with Needle Protection (Surshield) (K072894), manufactured by Terumo Europe N.V.
- 3. Surflo' Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547), manufactured by Terumo Europe N.V.
Any differences between the devices do not raise any significant issues of safety and effectiveness.
Note: For performance testing reference will also be made to the Surflo Winged Infusion Set with Filher and Needle Protection (K070547), as these sets are identical as the sets to be cleared by this submission except for the presence of the fitter, consequently their use is limited to infusion.
#### ઉં: Summary of Technological Modifications compared to Predicate Devices:
The technological modifications made to the subjected device compared to the predicate devices are summarized in the following table:
| Predicate device | Modification summary (Proposed device) |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Surflo Winged Infusion Set with<br>Needle Protection (Surshield)<br>(K052887) & (K072894) | a) Introduction of microbore tubing<br>b) Adapted design Surshield protector (the connection with<br>the tube remains the same) |
| Surflo Winged Infusion Set with<br>Filter and Needle Protection<br>(K070547) | a) Use of the same luer adaptor as predicate device BUT<br>without filter<br>b) Adapted design Surshield protector (the connection with<br>the tube remains the same) |
Page 11
{2}------------------------------------------------
## 7. Summary of Verification Activities:
All necessary verification and validation tests have been performed for the The Surflo Winged Infusion Sets with Needle Protection (Surshield) to assure substantial equivalence to the predicate devices. Summary of all verification activities including acceptance criteria is given in the following table:
| | ACCEPTANCE CRITERIA |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Flow rate | The flow rate for the set with a 23 G needles is ≥ 1.7ml/min & for a set with<br>a 25 G needles is ≥ 1.5 ml/min |
| 2. Dead space volume | Dead space volume for sets with 23 G & 25G needle ≤ 0.25 ml |
| 3. Air flow choke test | No obstruction of fluid through the set |
| 4. Air leakage set<br>(= Integrity of the set) | No air leakage from the set<br>(European Pharmacopoeia.3.2.6.) |
| 5. Air leakage adapter | No air leakage at the luer fitting (EN 1707) |
| 6. Liquid leakage<br>adapter | No liquid leakage at the luer fitting (EN 1707) |
| 7. Conical fitting | 6% luer (EN 20594-1) |
| 8. Fitting strength.<br>protector | The force to pull the protector from the wing hub is ≥ 0.3 N and ≤ 12N |
| 9. Torque resistance<br>cap/adapter | The torque force required to unscrew the cap from the adapter is ≤ 9N.cm. |
| 10.Bonding strength<br>Cannula | The bonding strength between the cannula and the wing hub is ≥ 20N |
| 11.Force at break | The force to separate the the tubing from the wing/connector is ≥ 15N |
| 12.Needle penetration<br>resistance | Needle point ≤ 0.14N<br>Drag ≤ 0.03N |
| 13.Break strength of<br>shield cover joint | The force to detach the Surshield protector from the wing hub is ≥ 4N |
| 14.Force to unlock the<br>safety feature | The force required to unlock the Surshield protector from the cannula is ≥ 2N |
| 15.Misalignment<br>Surshield Protector | The angle measured axially from the cannula between the horizontally<br>positioned wings and the vertically positioned Surshield protector does not<br>exceed 15° |
The Surflo Winged Infusion Sets with Needle Protection (Surshield) met all acceptance criteria as indicated in table above. None of the obtained data raises any new issue of safety and effectiveness.
Page 12
{3}------------------------------------------------
#### 8. Additional Safety Information
The sterility of the Surflo Winged Infusion Sets with Needle Protection (Surshield) is assured by using a validated sterilization method qualified in accordance with in accordance with the requirements of EN ISO 11135-1:2007 "Sterilization of health care products - Ethylene oxide - Part 1:Requirements for development, validation and routine control of a sterilization process for medical devices" to a sterilizy assurance level (SAL) of 10-6 as required by EN 556-1: "Sterilization of Medical Devices -Requirements for medical devices to be designated "STERILE" - Part - !: Requirements for terminally sterilized medical devices".
Ethylene oxide residual levels resulting from EtO sterilization are in compliance with EN ISO 10993-7: "Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals ...
The 23G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) are Externally Communicating devices, Circulating Blood, Limited Exposure (24 hrs). The devices' blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Mennorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing".
The expiration dating for the 25G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) have been established at 5 years which.is the same as the cleared Surflo Winged Infusion Sets with Needle Protection (Surshield) (K052887 and K072894).
- Conclusion 9.
The Surflo Winged Infusion Sets with Needle Protection (Surshield) manufactured by Terumo Europe N.V. and submitted in this 510(k) file are substantially equivalent in intended use. description, specifications, and technology/principles of operation, materials, and performance to the following cleared devices:
- 1. Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887), manufactured by Terumo Europe N.V.
- 2. Surflo Winged Infusion Set with Needle Protection (Surshield) (K072894), manufactured by Terumo Eurone N.V.
- 3. Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547), manufactured by Terumo Europe N.V.
Differences between the devices do not raise any new issues of safely or effectiveness.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is positioned to the right of the text, which is arranged in a circular fashion around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. M.J. Aerts Regulatory Affairs Manager Terumo Europe N.V. Interleuvenlaan 40 B-3001 Leuven-Belgie Belgium
AUG-0 2 2010
Re: K100946
Trade/Device Name: Surflo Winged Infusion Set with Needle Protection (Surshield™) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 1, 2010 Received: July 6, 2010
Dear Ms. Aerts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Ms. M.J. Aerts
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Pease
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indication for Use
510(k) Number (if known): K100946
Device Name: Surflo Winged Infusion Set with Needle Protection (Surshield")
Indication For Use:
The Surflo Winged Infusion Set with Needle Protection (Surshield) is intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.
Prescription Use __ X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. Chapman
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100946
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.