PRIMEWIRE PRESTIAGE PRESSURE GUIDE WIRE, MODELS 8185, 8185J, 8300, 8300J

K100930 · Volcano Corporation · DQX · Apr 30, 2010 · Cardiovascular

Device Facts

Record IDK100930
Device NamePRIMEWIRE PRESTIAGE PRESSURE GUIDE WIRE, MODELS 8185, 8185J, 8300, 8300J
ApplicantVolcano Corporation
Product CodeDQX · Cardiovascular
Decision DateApr 30, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The PrimeWire PRESTIGE™ Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Device Story

Steerable guide wire with pressure transducer mounted 3 cm proximal to tip; measures intravascular blood pressure. Used in conjunction with SmartMap, s5 Family, and ComboMap systems. Diameter 0.014" (0.36 mm); available in 185 cm and 300 cm lengths; straight or pre-shaped tips. Packaged preconnected to connector with torque device. Operated by physicians during diagnostic or interventional procedures in clinical settings. Provides hemodynamic data to assist in diagnosis and treatment of blood vessel disease.

Clinical Evidence

Bench testing only. Performance data and design verification plan confirm comparability to predicate devices.

Technological Characteristics

0.014" diameter steerable guide wire; pressure transducer located 3 cm proximal to tip. Compatible with SmartMap, s5, and ComboMap systems. Available in 185 cm and 300 cm lengths with straight or pre-shaped tips. Includes torque device.

Indications for Use

Indicated for patients undergoing diagnostic angiography or interventional procedures requiring blood pressure measurement in coronary or peripheral vessels.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ki00930 SmartWire Pressure Guide Wire Special 510(k) APR 3 0 2010 ## 510 (K) Summary # PrimeWire PRESTIGE™ Pressure Guide Wire Date Prepared: April 2, 2010 Submitted by: Volcano Corporation 3661 Valley Centre Dr. Suite 200 San Diego, CA 92130 Contact person: Marilyn Pourazar Director of Regulatory Affairs Phone number: (858) 720-4116 Facsimile number: (858) 720-0335 PrimeWire PRESTIGETM Pressure Guide Wire Device Name: | Classification name: | | Class | |----------------------|---------------------------------------------|-------| | | > 870.1330 Catheter guide wire | II | | | > 870.2870 Catheter tip pressure transducer | II | ### Predicate Device: The Volcano PrimeWire PRESTIGE™ Pressure Guide Wire is substantially equivalent to the following: | 510(k) Number | Product Name | Clearance Date | |----------------|------------------------------------|----------------| | K021209 | SmartWire/SmartMap Pressure System | May 17, 2002 | | K070487 | SmartWire II Pressure Guide Wire | March 16, 2007 | | Letter to File | PrimeWire Pressure Guide Wire | June 26, 2008 | | Letter to File | PrimeWire Pressure Guide Wire | April 6, 2009 | ## Device Description: The PrimeWire PRESTIGE™ Pressure Guide Wire is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The PrimeWire PRESTIGE™ guide wire measures pressure when used with the SmartMap, s5 Family, and ComboMap 19 {1}------------------------------------------------ systems. The PrimeWire PRESTIGE™ guide wire has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm and 300 cm, with straight or pre-shaped tips. The PrimeWire PRESTIGE™ guide wire is packaged preconnected to the connector with a torque device to facilitate navigation through the vasculature. ### Intended Use: The PrimeWire PRESTIGE™ Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease. # Device Technological Characteristics and Comparison to Predicate Device: The Volcano Corporation PrimeWire PRESTIGE™ Pressure Guide Wire Device is substantially equivalent to the predicate devices, Volcano SmartWire II/PrimeWire Pressure Guide Wires. The modified PrimeWire PRESTIGE™ Pressure Guide Wire Device uses the same fundamental scientific technology and has the same intended use as that of the predicate device. #### Performance Data: Applicable testing was performed in accordance with the Design Verification Plan including a Risk Analysis addressing the impact of modifications to the device and components. The test results indicate the revised product is comparable to the predicate device. #### Conclusion: The Volcano PrimeWire PRESTIGE™ Pressure Guide Wire Device has the same performance specifications, fundamental scientific technology and intended use as that of the predicate devices, Volcano SmartWire II/PrimeWire Pressure Guide Wires. 20 {2}------------------------------------------------ Modifications to the device do not raise any new questions regarding safety or efficacy. The performance data and a declaration of conformity with design controls support a determination of continuing substantial equivalence of the modified device to the predicate device. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized human figure or bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 APR 3 0 2010 Volcano Corporation c/o Ms. Marilyn Pourazar Director of Regulatory Affairs 3661 Valley Centre Dr., Suite 200 San Diego, CA 92130 Re: K100930 Device Name: PrimeWire PRESTIGE™ Pressure Guide Wire, Models 8185, 8185J, 8300, 8300J Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (Two) Product Code: DOX Dated: April 2, 2010 Received: April 5, 2010 Dear Ms. Pourazar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approvál application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Marilyn Pourazar Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Singerely yours. Bram D. Zuckerman, M.D. Directd Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K100930 PrimeWire PRESTIGETM Pressure Guide Wire Special 510(k) 510(k) Number (if known): PrimeWire PRESTIGE™ Pressure Guide Wire Device Device Name: Indications for Use: The PrimeWire PRESTIGE™ Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease. The intended use and indications for use of the modified device as described in its labeling have not changed. The fundamental scientific technology of the modified device has not changed. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X OR Over-the-Counter Use (Per 21 CFR 801.19) W.M.C. vision Sian-Off) Division of Cardiovascular Devices 510(k) Numbe
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...